BOSTON XO® (hexafocon A) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery). Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. Boston XO Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject: 1. cannot be adequately corrected with spectacle lenses 2. requires a rigid gas permeable contact lens surface to improve vision 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities. Common causes of corneal distortion include but are not limited to corneal infections as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error. The lenses may be disinfected using a chemical disinfection (not heat) system only.
Device Story
Fluoro silicone acrylate copolymer rigid gas permeable contact lenses; available in spherical, aspherical, toric, and scleral designs. Function as refracting medium to focus light on retina; scleral designs provide physical protection, aqueous hydrated environment, and saline bath for ocular surface disorders. Used in clinical settings by eye care professionals. Output is improved visual acuity or therapeutic protection for irregular/distorted corneas and ocular surface diseases. Benefits include vision correction and management of complex corneal conditions.
Clinical Evidence
Bench testing only. No new clinical data required as the application was for labeling revisions to existing cleared materials and designs.
Technological Characteristics
Fluoro silicone acrylate copolymer; lathe-cut; rigid gas permeable; water content <1%; oxygen permeability (Dk) 100 (XO) and 141 (XO2); refractive index 1.415-1.424; chemical disinfection only.
Indications for Use
Indicated for daily wear for correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic/non-aphakic persons. Also indicated for management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, post-refractive surgery) and therapeutic use for ocular surface disease (Graft-versus-Host, Sjögren's, dry eye, Filamentary Keratitis, limbal stem cell deficiency, skin disorders, neurotrophic keratitis, corneal exposure). XO2 additionally indicated for orthokeratology (myopia reduction up to 5.00D).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2017
BAUSCH + LOMB Incorporated % Ellen M. Beucler Vice President Foresight Regulatory Strategies, Inc. 187 Ballardvale Street, Suite A250 Wilmington, MA 01887
Re: K171404
Trade/Device Name: BOSTON XO® (hexafocon A), BOSTON XO2® (hexafocon B) Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: May 10, 2017 Received: May 12, 2017
Dear Ellen M. Beucler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K171404
#### Device Name
BOSTON XO® (hexafocon A) Rigid Gas Permeable Contact Lens
#### Indications for Use (Describe)
BOSTON XO® (hexafocon A) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
Boston XO Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
The Boston XO Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K171404
#### Device Name
BOSTON XO2® (hexafocon B) Rigid Gas Permeable Contact Lens
### Indications for Use (Describe)
BOSTON XO2® (hexafocon B) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
BOSTON XO2 Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
Boston XO2 Scleral Contact Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
The Boston XO2 Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
## 510(k) Summary
## 1. Applicant Information
Bausch + Lomb Inc. 1400 North Goodman Street Rochester, NY 14609 USA
| Contact: | Glenn Davies, O.D.<br>Senior Director Regulatory Affairs |
|----------------|----------------------------------------------------------|
| Phone: | (585) 338-8215 |
| Fax: | (585) 338-0702 |
| email: | glenndavies@bausch.com |
| Date Prepared: | July 11, 2017 |
### 2. Device Information
| Classification name: | Rigid Gas Permeable Contact Lens |
|------------------------|--------------------------------------------------------------------------------------------|
| Device classification: | Class II |
| Regulation number: | 21 CFR 886.5916 (Rigid Gas Permeable Contact Lenses) |
| Product code: | HQD |
| Proprietary name: | Boston XO® (hexafocon A) and Boston XO2®<br>(hexafocon B) Rigid Gas Permeable Contact Lens |
## 3. Predicate Devices
Bausch + Lomb Inc. claims substantial equivalence to Boston Scleral Contact Lens, PMA No. P860022/S40 approved March 1, 1994 and BostonSight PD Prosthetic Device cleared in K161461 on July 25, 2016.
## 4. Description of Device
The Boston XO and Boston XO2 Rigid Gas Permeable (RGP) Contact Lenses are fluoro silicone acrylate copolymer rigid gas permeable contact lenses available in a variety of tints and may contain an ultraviolet light absorber. The hemispherical shells are available in several lens designs including spherical, aspherical, toric, scleral and irregular cornea designs.
{7}------------------------------------------------
## 5. Indications for Use
Boston XO (hexafocon A) and Boston XO2 (hexafocon B) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and non aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
The Boston XO2 (hexafocon B) Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
The Boston XO and XO2 Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
{8}------------------------------------------------
## 5. Indications for Use (continued)
The Boston XO and XO2 Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment.
When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
### 6. Performance Data
There are no changes to the Boston XO (hexafocon A) and Boston XO2 (hexafocon B) Rigid Gas Permeable contact lens material or lens designs as part of this application. These materials have been previously cleared for scleral lens designs. This application is for labeling revisions. No performance data is required.
## 7. Substantial Equivalence
In support of the expanded indications for Boston XO and XO2 to include scleral lens indications the following statements apply.
- Boston XO and XO2 Rigid Gas Permeable Scleral Contact Lenses are fluoro silicone acrylate contact lenses that use similar actions and technological characteristics as the predicate devices (see Table 1 and Table 2).
- . The Boston XO and XO2 Rigid Gas Permeable Scleral Lenses present the same risks and benefits as the predicate devices and are as safe and effective as the predicate devices when used according to the labeled directions for use and requested indications.
Therefore, the Boston XO and XO2 Rigid Gas Permeable Scleral Lenses are substantially equivalent to the BostonSight PD Prosthetic Device and the Boston Scleral Contact Lens.
{9}------------------------------------------------
| omnannen<br>3<br>villum<br>C<br>) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| |
| <br>ાદ તા<br>ﺎ<br>) |
| l<br>8 |
| I |
| 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>œ<br>uly uncil<br>1 1 6 0 0 0 0 1 1<br>H<br>1 |
| D VEUT DUWUV<br>1<br>Thetmittel<br>P<br>All I<br>C<br>2 |
| 1<br>1<br>4<br>œ |
| | Boston XO & XO2 Scleral Lens<br>(hexafocon A and hexafocon B) | BostonSight PD<br>Prosthetic Device<br>(oprifocon A) | Boston Scleral Lens<br>(oprifocon A) | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| | K171404 | K161461 | P860022/S040 | |
| | Subject Device | Predicate Device | Predicate Device | |
| Indication for Use | BOSTON XO (hexafocon A) and XO2® (hexafocon B) Contact Lenses are indicated for daily wear for the<br>correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and non<br>aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes<br>that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus,<br>pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).<br>BOSTON XO2 Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the<br>temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the<br>orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.<br>Furthermore, eyes suffering fromcertain ocular surface disorders may benefit fromthe physical protection,<br>aqueous hydrated environment and the saline bath provided by scleral lens designs.<br>Boston Xo and XO2 Scleral Contact Lens designs for daily wear are indicated for therapeutic use for the<br>management of irregular and distorted corneal surfaces where the subject:<br>1. cannot be adequately corrected with spectacle lenses<br>2. requires a rigid gas permeable contact lens surface to improve vision<br>3. is unable to wear a comeal rigid gas permeable lens due to corneal distortion or surface irregularities<br>Common causes of corneal distortion include but are not limited to comeal infections, trauma, tractions as a<br>result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or comeal<br>transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid<br>marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy,<br>granular comeal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO and XO2 Scleral<br>Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g.<br>ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal<br>stemcell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal bums), disorders of the<br>skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial<br>Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit fromthe presence of an<br>expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use<br>for a distorted comea or ocular surface disease, the Boston Scleral Lenses may concurrently provide<br>correction of refractive erroг.<br>lenses may be disinfected using a chemical disinfection (not heat) system only. | | | |
| Actions/Operational<br>Principles | When placed on the eye the Rigid Gas Permeable Contact Lens acts as a<br>refracting mediumto focus light rays on the retina to improve visual acuity. | Indicated for therapeutic use in eyes with<br>ocular surface disease from dry eye (e.g. ocular<br>Graft-versus-Host disease, Sjögren's<br>syndrome, dry eye syndrome), limbal stem cell<br>deficiency (e.g.Stevens-Johnson syndrome,<br>chemical and thermal burns, radiation),<br>disorders of the skin (e.g. atopy, ectodermal<br>dysplasia), neurotrophic keratitis (e.g. Herpes<br>simplex, Herpes zoster, Familial Dysautonomia),<br>and corneal exposure (e.g. anatomic, paralytic)<br>that might benefit from the presence of an<br>expanded tear reservoir and protection against<br>an adverse environment. When prescribed for<br>therapeutic use for distorted cornea or ocular<br>surface disease, the BostonSight PD Prosthetic<br>Device may incidentally provide correction of<br>refractive erroг.<br>When placed on the eye the Rigid Gas<br>Permeable Contact Lens acts as a refracting<br>medium to focus light rays on the retina to<br>improve visual acuity. | Indicated for eyes having significantly reduced<br>vision due to the presence of a distorted<br>comeal surface that:<br>1) precludes satisfactory spectacle lens<br>correction 2) demonstrates significantly<br>improved rigid contact lens corrected vision 3)<br>is incapable of wearing traditional corneal<br>lenses because of the inability to achieve<br>adequate lens centration/stability and/or<br>tolerance to physical contact with a lens.<br>Furthermore, eyes suffering from certain ocular<br>surface disorders may benefit from the physical<br>protection and the saline bath provided by a<br>scleral lens.<br>When placed on the eye the Rigid Gas<br>Permeable Contact Lens acts as a refracting<br>medium to focus light rays on the retina to<br>improve visual acuity. | |
| Product Code | HQD | HQD | HQD | |
| Common name | Contact Lens, Rigid Gas Permeable | Contact Lens, Rigid Gas Permeable | Contact Lens, Rigid Gas Permeable | |
| Device Class | II | II | II* | |
| CFR Reference | 21 CFR 886.5916 | 21 CFR 886.5916 | 21 CFR 886.5916 | |
| Production Method | Lathe Cut | Lathe cut | Lathe cut | |
| FDA Group # | Group # 3 Fluom Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate | |
| | Boston XO & XO2<br>Contact Lens<br>K171404 |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.