ONSI -56 (ONSIFOCON A) RIGID GAS PERMEABLE SPHERICAL, ASPHERIC, TORIC AND BIFOCAL CONTACT LENS
Applicant
The Lagado Corp.
Product Code
HQD · Ophthalmic
Decision Date
Sep 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
The ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. The lenses may be disinfected only by using chemical disinfection
Device Story
ONSI™-56 is a rigid gas permeable (RGP) contact lens made of onsifocon A polymer. Available in spherical, aspheric, toric, and translating bifocal designs. Used for daily wear to correct visual acuity in myopic, hyperopic, or presbyopic patients. Input is the patient's refractive error; device acts as a corrective optical lens. Operated by the patient for daily wear. Healthcare providers prescribe and fit the lenses based on patient needs. Benefits include improved visual acuity and correction of astigmatism. Lenses are available in various tints and with UV-blocking options.
Clinical Evidence
Clinical trial of 37 subjects (ages 15-58, 73% female) over 3 months of daily wear. Primary endpoints included safety (adverse events, slit lamp findings) and efficacy (visual acuity, wear time). No adverse events reported. Final visual acuity was within 1 line of initial best-corrected acuity. Slit lamp findings showed no significant clinical concerns. Study concluded the device is safe and effective.
Technological Characteristics
Material: onsifocon A (trifluoroethyl methacrylate polymer with siloxane components). Water content <1%. Oxygen permeability: 56.2 (ANSI units). Refractive index: 1.452. Hardness: D/85 (Shore). Wetting angle: 7.25 degrees (sessile drop). Mechanical strength: 2698 psi. Available in spherical, aspheric, toric, and bifocal designs. Chemical disinfection required.
Indications for Use
Indicated for daily wear for correction of visual acuity in non-aphakic persons with non-diseased eyes who are myopic, hyperopic, or presbyopic, and may exhibit corneal astigmatism.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Oxycon (wilofocon A) Rigid Gas Permeable Contact Lenses
Submission Summary (Full Text)
{0}------------------------------------------------
K033599
# SEP 16 2004
## 510(k) SUMMARY FOR FREEDOM OF INFORMATION
### ONSI™-56 (onsifocon A) RIGID GAS PERMEABLE SPHERICAL, ASPHÈRIC, TORIC AND BIFOCAL CONTACT LENS FOR DAILY WEAR
| 1. Submitted by: | The Lagado Corporation<br>2890 South Tejon St<br>Englewood, CO 80110 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | John M. Szabocsik, Ph.D.<br>Official agent Szabocsik and Associates<br>203 N. Wabash, Ste 1200<br>Chicago, IL 60601<br>(312) 553-0828 |
| 2. Date prepared: | June 30, 2004 |
| 3. Device:<br>Common Name | ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric,<br>Toric and Bifocal Contact Lens |
| Trade Name | ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric,<br>Toric and Bifocal Contact Lens |
| 4. Classification | Class II (Performance Standards)<br>21 CFR 886.5916 (b) (1)<br>Rigid Gas Permeable Contact Lens for Daily Wear |
| 5. Substantial<br>Equivalence | This product is substantially equivalent to other currently marketed rigid lenses, such as Oxycon (wilofocon A) Rigid Gas Permeable Contact Lenses |
| 6. Device<br>Description | The ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric,<br>Toric and Bifocal Contact Lenses are available in spherical designs in the<br>clear untinted or blue, green, gray, or blue-violet tinted varieties and blue<br>tint with UV-blocker. Each is a shell of the following dimensions. |
| | Spherical Contact Lens:<br>Horizontal Lens Size: 6.5 mm to 11.50mm<br>Base Curve: 6.50mm to 9.50mm<br>Distance Powers: +12.00D to -20.00D<br>Center Thickness:<br>for low minus 0.05mm to 0.30mm<br>for plus 0.10mm to 0.70mm |
| | Aspheric Lens<br>Eccentricity 0 to 1.5<br>Peripheral Curves 0.1 to 1.0mm |
| | Toric Lens<br>Axis 1 to 180 degrees in 1 degree steps |
| | Cylinder power 0.50 to 4.00D |
FOI Summary, Pg 1
{1}------------------------------------------------
| Translating Bifocal Contact Lens: | | | |
|-----------------------------------|---------|----|---------|
| Horizontal Lens Size: | 8.00mm | to | 10.50mm |
| Base Curve: | 6.50mm | to | 8.50mm |
| Distance Power: | +12.00D | to | -20.00D |
| Add Power | 1.00 | to | 4.00D |
The lens material, onsifocon A, is trifluoroethyl methacrylate polymer with tris (trimethylsiloxy)methacryloxypropylsilane 3-trimethoxysily|propylmethacrylate methacrylic acid 1,3-bis(3- methacryloxypropyl)tetrakis(trimethylsiloxy)disiloxane ethylene glycol dimethacrylate 2-hydroxyethyl methacrylate N-vinylpyrrolidone. The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and CI Solvent Yellow 18; the blue-violet contain D&C Green No. 6 and D&C Violet No. 2; the blue-UV lenses contain D&C Green No 6 and a UV absorber, NORBLOC 7966. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
### 7. Intended use
The ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism.
The lenses may be disinfected only by using chemical disinfection
8. Comparison to predicate devices: see following table
### SUBSTANTIAL EQUIVALENCE
| Water Content | OXYCON (wilofocon A) | ONSITM-56 (onsifocon A) |
|-----------------------------------|-----------------------------------|----------------------------------------------|
| | <1% | <1.0% |
| Oxygen Permeability | $26x10^{-11}$<br>Fatt Units(1990) | 56.2<br>ANSI units |
| Refractive Index | 1.44 | 1.452 |
| Hardness | D/89 (Shore) | D/85 (Shore) |
| Specific Gravity | 1.25 | 1.206 |
| Residual Monomers | | no leachable monomers<br>detected |
| Wetting Angle | 23.0\$\nabla\$2<br>(CLMA method) | 7.25E\$\nabla\$1.55<br>(sessile drop method) |
| Mechanical (flexural)<br>Strength | | 2698psi |
{2}------------------------------------------------
ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses, FOI
| | Light Transmittance | OXYCON (wilofocon A) | ONSITM-56 (onsifocon A) |
|--|---------------------|----------------------|----------------------------------------|
| | clear | | >95% T |
| | blue | | >70% T |
| | green | 93% | >70% T |
| | gray | 87% | >70% T |
| | blue-violet | | >70% T |
| | blue UV | | >70% T (400-780nm)<br>0% T (200-380nm) |
#### Introduction
The ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical. Aspheric, Toric and Bifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in notaphakic persons with non-diseased eyes that are myopic, or presbyopic and which may exhibit corneal astigmatism.
Contained in the submission are comparisons of the product to the predicate device, information on the chemistry and manufacturing, results of toxicological tests, and the report of a clinical trial of 37 subjects, who have used the product over a period of three months.
#### I. Chemistry and Manufacturing
The lens material, onsifocon A, is trifluoroethyl methacrylate polymer with tris (trimethylsiloxy)methacryloxypropylsilane 3-trimethoxysily|propy|methacrylate methacrylic acid 1,3-bis(3- methacryloxypropyl)tetrakis(trimethylsiloxy)disiloxane ethylene glycol dimethacrylate 2-hydroxyethyl methacrylate N-vinylpyrrolidone.
The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and CI Solvent Yellow 18; the blue-violet lenses contain D&C Green No 6 and D&C Violet No 2; the blue-UV lenses contain D&C Green No 6 and a UV absorber, NORBLOC 7966. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
The ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical. Aspheric, Toric and Bifocal Contact Lenses are available in spherical designs in the clear untinted or blue, green, gray or blue-violet tinted varieties and blue with UV-blocker. Each is a shell of the following dimensions.
| Spherical Contact Lens: | | | |
|-------------------------|---------|----|---------|
| Horizontal Lens Size: | 6.5 mm | to | 11.50mm |
| Base Curve: | 6.50mm | to | 9.50mm |
| Distance Powers: | +12.00D | to | -20.00D |
| Center Thickness: | | | |
| for low minus | 0.05mm | to | 0.30mm |
| for plus | 0.10mm | to | 0.70mm |
| Aspheric Lens | | | |
| Eccentricity | 0 | to | 1.5 |
| Peripheral Curves | 0.1 | to | 1.0mm |
{3}------------------------------------------------
ONSITM-56 (onsifocon A) Rigid Gas Permeable Contact Lenses. FOI
Toric Lens Axis 1 to 180 degrees in 1 degree steps Cylinder power 0.50 to 4.00D
| Translating Bifocal Contact Lens: | | | |
|-----------------------------------|---------|----|---------|
| Horizontal Lens Size: | 8.00mm | to | 10.50mm |
| Base Curve: | 6.50mm | to | 8.50mm |
| Distance Power: | +12.00D | to | -20.00D |
| Add Power | 1.00 | to | 4.00D |
#### II. Toxicology
The toxicological testing is summarized below. The lens material was shown to be non-toxic in all tests.
A. Agar Overlay Cytotoxicity:
The ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lens material was tested in a direct contact cytotoxicity assay. The lenses were noncytotoxic.
#### B. Systemic toxicity:
Saline and cottonseed oil extracts of the ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lens material were evaluated for systemic toxicity by intraperitoneal (ip) or intravenous (iv) injection in healthy mice, 50ml/kg body weight. The animals were observed over a 72 hour period, and showed no difference from control animals. The lenses passed the test requirements, that there be no difference between the response of test and control animals.
C. Acute Ocular irritation:
Saline and cottonseed oil extracts of the ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lens material were evaluated for ocular irritation by instillation into the inferior ocular cul-de-sac of rabbits. The eyes were examined over a 72 hour period and showed no irritation.
#### III. Microbiology
The material has a water content less than 1%, and is therefore exempt from the microbiological requirements for hydrophobic contact lenses. Bioburden testing will be performed periodically on finished product.
### IV. Clinical Studies
A clinical trial of 3 months usage of the ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical Contact Lenses by 37 subjects, wearing lenses on a daily wear schedule, showed that the product is substantially equivalent to other lenses available on the market. Comparison was made to historical controls wearing wilofocon A. The clinical summary follows.
The study was conducted over 3 months of wear, subjects being seen initially, and after 1, 2 weeks and 1, 2 and 3 months.
{4}------------------------------------------------
ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses, FOI
Three (3) investigators enrolled a total of 37 test subjects. The age range of the test population was from 15 to 58, with 27 (73%) females and 10 (27%) males. Of the 37 subjects, 36 (97%) completed 3 months of wear, and 1 (3%) discontinued.
FINDINGS a. SAFETY:
(1) Adverse Reactions
The FDA regulations for medical devices (21 CFR 812.3) define an unanticipated adverse device effect as:
"any serious adverse effect on health or safety or any life-threatening problem or death caused by, or associated with, a device, if that effect, problem or death was not previously identified in nature, severity or degree of incidence in the investigational plan or application (including a supplementary plan or application), or any other unanticipated serious problem associated with a device that related to the rights, safety or welfare of the subject"
There were no adverse events during this study.
(2) Slit Lamp Findings:
A positive slit lamp finding is a routinely occurring complication that can be expected with or without the presence of contact lenses. The degree of severity may range from very slight, representing no medical concern, to serious, requiring medical treatment.
The following table shows the incidence of slit lamp findings at the initial and final visits.
| INCIDENCE OF SLIT LAMP FINDINGS | | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| FINDING FINDING | INITIAL ---------------------------------------------------------------------------------------------------------------------------------------------------------------------- | FINAL |
| | | |
| NO FINDINGSª | 90.5 | વેરે '8 |
| EDEMA® | 0.0 | 0.0 |
| NEOVASCUL. | 1.4 | 0.0 |
| STAINING | 5.4 | 3.2 |
| HYPEREMIA | 1.4 | 0.0 |
| PALPEBRAL | | |
| ABNORM. | 1.4 | 0.0 |
| OTHER | 0.0 | 0.0 |
| | | |
a Percent of eyes examined with no findings, initial or final scheduled follow-up visits only
b Percent of eyes with finding, initial or final scheduled follow-up visits only
#### (3) Symptoms, Problems and Complaints:
Symptoms, problems and complaints were reported by the investigators at each visit. "No symptoms" were reported at 76% of the scheduled follow up visits, and at all interim visits. The most frequent symptoms were of lens awareness and reading problems, followed by need for cleaning and dryness. The "other" symptoms in the population were: tired or fatigue (4 reports, 4 eyes of 2 subjects); cloudy or fogging (8 reports, 4 eyes of 2 subjects); scratchy sensation or itch (4 reports, 4 eyes of 2 subjects); difficult to remove (6 reports, 2 eyes of 1 subject).
{5}------------------------------------------------
#### SELECTED SYMPTOMS, PROBLEMS AND COMPLAINTS
| SYMPTOM | INCIDENCE |
|-------------------|-----------|
| NONEa | 75.7% |
| LENS AWARENESSb | 5.7% |
| READING PROBLEMS | 5.5% |
| NEED FOR CLEANING | 5.2% |
| DRYNESS | 4.6% |
- a Percent of eyes examined with no findings, regularly scheduled visits only b Percent of eyes with finding, regularly scheduled visits only
#### (4) Discontinuations:
Throughout the study, 1 subject (3%) was discontinued.
#### b. EFFICACY:
#### (1) Visual Acuity:
All eyes had a final visual acuity within 1 line of the initial best corrected acuity.
#### (2) Wear Time:
Wear time remained essentially unchanged over the three months of the study, indicating continuing comfort and cleanliness with the investigational lenses.
#### (3) Lens Replacements:
A total of 10 lenses were replaced during the study, 6 for visual acuity, 5 of the 6 within the first month. These results indicate that the lenses are effective in the correction of myopia.
#### Gender Comparisons:
The overall population was 73% female, 27% male. No further gender analysis was warranted.
### OVERALL CONCLUSION OF THE CLINICAL STUDY:
The data of the clinical trial demonstrate the safety and efficacy of the ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical Contact Lenses.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the upper half of the circle. Inside the circle is an abstract symbol of an eagle with three lines representing the wings and a stylized body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 6 2004
The Lagado Corporation c/o John M. Szabocsik, Ph.D. Szabocsik and Associates 203 North Wabash Avenue Suite 1200 Chicago, IL 60601
Re: K033599
KUSSJ77
Trade/Device Name: ONST™ – 56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HQD Dated: July 7, 2004 Received: July 8, 2004
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments, or to conniner of Price that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmeter for (10) and the device, subject to the general controls provisions of the Act. The I ou may, therefore, mains of the Act include requirements for annual registration, listing of general conable pro resuring practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassinos (tional controls. Existing major regulations affecting your device can may oe subject to back acceral Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Tease of advised that I Driseanter sever device complies with other requirements of the Act that 1 Dri has made a assoc regulations administered by other Federal agencies. You must of any I ederal statutes uncrements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set (21 CFR Part 607), laborning (21 CFR Part 820); and if applicable, the electronic forth in the quality of events (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
# Page 2 - John M. Szabocsik, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w you to organization of substantial equivalence of your device to a legally promance houried.com "consification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you active of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Kalipi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## 510(k) NUMBER (IF KNOWN) K033599
ONSI™-56 (onsifocon A) Rigid Gas Permeable Spherical, DEVICE NAME Aspheric, Toric and Bifocal Contact Lens
#### INDICATIONS FOR USE
The ONSIJ-56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism.
The lenses may be disinfected only by using chemical disinfection
Prescription Use
(Part 21CFR 801 Subpart D)
Over-The-Counter-Use OR (Part 21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE | |
|-------------------------------------------------------|--|
| Page 1 of 1 | |
| (Division Sign-Off) | |
| Division of Ophthalmic Ear, | |
Division of Ophthal
Nose and Throat Devises
510(k) Number K033599
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.