EQUESTRA FLUID DELIVERY SYSTEM

K040483 · Medtronic Sofamor Danek, Inc. · HRX · Jul 23, 2004 · Orthopedic

Device Facts

Record IDK040483
Device NameEQUESTRA FLUID DELIVERY SYSTEM
ApplicantMedtronic Sofamor Danek, Inc.
Product CodeHRX · Orthopedic
Decision DateJul 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EQUESTRA™ Delivery System is intended to provide surgeons with a percutaneous means of delivering legally cleared bone cement to the surgical site in orthopedic procedures.

Device Story

EQUESTRA™ Fluid Delivery System is a set of manual, single-use disposable surgical instruments; includes bone tamps and stylets. Designed for percutaneous delivery of commercially cleared bone cement to orthopedic surgical sites. System reservoirs are supplied empty; device does not contain bone cement. Operated by surgeons in clinical/surgical settings. Provides mechanical means for cement placement; facilitates orthopedic procedures. No electronic, software, or algorithmic components.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Manual, single-use disposable orthopedic surgical instruments. Includes bone tamps and stylets. No energy source, connectivity, or software. Materials are general surgical instrument grade.

Indications for Use

Indicated for surgeons requiring a percutaneous method to deliver legally cleared bone cement to surgical sites during orthopedic procedures.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Submission Summary (Full Text)

{0}------------------------------------------------ K04o4f3 [III]] 2 3 2004 # EQUESTRA™ Flaid Lelivery System 510(k) Summary June 2004 - Medtronic Sofamor Danek I. Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 Richard W. Treharne, PhD Contact: Senior Vice President Regulatory Affairs - Proprietary Trade Name: EQUESTRATM Fluid Delivery System II. - Classification Name: Orthopedic Manual Surgical Instrument III. - Regulation Number: 888.4540, 878.4800 and 878.4200 IV. ### V. Product Description The EQUESTRA™ Fluid Delivery System is designed to provide surgeons with a means to inject commercially cleared bone cement to the surgical site in orthopedic procedures. The EQUESTRA™ Fluid Delivery System does not contain any bone cement material, as this system only consists of general instruments. The EQUESTRA™ Fluid Delivery System consists of a variety of general surgical instruments. The system reservoirs are supplied empty as the system itself does not include any bone cement. Included in the system are bone tamps and stylets. All of the system components are manual, single-use disposable instruments. ### VII Indications The EQUESTRA™ Delivery System is intended to provide surgeons with a percutaneous means of delivering legally cleared bone cement to the surgical site in orthopedic procedures. ### VIII Substantial Equivalence Documentation was provided which demonstrated the EQUESTRA™ Delivery System to be substantially equivalent to various Class I exempt instruments including bone tamps, cannulas and stylets. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's logo, which is a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle. Public Health Service JUL 2 3 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132 Re: K040483 Trade/Device Name: EQUESTRA™ Fluid Delivery System Regulation Number: 21 CFR 888.1100, 21 CFR 888.4540 Regulation Name: Arthroscope; Orthopedic manual surgical instrument Regulatory Class: II Product Code: HRX, HXG Dated: June 22, 2004 Received: June 23, 2004 Dear Dr. Treharne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Richard W. Treharne, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Sincerely, yours, Mark N Milkenm Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): K040483 Device Name: EQUESTRA™ Fluid Delivery System ## Indications for Use: The EQUESTRA™ Delivery System is intended to provide surgeons with a percutaneous means of delivering legally cleared bone cement to the surgical site in orthopedic procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for Mark N Millkum Division of General, Restorative, and Neurological Devices **510(k) Number** K040983
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...