← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K072706

# MICROCHOICE DRILLS AND SAWS, MICROPOWER DRILLS AND SAWS, MICROPOWER ORALMAX HIGH SPEED DRILL (K072706)

_Conmed Linvatec · HRX · Oct 24, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K072706

## Device Facts

- **Applicant:** Conmed Linvatec
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Oct 24, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The MicroPower® Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The fields of application include: orthopedic, arthroscopic, neurosurgical, otolaryngological, plastic/reconstructive, and oral/maxillofacial procedures. The MicroPower® OralMax Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The field of application includes only oral/maxillofacial. The MicroChoice® Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The fields of application include: orthopedic, arthroscopic, otolaryngological, oral/maxillofacial, hand, foot, neuro, and plastic/reconstructive surgical procedures.

## Device Story

Electric pencil-grip handheld surgical drill/saw systems; used with Advantage®, E9000®, and PowerPro® controllers. Device connects to controller via cord for power; actuated by lever or footswitch. Accepts various attachments (bur guards, drill bits, burs, saw blades). Used by surgeons in OR settings for bone and soft tissue cutting. Provides mechanical power to surgical cutting accessories; facilitates precise bone/tissue removal. Updated handpiece identification mechanism compared to predicates.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Electric powered surgical handpiece system; pencil-grip form factor; corded connection to external controller (Advantage®, E9000®, PowerPro®); lever or footswitch actuation; compatible with various mechanical drill/saw attachments.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Universal Drive System ([K971059](/device/K971059.md))
- MicroPower Hand Piece: Medium Speed Drill, Sagittal Saw, Reciprocating Saw, and Oscillating Saw ([K060198](/device/K060198.md))
- MicroPower Hand Piece: High Speed Drill ([K060260](/device/K060260.md))
- MicroPower Hand Piece: Oral Max High Speed Drill ([K060270](/device/K060270.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT 2 4 2007

#### 510(k) SUMMARY

### MicroPower®, MicroPower® OralMax, and MicroChoice® Handpiece Systems

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Linvatec is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number __ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

#### A. Submitter

ConMed Linvatec 11311 Concept Boulevard Largo, Florida 33773-4908 Registration Number: 1017294

#### B. Company Contact

Sue F. Dauterman Regulatory Affairs Specialist (727) 399-5321 Telephone (727) 399-5264 FAX

# C. Device Name

Common Name: Surgical Drill System

Proposed Class/Device: Class II

| Trade Name                                | Classification<br>Panel | Classification Name                                  | Product<br>Code |
|-------------------------------------------|-------------------------|------------------------------------------------------|-----------------|
| MicroChoice Drills<br>and Saws            | Orthopedic              | Arthroscope, 888.1100                                | HRX             |
| MicroPower Drills<br>and Saws             | Neurology               | Electric cranial drill motor,<br>882.4360            | HBC             |
| MicroPower<br>OralMax High<br>Speed Drill | Dental                  | Bone cutting instrument and<br>accessories, 872.4120 | DZI             |

{1}------------------------------------------------

510(k) Summarv

MicroPower®, MicroPower® OralMax, and MicroChoice® Handpiece Systems

### D. Predicate/Legally Marketed Devices

| 510(k) Name                                                                                           | 510(k) # | Owner                   |
|-------------------------------------------------------------------------------------------------------|----------|-------------------------|
| Universal Drive System                                                                                | K971059  | Linvatec<br>Corporation |
| MicroPower Hand Piece: Medium Speed<br>Drill, Sagittal Saw, Reciprocating Saw, and<br>Oscillating Saw | K060198  | Linvatec<br>Corporation |
| MicroPower Hand Piece:High Speed Drill                                                                | K060260  | Linvatec<br>Corporation |
| MicroPower Hand Piece:Oral Max High<br>Speed Drill                                                    | K060270  | Linvatec<br>Corporation |

# E. Device Description

The MicroPower®, MicroPower® OralMax, and MicroChoice® Handpiece Systems are electric pencil grip handhelds that are used in conjunction with the Advantage®, E9000® and PowerPro® controllers. A cord connects the handpiece to the controller that supplies power to the device. The handpieces have a lever or footswitch that is used to actuate the device. The drills use a wide variety of attachments, including bur guards, drill bits and burs. The saws use a wide variety of blades.

# F. Intended Use

| Name                               | Intended Use                                                                                                                                                                                                                                                                                                                                            |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MicroPower®<br>Handpiece<br>System | The MicroPower® Handpiece System functions as a<br>powered instrument system consisting of drills, saws<br>and associated handpieces to perform cutting of soft<br>tissue and bone.  The fields of application include:<br>orthopedic, arthroscopic, neurosurgical,<br>otolaryngological, plastic/reconstructive, and<br>oral/maxillofacial procedures. |

{2}------------------------------------------------

| Name                                          | Intended Use                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MicroPower®<br>OralMax<br>Handpiece<br>System | The MicroPower® OralMax Handpiece System<br>functions as a powered instrument system consisting<br>of drills, saws and associated handpieces to perform<br>cutting of soft tissue and bone. The field of application<br>includes only oral/maxillofacial.                                                                                                            |
| MicroChoice®<br>Handpiece<br>System           | The MicroChoice® Handpiece System functions as a<br>powered instrument system consisting of drills, saws<br>and associated handpieces to perform cutting of soft<br>tissue and bone. The fields of application include:<br>orthopedic, arthroscopic, otolaryngological,<br>oral/maxillofacial, hand, foot, neuro, and<br>plastic/reconstructive surgical procedures. |

#### G. Technological Characteristics

The MicroPower®, MicroPower® OralMax, and MicroChoice® Handpiece Systems are identical to the predicate devices cleared in the original submissions except for an update to the handpiece identification mechanism as detailed in Section 9 of this submission. These modifications do not affect the device's intended use or performance specifications in a manner that raises any new issues regarding safety and effectiveness.

#### H. Substantial Equivalence

The MicroPower®, MicroPower® OralMax, and MicroChoice® Handpiece · Systems are substantially equivalent in intended use, design and technological characteristics to the below listed systems.

| Proposed Name                           | Predicate 510(k) Name                                                                                    | 510(k) # |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------|----------|
| MicroChoice® Handpiece<br>System        | Universal Drive System                                                                                   | K971059  |
| MicroPower® Handpiece<br>System         | MicroPower Hand Piece: Medium<br>Speed Drill, Sagittal Saw,<br>Reciprocating Saw, and<br>Oscillating Saw | K060198  |
|                                         | MicroPower Hand Piece:High<br>Speed Drill                                                                | K060260  |
| MicroPower® OralMax<br>Handpiece System | MicroPower Hand Piece:Oral Max<br>High Speed Drill                                                       | K060270  |

Testing conducted prior to product release assures that the new devices do not raise any new issues of safety and efficacy.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, featuring four parallel lines that curve and taper towards the right side of the image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 4 2007

ConMed Livatec % Ms. Sue F. Dauterman Regulatory Affairs Specialist 11311 Concept Boulevard Largo, Florida 33773-4908

Re: K072706

Trade/Device Name: MicroPower® Handpiece System MicroPower® OralMax Handpiece System MicroChoice® Handpiece System

Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX Dated: September 21, 2007 Received: September 24, 2007

Dear Ms. Dauterman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Sue F. Dauterman

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### K072700 510(k) Number (if known):

Device Name: MicroPower® Handpiece System

Indications for Use:

The MicroPower® Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The fields of application include: orthopedic, arthroscopic, neurosurgical, otolaryngological, plastic/reconstructive, and oral/maxillofacial procedures.

Prescription Use_X__ AND/OR (Part 21 CFR 801 Subpart D)

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

of (Division Sign-Off)

(Division Degand Neurological Devices

**510(k) Number** 16072706

{6}------------------------------------------------

KO72706 510(k) Number (if known):

Device Name: MicroPower® OralMax Handpiece System

Indications for Use:

i

The MicroPower® OralMax Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The field of application includes only oral/maxillofacial.

Prescription Use_X__ AND/OR (Part 21 CFR 801 Subpart D)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{7}------------------------------------------------

510(k) Number (if known): K072706

Device Name: MicroChoice® Handpiece System

Indications for Use:

The MicroChoice® Handpiece System functions as a powered instrument system consisting of drills, saws and associated handpieces to perform cutting of soft tissue and bone. The fields of application include: orthopedic, arthroscopic, otolaryngological, oral/maxillofacial, hand, foot, neuro, and plastic/reconstructive surgical procedures.

Prescription Use_X____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of General, Restorative, and Neurological Devices

**510(k) Number**

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K072706](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K072706)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
