The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.
Device Story
System comprises indirect ophthalmoscope headpiece, headset laser enclosure, wireless user interface tablet, and footswitch. Physician wears headset to visualize target tissue; uses wireless tablet to set treatment parameters; targets tissue via visible red aiming beam. Physician presses footswitch to deliver therapeutic green laser light (520 nm) to tissue. Built-in eye safety filters protect physician. Used in clinical settings by ophthalmologists. Benefits include compact, lightweight, headset-mounted design for ophthalmic photocoagulation, iridotomy, and trabeculoplasty.
Clinical Evidence
No clinical data. Evidence consists of bench testing for electrical/mechanical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22), laser performance (IEC 60825-1, ANSI Z136-1/3), ophthalmic-specific requirements (ISO 15004-2), software verification/validation (IEC 62304), and human factors testing in simulated use environments.
Technological Characteristics
Green semiconductor (diode) laser, 520 nm wavelength, continuous wave mode. Headset-mounted form factor. Wireless tablet interface. Footswitch-controlled emission. Integrated eye safety filters. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 15004-2, ISO 14971, ISO 62366-1, and IEC 62304.
Indications for Use
Indicated for ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty in patients requiring such treatments.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Predicate Devices
Norlase LEAF Green Laser Photocoagulator (K190083)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 10, 2021
Norlase ApS % Sheila Pickering Consultant Sheila Pickering Consulting Group 2081 Longden Circle Los Altos, CA 94024
Re: K201623
Trade/Device Name: Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF, GEX
Dear Sheila Pickering:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter sent August 11, 2020. Specifically, FDA is updating this SE Letter because the original letter sent was missing the date of clearance. This is an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact LT Charles Chiang, OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 301-796-6459, Charles.Chiang@fda.hhs.gov.
Sincerelv.
# Charles Chiang -S
LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Norlase ApS % Sheila Pickering Consultant Sheila Pickering Consulting Group 2081 Longden Circle Los Altos, CA 94024
Re: K201623
Trade/Device Name: Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF, GEX Dated: July 6, 2020 Received: July 13, 2020
Dear Sheila Pickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer N. Brown -S
for LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K201623
Device Name
Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator
Indications for Use (Describe)
The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
## 510(k) Summary K201623 Prepared June 12, 2020
| Sponsor: | Norlase ApS<br>Brydehusvej 30<br>2750 Ballerup<br>Denmark |
|-----------------------|-------------------------------------------------------------|
| Contact Person: | Jan Forstberg |
| Telephone: | +4593977472 |
| Submission Date: | June 12, 2020 |
| Device Name: | Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator |
| Common Name: | Photocoagulator |
| Trade Name: | Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator |
| Classification: | |
| Regulatory Class: | II |
| Review Category: | Ophthalmology Photocoagulator<br>21CFR 886.4390 (HQF) |
| Classification Panel: | Ophthalmology |
#### A. Legally Marketed Predicate Devices
The primary predicate device is the Norlase LEAF Green Laser Photocoagulator (K190083). The reference predicate is the Iridex TruFocus LIO (K170718).
#### B. Device Description:
The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system was developed to utilize advancements in visible green semiconductor (diode) laser technology and to provide a compact design to be incorporated in a lightweight headset worn by the physician. The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system consists of an indirect ophthalmoscope headpiece, a headset laser enclosure/casing, a wireless user interface tablet, a footswitch with a battery, and a cable extending from the footswitch to the headset. The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system is controlled by the wireless user interface tablet that allows the physician to set the desired parameters for treatment. A shrouded footswitch is connected to the headset to control the emission of laser light and the battery to power the device illumination and laser source. The physician uses the headset to locate the area to be treated and after setting the appropriate parameters, targets the desired tissue
{5}------------------------------------------------
utilizing a visible red aiming beam. The foot switch is pressed by the physician to deliver therapeutic green laser light to the targeted tissue. Eye safety filters are built into the laser unit to protect the physician's eyes while providing a clear view of the target area.
#### C. Intended Use / Indications for Use
The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy and trabeculoplasty.
## D. Substantial Equivalence
The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system is substantially equivalent to the original device. the Norlase Leaf Green Laser Photocoagulator cleared per K190083, as well as the Iridex TruFocus LIO (K170718) They are both laser photocoagulators for the purpose of Ophthalmology treatment. The indications for use and the core technology of the primary Norlase predicate have not changed. The core difference is the change of device configuration from a stationary slit lamp mounted device to a headset mounted LIO configuration. The reference predicate has a broader indication for use for multiple applications but has substantially equivalent technological characteristics.
| Device Name | Primary Predicate<br>Norlase LEAF<br>Photocoagulator<br>K190083 | Reference Predicate<br>Iridex TruFocus LIO<br>K170718 | Subject Device<br>Norlase Laser<br>Indirect<br>Ophthalmoscope<br>(LIO)<br>Photocoagulator | SE |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Indications for<br>Use | Intended to be used<br>in ophthalmic laser<br>procedures<br>including retinal<br>and macular<br>photocoagulation,<br>iridotomy, and<br>trabeculoplasty. | The Iridex TruFocus LIO<br>Premiere Laser Indirect<br>Ophthalmoscope with the<br>Family of Iridex® IQ Laser<br>Systems is intended for soft<br>and fibrous tissue, including<br>osseous tissue incision,<br>excision, coagulation,<br>vaporization, ablation and<br>vessel hemostasis in the<br>medical specialties of,<br>dermatology, ear, nose and<br>throat (ENT)/<br>otolaryngology, and<br>ophthalmology | Intended to be<br>used in ophthalmic<br>laser procedures<br>including retinal<br>and macular<br>photocoagulation,<br>iridotomy, and<br>trabeculoplasty. | Yes |
| Manufacturer | Norlase | Iridex Corp. | Norlase | N/A |
| 510(k) number | K190083 | K170718 | TBD | N/A |
| Regulation<br>medical<br>specialty | Ophthalmic | General and Plastic Surgery | Ophthalmic | YES |
| Review panel | Ophthalmic | Ophthalmic | Ophthalmic | YES |
| Product code | HOF | GEX | HOF | YES |
{6}------------------------------------------------
| Regulation number | 886.4390 | 878.4810 | 886.4390 | YES |
|------------------------|----------------------------|----------------------------|----------------------------|-----|
| Regulation description | Ophthalmic photocoagulator | Ophthalmic photocoagulator | Ophthalmic photocoagulator | YES |
| Classification | II | II | II | YES |
| <b>Treatment beam</b> | | | | |
| Laser type | Laser diode | Laser diode | Laser diode | YES |
| Wavelength | 520 nm | 532mm | 520 nm | YES |
| Laser mode | Continuous wave | Continuous wave | Continuous wave | YES |
Based on the comparison of indications for use and technological characteristics, the subject device is substantially equivalent to the primary predicate device as well as the reference predicate. Based on the performance data provided in the submission the differences do not introduce new issues related to safety and efficacy.
# E. Performance Data
Every specification of the Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system has been verified and validated as required by the risk analysis. All design verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met.
| ISO 14971 | Application of risk management to medical devices |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 15004-2 | Ophthalmic instruments - Fundamental requirements and test methods -<br>Part 1: Light hazard protection |
| IEC 60601-1 | Medical electrical equipment- General requirements for basic safety and<br>essential performance |
| IEC 60601-1-2 | Electromagnetic disturbances |
| IEC 60601-2-22 | Medical electrical equipment - Part 2: Particular requirements for the<br>safety of diagnostic and therapeutic laser equipment |
| IEC 60825-1 | Safety of laser products - Part 1: Equipment classification, and<br>requirements [Including: Technical Corrigendum 1 (2008),<br>Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)] |
| IEC 62304 | Medical device software - Software life-cycle processes |
| IEC 62366-1 | Application of usability engineering to medical devices |
| ANSI Z136-1 | American National Standard for Safe Use of Lasers |
| ANSI Z136-3 | American National Standard for Safe Use of Lasers in Health Care |
The verification and validation testing included testing to the following applicable standards:
Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30. Norlase certifies that all verification and validation activities provided in this submission were performed by designated individuals and
{7}------------------------------------------------
results demonstrate that predetermined acceptance criteria were met. Successful results for the following tests were included in the submission as performance data supporting substantial equivalence:
- 1. Bench testing for electrical and mechanical safety in compliance with the standards cited above
- Bench testing for laser performance in compliance with the laser standards cited above 2.
- Bench testing specific to ophthalmic devices in compliance with the standard cited above 3.
- 4. Software testing, consisted of verification and validation testing in compliance with ISO 62304, including test cases related to off the shelf software, as well as cybersecurity features
- 5. Human factors testing to demonstrate usability in a simulated use environment when used by health care professionals
Clinical data was not required for this type of device.
#### F. Conclusion
Potential risks were identified according to the ISO 14971 and ISO 62366 Standards. The risks were analyzed with regard to risk/benefit category and mitigations were implemented and tested as part of the performance testing described above. All risk mitigations were satisfactorily verified and validated. Where there were technological differences from the predicates, these were shown not to result in any new issues of safety or efficacy according to the performance data submitted.
Therefore, the Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator is substantially equivalent to the predicate devices with regards to intended use and technological characteristics. Results of performance testing demonstrated that the device met the design requirements and as well as the user needs.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.