# ARTHREX CONTINUOUS WAVE III ARTHOSCOPY PUMP, MODEL AR-6475 (K024291)

_Arthrex, Inc. · HRX · Mar 20, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K024291

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Mar 20, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Arthrex Continuous Wave III Arthroscopy Pump is intended to provide consistent, non-pulsing control of intra-articular irrigation and distention pressuring during all phases of arthroscopic surgery.

## Device Story

The Arthrex Continuous Wave III Arthroscopy Pump is a roller, peristaltic pump used in arthroscopic surgery to maintain intra-articular irrigation and distention pressure. It features a universal input power supply, membrane switch user inputs, and a vacuum fluorescent/dot matrix display. The device includes an added flush function and an autoclavable remote. Operated by surgical staff in an OR setting, the pump provides consistent, non-pulsing fluid flow to the surgical site. The system uses upgraded software to manage pump operation. By maintaining stable joint distention, the device assists the surgeon in visualizing the joint space during procedures, potentially improving surgical outcomes.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on functional comparison and design characteristics.

## Technological Characteristics

Roller, peristaltic pump; universal input grade switching power supply; vacuum fluorescent and dot matrix display; membrane switch overlays; autoclavable remote; software-controlled operation.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Arthrex AR-6400
- Arthrex AR-6300

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Ko24291

### MAR 2 0 2003

### PREMARKET NOTIFICATION SUMMARY OF SUBTANTIAL EQUIVALENCE Arthrex Continuous Wave III Arthroscopy Pump

| NAME OF SPONSOR: | Arthrex, Inc.<br>2885 S. Horseshoe Drive<br>Naples, Florida 34104                                                                                              |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) CONTACT:  | Sally Foust, RAC<br>Requlatory Affairs Specialist<br>Arthrex, Inc.<br>Telephone: (239) 643-5553 ext. 1251<br>FAX: (239) 430-3494<br>E-mail: sfoust@arthrex.com |
| TRADE NAME:      | Arthrex Continuous Wave III Arthroscopy<br>Pump                                                                                                                |
| COMMON NAME:     | Pump                                                                                                                                                           |
| CLASSIFICATION:  | Arthroscope<br>21 CFR 888 1111                                                                                                                                 |

#### DEVICE PRODUCT CODE: HRX

### DEVICE DESCRIPTION AND INTENDED USE:

The Arthrex Continuous Wave III Arthroscopy Pump is a roller, peristaltic, arthroscopic pump designed with a universal input grade switching power supply allowing the pump to function automatically within voltages ranges found in Europe and in the Americas. The pump is designed with upgraded software, employs a combination vacuum fluorescent and dot matrix display for high visibility, uses membrane type switch overlays for user inputs, and has an added flush function.

The Arthrex Continuous Wave III Arthroscopy Pump is intended to provide consistent, non-pulsing control of intra-articular irrigation and distention pressuring during all phases of arthroscopic surgery.

### SUBSTANTIAL EQUIVALENCE

The Arthrex Continuous Wave III Arthroscopic Pump is a functional equivalent of the currently marketed AR-6400 and the discontinued AR-6300 Arthrex arthroscopic pumps. The subject pump retains current functions and is determined by Arthrex, Inc. to be substantially equivalent to its Arthrex pump predecessors and other currently marketed predicate devices.

The addition of a universal input grade switching power supply, of an improved visual display, of an autoclavable remote, and of an upgrade software package does not affect the safety and effectiveness of the subject device when compared to its Arthrex pump predecessors and other predicate devices.

Image /page/0/Picture/11 description: The image shows a sequence of numbers. The numbers are "000005". The numbers are black and the background is white. 

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human form, represented by three abstract shapes that resemble profiles facing right. The shapes are arranged in a way that suggests a sense of unity and forward movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 0 2003

Ms. Sally Foust, RAC Regulatory Affairs Specialist Arthrex, Inc. 2885 South Horseshoe Drive Naples, Florida 34104

Re: K024291

Trade/Device Name: Arthrex Continuous Wave III Arthroscopy Pump Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope and accessories Regulatory Class: II Product Code: HRX Dated: December 20, 2002 Received: December 24, 2002

Dear Ms. Foust:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Sally Foust, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Mark A. Milken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K024291 510(k) Number (if known) _

INDICATIONS FOR USE:

The Arthrex Continuous Wave III Arthroscopy Pump is intended to provide consistent, non-pulsing control of intra-articular irrigation and distention pressuring during all phases of arthroscopic surgery.

Concurrence of CDRH, Office of Device Evaluation

-----------------------------------------------------------------------------------------------------------------------

Prescription Use _____________________________________________________________________________________________________________________________________________________________

Over-The-Counter

Use

(Per 21 CFR 801.109)

Mark A Milliken

vision Sign-Off on of General, Restorative cological Devices

K024291

## 00004

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K024291](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K024291)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com