← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K062747

# ARTHREX RETROBUTTON, 15MM, 60MM, RETROBUTTON, LONG, 15MM, 60MM (K062747)

_Arthrex, Inc. · HRS · Sep 29, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K062747

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Sep 29, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Arthrex RetroButton™ for fixation of bone to bone or soft tissue to bone, and is intended as a fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrex will be offering this for Anterior Cruciate Ligament (ACL) Repair.

## Device Story

Arthrex RetroButton™ is a pre-threaded button construct used for orthopedic fixation. It functions as a fixation post or distribution bridge to manage suture tension during ligament or tendon repair, specifically ACL reconstruction. The device is used by surgeons in an operating room setting. By providing a stable anchor for sutures, it facilitates secure tissue-to-bone or bone-to-bone attachment, aiding in the healing and stabilization of the repaired site.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Pre-threaded button construct for bone fixation. Materials include synthetic nonabsorbable suture (polyethylene or silk). Device is a metallic bone fixation appliance.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex FiberWire® Button Repair Kit ([K031666](/device/K031666.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K062747 pg 1 of 1

# VIII. 510(k) Summary of Safety and Effectiveness

## Arthrex RetroButton™

| Manufacturer / Sponsor                 | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, Florida 34108-1945                                                 |  |  | SEP 29 2000 |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--|--|-------------|
| 510(k) Contact:                        | Ann Waterhouse, RAC<br>Regulatory Affairs Project Manager<br>Telephone: (239) 643-5553 ext. 1179<br>FAX: (239) 598-5508 |  |  |             |
| Trade Name:                            | Arthrex RetroButton™                                                                                                    |  |  |             |
| Common Name:                           | Plate, fixation, bone<br>Suture, synthetic, nonabsorbable, polyethylene<br>or silk                                      |  |  |             |
| Product Code /<br>Classification Name: | HRS, 21 CFR 888.3030<br>Plate, Fixation, bone                                                                           |  |  |             |
|                                        | GAT, 21 CFR 878.5000<br>Suture, synthetic, nonabsorbable, polyethylene                                                  |  |  |             |
|                                        | GAP, 21 CFG 878.5030<br>Suture, nonabsorbable, silk                                                                     |  |  |             |
| Predicate Device:                      | Arthrex FiberWire® Button Repair Kit K031666                                                                            |  |  |             |

# Device Description and Intended Use:

The Arthrex RetroButton™ is a pre-threaded button construct.

The Arthrex RetroButton™ for fixation of bone to bone or soft tissue to bone, and is intended as a fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrey will he offering this for Anterior Cruciate Ligament (ACL) Repair.

### Substantial Equivalence:

The Arthrex RetroButton™ is substantially equivalent to the predicate Arthrex FiberWire® Button Repair Kit in which the basic features and intended uses are the same. Any differences between the Arthrex RetroButton™ and the prodicate are considered minor and do not raise questions concerning safety and effectiveness. Based on the information submitted, Arthrex, Inc. has deterrined that the RetroButton™ is substantially equivalent to the currently marketed predicate device.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

od and Drug Administration 200 Corporate Boulevard Rockville MD 20850

Arthrex, Inc. % Ms. Ann Waterhouse, RAC Regulatory Affairs Project Manager 1370 Creekside Boulevard Naples, Florida 34108-1945

SEP 29 2006

Re: K062747

============================================================================================================================================================================== Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS, GAT, GAP Dated: September 13, 2006 Received: September 14, 2006

Dear Ms. Waterhouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

# Page 2 - Ms. Ann Waterhouse, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) rmarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use Form ll.

510(k) Number (if known):

510(k) Number (if known): 《062747

Device Name: Arthrex RetroButton™

Indications for Use:

The Arthrex RetroButton™ for fixation of bone to bone or soft tissue to bone, and is intended as a fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrex will be offering this for Anterior Cruciate Ligament (ACL) Repair.

Prescription Use_ × (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1 -

(Division b Division of General, Restorative, and Neurological Devices

**510(k) Number** IL062797

9

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K062747](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K062747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
