OPTIMUM GP (roflufocon D, roflurocon E) Daily Wear Contact Lens, HEXA100 (hexafocon A) Daily Wear Contact Lens
Applicant
Contamac, Ltd.
Product Code
HQD · Ophthalmic
Decision Date
Apr 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OPTIMUM GP (roflufocon D. roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for the correction of refractive ametropia (myopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive surgery. Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. OPTIMUM GP (roflufocon D) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject: 1. cannot be adequately corrected with spectacle lenses 2. requires a rigid gas permeable contact lens surface to improve vision 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses may concurrently provide correction of refractive error. The lenses may be disinfected using a chemical disinfection (not heat) system only.
Device Story
Rigid gas permeable (RGP) contact lenses; lathe-cut from fluoro-silicone acrylate materials (roflufocon D, roflufocon E, hexafocon A). Optional Tangible Hydra-PEG coating (polyethylene glycol-based polymer) for enhanced wettability. Used by patients with refractive errors or irregular corneal conditions/ocular surface diseases. Lenses provide refractive correction, physical protection, and an aqueous hydrated environment (scleral design). Prescribed by eye care professionals. Shipped dry or wet in polypropylene cases. Chemical disinfection only.
Clinical Evidence
No clinical data presented in this submission; safety and effectiveness of the materials were addressed in previous applications. The submission focuses on expanding the labeling to include therapeutic indications.
Technological Characteristics
Materials: roflufocon D, roflufocon E, hexafocon A (FDA Group #3 fluoro-silicone acrylate). Surface: Optional covalent Tangible Hydra-PEG coating. Oxygen permeability (Dk): 100-125 x 10^-11. Hardness: 75-80 Shore D. Water content: <1%. UV blocking: >98% UVB, >84-95% UVA. Production: Lathe-cut. Disinfection: Chemical only.
Indications for Use
Indicated for correction of refractive ametropia (myopia, astigmatism, presbyopia) in aphakic/non-aphakic persons with non-diseased eyes; management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-surgical); therapeutic use for ocular surface disorders (Graft-versus-Host, Sjögren's, dry eye, Filamentary Keratitis, limbal stem cell deficiency, neurotrophic keratitis, corneal exposure) requiring expanded tear reservoir/protection.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Boston XO® (hexafocon A) and Boston XO2® (hexafocon B) Rigid Gas Permeable Contact Lens (K171404)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 25, 2018
Contamac, Ltd. % Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln West Linn, OR 97068
Re: K180616
Trade/Device Name: OPTIMUM GP (roflufocon D, roflurocon E) Daily Wear Contact Lens, HEXA100 (hexafocon A) Daily Wear Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: March 5, 2018 Received: March 8, 2018
Dear Bret Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J Angelo Green -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180616
### Device Name
OPTIMUM GP (roflufocon D. roflufocon E) Daily Wear Contact Lens; HEXA100 (hexafocon A) Daily Wear Contact Lens
### Indications for Use (Describe)
OPTIMUM GP (roflufocon D. roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for the correction of refractive ametropia (myopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive surgery.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. OPTIMUM GP (roflufocon D) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K180616
#### I. SUBMITTER
- April 10th, 2018 Date Prepared:
- Name: Contamac Ltd. Address: Carlton House Shire Hill Saffron Walden Essex CB11 3AU Contact Person: Robert McGregor Managing Director Phone number: 01799 514800 Consultant: Bret Andre EyeReg Consulting, Inc. 6119 Canter Ln. West Linn, OR 97068
(503) 372-5226
#### II. DEVICE
Phone number:
| Trade Name: | OPTIMUM GP (roflufocon D, roflufocon E) Daily Wear Contact Lens;<br>HEXA100 (hexafocon A) Daily Wear Contact Lens |
|-------------------------|-------------------------------------------------------------------------------------------------------------------|
| Common<br>Name: | Daily wear rigid gas permeable contact lens |
| Classification<br>Name: | Rigid gas permeable contact lens. (21 CFR 886.5916) |
| Regulatory<br>Class: | Class II |
| Product Code: | HQD |
{5}------------------------------------------------
#### III. PREDICATE DEVICE
The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are substantially equivalent to the following predicate device:
- 트 "Boston XO® (hexafocon A) and Boston XO2® (hexafocon B) Rigid Gas Permeable Contact Lens" By Bausch + Lomb Inc. 510(k) number; K171404
#### IV. DEVICE DESCRIPTION
### The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear
Contact Lenses are lathe cut from one of the following hydrophobic, FDA Group #3 fluoro-silicone acrylate materials: roflufocon D, roflufocon E, or hexafocon A-with the following properties:
| | ROFLUFOCON D | ROFLUFOCON E | HEXAFOCON A |
|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Refractive Index | 1.4333 | 1.4332 | 1.414 |
| Light Transmission (clear) | >97% | >97% | - |
| Light Transmission (tinted) | >90% | >90% | >91% |
| Water Content | <1% | <1% | <1% |
| Specific Gravity | 1.166 | 1.155 | 1.266 |
| Shore D Hardness | 75 | 77 | 80 |
| Oxygen Permeability<br>(Dk) ISO/FATT Method | $100 \times 10^{-11} \frac{cm^2}{sec} \frac{ml O_2}{ml \times mm Hg @ 35^{\circ}C}$ | $125 \times 10^{-11} \frac{cm^2}{sec} \frac{ml O_2}{ml \times mm Hg @ 35^{\circ}C}$ | $113 \times 10^{-11} \frac{cm^2}{sec} \frac{ml O_2}{ml \times mm Hg @ 35^{\circ}C}$ |
| Contain one or more of the following color<br>additives conforming to:<br>21 CFR Part 73 & 74, Subpart D | D & C Green No. 6, FD & C Red<br>No. 17,<br>CI Solvent Yellow 18 | D & C Green No. 6, FD & C Red<br>No. 17,<br>CI Solvent Yellow 18 | D&C Green No. 6, C.I.<br>Solvent Yellow No. 18,<br>D&C Violet No. 2 and<br>D&C Red No. 17 |
| UV Light Blocking<br>(UVB - 280nm – 315nm; UVA 316nm<br>– 380nm) | >98% UVB<br>>95% UVA | >98% UVB<br>>95% UVA | >98% UVB<br>>84% UVA |
### The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear
Contact Lenses are available with or without Tangible™ Hydra-PEG-which is a thin polyethylene glycol (PEG)-based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible™ Hydra-PEG, the underlying material is encapsulated in a thin layer of polymer that results in measurable improvement of wettability compared to untreated lenses.
{6}------------------------------------------------
The surface properties of roflufocon D, roflufocon E, and hexafocon A materials uncoated and coated with Tangible Hydra-PEG are depicted in the following table:
| | roflufocon D | roflufocon E | hexafocon A |
|---------------------------------------------------------|------------------------------------------------|------------------------------------------------|----------------------------------------------|
| Captive Bubble Contact<br>Angle ± Standard<br>Deviation | Coated: 40.40°±5.05°<br>Uncoated: 93.28°±2.13° | Coated: 36.90°±8.04°<br>Uncoated: 93.64°±3.47° | Coated: 49.1°±5.81°<br>Uncoated: 96.4°±2.79° |
The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses may be packaged and shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses is the PolyVial Contact Lens Case. When shipped "wet", the lenses may be packaged and shipped in the Unique pH contact lens care system by Menicon Co., Ltd. The active ingredients in Unique pH solution are Edetate Disodium 0.01% and Polyquaternium 10.0011%.
The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are available in the following lens parameters:
| Parameter | Range | Tolerance |
|-------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve | 5.00mm to 8.00mm | ± 0.05 mm |
| Center Thickness | 0.20mm to 0.50mm | ± 0.02 mm |
| Diameter | 7.00mm to 22.00mm | ± 0.10mm |
| Spherical Power | -20.00 D to +20.00 D<br>(in 0.25D steps) | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
| Cylindrical Power | Up to -10.00 D<br>(in 0.25 D steps) | ± 0.25 (0 to = 2D)<br ± 0.37 (2 to = 4D)<br ± 0.50 (over 4D) |
| Cylindrical Axis | 1° to 180° (in 1° steps) | ± 5° |
| Multifocal Power | +1.00 D to 4.00 D<br>(in 0.25 D steps) | ± 0.25D |
{7}------------------------------------------------
#### V. INDICATIONS FOR USE
OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for the correction of refractive ametropia (myopia, astigmatism and presbyopia) in aphakic and non aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive surgery.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The OPTIMUM GP (roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease. Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic. paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the OPTIMUM GP (roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
{8}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
# The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear
Contact Lenses are substantially equivalent to the Boston XO® (hexafocon A) and Boston XO2®(hexafocon B) Rigid Gas Permeable Contact Lenses (cleared under K171404) in terms of the following:
- Intended use - daily wear contact lenses
- 트 Indications for use - therapeutic (scleral)
- 트 Actions
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group – group # 3 fluoro silicone acrylate
- 트 Production method - lathe cut
The following matrix illustrates the production method, lens function and material characteristics of the OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses, as well as the predicate device.
| | OPTIMUM GP and<br>HEXA100 Daily Wear<br>Contact Lenses | Boston XO® and Boston XO2® RGP<br>Contact Lenses |
|-----------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device (K171404) |
| Classification | Same as predicate | Class II<br>Lenses, Rigid Gas Permeable, Daily Wear<br>21 CFR 886.5916 |
| Product Code | Same as predicate | HQD |
| FDA Group # | Same as predicate | Group # 3 Fluoro Silicone Acrylate |
| USAN | roflufocon D, roflufocon E,<br>hexafocon A | hexafocon A, hexafocon B |
| Production Method | Same as predicate | Lathe-Cut |
| Actions | Same as predicate | The contact lenses act as a refractive medium<br>that focus light rays from near and distant<br>objects on the retina. When placed on the<br>eye for therapeutic use, the lens replaces or<br>supports impaired ocular surface function. |
| Intended Use | Same as predicate | Daily Wear |
| Indication for Use | Same as predicate | Indicated for therapeutic<br>use for the management of irregular and<br>distorted corneal surfaces where the subject:<br>1. cannot be adequately corrected with<br>spectacle lenses<br>2. requires a rigid gas permeable contact lens<br>surface to improve vision<br>3. is unable to wear a corneal rigid gas<br>permeable lens due to corneal distortion or<br>surface irregularities Common causes of<br>corneal distortion include but are not limited<br>to corneal infections, trauma, tractions as a<br>result of scar formation secondary to<br>refractive surgery or corneal transplantation.<br>Causes may also include corneal<br>degeneration and corneal dystrophy. Also<br>indicated for therapeutic use in eyes with<br>ocular surface, limbal stem cell deficiency,<br>disorders of the skin, neurotrophic keratitis,<br>and corneal exposure that might benefit from<br>the presence of an expanded tear reservoir<br>and protection against an adverse<br>environment. |
| Specific Gravity | roflufocon D: 1.166<br>roflufocon E: 1.155<br>hexafocon A: 1.266 | hexafocon A: 1.266<br>hexafocon B: 1.190 |
| Shore D Hardness | roflufocon D: 75<br>roflufocon E: 77<br>hexafocon A: 80 | hexafocon A: 80<br>hexafocon B: 78 |
| Oxygen Permeability<br>x 10 11 (cm2/sec) (ml O2/ml x<br>mm Hg @ 350C) | roflufocon D: 100<br>roflufocon E: 125<br>hexafocon A: 111 | hexafocon A: 111<br>hexafocon B: 141 |
| Refractive Index | roflufocon D: 1.433<br>roflufocon E: 1.433<br>hexafocon A: 1.415 | hexafocon A: 1.415<br>hexafocon B: 1.424 |
| Water Content (%) | <1% | <1% |
| UV Absorber Available | Yes | Yes |
| Hydra-PEG Surfacing<br>Available | Yes | No |
# Substantial Equivalence Matrix
{9}------------------------------------------------
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
### ~ Non-Clinical Studies ~
The purpose of this application is to modify the labeling of previously FDA cleared RGP contact lenses/materials to include a therapeutic indication for use. Non-clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D. roflufocon E, and hexafocon A materials has been addressed in previous applications.
### ~ Clinical Studies ~
Clinical performance data to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D, roflufocon E, and hexafocon A has been previously addressed.
## VIII. CONCLUSIONS
### Substantial Equivalence
Information presented in this Premarket Notification establishes that OPTIMUM GP (rofflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact Lenses are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.
### Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.