← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K052152

# B-BOP PLATE (K052152)

_New Deal, S.A. · HRS · Sep 13, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052152

## Device Facts

- **Applicant:** New Deal, S.A.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Sep 13, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The B-BOP® plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include: - Moderate to severe hallux valgus - -Hallux varus

## Device Story

B-BOP® plate is a metallic bone fixation device designed for basal osteotomy of the first metatarsal. Device features anatomical, pre-bent shape for plantar positioning to enhance stability and reduce bulk; available in right/left configurations with 5° or 10° angulations. Implanted by surgeons in clinical settings to stabilize bone segments following osteotomy. Fixation achieved using four Snap-Off screws (2.7mm or 3.0mm diameter). Plate surface is sandblasted to improve bone adherence. Benefits include improved correction of significant forefoot deformities like hallux valgus and hallux varus.

## Clinical Evidence

Bench testing only. Mechanical performance validated through bending strength and fatigue testing. No clinical data provided.

## Technological Characteristics

Material: Titanium alloy (Ti-6AI-4V ELI). Design: Anatomical, pre-bent plate for plantar positioning; sandblasted bone-contact surface. Fixation: 4-hole screw configuration using 2.7mm/3.0mm diameter Snap-Off screws. Sterilization: Provided sterile. Connectivity: None (mechanical implant).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex Small Fragment Plates and Screws ([K040907](/device/K040907.md))
- Acumed Congruent Bone Plate System ([K012655](/device/K012655.md))
- Synthes Modular Foot System ([K001941](/device/K001941.md))

## Reference Devices

- HALL 0-Fix system ([K021626](/device/K021626.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K052152 p1/2

Image /page/0/Picture/1 description: The image shows the logo for Newdeal, an Integra LifeSciences Company. The logo features the word "newdeal" in bold, black font, with a circle to the left of the word. The circle has four black dots and one gray dot. Below the word "newdeal" is the text "An Integra LifeSciences Company".

SEP 1 3 2005

## 510(k) SUMMARY

Submitter's Name and Address A. Newdeal SAS 10, place d'Helvétie 69006 LYON FRANCE Tel .: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52 ESTABLISHMENT REGISTRATION NUMBER: 9615741

#### Contact Person B.

Morgane GRENIER Regulatory and Clinical Affairs Manager Newdeal SAS 10, place d'Helvétie 69006 LYON FRANCE Tel: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52

- ं Date Summary Prepared August 1, 2005
- Name of Device D. Proprietary Name: B-BOP® plate

Common Name: Plate, fixation, bone

## Classification Name and Reference

Single/multiple component metallic fixation appliances and bone accessories (21 CFR 888.3030)

Device Product Code: HRS

### Proposed Regulatory Class: Class II

Panel: Orthopedic

#### Device Description E.

In the case of significant deformities (Hallux Valgus, Hallux Varus) of the forefoot, the basal or proximal osteotomy is generally more indicated than the distal osteotomy allowing greater corrections of the 1st metatarsal.

{1}------------------------------------------------

\$\cos 2/52 \ p^{2}/_{2}\$

Newdeal has developed a plate dedicated to the fixation of basal osteotomy of the 1st metatarsal by designing the B-BOP® plate which allow to meet at best the specifications of that kind of surgical procedure.

Characteristics of the plate have taken into account all the requirements associated with basal osteotomy:

- Plantar positioning to obtain the best stability and resistance and allowing to reduce bulk
- Anatomical shape adapted to the plantar curve of the 15 metatarsal । (pre-bent), available in right or left side and in two angulations (5° and 10°) for medium or large deformation.

The B-BOP® plate is made from titanium alloy (Ti-6AI-4V ELI) and the part of the plate in contact with bone is sandblasted for a better adherence of the plate.

Plates and screws are also available which are color-coded for ease of identification.

Fixation of the plate is provided by four Snap-Off screws, already present with the HALL 0 -Fix system (cleared under K021626).

With the B-BOP® plate, screws are available in 2 diameters (2.7mm and 3.0mm) and in a length range from 14mm to 26mm (2 mm increment). The screws are provided sterile with the B-BOP® plate.

#### F. Indications for Use

The B-BOP® plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include:

- Moderate to severe hallux valgus -
- -Hallux varus

#### Substantial Equivalence G.

The B-BOP® plate is substantially equivalent in terms of design, material, indications for use and fixation with the following predicate devices:

| Arthrex | Small Fragment Plates and Screws | K040907 |
|---------|----------------------------------|---------|
| Acumed  | Congruent Bone Plate System      | K012655 |
| Synthes | Modular Foot System              | K001941 |

#### Comparison of Technological Characteristics H.

The technological characteristics of the B-BOP® plate are the same as the characteristics of predicate devices in terms of intended use and design. All of these plates have the following characteristics:

- Holed for screwed fixation -
- Made from Titanium alloy or stainless steel -

{2}------------------------------------------------

- Equivalent size range -
- Intended to be implanted for fixation of osteotomies, including basal osteotomy of the metatarsal.

#### I. Summary of Studies

Bending strength test and fatigue testing have been carried out. The B-BOP® plate meets our acceptance criteria.

#### J. Conclusion

The B-BOP® plate is substantially equivalent to commercially marketed devices, the Arthrex Small Fragment Plates and Screws (K040907), the Acumed Congruent Bone Plate System (K012655) and the Synthes Modular Foot System (K001941).

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

SEP I 3 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Newdeal SAS c/o Ms. Judith E. O'Grady Integra Lifesciences 311 Enterprise Drive Plainsboro, New Jersey 08536

Re: K052152 Trade/Device Name: B-BOP® plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: August 1, 2005 Received: August 9, 2005

Dear Ms. O'Grady:

We have reviewed your Section 510(k) premarket notification of intent to market the device w for have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can haf our d in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Judith E. O'Grady

This letter will allow you to begin marketing your device as described in your Section 510(k) I his icter will anow you to ough mating of substantial equivalence of your device to a legally premarket notification: "The PDF intellight for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at not 10-10-10-120. Also, please note the regulation entitled, Colliation of Compulars as (21 the Part 807.97). You may obtain Misoranding of reference to premessionsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

signature

or Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

s2152

Device Name: B-BOP® plate

Indications For Use:

The B-BOP® plate is intended for fixation of osteotomy of the basis of the first metatarsal. Examples include:

- Moderate to severe hallux valgus -
- Hallux varus -

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

・・

AND/OR

Over-The-Counter Use __ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of 1

14052152 510(k) Number_

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052152](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K052152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
