← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K213978

# TELIGEN System (K213978)

_Medos International SARL · HRX · Oct 18, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K213978

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Oct 18, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.

## Device Story

TELIGEN System is a sterile, single-use procedure kit for minimally invasive spinal surgery; includes camera, ports, port holder, TELIGEN Clear, soft tissue retractor, port cutter, and bone graft delivery instruments. Camera collects images of surgical site; images transferred to HD Camera Control System (CCS) for display on monitor. Used by surgeons in clinical settings to facilitate visualization and discectomy. System provides magnification and illumination to assist surgeon in surgical site access and tissue removal; benefits patient through minimally invasive approach.

## Clinical Evidence

Bench testing only. Evidence includes design verification and validation, human factors/usability, electrical safety, thermal safety, EMC testing, and software/systems testing.

## Technological Characteristics

Materials: glass, polycarbonate, ABS, PVC, PEEK, stainless steel, polypropylene, glass fiber reinforced polyarylamide. Components: camera, ports, port holder, retractor, cutter, graft delivery instruments. Energy: electrical (CCS). Connectivity: camera to CCS to monitor. Sterilization: sterile, single-use.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Voyant System ([K191579](/device/K191579.md))
- XTool™ MIS Discectomy Device ([K122861](/device/K122861.md))

## Reference Devices

- OVS1 Video System ([K123359](/device/K123359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 18, 2022

Medos International, SARL Daria Bochenek Senior Regulatory Affairs Specialist Chemin-Blanc 38 Le Locle, CH-NE 2400 Switzerland

Re: K213978

Trade/Device Name: TELIGEN Procedure Kit Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: September 16, 2022 Received: September 16, 2022

Dear Daria Bochenek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213978

Device Name TELIGEN System

Indications for Use (Describe)

The TELIGEN System is indicated to provide minimally invasive access, visualization, magnification and discectomy of the surgical area of the spine.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| A.                                    | Submitter Information                                                                                                                                                                                                                                                                                                                                                                           |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------------------------|-------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------|----------------------------------------|---------------------------------------|------------------|--------|----------------------|
|                                       | <table><tr><td>510(k) Sponsor:</td><td>Medos International, SARL</td></tr><tr><td>Contact Person:</td><td>Daria Bochenek, Senior Regulatory Affairs Specialist<br/>Eimattstrasse 3<br/>4436 Oberdorf<br/>Switzerland</td></tr><tr><td>Telephone:</td><td>+41 61 965 61 54</td></tr><tr><td>Fax:</td><td>+41 61 965 66 16</td></tr><tr><td>Email:</td><td>dbochene@its.jnj.com</td></tr></table> | 510(k) Sponsor:           | Medos International, SARL     | Contact Person:               | Daria Bochenek, Senior Regulatory Affairs Specialist<br>Eimattstrasse 3<br>4436 Oberdorf<br>Switzerland | Telephone:                            | +41 61 965 61 54                       | Fax:                                  | +41 61 965 66 16 | Email: | dbochene@its.jnj.com |
| 510(k) Sponsor:                       | Medos International, SARL                                                                                                                                                                                                                                                                                                                                                                       |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Contact Person:                       | Daria Bochenek, Senior Regulatory Affairs Specialist<br>Eimattstrasse 3<br>4436 Oberdorf<br>Switzerland                                                                                                                                                                                                                                                                                         |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Telephone:                            | +41 61 965 61 54                                                                                                                                                                                                                                                                                                                                                                                |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Fax:                                  | +41 61 965 66 16                                                                                                                                                                                                                                                                                                                                                                                |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Email:                                | dbochene@its.jnj.com                                                                                                                                                                                                                                                                                                                                                                            |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| B.                                    | Date Prepared                                                                                                                                                                                                                                                                                                                                                                                   |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
|                                       | 15 September 2022                                                                                                                                                                                                                                                                                                                                                                               |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| C.                                    | Device Name                                                                                                                                                                                                                                                                                                                                                                                     |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
|                                       | <table><tr><td>Trade/Proprietary Name:</td><td>TELIGEN System</td></tr><tr><td>Common/Usual Name:</td><td>HRX – Arthroscope;</td></tr><tr><td>Device Classification and Regulation:</td><td>Class II<br/>HRX – 21 CFR §888.1100</td></tr><tr><td>Classification Product and Panel Code</td><td>HRX – Orthopedic</td></tr></table>                                                               | Trade/Proprietary Name:   | TELIGEN System                | Common/Usual Name:            | HRX – Arthroscope;                                                                                      | Device Classification and Regulation: | Class II<br>HRX – 21 CFR §888.1100     | Classification Product and Panel Code | HRX – Orthopedic |        |                      |
| Trade/Proprietary Name:               | TELIGEN System                                                                                                                                                                                                                                                                                                                                                                                  |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Common/Usual Name:                    | HRX – Arthroscope;                                                                                                                                                                                                                                                                                                                                                                              |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Device Classification and Regulation: | Class II<br>HRX – 21 CFR §888.1100                                                                                                                                                                                                                                                                                                                                                              |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Classification Product and Panel Code | HRX – Orthopedic                                                                                                                                                                                                                                                                                                                                                                                |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| D.                                    | Predicate Device Names                                                                                                                                                                                                                                                                                                                                                                          |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
|                                       | <table><tr><td>Primary Predicate Device:</td><td>Voyant System (K191579) – HRX</td></tr><tr><td>Additional Predicate Devices:</td><td>XTool™ MIS Discectomy Device (K122861) – HRX</td></tr><tr><td>Reference device:</td><td>OVS1 Video System (K123359) – FCW, GCJ</td></tr></table>                                                                                                          | Primary Predicate Device: | Voyant System (K191579) – HRX | Additional Predicate Devices: | XTool™ MIS Discectomy Device (K122861) – HRX                                                            | Reference device:                     | OVS1 Video System (K123359) – FCW, GCJ |                                       |                  |        |                      |
| Primary Predicate Device:             | Voyant System (K191579) – HRX                                                                                                                                                                                                                                                                                                                                                                   |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Additional Predicate Devices:         | XTool™ MIS Discectomy Device (K122861) – HRX                                                                                                                                                                                                                                                                                                                                                    |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |
| Reference device:                     | OVS1 Video System (K123359) – FCW, GCJ                                                                                                                                                                                                                                                                                                                                                          |                           |                               |                               |                                                                                                         |                                       |                                        |                                       |                  |        |                      |

{4}------------------------------------------------

#### E. Device Description

The TELIGEN Procedure Kit is part of the TELIGEN System. It is a sterile, single use kit intended for use in surgical spinal procedures to allow for access, visualization, discectomy and graft delivery.

The TELIGEN Procedure Kit includes a camera, ports and port holder, TELIGEN Clear, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments.

The TELIGEN HD Camera Control System (CCS) is part of the TELIGEN System. It operates the TELIGEN Camera which is used for illumination and visualization of the surgical site. The image collected at the camera head, is transferred to the CCS and subsequently displayed on the monitor.

#### F. Indications for Use

The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.

### Summary of Similarities and Differences in Technological Characteristics, G. Performance, and Intended Use

The technological characteristics, including design, material and performance as well as intended use of the TELIGEN System are consistent with those of the predicate devices.

#### H. Materials

- . TELIGEN Camera: glass lens, plastic (polycarbonate, acrylonitrile butadiene styrene, polyvinyl chloride, PEEK), stainless steel.
- . TELIGEN Ports: plastic (polypropylene).
- TELIGEN Port Holder: plastic (polycarbonate, glass fiber reinforced . polyarylamide).
- . TELIGEN Clear: stainless steel.

#### I. Performance Data

The performance data for the subject devices consists of the following evaluations:

Design Verification and Validation Testing, Human Factor/Usability, Electrical Safety, Thermal Safety & EMC Testing, Software and Systems Testing.

{5}------------------------------------------------

#### J. Conclusion

The indications for use of the TELIGEN System are consistent with those of the predicate devices. The technological characteristics of the TELIGEN System in terms of design, materials and performance are consistent with those of the predicate devices. The TELIGEN System is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K213978](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K213978)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
