K181682 · Carl Zeiss Meditec, AG · HQF · Mar 6, 2019 · Ophthalmic
Device Facts
Record ID
K181682
Device Name
VISULAS green
Applicant
Carl Zeiss Meditec, AG
Product Code
HQF · Ophthalmic
Decision Date
Mar 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye including trabeculoplasty and iridotomy.
Device Story
VISULAS green is a 532 nm frequency-doubled solid-state ophthalmic laser system. It delivers thermal energy to ocular tissue via single-spot or multi-spot modes to perform photocoagulation. The system comprises a laser console, touch control panel, foot switch, instrument table, and various applicators (laser slit lamp, laser indirect ophthalmoscope, endoprobes). Operated by clinicians in clinical settings, the device allows for precise control of laser power (50-1500 mW) and pulse duration (10-2500 ms). The output is used to treat ocular diseases by inducing thermal coagulation. The device provides therapeutic benefits by enabling targeted tissue treatment in the eye.
Clinical Evidence
No clinical data, animal testing, or bench testing were required to establish substantial equivalence. The device relies on software verification and validation testing and compliance with recognized safety standards (ANSI/AAMI ES60601-1, IEC 60601-1-2, ANSI Z80.36, IEC 60825-1, IEC 60601-2-22, IEC 62133).
Technological Characteristics
532 nm frequency-doubled solid-state laser; 50-1500 mW power; 10-2500 ms pulse duration; 50-1000 μm spot size. Includes laser console, touch control panel, foot switch, and applicators (LSL, VISULINK 532/U, Endoprobes, LIO). Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, ANSI Z80.36, IEC 60825-1, IEC 60601-2-22, and IEC 62133.
Indications for Use
Indicated for photocoagulation of ocular tissues in the posterior and anterior segments of the eye, including trabeculoplasty and iridotomy.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Predicate Devices
VISULAS 532s Laser System with the VITE option (K100035)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 6, 2019
Carl Zeiss Meditec AG % Lisa Granev Senior Consultant Biologics Consulting Group, Inc. 1555 King Street. Suite 300 Alexandria, VA 22314
Re: K181682
Trade/Device Name: VISULAS green Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: January 31, 2019 Received: January 31, 2019
Dear Lisa Graney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely yours, Alexander Beylin -S 2019.03.06 10:50:35 -05'00' for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181682
Device Name VISULAS green
Indications for Use (Describe)
VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye including trabeculoplasty and iridotomy.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92), the 510(k) Summary for VISULAS green is provided below.
#### 1. SUBMITTER
| Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>D-07745 Jena<br>Germany<br>Phone: +49 3641 2200 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Dr. Christian Münster<br>Director Regulatory and Clinical Affairs<br>Phone: +49 7364 20 6985<br>christian.muenster@zeiss.com |
| Submission Correspondent: | Lisa Graney<br>Senior Consultant<br>Biologics Consulting Group<br>1555 King Street, Suite 300<br>Alexandria, VA 22314<br>lgraney@biologicsconsulting.com)<br>Phone: (571) 777-9518<br>Fax: (703) 548-7457 |
| Date Prepared: | June 25 2018 |
### 2. DEVICE
| Device Trade Name: | VISULAS green |
|---------------------|-----------------------------------|
| Device Common Name: | Laser, Ophthalmic |
| Classification Name | 21 CFR 886.4390, Ophthalmic Laser |
| Regulatory Class: | Class II |
| Product Code: | HQF |
### 3. PREDICATE DEVICE
Predicate Device:
VISULAS 532s Laser System with the VITE option (K100035)
{4}------------------------------------------------
### DEVICE DESCRIPTION 4.
VISULAS green is an ophthalmic laser with a 532 nm wavelength used for standard photocoagulation treatments of ocular tissue. VISULAS green is operated in single-spot mode or in multi-spot mode.
VISULAS green consists of a laser console, touch control panel, optional laser light applicators such as laser slit lamp or laser indirect ophthalmoscope, foot switch and instrument table. The device can also be used with various accessories, such as laser endoprobes, or the SL cam 5.0.
### ડ. INTENDED USE/INDICATIONS FOR USE
VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye, including trabeculoplasty and iridotomy.
### TECHNOLOGICAL COMPARISON 6.
A technological comparison between the subject and predicate devices is shown in Table 1. The cited predicate device is VISULAS 532s Laser System with the VITE option (herein after called VISULAS 532s), cleared under K100035.
{5}------------------------------------------------
| | VISULAS green | VISULAS 532s |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | Subject Device | K100035 |
| Legal manufacturer | Carl Zeiss Meditec AG | Carl Zeiss Meditec AG |
| Indications for Use Statement | VISULAS green is intended for use in photocoagulating ocular<br>tissues in the treatment of diseases of the eye, including<br>trabeculoplasty and iridotomy. | The VISULAS 532s Laser System with the VITE option is<br>intended for use in retinal, pan-retinal, focal and grid<br>photocoagulation of ocular tissues in the treatment of diseases of<br>the eye including:<br>-Proliferative and non-proliferative diabetic retinopathy<br>-Macular edema<br>-Branch and central retinal vein occlusion<br>-Lattice degeneration<br>-Retinal tears and detachments |
| | General Laser Specifications | |
| Laser type | solid state laser, frequency-doubled | solid state laser, frequency-doubled |
| Wavelength | 532 nm | 532 nm |
| Power (cw laser) | 50 to 1500 mW | 50 to 1500 mW |
| Length of pulse | 10 - 2500 ms and cw | 10 - 2500 ms and -cw |
| Repeat Mode | Yes | Yes |
| Laser spot size settings | 50 μm to 1000 μm | 50 μm to 1000 μm |
| Aiming beam | 620-650 nm | 620-650 nm |
| Types of laser applicators | LSL, VISULINK 532/U, Endoprobes, LIO | LSL, VISULINK 532/U, Endoprobes, LIO |
| Multi spot treatment mode | Yes | Yes |
| Conditions of Use | | |
| Site in the body | Eye | Eye |
| Principle of operation | Photocoagulation of ocular tissue | Photocoagulation of ocular tissue |
### Table 1. Technological characteristics comparison table of VISULAS green and VISULAS 532s
{6}------------------------------------------------
VISULAS green and the predicate device are intended as ophthalmic lasers for photocoagulation. The indications for use are within the same intended use as the predicate device and do not raise different questions of safety and effectiveness.
VISULAS green and the predicate laser system have minor parameter differences but share the same fundamental principle of operation:
Photocoagulation: Continuous wave operation in which thermal energy created by the absorption of the laser energy by ocular tissue causes coagulation.
VISULAS green is substantially equivalent to the predicate laser system presented in the 510(k) premarket notification with respect to indications for use and technological characteristics. Differences between the subject device and the predicate device do not raise any new issues of safety or effectiveness.
### 7. PERFORMANCE DATA
# Biocompatibility Testing
The materials and manufacturing of the subject device is identical to the predicate device and currently marketed devices. Therefore, no new biocompatibility testing is needed for this device.
# Laser safety, electrical safety and electromagnetic compatibility (EMC)
VISULAS green was evaluated against the following requirements and was found to comply with:
- ANSI/AAMI ES60601-1:2005/(R) 2012 ●
- IEC 60601-1-2:2014 ●
- ANSI Z80.36:2016 ●
- . IEC 60825-1:2007
- IEC 60601-2-22:2012
- IEC 62133: 2012 ●
# Software Verification and Validation Testing
The software of this device is considered as a "Major" level of concern. Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification and validation of VISULAS green demonstrated that the product works as designed.
# Bench Testing
Not Applicable. Bench testing was not necessary to establish the substantial equivalence of this device.
{7}------------------------------------------------
# Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
# Clinical Data
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
### CONCLUSION 8.
The VISULAS green is substantially equivalent to the predicate device, VISULAS 532s (K100035).
The VISULAS green is similar in technological characteristics, performance, principles of operation (photocoagulation) and has similar indications for use as the predicate device. Any differences between the proposed device and the predicate device do not raise any new issues of safety or effectiveness. Thus VISULAS green is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.