STRYKER TEMPEST ORTHOPEDIC PUMP

K963304 · Stryker Endoscopy · HRX · Jan 22, 1997 · Orthopedic

Device Facts

Record IDK963304
Device NameSTRYKER TEMPEST ORTHOPEDIC PUMP
ApplicantStryker Endoscopy
Product CodeHRX · Orthopedic
Decision DateJan 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2

Intended Use

The Stryker Tempest Orthopedic Pump is intended to be used by surgeons in Orthopedic joints, including the knee, shoulder, ankle, elbow, wrist, and temporomandibular joints. It will be used to distend joints. The distension and saline circulation allow better viewing of the joint and tissues, as well as, irrigation of loose debris. The Tempest Orthopedic Pump replaces currently used manual infusion systems, utilizing either a bulb syringe or gravity flow. The Stryker Orthopedic Pump is to be used with Stryker Arthroscopes, and Pressure Sensing Cannula, found substantially equivalent under 510(k) number K771200, Arthroscopes, and K944593, Pressure Sensing Cannula, respectively.

Device Story

Stryker Tempest Orthopedic Pump system; consists of pump console, disposable tube set, and pressure sensing cannula. Used in orthopedic surgery to distend joints and irrigate debris via saline circulation. Replaces manual bulb syringe or gravity flow systems. Operated by surgeons in sterile field using reusable hand control. Pump monitors joint pressure; features automatic motor stop, outflow solenoid valve activation, and audible/visual alarms if pressure exceeds 150 mmHg or if pressure change is not detected within four seconds. Provides controlled joint distension for improved visualization during arthroscopy.

Clinical Evidence

Bench testing only. Validation includes electrical safety (CSA C22.2 No. 125, UL 544), sterilization validation (AAMI ST 27, AAMI ST 32), and biocompatibility testing (USP Intracutaneous, Systemic Toxicity, Maximization, MEM Elution, Hemolysis).

Technological Characteristics

Electrically powered pump console; disposable tube set; pressure sensing cannula. Safety features: over-pressure control (150 mmHg limit), outflow solenoid valve, audible/visual alarms. Sterilization: ETO (AAMI ST 27) and Gamma irradiation (AAMI ST 32). Electrical standards: CSA C22.2 No. 125, UL 544.

Indications for Use

Indicated for surgeons performing orthopedic joint procedures (knee, shoulder, ankle, elbow, wrist, temporomandibular) requiring joint distension and saline irrigation for visualization and debris removal.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} JAN 22 1997 K96 3304 2590 Walsh Avenue, Santa Clara, CA 95051 (408)567-9100 FAX (408)567-2507 # SUMMARY OF SAFETY AND EFFECTIVENESS — 510(K) # K963304 | Device Name : | | | --- | --- | | Current Classification Name | : Arthroscope and accessories CFR 888.1100 Class II | | Common and Usual Name | : Arthroscopy Pump | | Propriety Name | : Stryker Tempest Orthopedic Pump | | Device Sponsor | : Stryker Endoscopy 2590 Walsh Avenue, Santa Clara, CA 95051 FDA Registration # 2936485 | | Regulatory Classification | : Class II | This summary of 510(k) safety and effectiveness being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92. The Stryker Tempest Orthopedic Pump System, consisting of Pump Console which powers disposable Tube Set, and Pressure Sensing Cannula, 510(k) # K944593, during orthopedic surgery is equivalent in intended use, safety and effectiveness to existing Stryker Infusion Pump, 510(k) # K910858, used in similar applications, by Stryker Corporation. The disposable tube set utilized in Stryker Tempest Orthopedic Pump, will be provided sterile for single applications. The material of constructions, maximum pressure and flow rate are equivalent to currently marketed products. As proven in Orthopedic applications, these design offer the required precision for controlling the joints distension. The Tempest Orthopedic Pump technology utilized in this device is equivalent to existing marketed products. Power modality, intended use, methods of sterilization and safety risks are substantially equivalent. The Stryker Tempest Orthopedic Pump is electrically powered and is designed to meet CSA Electromedical Standard Number C22.2 Number 125, and UL 544 Standards. The disposable tube set ETO sterilization processes are validated per AAMI standard ST 27, to a S.A.L. of $10^{-6}$. Gamma irradiation validation is per AAMI standard ST 32, method 1, section 5.2.3.1., for sterile packaged, single use disposable tube set. The minimum dose for an S.A.L. of $10^{-6}$ is 1.66 Mrads. The biocompatibility testing will include USP Intracutaneous, USP Systemic Toxicity Test, Maximization, USP MEM Elution, and Hemolysis. These validation are equivalent to the existing marketed products. The Tempest Orthopedic Pump is designed with appropriate safety features for over pressure controls and alarm. First, under normal Run/Stop mode, the pump prevents the pressure sensed at the joint or body cavity from exceeding 150 mmHg. The pump motor will stop, a red light emitting diode illuminates, and outflow solenoid valve opens to relieve excess pressure. Under Lavage mode, any pressure above 150 mmHg for 3 seconds will trigger an uninterrupted audible alarm. Second, if the pump will not function if the pressure does not detect a change in actual pressure within four seconds, the pump motor will stop, followed by an audible alarm. The ONLY way to re-activate the pump is to depress the RUN/STOP switch. Third, the Lavage feature has an over pressure alarm. When using Lavage feature, and joint pressure go above 150 mmHg for 3 seconds, the uninterrupted audible alarm will sound. A reusable hand control will be offered to allow surgeons to perform surgery in the sterile field. Overall, the Stryker Tempest Orthopedic Pump System does not raise any new safety and efficacy concerns when compared to Infusion pump. Therefore, the Stryker Pump System is substantially equivalent to Stryker Infusion Pump System. {1} **Indications For Use**: The Stryker Tempest Orthopedic Pump is intended to be used by surgeons in Orthopedic joints, including the knee, shoulder, ankle, elbow, wrist, and temporomandibular joints. It will be used to distend joints. The distension and saline circulation allow better viewing of the joint and tissues, as well as, irrigation of loose debris. The Tempest Orthopedic Pump replaces currently used manual infusion systems, utilizing either a bulb syringe or gravity flow. The Stryker Orthopedic Pump is to be used with Stryker Arthroscopes, and Pressure Sensing Cannula, found substantially equivalent under 510(k) number K771200, Arthroscopes, and K944593, Pressure Sensing Cannula, respectively. ![img-0.jpeg](img-0.jpeg) Roy S. Chin Associate Project Engineer
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...