← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K130430

# IVAS 10MM AND 15MM BALLOON CATHETER (K130430)

_Stryker Corporation · HRX · May 23, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K130430

## Device Facts

- **Applicant:** Stryker Corporation
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** May 23, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Stryker iVAS Inflatable Vertebral Augmentation System) is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

## Device Story

Stryker iVAS is an inflatable bone tamp used in percutaneous vertebral augmentation. Device consists of a catheter with an inflatable balloon at the distal end. Physician inserts catheter into vertebral body; inflates balloon to create a void or reduce fractures in cancellous bone. Once void is created, balloon is deflated and removed; space is filled with cleared PMMA bone cement or Cortoss material. Used in clinical settings by physicians. Modifications in this version include changes to balloon wall thickness, spiral cut on stainless steel catheter tube, and color of proximal reflow and luer hub.

## Clinical Evidence

No clinical testing was deemed necessary for this 510(k). Substantial equivalence supported by non-clinical bench testing confirming device meets specifications and performance characteristics.

## Technological Characteristics

Inflatable bone tamp; stainless steel catheter tube with spiral cut; inflatable balloon at distal end. Dimensions: 11 Gauge. Materials: Stainless steel, balloon material. Sterilization: Not specified. Connectivity: None. Software: None.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Stryker® iVAS 20mm (10 Gauge) Balloon Catheter ([K113477](/device/K113477.md))
- Stryker® iVAS 10mm (10 Gauge) Balloon Catheter ([K103807](/device/K103807.md))
- Stryker® iVAS 15mm (10 Gauge) Balloon Catheter ([K093419](/device/K093419.md))

## Reference Devices

- Kyphx Xpander Inflatable Bone Tamp

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K130430

MAY 2 3 2013

4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com

# stryker

Instruments

# 510(k) Summary

# 1. Contact Details

Stryker Instruments 4100 E. Milham Avenue Kalamazoo, MI 49001 (p) 269-323-7700 (f) 269-389-5412

Christina McKee Christina.McKee@Stryker.com

February 13, 2013

# 2. Device Name

Trade Name: Stryker Inflatable Vertebral Augmentation System (iVAS)

Common Name: Inflatable Bone Tamp

Classification Name: Arthroscope Cement, Bone, Vertebroplasty

# 3. Legally Marketed Predicate Device(s)

| 510(k) Number | Product Code | Trade Name                                     | Manufacturer        |
|---------------|--------------|------------------------------------------------|---------------------|
| K113477       | HRX          | Stryker® iVAS 20mm (10 Gauge) Balloon Catheter | Stryker Instruments |
|               | NDN          |                                                |                     |
| K103807       | HRX          | Stryker® iVAS 10mm (10 Gauge) Balloon Catheter | Stryker Instruments |
|               | NDN          |                                                |                     |
| K093419       | HRX          | Stryker® iVAS 15mm (10 Gauge) Balloon Catheter | Stryker Instruments |
|               | NDN          |                                                |                     |

{1}------------------------------------------------

# 4. Device Description

The Stryker® iVAS balloon catheter is a bone tamp with an inflatable component (balloon) at the distal end. The balloon is inflated to create a void within the vertebral hody.

# 5. Intended Use/Indications for use

The Stryker iVAS Inflatable Vertebral Augmentation System) is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements and Cortoss® Bone Augmentation Material indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

# 6. Substantial Equivalence Comparison

The Stryker® iVAS 11 Gauge Balloon Catheter has the same intended use/indications for use as the predicate Stryker® iVAS 10 Gauge Balloon Catheter.

Although the device incorporates modifications to the balloon wall thickness, the spiral cut on the stainless steel catheter tube, and the color of the proximal reflow and luer hub, the modifications do not alter the intended use, indications for use, or the fundamental scientific technology of the predicate devices.

Risk assessment of the modifications in the form of design and use failure modes and effects analysis (design and use FMEAs) has been conducted in accordance with EN ISO 14971. Stryker has determined that the modifications to the predicate device raise no new questions of safety or effectiveness.

# 7. Non-clinical Testing

The Stryker® iVAS balloon catheter meets the specification and performance characteristics and are substantially equivalent to the predicate devices. The modified device is not technologically different than the predicate device.

### 8. Clinical Testing

No clinical testing was deemed necessary for this 510(k).

{2}------------------------------------------------

# 9. Conclusions

The Stryker® iVAS (10mm and 15mm) 11 Gauge balloon catheters are substantially equivalent in intended use, technological characteristics, safety, and effectiveness to the previously cleared Stryker® iVAS Balloon Catheters and the Kyphx Xpander Inflatable Bone Tamp. The products have the same fundamental scientific technology, basic design, functional characteristics and the same clinical applications. The Stryker® iVAS balloon catheter does not raise any new safety and efficacy concerns when compared to similar devices already legally marketed. Therefore, the Stryker® (10mm and 15mm) 11 Gauge iVAS balloon catheters are equivalent to the existing predicate `devices.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and features an abstract image of an eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Letter dated: May 23, 2013

Stryker Corporation % Ms. Christina McKee Regulatory Affairs Associate Analyst 4100 East Milham Avenue Kalamazoo, Michigan 49001

Re: K130430

Trade/Device Name: Stryker® Inflatable Vertebral Augmentation System (iVAS) Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: April 15, 2013 Received: April 19, 2013

Dear Ms. McKee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

### Page 2 - Ms. Christina McKee -

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark Nimelkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(K) Number (if known):

Device Name: Stryker Inflatable Vertebral Augmentation System (iVAS)

### Indications for Use

The Stryker® iVAS Inflatable Vertebral Augmentation System (system) is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

Prescription Use ___________X

and/or

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

# Laurence D! (Soyne -A

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130430

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K130430](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K130430)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
