← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K093717

# C-MOR VISUALIZATION DEVICE (K093717)

_Axis Surgical Technologies, Inc. · HRX · Mar 2, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K093717

## Device Facts

- **Applicant:** Axis Surgical Technologies, Inc.
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Mar 2, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Axis Surgical Technologies C-MOR Visualization Device is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.

## Device Story

C-MOR Visualization System is a portable, battery-operated arthroscope for diagnostic and operative procedures. Device features integrated LCD monitor, handle, user controls, and LED light source. Rigid probe shaft contains CMOS camera and fiber optic lighting at distal tip. Operated by clinician in one hand; LCD monitor displays real-time images from probe. Viewing angle of LCD is adjustable. Device provides illumination and visualization of interior body cavities. Benefits include portability and integrated visualization for arthroscopic/endoscopic access.

## Clinical Evidence

No clinical data. Performance evidence is based on bench testing, including compliance with international standards for EMC, Medical Device Electrical Safety, and biocompatibility.

## Technological Characteristics

Battery-operated, portable arthroscope. Rigid probe (150 mm) with CMOS camera and fiber optic lighting. Integrated LCD monitor. Total length 342 mm; weight <300g. LED light source. Complies with international standards for EMC, electrical safety, and biocompatibility.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Surgview integrated visualization system ([K082293](/device/K082293.md))
- S&N Video Arthroscope ([K043395](/device/K043395.md))
- InnerVue Diagnostic Scope System ([K072879](/device/K072879.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Premarket Notification December 1, 2009

K0937+7

# SECTION 5 - 510(K) SUMMARY

# Axis Surgical Technologies, Inc.

## Submitter / Holder:

MAR - 2 2010

Axis Surgical Technologies, Inc. 325 E. Middlefield Road Mountain View, CA 94043

# Contact Person:

James Farnworth RA/QA Tel: 650-968-3800 x5610 Fax: 650-968-3803 Email: jfarnworth@axissurgical.com

# Date prepared:

December 1, 2009

# Device Information:

| Trade name:          | C-MOR Visualization System |
|----------------------|----------------------------|
| Common name:         | Arthroscope                |
| Classification name: | Arthroscope                |
| Panel:               | Orthopedic                 |
| Product Code:        | HRX                        |
| Device Class:        | Class 2                    |

# Predicate Devices:

| 510(k) Number | Trade Name                               | Manufacturer                   |
|---------------|------------------------------------------|--------------------------------|
| K082293       | Surgview integrated visualization system | Biovision Technologies,<br>LLC |
| K043395       | S&N Video Arthroscope                    | Smith & Nephew                 |
| K072879       | InnerVue Diagnostic Scope System         | Biomet                         |

#### Intended Use / Indications for Use:

The Axis Surgical Technologies C-MOR Visualization Device is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.

{1}------------------------------------------------

# Device Description:

The C-MOR Visualization Device is a battery operated, portable, visualization device including an integrated LCD monitor and probe. The C-MOR Visualization Device is easily handled in one hand and includes a user interface, electronics, user controls, and LED light source. The LCD monitor displays real-time images from the probe and is flexibly mounted on the handle - allowing adjustment of the LCD viewing angle. The probe extends from the handle as a rigid shaft seating a CMOS camera and fiber optic lighting located at the distal tip.

The overall length of the C-MOR Device 342 mm (approx. 13.5 inches); including the probe portion and handle. The probe length, extending from the handle, is 150 mm (approx. 6 inches). The entire unit weighs less than 300 grams (approx. 0.67 lbs.).

#### Substantial Equivalence:

The C-MOR Visualization Device has the same indications and intended use as the predicate devices. The subject and predicate devices utilize the same basic technology and the same operating principles. Performance data supporting the claim of substantial equivalence is included in the body of the premarket notification.

# Performance:

The C-MOR Visualization Device complies with the international standards for EMC, Medical Device Electrical Safety and biocompatibility.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird, with its wings forming a human profile.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Axis Surgical Technologies, Inc. % Mr. James Farnworth Director, RA/QA 325 E. Middlefield Road Mountain View, California 94043

MAR - 2 2010

Re: K093717

Trade/Device Name: C-MOR Visualization System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: January 28, 2010 Received: January 29, 2010

Dear Mr. Farnworth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not mislcading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. James Farnworth

CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Mark A. Milkean

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Axis Surgical Technologies, Inc. C-MOR Visualization Device

# SECTION 4 - INDICATIONS FOR USE STATEMENT

#### Indications for Use

K093717 510(k) Number (if known):

Device Name: C-MOR Visualization System

Indications For Use:

The Axis Surgical Technologies C-MOR Visualization Device is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, Page 1 of and Restorative Devices

510(k) Number K093717 Page 9

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K093717](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K093717)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
