← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K040588

# SYNTHES (USA) 2.4 MM LCP WRIST PLATE (K040588)

_Synthes (Usa) · HRS · May 7, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K040588

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** May 7, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Synthes (USA) 2.4mm LCP Wrist Plate is intended for Colles' and distal radius fractures with dorsal angulation of the distal fragment.

## Device Story

The 2.4 mm LCP Wrist Plate is a pre-contoured stainless steel orthopedic implant designed for internal fixation of distal radius fractures. The plate is positioned dorsolaterally on the distal radius and affixed to the second metacarpal to span the wrist. It features combination and DCU holes to accommodate 2.4 mm cortex and 2.4 mm locking screws. Used by orthopedic surgeons in an operating room setting, the device provides mechanical stabilization of the fracture site to facilitate bone healing. The low-profile design (2.6 mm thickness, 7.5 mm width) aims to minimize soft tissue irritation while maintaining structural integrity.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Material: Stainless steel. Dimensions: 170 mm length, 2.6 mm thickness, 7.5 mm width. Design: Pre-contoured, low-profile plate with combination and DCU holes for 2.4 mm cortex and locking screws. Fixation: Metallic bone fixation appliance.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Biomet® Colles Fracture Plate

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY - 7 2004

040588
page 1 of 1

## Summary of Safety and Effectiveness Information - 510(k) Summary 3.0

| SPONSOR:            | Synthes (USA)<br>1690 Russell Road<br>Paoli, PA 19301<br>(610) 647-9700<br>Contact: Lisa M. Boyle                                                                                                                                                                                                                                                                                                                                                                      |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE NAME:        | 2.4 mm LCP Wrist Plate                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CLASSIFICATION:     | Class II, § 888.3030 – Single / multiple component metallic bone fixation<br>appliance and accessories                                                                                                                                                                                                                                                                                                                                                                 |
| PREDICATE DEVICE:   | • Biomet® Colles Fracture Plate                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| DEVICE DESCRIPTION: | The 2.4 mm LCP Wrist Plate is a pre-contoured plate which is positioned<br>dorsolaterally on the distal radius and affixed to the second metacarpal to<br>span the wrist. A low profile construct is maintained by incorporating a 2.6<br>mm thickness, and a width of 7.5 mm. The plate has combination and DCU<br>holes which accept 2.4 mm cortex and 2.4 mm locking screws. The plate is<br>made out of Stainless Steel with tapered ends and is 170 mm in length. |
| INTENDED USE:       | The Synthes (USA) 2.4mm LCP Wrist Plate is intended for Colles' and<br>distal radius fractures with dorsal angulation of the distal fragment.                                                                                                                                                                                                                                                                                                                          |

.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem with three wavy lines representing the feathers.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY = 7 2004

Lisa M. Boyle Regulatory Associate Synthes USA 1690 Russel Road P.O. Box 1766 Paoli. Pennsylvania 19301

Re: K040588

Trade/Device Name: Synthes (USA) 2.4 mm LCP Wrist Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: March 04, 2004 Received: March 05, 2004

Dear Ms. Boyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Lisa M. Boyle

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark N. Millkann

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use Statement 2.0

| Page | 1 of 1 |
|------|--------|
|------|--------|

| 510(k) Number (if known):      | K040588                              |
|--------------------------------|--------------------------------------|
| Device Name:                   | Synthes (USA) 2.4 mm LCP Wrist Plate |
| Indications/Contraindications: |                                      |

The Synthes (USA) 2.4mm LCP Wrist Plate is intended for colles' and distal radius fractures with dorsal angulation of the distal fragment.

*Division Sign-Off*
  

 Division of General, Restorative,
   

 and Neurological Devices

| 510(k) Number | K040588 |
|---------------|---------|
|---------------|---------|

| Prescription Use <b>X</b> (Per 21 CFR 801.109) | OR | Over-The-Counter Use __________ |
|------------------------------------------------|----|---------------------------------|
|------------------------------------------------|----|---------------------------------|

: ・

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

:

Confidential

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K040588](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K040588)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
