← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K012630

# REPROCESSED LINVATEC ARTHROSCOPIC BURS (K012630)

_Alliance Medical, Inc. · HRX · Nov 9, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K012630

## Device Facts

- **Applicant:** Alliance Medical, Inc.
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Nov 9, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Reprocessed Arthroscopic Shavers are intended to resect tissue and bone found in articular body cavities during orthopedic, maxillofacial, hand, foot and plastic surgery in patients requiring arthroscopic or orthopedic surgery. Reprocessed arthroscopic shavers are indicated for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.

## Device Story

Reprocessed arthroscopic burs consist of a bur or blade at the end of a long rod rotating within a hollow stainless steel housing. The housing features a distal window for tissue/bone resection while protecting adjacent structures. The device attaches to a motorized handpiece that drives the internal blade and provides suction to remove debris from the surgical site. Used by surgeons in OR settings for orthopedic, maxillofacial, hand, foot, and plastic surgeries. The device functions by mechanical abrasion and excision of tissue/bone.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, validation of reprocessing, and functional performance tests to demonstrate that the reprocessed burs perform as originally intended.

## Technological Characteristics

Stainless steel housing with internal rotating bur/blade. Mechanical actuation via motorized handpiece. Suction-assisted debris removal. No software or electronic components.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Linvatec® Merlin Polyblade Shavers ([K940515](/device/K940515.md))
- Linvatec® Universal Drive System ([K971059](/device/K971059.md))
- Linvatec® Universal Drive System ([K981269](/device/K981269.md))
- Linvatec® Integrated Drive/Pump System ([K981636](/device/K981636.md))
- Linvatec® E9000 System ([K990524](/device/K990524.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K012630

Image /page/0/Picture/1 description: The image shows the logo for Alliance Medical Corporation. The logo consists of three curved lines above the word "ALLIANCE" in capital letters. Below the word "ALLIANCE" is the phrase "MEDICAL CORPORATION" in smaller capital letters. The logo is black and white.

NOV 0 9 2001

## PART B: 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

10232 South 5 1st Street Phoenix, Arizona 85044

TEL 480.763.5300 FAX 480.763.5310 Toll Free 888.888.3433 ww.alliance-medical.com

Submitter:

Alliance Medical Corporation 10232 South 51st Street Phoeniz, Arizona 85044

Contact:

Don Selvev Vice President, Regulatory Affairs and Quality Assurance (480) 763-5300

Date of preparation:

August 10, 2001

Name of device:

Trade/Proprietary Name: Reprocessed Arthroscopic Burs Common or Usual Name: Arthroscopic Bur Classification Name: Arthroscope

## Reprocessed devices:

| Manufacturer         | Description            | Model |
|----------------------|------------------------|-------|
| Linvatec Corporation | Oval Bur               | C9101 |
| Linvatec Corporation | Oval Bur               | C9102 |
| Linvatec Corporation | Oval Bur Left Helix    | C9106 |
| Linvatec Corporation | Spherical Bur          | C9110 |
| Linvatec Corporation | Spherical Bur          | C9111 |
| Linvatec Corporation | Spherical Bur          | C9112 |
| Linvatec Corporation | Vortex Router          | C9131 |
| Linvatec Corporation | Vortex Router Unhooded | C9134 |
| Linvatec Corporation | Oval Bur               | H9101 |
| Linvatec Corporation | Oval Bur               | H9102 |
| Linvatec Corporation | Spherical Bur          | H9110 |
| Linvatec Corporation | Vortex Router Hooder   | H9131 |
| Linvatec Corporation | Vortex Router          | H9132 |

Predicate device(s):

K940515 K971059 K981269 K981636 K990524

Linvatec® Merlin Polyblade Shavers Linvatec® Universal Drive System Linvatec® Universal Drive System Linvatec® Integrated Drive/Pump System Linvatec® E9000 System

Device description:

Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and, shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses.

The arthroscopic shaver components reprocessed by Alliance Medical Corporation include a bur or blade at the end of a long rod that rotates within a long hollow stainless steel housing.

{1}------------------------------------------------

KOi 2630

Image /page/1/Picture/1 description: The image shows three curved lines that are stacked on top of each other. The lines are black and have a thick, bold appearance. The lines are arranged in a way that creates a sense of depth and movement. The overall impression is one of simplicity and elegance.

## ALLIANCE MEDICAL CORPORATION

The housing has a window cut out on one side of the distal end, allowing the bur to cut one structure, while the adjacent one is still protected by the housing on the opposite side of the bur or blade. This system attaches to a motorized handpiece that drives the internal bur or blade inside the outer housing and provides suction to pull the cut tissue and/or bone away from the surgical site.

Reprocessed Arthroscopic Shavers are intended to resect tissue and bone found in articular body cavities during orthopedic, maxillofacial, hand, foot and plastic surgery in patients requiring arthroscopic or orthopedic surgery.

Reprocessed arthroscopic shavers are indicated for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.

The design, materials, and intended use of the Reprocessed Arthroscopic Burs are identical to the predicate devices. The mechanism of action of the Reprocessed Arthroscopic Bur is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Arthroscopic Burs.

- Biocompatibility ●
- Validation of reprocessing ●
- Function Test(s) ●

Performance testing demonstrates that Reprocessed Arthroscopic Burs perform as originally intended.

In accordance with the Federal Food, Drug and Cosmetic Act 21 CFR Part 807 and based on the information provided in this premarket notification, Alliance Medical Corporation concludes that the modified device (the Reprocessed Arthroscopic Bur) is safe, effective and substantially equivalent to the predicate devices as described herein.

Intended use:

Indications statement:

Technological characteristics:

Performance data:

Conclusion:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three curved shapes that resemble a human figure.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 9 2001

Mr. Don Selvey Regulatory Affairs and Quality Assurance Alliance Medical Corporation, Inc. 10232 South 51st Street Phoenix, Arizona 85044

Re: K012630

Trade/Device Name: Reprocessed Linvatec Arthroscopic Burs Regulation Number: 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX Dated: August 10, 2001 Received: August 13, 2001

Dear Mr. Selvey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Don Selvey

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Styph Rhodes

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## II. Indications for Use Statement

NOV 0 9 2001

KO12630 510(k) Number (if known):

Device Name: Alliance Medical Corporation Reprocessed [device name]

Indications for Use: Reprocessed Arthroscopic Shavers are indicated for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.

| Manufacturer         | Description            | Model |
|----------------------|------------------------|-------|
| Linvatec Corporation | Oval Bur               | C9101 |
| Linvatec Corporation | Oval Bur               | C9102 |
| Linvatec Corporation | Oval Bur Left Helix    | C9106 |
| Linvatec Corporation | Spherical Bur          | C9110 |
| Linvatec Corporation | Spherical Bur          | C9111 |
| Linvatec Corporation | Spherical Bur          | C9112 |
| Linvatec Corporation | Vortex Router          | C9131 |
| Linvatec Corporation | Vortex Router Unhooded | C9134 |
| Linvatec Corporation | Oval Bur               | H9101 |
| Linvatec Corporation | Oval Bur               | H9102 |
| Linvatec Corporation | Spherical Bur          | H9110 |
| Linvatec Corporation | Vortex Router Hooder   | H9131 |
| Linvatec Corporation | Vortex Router          | H9132 |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Styck Rurder

Prescription Use (per 21 CFR 801.109)

CONFIDENTIAL

Sign-Off) Division of General, Restorativer-the-Counter Use ____________________________________________________________________________________________________________________________ and Neurological Devices

510(k) Number
Alliance Medical Corporation
Reprocessed Arthroscopic Burs
Traditional 510(k)

12

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K012630](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K012630)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
