← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K022325

# NORMED TITANIUM OSTEOTOMY PLATING SYSTEM (K022325)

_Osteomedics, Inc. · HRS · Sep 9, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K022325

## Device Facts

- **Applicant:** Osteomedics, Inc.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Sep 9, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Normed Titanium Osteotomy Plating System intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes skeleton. The system can be used in both adult and pediatric patients.

## Device Story

System consists of titanium plates (wedge and step configurations) and screws for bone fixation. Plates attached to bone via screw fixation; screw holes threaded to accept locking screws with threaded or non-threaded heads. Used by surgeons for stabilization/fixation of small bone fractures, revisions, joint fusions, and reconstructions in hands, feet, wrists, ankles, fingers, and toes. Provides mechanical support to bone segments to facilitate healing.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Titanium plates (wedge and step configurations) and screws. Screw diameters: 2.3 mm and 3.5 mm; lengths: 10-28 mm. Threaded plate holes for locking screws. Mechanical fixation device.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 9 2002 SEP

Kodd 325
page 1 of 1

## 510 (k) Summary

Device Name: Normed Titanium Osteotomy Plating System

Device Identification: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Class II

Product Code: 87 HRS (21 CFR - 888.3030)

The Normed Titanium Osteotomy Plating System consists of titanium plates in two configurations, wedge and step, which are attached to the bone using screw fixation. The wedge plates are available in six lengths, namely, 3,4,5,6,7and 8 mm with 2.3 mm diameter screws extends from 10 to 28 mm long. The step plates are available in four lengths, namely, 3,4,5 and 6 mm with 3.5 mm diameter screws extends from 10 to 28 mm long. The plates screw holes are threaded and can accept locking screws with threaded and non-threaded screw head.

The Normed Titanium Osteotomy Plating System intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes skeleton. The system can be used in both adult and pediatric patients.

Official Contact Person: Albert Enayati President Osteomedics Inc. 809 Carter Lane Paramus, NJ 07652 Tel: (201) 444-7306 Fax: (201) 444-7395 E-mail: osteomedics@aol.com

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

9 2002 SEP

Albert Enayati President Osteomedics, Inc. 809 Carter Lane Paramus, New Jersey 07652

Re: K022325

Trade/Device Name: Normed Titanium Osteotomy Plating System Regulation Number: 888.3030 Regulation Name: Single/Multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: July 7, 2002 Received: July 17, 2002

Dear Mr. Enayati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -- Mr. Albert Enayati

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely, yours.

M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for use

510 (k) Number ( if known): k02035

Device Name: Normed Titanium Osteotomy Plating System

Indications for use:

The Normed Titanium Osteotomy Plating System intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes skeleton. The system can be used in both adult and pediatric patients.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Prescription use X

(Per 21 CFR 801.109)

OR OVER – THE – COUNTER USE

Styp Purdy

(Division Sign-Off) Division of General. Restorative and Neurological Devices

510(k) Number k022325

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K022325](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K022325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
