Oxycon (wilofocon A) Rigid Gas Permeable, Spherical, Aspherical, Toric, and Bifocal Contact Lenses are indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and nonaphakic persons with non-diseased eyes. The lens may be disinfected using a chemical disinfection system only.
Device Story
Oxycon is a rigid gas permeable (RGP) contact lens manufactured from wilofocon A, a fluoro silicone acrylate polymer. Designed as a hemispherical shell, the lens corrects refractive errors including myopia, hyperopia, astigmatism, and presbyopia. It is available in spherical, aspherical, toric, and bifocal configurations. The device is intended for daily wear by patients with non-diseased eyes. It requires chemical disinfection. Clinicians fit the lens based on patient keratometry and refractive needs. The lens provides visual acuity correction, benefiting patients by compensating for refractive ametropia. It is used in clinical settings for fitting and by patients for daily vision correction.
Clinical Evidence
Clinical study enrolled 122 eyes (110 completed 3 months). Primary endpoints included slit-lamp findings, symptoms/problems/complaints (SPCs), keratometry, refraction, and visual acuity. Results: No adverse reactions or SLFs requiring treatment. 99.1% of eyes showed refraction changes ≤ 1.00D. No patient experienced visual acuity drop > 1 line. Average wearing time 14.9 hours. Preclinical testing included systemic injection (mice), eye irritation (rabbits), and agar overlay cytotoxicity (USP/NF), all showing non-toxic/non-cytotoxic results.
Technological Characteristics
Material: wilofocon A (fluoro silicone acrylate polymer) with D&C Green No. 6. Oxygen permeability (Dk): 26 x 10^-11 (cm²/sec)(ml O₂/ml x mm Hg) at 35°C. Refractive index: 1.44. Water content: <1%. Hardness (Shore): 80. Dimensions: 6.5-11.5 mm diameter, 6.50-9.00 mm base curve, -20.00 to +20.00 diopters. Surface: Hydrophobic. Wetting angle: 23.0 ± 2. Sterilization: Chemical disinfection system.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic and non-aphakic persons with non-diseased eyes.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{0}------------------------------------------------
Submitted by:
Daniel J. Manelli Farkas & Manelli, P.L.L.C. 2000 M Street NW (Suite 700) Washington, DC 20036
On behalf of Con-Cise Contact Lens Co., Inc. 510(k) Submission: Oxycon™ RGP Contact Lens
| Device Classification: | Rigid Gas Permeable Contact Lens (21 CFR 886.5916) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device: | Oxycon Rigid Gas Permeable Contact Lens (K974636) |
| | * *<br>*<br>*<br>*<br>* |
| NAME AND USE OF DEVICE | |
| Device Generic Name: | wilofocon A |
| Device Trade Name: | Oxycon rigid gas permeable contact lens |
| Indications: | Oxycon rigid gas permeable contact lenses<br>are indicated for daily wear correction of<br>visual acuity in aphakic and non-aphakic<br>persons with non-diseased eyes who are<br>nearsighted (myopic), farsighted (hyperopic),<br>presbyopic or have corneal astigmatism. |
| Airperm Ranges: | Chord diameter 6.5 to 11.5 mm<br>Center thickness 0.05 to 0.70 mm<br>Base Curve 6.50 to 9.00 mm<br>Powers -20.00 to +20.00 diopters |
| Cleaning and Disinfection: | The Oxycon rigid gas permeable contact lens<br>must be disinfected using only a chemical<br>(not heat) disinfection system as<br>recommended. |
{1}------------------------------------------------
## DEVICE CHARACTERISTICS
The Oxycon (wilofocon A) Rigid Gas Permeable Contact Lens is available as a spherical, aspherical, and astigmatic (toric) lens. The lens material, wilofocon A, is a fluoro silicone acrylate polymer which contains D&C Green No. 6 as a color additive. The Oxycon (wilofocon A) Contact Lens is a hemispherical shell of the following dimensions:
### LENS PARAMETERS AVAILABLE
| * Diameter | 6.5 to 11.5 mm |
|-------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| * Center Thickness<br>for Low Minus Lens:<br>for Plus Lens: | 0.05 to 0.30 mm<br>0.10 to 0.70 mm |
| * Base Curve | 6.50 to 9.00 mm |
| * Powers | -20.00 to +20.00 Diopters |
| *Aspheric Lens Eccentricity<br>Oblate to Prolate | - 1.5 to 1.5 |
| Peripheral Curves | 0.1 to 10 mm. Flatter or Steeper than Base Curve |
| * Toric Lens<br>Axis<br>Cylinder power | 1 to 180 degrees in 1 degree steps<br>0.50 to 4.00 Diopters |
| The physical/optical properties of the lens are: | |
| * Specific Gravity | 1.25 |
| * Refractive Index (589 nm at 25°C) | 1.44 |
| * Light Transmittance<br>Clear (370-760 nm)<br>Tinted (D&C Green No. 6) (380-780 nm) | 93%<br>87% |
| * Surface Character | Hydrophobic |
| * Wetting Angle (CLMA Method) | 23.0 ± 2 |
| * Water Content (Gravimetric method at 22°C) | <1% |
| * Oxygen Permeability<br>(cm²/sec) (ml O₂/ml x mm Hg) Polarographic method of Fatt (1990) | Dk 26 x 10-11 at 35°C |
| * Hardness (Shore) | 80 |
{2}------------------------------------------------
# Exhibit 4 Page 3
# SUMMARY OF STUDIES Oxycon (wilofocon A) Rigid Gas Permeable Contact Lens
### PRECLINICAL STUDIES
#### Systemic Injection Test in Mice 1.
The test is based on the FDA guidelines entitled "Testing Guidelines for Class II Contact Lenses. April 1989," and the United States Pharmacopeia XXII, National Formulary XVII. None of the test and control animals exhibited overt signs of toxicity at any of the observation points and hence the test material is considered non-toxic.
#### 2. Eye Irritation Test in Rabbits
The test is based on the FDA guidelines entitled "Testing Guidelines for Class III Contact Lenses, April 1989." The extract of test contact lenses was evaluated for the potential to produce ocular irritation over a 72 hour period. The test article did not cause any significant irritation to the ocular tissue of the test animals.
#### က် Tissue Culture-Agar Overlay Cytotoxicity Assay
The assay is based on that described in USP and National Formulary. The contact lens blanks did not cause any observable cytotoxic effects when tested according to the procedures outlined in the study report and is therefore considered non-cytotoxic.
Based on the results of the preclinical testing it is reasonable to assume that the Oxycon (wilofocon) Rigid Gas Permeable Contact Lens Polymer is non-toxic.
### CLINICAL STUDIES
All eyes enrolled in the study were accounted for. A total of 122 eves were enrolled in the study. A total of 110 eyes completed a minimum of three months of contact lens wear, 0 eyes were incomplete and 12 eyes were discontinued. The largest number of patients enrolled in this study (92.7%) were previous wearers of rigid gas permeable contact lenses. The average age was 37 years. During the three-month course of this study there were a total of 428 scheduled and 38 unscheduled eye visits.
Slit-Lamp: There was a total of 99 reports of Grade 1 SLFs of which the largest number was for corneal staining (77 reports or 18% of all visits). The staining at
{3}------------------------------------------------
the follow-up visits was nearly all of the type three and nine o'clock staining, which was also present at the initial visit. The next most common slit lamp findings were for injection, which included 15 reports or 3.5% of the visits. There were five reports of grade 2 findings, all for staining and which were resolved by lens changes.
The total number of slit-lamp findings were within expecteds for wearers of an RGP contact lens. The large number of reports for corneal staining can be attributed to the fact hat most patients were previous rigid lens wearers who entered the study with three and nine o'clock staining. During the time that lenses were worn by the discontinued patients there were no slit lamp findings reported.
SPCs: There were a total of 135 reports of symptoms, problems or complaints for the 356 eve visits during the three-month course of this study. The largest number of complaints (8.7%) was for variable vision, followed by lens awareness (6.2%). All symptoms, problems or complaints were reduced over time. None of the categories for symptoms, problems, and complaints can be considered outside of the normal expecteds for a rigid gas permeable contact lens.
Keratometry: The majority of the eve meridians (90.4%) showed no change to 0.50 diopters change over the course of this investigation. Changes of 1.00 diopter or less were found in 90.6% of the eve meridians. The number of changes were nearly equal for those who had increased or decreased power which resulted in an average change of 0.08 diopters flattening. In the three patients who experienced a change of greater than 1.00 diopter, two eye meridians showed a decrease of 1.25 diopters and one meridian showed an increase of 1.25 diopters. The changes in keratometry were generally low and may be attributed to adaptation to a new lens type or usual clinical measurement variation. The three patients who showed changes in one eye meridian of 1.25 diopters all had no change in their normal visual acuity of 20/20 and no significant signs or symptoms.
Refraction: the majority of the eyes (99.1%) showed no change to 1.00 diopters change over the course of this investigation. The number of spherical and cylindrical changes were nearly equal for those who had increased or decreased power, which resulted in an average change of near zero. In the two patients who experienced a change of greater than 1.00 diopter, one (306) showed a spherical change in the left eye of 1.50 diopters. One patient (505) showed a cylindrical decrease in both eyes of 1.50 diopters, which was attributed to a better fit of the test lens than was the case for the lens worn prior to the study. The changes in refraction were generally low and may be attributed to a better fitting lens or usual clinical measurement variation. The three eyes that showed changes of 1,50 diopters all had no change in their normal visual acuity of 20/20 and no significant
{4}------------------------------------------------
signs or symptoms. For the patients who discontinued, the 12 test eyes showed no change to 0.50 diopters change over the course of this investigation. The number of spherical and cylindrical changes were nearly equal for those who had increased or decreased power. The changes in refractive sphere and cylinder were generally low and may be attributed to normal variation in the findings.
Visual Acuity: A comparison was made between the best corrected visual acuity at the initial examination to the visual acuity when wearing contact lenses at the final examination. There were no patients who experienced a drop of visual acuity that was greater than one line during the course of this study. Two patients had reduced visual acuity both initially and at the end of this study and merit comment. Visual acuity results when wearing Oxycon Rigid Gas Permeable Contact Lenses during the course of this investigation were found to be within expecteds. No patients experienced a drop of visual acuity that was greater than one line.
The average wearing time achieved at the end of the study was 14.9 hours.
### Conclusions:
There were no eyes that experienced adverse reactions, SLFs requiring treatment or SPCs requiring treatment. The Oxycon rigid gas permeable contact lens was shown to be worn safely by all patient eyes in this study including those that completed (110 eyes) and discontinued (12 eyes) the study.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 4 1999
Mr. Daniel J. Manelli Representative for Con-Cise Contact Lens Co. 2000 M Street NW, Suite 700 Washington, DC 20036
Re: K990895 Trade Name: Oxycon™ Lens Regulatory Class: II Product Code: 86 HQD Dated: March 16, 1999 Received: March 17, 1999
Dear Mr. Manelli:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{6}------------------------------------------------
Page 2 -- Mr. Daniel J. Manelli
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# EXHIBIT 2
Page 1 of 1
510(k) Number (if known): K 99 0895
Oxycon (wilofocon A) Rigid Gas Permeable Spherical, Aspherical, Device Name: Toric, and Bifocal Contact Lenses
Indications for use:
Oxycon (wilofocon A) Rigid Gas Permeable, Spherical, Aspherical, Toric, and Bifocal Contact Lenses are indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and nonaphakic persons with non-diseased eyes. The lens may be disinfected using a chemical disinfection system only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Zorfele
(Division Sign-Off) Division of Ophthalmic Devices 510(k) Number_k99 o 895
Prescription use
(Per 21 CFR 801.109)
OR
Over-The Counter Use
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.