← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K132883

# HEMODIA ARTHROSCOPY PUMP TUBE SETS (HATS) (K132883)

_Hemodia Sas · HRX · Nov 1, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K132883

## Device Facts

- **Applicant:** Hemodia Sas
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Nov 1, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** 3rd-Party Reviewed

## Intended Use

The device intended to be used in conjunction with the FMS/Depuy/Mitek arthroscopic pump systems and delivers irrigation fluid both to and from the pump and to and from the operative site during arthroscopic procedures.

## Device Story

HEMODIA ARTHROSCOPY PUMP TUBE SETS (HATS) are single-use disposable tube sets; used with FMS/Depuy/Mitek arthroscopic pump systems; deliver irrigation fluid to/from pump and operative site during arthroscopic procedures. Operated by surgical staff in clinical settings. Device is not self-powered; relies on external pump energy. Output is fluid flow management; facilitates visualization and maintenance of operative field. Benefits include consistent irrigation during surgery.

## Clinical Evidence

Bench testing only. Testing included functional equivalence, tube set leak testing under pressure, and bond strength. Clinical evaluation report concluded clinical testing was not necessary to support substantial equivalence.

## Technological Characteristics

Single-use disposable tube sets. Non-powered; utilizes external pump energy source. Materials, design, and function are similar to predicate devices. Sterilization method is consistent with predicate devices.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- OrthoConcept (FMS 4+ & disposables) ([K925160](/device/K925160.md))
- FMS (FMS DUO® + & disposables) ([K954465](/device/K954465.md))
- FMS (REF. 4102CV INTERMEDIARY TUBING, REF. 4509CV STERILE ZONE KIT) ([K951843](/device/K951843.md))
- FMS (FMS SOLO® & disposables) ([K002040](/device/K002040.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

·

Image /page/0/Picture/2 description: The image shows a stylized letter H. The letter is drawn with thick, black lines and has a rounded, almost cartoonish appearance. The bottom of the letter has some text, but it is too blurry to read.

## 5. 510(K) SUMMARY

## HEMODIA ARTHROSCOPY PUMP TUBE SETS - 510(k) Summary

| OWNER:                               | Hemodia S.A.S.<br>85 du Chêne Vert, 31670 Labège FRANCE<br>Tel: +33 5 61 00 71 81 - Fax: +33 5 61 00 47 40<br>remi.teuliere@hemoda.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                             | Mr. Rene van de Zande, President & CEO<br>Emergo Group, Inc.<br>Phone: 512.327.9997 Fax: 512.327.9998<br>usagent@emergogroup.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Summary Prepared:               | October 15th 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Trade Name:                   | HEMODIA ARTHROSCOPY PUMP TUBE SETS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Common Name:                         | Arthroscopic pump tube sets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Name:                 | Arthroscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Code:                 | 21 CFR Part 888.1100<br>87 Orthopedic<br>Product Code: HRX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Equivalent Device(s):                | OrthoConcept (FMS 4+ & disposables) K925160 FMS K954465 (FMS DUO ® + & disposables) FMS K951843 (REF. 4102CV INTERMEDIARY<br>TUBING, REF. 4509CV STERILE ZONE KIT) FMS K002040 (FMS SOLO® & disposables)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description:                  | HEMODIA ARTHROSCOPY PUMP TUBE SETS<br>(HATS) are tube sets that deliver irrigation fluid both to<br>and from the pump and to and from the operative site<br>during arthroscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended and Indications<br>for Use: | The device intended to be used in conjunction with the<br>FMS/Depuy/Mitek arthroscopic pump systems and<br>delivers both irrigation fluid to and from the pump and to<br>and from the operative site during arthroscopic<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Technological<br>Characteristics     | The TUBE SETS are similar in material (including<br>packaging material), design, function, and application to<br>the single use predicate devices. The device is not self-<br>powered but uses the same energy source (pumps) as<br>the predicate devices. The technological characteristics<br>are the same as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Performance Data:                    | Performance testing was performed to demonstrate<br>equivalence to the tube sets not the hardware<br>referenced in the predicate devices. Hemodia's testing<br>included bench testing for functional equivalence, tube<br>set leak testing under pressure, bond strength. These<br>test support that the design, packaging, sterilization &                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Clinical Performance Data:           | A clinical evaluation report was undertaken, supporting<br>that clinical testing was not necessary to support<br>substantial equivalence to predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Substantial Equivalence summary:     | Based on the comparison of the intended use, the<br>technological characteristics and performance data,<br>Hemodia S.A.S has determined that the proposed<br>HEMODIA ARTHROSCOPY PUMP TUBE SETS<br>(HATS) are substantially equivalent to the currently<br>marketed single use predicate devices, OrthoConcept<br>(FMS 4+ & disposables) K925160, FMS K954465 (FMS<br>DUO® + & disposables) FMS K951843 (REF. 4102CV<br>INTERMEDIARY TUBING, REF. 4509CV STERILE<br>ZONE KIT), FMS K002040 (FMS SOLO® &<br>disposables). HATS are tube sets that deliver irrigation<br>fluid both to and from the pump and to and from the<br>operative site during arthroscopic procedures. The<br>proposed devices have the same intended use, are<br>similar in material (including packaging material), same<br>sterilization method, design, function, and application to<br>the single use predicate devices. The device is not self-<br>powered but use the same energy source (pump) as the<br>predicate device. Performance testing demonstrates<br>equivalence to the predicate single use tube sets.<br>Hemodia included bench testing for functional<br>equivalence, tube set leak testing under pressure, bond<br>strength. These test support that the design, packaging,<br>sterilization & labeling of the TUBE SETS are<br>substantially equivalent to currently marketed single use<br>predicate devices. Any differences between the HATS<br>tube sets and the predicates single use sets are<br>considered minor and do not raise questions concerning<br>safety and effectiveness |
|                                      | labeling of the TUBE SETS are substantially equivalent<br>to currently marketed single use predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{1}------------------------------------------------

annery

・

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, black font. The text is horizontally oriented and centered. The text appears to be a header or title, possibly from a document or website.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 1, 2013

Hemodia S.A.S % Mr. Dave Yungvirt Parmalink Technical Group, LLC 20 F Street NW. Suite 700 Washington, DC 20001

Re: K132883

Trade/Device Name: HEMODIA ARTHROSCOPY PUMP TUBE SETS Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: October 14, 2013 Received: October 18, 2013

Dear Mr. Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Dave Yungvirt

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

## Mark N. Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K132883

## 4. INDICATIONS FOR USE STATEMENT

510(k) Number if known: N/A

Device Name: HEMODIA ARTHROSCOPY PUMP TUBE SETS

Indications for use:

The device intended to be used in conjunction with the FMS/Depuy/Mitek arthroscopic pump systems and delivers irrigation fluid both to and from the pump and to and from the operative site during arthroscopic procedures.

Prescription Use _✔__ Use (Per 21 CFR 801 Subpart D) Subpart C)

AND/OR

Over-the-Counter

(Per 21 CFR 807

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) | Page 1 of 1 |
|--------------------------------------------------------|-------------|
| Long H. Chen-A                                         | for MXM     |

![Signature](signature.png)

(Division Sign-off)

| Division of Surgical Devices | Page 14 |
|------------------------------|---------|
| 510(k) Number:               | K132883 |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K132883](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K132883)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
