← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K082527

# ZIMMER UNIVERSAL LOCKING SYSTEM: 3.5 MM LOCKING PLATES AND SCREWS (K082527)

_Zimmer, Inc. · HRS · Dec 18, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K082527

## Device Facts

- **Applicant:** Zimmer, Inc.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Dec 18, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Universal Locking System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including: ○ Comminuted fractures ○ Supracondylar fractures ○ Extra-articular fractures ○ Fractures in osteopenic bone ○ Nonunions ○ Malunions

## Device Story

Zimmer Universal Locking System consists of 3.5 mm locking plates and screws for internal fracture fixation. Plate configurations include dual compression, reconstruction, tubular, straight 'T', and oblique 'T'. Plates feature figure-8 shaped holes accommodating standard or locking screws. Used by surgeons in clinical settings for stabilization of bone fractures and osteotomies. Device provides mechanical support to bone segments during healing process; aids in alignment and stabilization of complex fracture types including osteopenic bone. Output is physical fixation of bone fragments.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Materials: Tivanium (Ti-6Al-4V alloy) and CP Grade Titanium. System includes 3.5 mm locking plates and screws. Plate designs: dual compression, reconstruction, tubular, straight 'T', and oblique 'T'. Features figure-8 shaped holes for locking or standard screw accommodation. Mechanical fixation device; no energy source or software.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Zimmer Universal Locking System, 3.5mm Plates and Screws ([K060710](/device/K060710.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K082527

# DEC 1 8 2008

# Summary of Safety and Effectiveness

| Submitter:                            | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708                                                                                                                                                                                                                                                                              |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                       | Anthony Francalancia, RAC<br>Senior Associate, Regulatory Affairs<br>Telephone: (574) 372-4570<br>Fax: (574) 372-4605                                                                                                                                                                                                              |
| Date:                                 | August 29, 2008                                                                                                                                                                                                                                                                                                                    |
| Trade Name:                           | Zimmer® Universal Locking System: 3.5 mm<br>Locking Plates and Screws (Tivanium® Ti-6Al-4V<br>Alloy, CP Grade Titanium)                                                                                                                                                                                                            |
| Common Name:                          | 3.5mm TiULS Locking Plate System                                                                                                                                                                                                                                                                                                   |
| Classification Name<br>and Reference: | Plate, Fixation, Bone (21 CFR § 888.3030)<br>Screw, Fixation, Bone (21 CFR § 888.3040)                                                                                                                                                                                                                                             |
| Predicate Device:                     | Zimmer Universal Locking System, 3.5mm Plates<br>and Screws, manufactured by Zimmer, Inc.,<br>K060710, cleared April 26, 2006                                                                                                                                                                                                      |
| Device Description:                   | The Zimmer Universal Locking System is a plate<br>and screw system intended for internal fracture<br>fixation. The plate selection consists of dual<br>compression, reconstruction, tubular, straight "T"<br>and oblique "T" configurations. Plates<br>accommodate either standard or locking screws via<br>figure-8 shaped holes. |
| Intended Use:                         | The Universal Locking System is indicated for<br>temporary internal fixation and stabilization of<br>osteotomies and fractures, including:<br>○ Comminuted fractures<br>○ Supracondylar fractures<br>○ Extra-articular fractures<br>○ Fractures in osteopenic bone<br>○ Nonunions<br>○ Malunions                                   |

·

{1}------------------------------------------------

Comparison to Predicate Device:

The Zimmer Universal Locking System: 3.5mm Locking Plates and Screws (Tivanium Ti-6A1-4V Alloy, CP Grade Titanium) have the same intended use, similar performance characteristics and are similar in design to the prodicate devices.

Performance Data (Nonclinical and/or Clinical):

Non-Clinical Performance and Conclusions:

The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and cffective.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services, USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Zimmer, Inc. % Mr. Anthony Francalancia Senior Regulatory Affairs Associate P.O. Box 708 Warsaw, Indiana 46581

DEC 1 8 2008

Re: K082527

Trade/Device Name: Zimmer® Universal Locking System: 3.5 mm Locking Plates and Screw (Tivanium® Ti-6Al-4V Alloy, CP Grade Titanium) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS, HWC Dated: December 8, 2008 Received: December 9, 2008

Dear Mr. Francalancia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Anthony Francalancia

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M. Millman

Mark N. Melkerson Director Division of General; Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

## 510(k) Number (if known):

#### Device Name:

Zimmer® Universal Locking System: 3.5 mm Locking Plates and Screws (Tivanium® Ti-6Al-4V Alloy, CP Grade Titanium)

### Indications for Use:

The Universal Locking System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including:

- Comminuted fractures o
- Supracondylar fractures .0
- Extra-articular fractures o
- O Fractures in osteopenic bone
- Nonunions ಂ
- Malunions 0

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

| Concurrence of CDRH, Office of Device Evaluation (ODE)     |             |
|------------------------------------------------------------|-------------|
|                                                            |             |
| (Division Sign-Off)                                        |             |
| Division of General, Restorative, and Neurological Devices | Page 1 of 1 |
| 510(k) Number                                              | K082527     |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K082527](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K082527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
