The Scout Vessel Guard System is indicated as a cover for vessels during anterior vertebral surgery.
Device Story
Synthes Scout Tack Fixation is a single-use titanium alloy (TAN) tack used to secure the Scout Vessel Guard to lumbar vertebral bone. The tack features a low-profile head and helical geometry for bone purchase. It is inserted via light impaction through the vessel guard into the bone. Used by surgeons during anterior vertebral surgery as an alternative or complement to suturing. The device provides mechanical fixation of the vessel guard, ensuring it remains in place over vessels during the procedure. Benefits include simplified, secure attachment of the protective guard.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including pullout testing, dynamic mechanical testing, and dynamic biomechanical testing, supplemented by usability testing in cadaver labs.
Technological Characteristics
Material: Titanium alloy (TAN) per ASTM F1295-05. Design: Low-profile head, helical geometry for bone purchase. Principle: Mechanical fixation via light impaction. Single-use implant.
Indications for Use
Indicated for use as a vessel cover during anterior vertebral surgery in patients undergoing such procedures.
Regulatory Classification
Identification
An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.
K061727 — GORE PRECLUDE VESSEL GUARD · W. L. Gore & Associates, Inc. · Aug 7, 2006
K082782 — REPLICATION MEDICAL VESSEL GUARD · Replication Medical, Inc. · Apr 2, 2009
K173117 — NuVasive® VersaTie System · Nuvasive, Inc. · Dec 7, 2017
K030667 — ENCORE ANTERIOR SPINAL STAPLE SYSTEM · Encore Medical L.P. · May 28, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 1 4 2012
K111048
## 510(k) Summary
.
| 510(k) Summary | 510(k) Summary |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name of Firm: | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380 |
| 510(k) Contact: | Heather Guerin<br>Senior Regulatory Affairs Specialist<br>Telephone: 610-719-5432<br>Facsimile: 610-719-5102<br>Email: guerin.heather@synthes.com |
| Date Prepared: | January 3, 2012 |
| Trade Name: | Synthes Scout Tack Fixation |
| Classification: | 21 CFR § 870.3470<br>Class II<br>Cardiovascular Devices Panel<br>Product Code: OMR (vessel guard or cover)<br><br>21 CFR § 888.3040<br>Class II<br>Orthopaedic Devices Panel<br>Product Code: NDM (pin, fixation, threaded, metallic) |
| Predicate<br>Devices: | Synthes Scout Vessel Guard, K103558<br>Synthes Arch Fixation System, K032534<br>Replication Medical EnGuard Vessel Guard, K082782<br>Covidien Autosuture Tacker System, K090470 |
| Device<br>Description: | The Synthes Scout Tack Fixation is a method of fixation of the Scout Vessel<br>Guard to bone. The single-use Scout Tack implant is made of a titanium alloy<br>(TAN, per ASTM F1295-05, "Standard Specification for Wrought Titanium-<br>6Aluminum-7Niobium Alloy for Surgical Implant Applications (UNS R56700),"<br>January 1, 2005). It has a low profile head. The tack is inserted by light<br>impaction through the Scout Vessel Guard into lumbar vertebral bone. The<br>Scout Tack implant can be used as an alternative to, or a complement to, suturing<br>the Scout Vessel Guard in place. |
| Intended Use/<br>Indications for<br>Use: | The Scout Vessel Guard System is indicated as a cover for vessels during<br>anterior vertebral surgery. |
| Comparison of<br>the device to<br>predicate<br>device(s): | The Tack Fixation for the Synthes Scout Vessel Guard is substantially<br>equivalent to suture fixation of the Scout Vessel Guard (as cleared in<br>K103358) as demonstrated by pullout testing and dynamic mechanical<br>and biomechanical testing and by usability testing in cadaver labs. The Tack Fixation for the Synthes Scout Vessel Guard is substantially<br>equivalent to Synthes Arch Fixation System (K032534) in terms of<br>material of manufacture (TAN), biocompatibility, intended use (fixation<br>in the spine) and as demonstrated by pullout testing. The Tack Fixation for the Synthes Scout Vessel Guard is substantially<br>equivalent to Replication Medical EnGuard Vessel Guard (K082782), as<br>this device is marketed for fixation by means including tacks or staples.<br>Hence the intended use of the device is the same as Scout Tack fixation<br>of the Scout Vessel Guard. The Tack Fixation for the Synthes Scout Vessel Guard is substantially<br>equivalent to the Covidien Autosuture Tacker System (K090470) in |
| 510(k) Summary | |
| | material of manufacture, design (e.g. low profile, and helical geometry<br>for bone purchase), pullout testing, and in intended use (fixation of<br>prosthetic material during surgical procedures). |
| Performance<br>Data<br>(Non-Clinical<br>and/or Clinical): | Synthes conducted the following non-clinical tests to support a determination of<br>substantial equivalence: pullout testing, dynamic mechanical testing, dynamic<br>biomechanical testing, and usability (cadaver lab) testing. Clinical testing was<br>not required. |
| Conclusions: | The results of the non-clinical tests listed above support a determination of<br>substantial equivalence because the Synthes Tack fixation performed<br>equivalently to or superior to the predicate devices in these tests. |
.
{1}------------------------------------------------
-
:
. . . . .
·
:
. . . . .
.
:
: .
and the comments of the comments of
K111048 Response to Request for Additional Information CONFIDENTIAL
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the caduceus.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
JUN 1 4 2012
Synthes Spine c/o Heather Guerin Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, PA 19380
Re: K111048
Trade/Device Name: Synthes Scout Vessel Guard System Regulation Number: 870.3470 Regulation Name: Intracardiac patch or pledget Regulatory Class: II Product Code: OMR, NDM Dated: January 04, 2012 Received: January 05, 2012
Dear Ms. Guerin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of this device for reducing the incidence, severity, and extent of post-operative adhesion formation have not been established.
{3}------------------------------------------------
Page 2 - Ms. Heather Guerin
Furthermore, the indication for use as a cover for vessels during anterior vertebral surgery must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{4}------------------------------------------------
Page 3 – Ms. Heather Guerin
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Christy Foreman
Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health Food and Drug Administration
Enclosure
{5}------------------------------------------------
## Indications for Use Statement 5
K 111048 510(k) Number: (if known)
Synthes Scout Tack Fixation Device Name:
Indications for Use:
The Scout Vessel Guard System is indicated as a cover for vessels during anterior vertebral surgery.
X Prescription Use (21 CFR 801 Subpart D) AND / OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
.y. Killchem
(Division Sign-Off) (Division Sign-Oh)
Division of Cardiovascular Devices
KIII048 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.