BOSTON XO (HEXAFOCON A), BOSTON EO (ENFLUFOCON B) AND BOSTON ES (ENFLUFOCON A) RIGID GAS PERMEABLE CONTACT LENSES
K071043 · Bausch & Lomb, Inc. · HQD · Aug 2, 2007 · Ophthalmic
Device Facts
Record ID
K071043
Device Name
BOSTON XO (HEXAFOCON A), BOSTON EO (ENFLUFOCON B) AND BOSTON ES (ENFLUFOCON A) RIGID GAS PERMEABLE CONTACT LENSES
Applicant
Bausch & Lomb, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Aug 2, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
The Boston XO (hexafocon A), Boston EO (enflufocon B), and Boston ES (enflufocon A) Rigid Gas Permeable Contact Lenses are indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not-aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.
Device Story
Rigid gas permeable contact lenses composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer; available with or without UV absorber. Designed for daily wear to correct refractive errors or manage irregular corneal conditions. Prescribed by eye care professionals for patients with non-diseased eyes. Lenses require chemical disinfection systems. Device provides vision correction through physical refractive properties of the lens material.
Clinical Evidence
No clinical data provided. Clinical studies were deemed unnecessary as the modification (addition of UV blocker) did not raise new questions of safety or effectiveness.
Technological Characteristics
Rigid gas permeable contact lenses; material: aliphatic fluoroitaconate siloxanyl methacrylate copolymer. Properties: water content <1%, refractive index 1.425-1.443, UV blocker (UV 49). Oxygen permeability (Dk) ranges from 18 to 100 depending on material. Chemical disinfection required.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, astigmatism, and presbyopia in aphakic and non-aphakic persons with non-diseased eyes. Also indicated for management of irregular corneal conditions including keratoconus, pellucid marginal degeneration, or post-penetrating keratoplasty or refractive surgery.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{0}------------------------------------------------
K071043
510(K) SUMMARY BOSTON XO (hexafocon A), BOSTON EO (enflufocon B), and BOSTON ES (enflufocon A) Rigid Gas Permeable Contact Lenses for Daily Wear
#### 1. Applicant's Name and Address
Bausch & Lomb, Inc. 1400 North Goodman Street Rochester. NY 14609
- 2. Contact Person Debra Ketchum Manager,Global Regulatory Affairs Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609 (585) 338-8638
#### 3. ldentification of Device
| Common Name: | contact lens-rigid gas permeable (hydrophobic) | |
|------------------------|---------------------------------------------------------------------------------------------------------------|--|
| Trade Name: | Boston XO (hexafocon A), Boston EO<br>(enflufocon B), and Boston ES (enflufocon A)<br>Daily Wear Contact Lens | |
| Classification: | Class II ophthalmic (21CFR 886.5916) | |
| Device classification: | Class II (21 CFR 886.5916) | |
| Pro Code: | HQD | |
Predicate Devices:
K053124 and K013762: Boston XO (hexafocon A), Boston EO (enflufocon B), Boston ES (enflufocon A)
| 510(k) | Clearance Date | Device Description |
|---------|----------------|------------------------------------|
| K053124 | 1/30/2006 | Boston XO, Boston EO,<br>Boston ES |
| K013762 | 4/3/2002 | Boston XO, Boston EO,<br>Boston ES |
{1}------------------------------------------------
# 4. Description of device
Boston XO (hexafocon A) is a rigid gas permeable material, composed of siloxanyl fluoromethacrylate copolymer available with or without an ultraviolet absorber.
The Boston XO (hexafocon A), Boston EO (enflufocon B) and Boston ES (enflufocon A) are rigid gas permeable contact lens materials composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer available with or without an ultraviolet absorber.
| Property | BOSTON XO | BOSTON EO | BOSTON ES |
|--------------------------------|-------------|-------------|-------------|
| Specific Gravity | 1.27 | 1.23 | 1.22 |
| Refractive Index | 1.425 | 1.429 | 1.443 |
| Visible Light<br>Transmittance | 92% average | 92% average | 92% average |
| UV Blocker | UV 49 | UV 49 | UV 49 |
| Water Content | <1% | <1% | <1% |
| Wetting Angle | 49° | 49° | 52° |
| Oxygen | 140* | 82* | 36* |
| Permeability<br>(Dk)*** | 100** | 58** | 18** |
The physical and optical properties of the lenses are:
*gas to gas method
**polarographic method (ISO)
***polarographic method (FATT) (x 10-11 (cm202 x cm)/cm² x sec x mmHg)@35°C
#### Intended use 5.
The Boston XO (hexafocon A), Boston EO (enflufocon B), and Boston ES (enflufocon A) Rigid Gas Permeable Contact Lenses are indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not-aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.
{2}------------------------------------------------
#### 6. Description of Safety and Substantial Equivalence
The safety and efficacy of the Boston XO (hexafocon A), Boston EO (enflufocon B), and Boston ES (enflufocon A) was demonstrated in original 510(k) Premarket Notifications as follows:
- . K000795, cleared May 25, 2000
- . K013762, cleared April 3, 2002
- K053124, cleared January 30, 2006 .
Safety and efficacy for all subject devices were determined in prior 510(k) clearances. The addition of the UV blocker to the BOSTON XO material has been demonstrated as safe as evidenced by preclinical testing as required in current Pre-Market Notification Guidance for Daily Wear Contact Lenses, May 12, 1994. All three materials will now use the same UV blocker.
## 7. Clinical data:
Clinical studies for the BOSTON XO (hexafocon A), BOSTON EO (enflufocon B), and BOSTON ES (enflufocon A) materials have been deemed as not necessary in support of their clearance as no new or additional questions of safety or effectiveness have been raised as a result of the change of UV blocker in BOSTON XO (hexafocon A). BOSTON EO (enflufocon B) and BOSTON ES (enflufocon A) already incorporate the same UV blocker.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 2 2007
Bausch & Lomb, Inc. c/o Debra Ketchum 1400 North Goodman Street Rochester NY 14609-3547
Re: K071043
Trade/Device Name: Boston XO (hexafocon A), Boston EO (enflufocon B), and Boston ES (enflufocon A) Daily Wear Rigid Gas Permeable Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HOD Dated: July 17, 2007 Received: July 18, 2007
Dear Ms. Ketchum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M.B. Egelman SimD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
K 071043
Device Name:
510(k) Number (if known):
Boston XO (hexafocon A), Boston EO (enflufocon B), and Boston ES (eflufocon A) Daily Wear Contact Lens
### Indication for Use
Boston XO (hexafocon A), Boston EO (enflufocon B) and Boston ES (enflufocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | | |
|--------------------------------------------------------|---------|----|-----------------------|
| Prescription Use | X | OR | Over-the -counter-use |
| (Division Sign-Off) | | | |
| Division of Ophthalmic Ear, Nose and Throat Devises | | | |
| 510(k) Number | K071043 | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.