← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K160249

# ULTRA Telescopes (K160249)

_Olympus Winter & Ibe GmbH · HRX · Apr 21, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K160249

## Device Facts

- **Applicant:** Olympus Winter & Ibe GmbH
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Apr 21, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Intended Use

These devices are indicated for use in arthroscopic procedures to provide visualization during endoscopic diagnosis and treatment of joints such as the knee, shoulder, hip, ankle, foot, elbow.

## Device Story

ULTRA Telescopes are rigid endoscopes used for joint visualization during arthroscopic procedures. Device utilizes rod-lens image relay system and optical fiber bundle for illumination via external light source. Operated by physicians in clinical settings (OR/clinic). Device is reusable, non-sterile, and fully autoclavable. Provides direct optical visualization to surgeon to facilitate diagnosis and treatment of joint pathologies. Benefits include improved visualization via wider field of view compared to predicate.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included optical and mechanical validation, shipping tests, service life evaluation, illumination system testing, and design validation/usability studies. Biocompatibility, thermal safety, and electrical safety (per IEC 60601-2-18) were verified.

## Technological Characteristics

Rigid endoscope; rod-lens image relay; fiber-optic illumination; reusable; autoclavable. Standards applied: ISO 8600-1, 8600-3, 8600-4, 8600-5, 8600-6; AAMI/ANSI ES60601-1; IEC 60601-2-18; ISO 10993-1, 10993-5, 10993-12; ISO 14971; ISO 17665-1.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Olympus Arthroscopy System ([K951354](/device/K951354.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 21, 2016

Olympus Winter & Ibe Gmbh % Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation Of The Americas 3500 Corporate Parkway Center Vallev, Pennsylvania 18034

Re: K160249

Trade/Device Name: "ULTRA" Telescopes Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: January 29, 2016 Received: February 1, 2016

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K160249

Device Name

"ULTRA" Telescopes

Indications for Use (Describe)

These devices are indicated for use in arthroscopic procedures to provide visualization during endoscopic diagnosis and treatment of joints such as the knee, shoulder, hip, ankle, foot, elbow.

|  | <span> <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--|------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
|--|------------------------------------------------------------------------------------------------------|------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete.

e time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

January 29, 2016

## 1. General information

- Manufacturer/Holder Olympus Winter & Ibe GmbH Kuehnstr. 61 22045 Hamburg Germany
- 트 Establishment Registration No.: 9610773
- . Official Correspondent: Sheri L. Musqnung Regulatory Affairs & Quality Assurance Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-3147 484-896-7128 FAX: Email: sheri.musqnunq@olympus.com Establishment Registration No.: 2429304

## 2. Device identification

- 트 Proprietary name:
- 트 Common name:
- 트 Regulation Number:
- 트 Regulation Description:
- I Device Class:
- Product Code:
- Review Panel:

"ULTRA" Telescopes Arthroscopes 21 CFR 888.1100 Arthroscope ll HRX Orthopedic

{4}------------------------------------------------

#### 3. Predicate device

K951354. Olympus Arthroscopy System

#### 4. Description of device

The "ULTRA" Telescopes are rigid endoscopes. An image relay system of rod lenses transmits the endoscopic image. A bundle of optical fibers transmits light from an external light source to illuminate the endoscopic image.

The "ULTRA" Telescopes are delivered non-sterile. They are reusable and fully autoclavable. Before first and each subsequent use the device must be inspected and reprocessed according to defined reprocessing methods in the Instructions for Use.

The "ULTRA" Telescopes are available with different directions of view to allow use for various applications in accordance with the intended use as submitted with this 510(k).

#### 5. Indications of use

These devices are indicated for use in arthroscopic procedures to provide visualization during endoscopic diagnosis and treatment of joints such as the knee, shoulder, hip, ankle, foot, elbow.

## 6. Comparison of Technological characteristics

The subject and predicate devices are based on the same technological principles:

- 트 Rigid endoscopes consisting of insertion tube, locking cone, light-guide connector, eyepiece cup, rod-lens and light-guide system.
- 트 A bundle of optical fibers transmits light from an external light source to illuminate the endoscopic image.
- I The image relay system of rod lenses transmits the endoscopic image
- Identical portion width and optical system diameter
- I Telescopes are used with 4 mm trocar tubes
- I Identical Direction of Views (DOVs)
- The Field of View (FOV) of subject devices is wider (Ø Factor 1.2)
- I Maximum working length of subject device is minimally longer
- I The same materials in patient contact are used in predicate and subject device

## 7. Performance Data

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO-14971:2007.

{5}------------------------------------------------

#### 510(k) Summary for K160249

Performance tests were carried out to ensure that the system functions as intended and meets design specifications. The following performance tests and usability studies were conducted:

#### Biocompatibility testing

The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995.

The testing included the following tests:

- . Biological Safety - toxicology: AAMI ANSI ISO 10993-1:2009; Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process
- . Cytotoxicity: AAMI ANSI ISO 10993-5:2009; Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- . Chemical Analysis: AAMI ANSI ISO 10993-12:2012; Biological evaluation of medical devices - Part 12: Sample Preparation and Reference Materials

### Electrical safety and electromagnetic compatibility (EMC)

Electrical Safety was tested according to AAMI/ANSI ES 60601-1:2005 and C1:2009 and A2:2010; Medical Electrical Equipment - Part 1.1 General requirements for safety and essential performance. Electromagnetic compatibility (EMC) is not applicable. The devices are not electrically powered and are just used with light from a light source.

#### Thermal Safety

Tested according to IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: IEC 60601-2-18 Edition 3.0 2009-08, medical electrical equipment - part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment.

## Performance Testing Bench

Conducted tests cover optical and mechanical performance testing, shipping tests, tests related to the expected service life and testing of the illumination system. Additionally, design validation/usability tests were conducted.

{6}------------------------------------------------

The following standards have been applied to the "ULTRA" Telescopes:

## Rec. No. Standard

| 19-5 | AAMI / ANSI ES60601-1: 2005/(R)2012 and C1:2009/(R)2012 |
|------|---------------------------------------------------------|
|------|---------------------------------------------------------|

- 9-61 IEC 60601-2-18:2009
- 5-89 IEC 60601-1-6:2013
- 9-83 ISO 8600-1:2013
- 9-84 ISO 8600-3:1997 + AM1(2003)
- 9-94 ISO 8600-4:2014
- 9-39 ISO 8600-5:2005
- 9-40 ISO 8600-6:2005
- 2-156 AAMI ANSI ISO 10993-1: 2009
- 2-153 AAMI ANSI ISO 10993-5: 2009
- 2-191 ISO 10993-12: 2012
- 5-40 ISO 14971:2007
- ISO 17665-1:2006 14-333

### 8 Conclusions

The performance data support the safety of the device and demonstrate that the subject devices comply with the intended use as specified.

In summary, we believe the "ULTRA" Telescopes are substantially equivalent to the predicate devices with respect to the general design approach, function, and the intended use. The "ULTRA" Telescopes raise no new concerns of safety or effectiveness compared to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K160249](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K160249)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
