← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K090132

# MAXMORESPINE SPINAL SYSTEM (K090132)

_Hoogland Spine Products, GmbH · HRX · Jun 9, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K090132

## Device Facts

- **Applicant:** Hoogland Spine Products, GmbH
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Jun 9, 2009
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The maxMorespine System is intended to visualize the inside of the patient through a cannulated incision for diagnostic and surgical procedures such as: arthroplasty, nucleotomy, discectomy and foraminotomy.

## Device Story

The maxMorespine System is a set of manual surgical instruments used to facilitate spinal endoscopy. The system includes cannulas, sheaths, dilators, drills, trephines, forceps, punches, mallets, and guidewires. These instruments are used by surgeons to access and treat spinal anatomy and pathology through a cannulated incision. The system is designed to complement the maxMorespine Endoscope (K083552). The instruments are primarily non-sterile and require user sterilization, with the exception of a disposable manual bone drill supplied sterile via gamma irradiation. The device provides physical access to the surgical site, allowing the clinician to perform procedures such as discectomy or foraminotomy under endoscopic visualization.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Manual surgical instruments including cannulas, sheaths, dilators, drills, trephines, forceps, punches, mallets, and guidewires. Non-sterile (except disposable bone drill). Sterilization by user required for reusable components. Disposable bone drill sterilized via Gamma Irradiation.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Joimax THESSYS Multiscope ([K051827](/device/K051827.md))
- Richard Wolf Yeung Endoscopic Spine System ([K973405](/device/K973405.md))
- Arthro Kinetics' Endoscopic Spine System ([K061246](/device/K061246.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Section 1.0 510(k) Summary

as required by section 807.92(c)

JUN - 9 2009

# maxMorespine System

# 1.1

# Submitter Information

Hoogland Spine Products GmbH Feringastr. 7a D-85774 Unterföhring - Germany Phone No .: +40 (0) 89 21 58 960 - 20 +49 (0) 89 21 58 960 - 98 Fax No.:

# USA Agent:

Spine Now, LLC 94 Gardiners Ave, #307 Levittown, NY 11756 Phone No .: (888) 994-3335 Fax No.: (516) 277-0065

# Contact Information

Joan L. Carter, President, Spine Now, LLC. Phone No.: (516) 721-7442 E-mail: jcarter@spinenow.com

### 1.2 Establishment / Owner Information

Hoogland Spine Products GmbH Feringastr. 7a D-85774 Unterföhring - Germany

FDA Establishment Registration Number 30065611611

- 1.3 Proprietary Name maxMorespine System
## 1.4 Common Name Spinal Endoscopy System

### 1.5 Classification Name Accessories, Arthroscopic .

### 1.6 Regulation Number 21 CFR 888.1100

{1}------------------------------------------------

### 1.7 Regulatory Class

Class II

- Classification Product Code 1.8 HRX
### Intended Use 1.9

The maxMorespine System is intended to visualize the inside of the patient through a cannulated incision for diagnostic and surgical procedures such as: arthroplasty, nucleotomy, discectomy and foraminotomy.

### 1.10 Device Description

The maxMorespine System consists of basic cannulas, sheaths, dilators, drills, trephines, forceps, punches, mallet and quidewires which facilitate or compliment the maxMorespine Endoscope (submitted separately under K083552). The System is designed to provide the access and treatment of spinal anatomy and pathology.

#### 1.11 Substantial Equivalence

The maxMorespine System is substantially equivalent in purpose, design, materials and function to the following marketed products;

- Joimax THESSYS Multiscope (K051827)

- RIchard Wolf Yeung Endoscopic Spine System (K973405)
- Arthro Kinetics' Endoscopic Spine System (K061246)

#### 1.12 Sterilization

With the exception of the Disposable 4mm Manual Bone Drill, maxMorespine System will be sold non-sterile, to be sterilized prior to each procedure by the user. Sterilization instructions are provided in the Instructions for Use (see Section 2.5)

The Disposable Manual Bone Drill will be supplied sterile via Gamma Irradiation (see Bioburden Test Protocol - Section 3.2 and Certificate of Gamma Irradiation - Section 3.3).

#### 1.13 Software

No Software is needed for the maxMorespine System.

### 1.14 Conclusion

The specifications and intended use of the maxMorespine System is the same as those of the claimed predicate devices. There are no significant differences in design or manufacturing materials between the maxMorespine System and predicate devices. In all aspects, the maxMorespine System is substantially equivalent to products actively marketed as demonstrated in Predicate Devices Comparison Chart (see Section 4.1)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, vectorized design.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Hoogland Spine Products GmbH % Mr. Boris Miklitz Arabellastr 4 Muenchen GERMANY D-81925

Re: K090132

Trade/Device Name: maxMorespine System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX Dated: May 29, 2009 Received: June 2, 2009

Dear Mr. Miklitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

JUN - 9 2009

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

{3}------------------------------------------------

Page 2-Mr. Boris Miklitz

practice requirements as set torth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdirh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sincerely yours,

Mark N. Melkerson

Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

maxMorespine System

# Indications for Use

510(k) Number; K090132

maxMorespine System Device Name:

# Indications for Use:

The maxMorespine System is intended to visualize the inside of the patient through a cannulated incision for diagnostic and surgical procedures such as; arthroplasty, nucleotomy, discectomy and foraminotomy.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nilreorden Forman
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090132

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K090132](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K090132)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
