← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K052248

# DORSAL NAIL PLATE, ANATOMICAL (K052248)

_Hand Innovations, LLC · HRS · Sep 14, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052248

## Device Facts

- **Applicant:** Hand Innovations, LLC
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Sep 14, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Dorsal Nail Plate™, Anatomical is intended for the fixation of fractures and osteotomies involving the distal Radius.

## Device Story

Dorsal Nail Plate™, Anatomical is a metallic bone fixation device for distal radius fractures and osteotomies. Device consists of a distal segment for attachment to the dorsal surface of the distal radius and a proximal segment for insertion into the radial shaft. Used by surgeons in clinical settings to stabilize bone segments during healing. Output is physical fixation of bone. Benefits include structural support for fracture repair.

## Clinical Evidence

No clinical data; equivalence confirmed through pre-clinical testing.

## Technological Characteristics

Titanium alloy meeting ASTM F 136 requirements. Metallic bone fixation plate. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Dorsal Nail Plate of the Distal Radius Fracture Repair System ([K023007](/device/K023007.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 1 4 2005

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

K052248 pi/

## 510(k) Summary of Safety and Effectiveness

| General<br>Provisions                    | The name of the device is:                                                                                                                                                                                                                                                                                     |                      |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
|                                          | Proprietary Name                                                                                                                                                                                                                                                                                               | Common or Usual Name |
|                                          | Dorsal Nail Plate™, Anatomical                                                                                                                                                                                                                                                                                 | Plate Fixation Bone  |
| Name of<br>Predicate<br>Devices          | The device is substantially equivalent to:                                                                                                                                                                                                                                                                     |                      |
|                                          | • Dorsal Nail Plate of the Distal Radius Fracture Repair System (510(k) #<br>K023007 - December 5, 2002) - Hand Innovations, Inc.                                                                                                                                                                              |                      |
| Classification                           | Class II.                                                                                                                                                                                                                                                                                                      |                      |
| Performance<br>Standards                 | Performance standards have not been established by the FDA under section 514 of<br>the Food, Drug and Cosmetic Act.                                                                                                                                                                                            |                      |
| Indications for<br>Use                   | The Dorsal Nail Plate™, Anatomical is intended for the fixation of fractures<br>and osteotomies involving the distal Radius.                                                                                                                                                                                   |                      |
| Device<br>Description                    | The Dorsal Nail Plate™, Anatomical is intended to treat fractures in both left and<br>right Radius. The distal segment of the nail plate is intended to be attached to the<br>dorsal surface of the distal Radius. The proximal segment of the nail plate is<br>intended to be inserted into the Radial shaft. |                      |
| Biocompati-<br>bility                    | The Dorsal Nail Plate™, Anatomical do not require biocompatibility testing<br>because the Titanium Alloy used in fabrication meets the requirements of ASTM<br>F 136.                                                                                                                                          |                      |
| Summary of<br>Substantial<br>Equivalence | The Dorsal Nail Plate™, Anatomical is substantially equivalent to the predicate<br>Dorsal Nail Plate in regards to the intended use, materials, biocompatibility, and<br>overall performance characteristics. The equivalence was confirmed through pre-<br>clinical testing.                                  |                      |

:

.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing left and has its wings spread. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

SEP 1 4 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Ernesto Hernandez Vice President RA/QA Hand Innovations, LLC 8905 SW 87th Avenue, Suite 220 Miami, Florida 33176

Re: K052248 Trade/Device Name: Dorsal Nail Plate™, Anatomical Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: August 16, 2005 Received: August 18, 2005

Dear Mr. Hernandez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1.050.

{2}------------------------------------------------

Page 2-Mr.Ernesto Hernandez

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Metkerson

Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure 

{3}------------------------------------------------

510(k) Number (if known): K052348

Device Name: Dorsal Nail Plate™, Anatomical

## Indications for Use Statement

The Dorsal Nail Plate™, Anatomical is intended for the fixation of fractures and osteotomies involving the distal Radius.

Prescription Use______________________________________________________________________________________________________________________________________________________________

.

OR

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Dorsal Nail Plate™, Anatomical 510(k) Premarket Notification

510(k) Number_LOS UZ48 0024

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052248](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
