The NGL is indicated for use in photocoagulation of both anterior and posterior segments of the eye including: - Retinal Photocoagulation, panretinal photocoagulation and intravitreal . endophotocoagulation of vascular and structural abnormalities of the retina and choroid including: - Proliferative and nonproliferative retinopathy (including diabetic); । - I romeral re and many to age-related macular to age-related macular degeneration; ー - Retinal tears and detachments; - - Macular Edema; - - Retinopathy of prematurity; - - Choroidal neovascularization; - - Leaking microaneurysms. - Iridotomy/Iridectomy for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG), Acute Angle Closure Glaucoma (AACG), and Refractory Glaucoma. - Trabeculoplasty for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG) and Refractory Glaucoma. - And other laser treatments including: . - Internal sclerostomy; - - Lattice degeneration; - - Central and Branch Retinal Vein Occlusion; - - Suturelysis; - - Vascular and pigmented skin lesions. -
Device Story
Diode-pumped Neodymium-doped solid state laser; generates 532 nm green laser beam via frequency-doubled crystal; includes 635 nm diode aiming beam. Operated by physician via footswitch; laser energy delivered to patient eye via slit lamp, laser probe, or laser indirect ophthalmoscope (LIO). Laser engine module converts diode excitation of Neodymium atoms into amplified light via stimulated emission; frequency doubled by non-linear optical crystal. Used in clinical settings for ophthalmic photocoagulation; output enables treatment of vascular/structural ocular abnormalities. Benefits include precise tissue coagulation for retinal/glaucoma conditions.
Clinical Evidence
Bench testing only. Compliance with 21 CFR 1040 and international standards (ISO 14971, ISO 10993 series, EN 60601 series, IEC 60825-1) demonstrated. No clinical data provided.
Technological Characteristics
Diode-pumped Neodymium-doped solid state laser; 532 nm therapeutic beam; 635 nm diode aiming beam. Non-linear optical crystal for frequency doubling. Console non-sterile; laser probes sterile (EtO, 10^-6 SAL). Packaging: Tyvek lidding on HIPS trays. Complies with EN 60601-1, EN 60601-1-2, EN 60601-2-22, EN 60601-1-4, UL 60601-1, IEC 60825-1.
Indications for Use
Indicated for patients requiring photocoagulation of anterior/posterior eye segments, including treatment for diabetic/non-diabetic retinopathy, macular degeneration, retinal tears/detachments, macular edema, retinopathy of prematurity, choroidal neovascularization, microaneurysms, glaucoma (COAG, POAG, AACG, refractory), internal sclerostomy, lattice degeneration, retinal vein occlusion, suturelysis, and vascular/pigmented skin lesions.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
## 5. 510(K) SUMMARY
This summary is in accordance with 21 CFR 807.92(c).
KO 62624
The submitter of the 510(k) is:
Terry J. Dagnon . Director, Regulatory Affairs Alcon, Inc. 6201 South Freeway Fort Worth, Texas 76134 Phone: (817) 551-4325 Fax: (817) 551-4630
NOV -- 5 2007
Date Prepared: August 31, 2006
Device Subject to this 510(k):
Next Generation Laser Trade Name: Common Name: Ophthalmic Laser or, Frequency Doubled (532nm) Laser Classification Name: Laser, Ophthalmic (21 CFR 886.4390)
#### 5.1. Predicate Devices
510(k) Number Device
| K962592 | Alcon Eyelite® |
|---------|------------------------|
| K041598 | Laserex™ Solitaire |
| K031665 | Iris Medical Oculight™ |
#### 5.2. Device Description
The NGL Ophthalmic Laser, like its predicate device Alcon Ophthalas® 532 EyeLite® (EyeLite®), is a diode pumped Neodymium-doped solid state laser designed for ophthalmic use. LASER is an acronym for "Light Amplification by Stimulated Emission of Radiation." This laser delivers a visible 532 nm green laser beam (frequency doubled), and a visible 635 mm Diode Laser aiming beam (635 nm is an approximate value between 630-640 nm).
The laser firing is actuated by footswitch. The laser is delivered to the patient through any of the following accessories (a) Slit Lamp, (b) Laser Probe or (c) Laser Indirect Ophthalmoscope (LIO).
In case of the NGL, like its predicate device Alcon Ophthalas® 532 EyeLite®, the laser beam is generated when a laser diode beam excites the Neodymium atoms in the rod material. When one atom which has been excited by the beam returns to its initial stable level, the energy difference between the two states is emitted as radiation in the form of a photon. When this photon meets another excited atom, emission of a second photon occurs. The second photon has the same phase, wavelength, and direction as the first photon. The light emitted in this manner oscillates between two end mirrors. The light is amplified by this
{1}------------------------------------------------
stimulated emission process and a laser beam is produced. This laser beam is frequency doubled in a special crystal. The crystal is an optical dielectric that exhibits a non-linear optical response. The 532 nm wavelength is produced by harmonic generation. These steps leading to generation of the output laser beam generation is done in the laser engine module of the device.
A red 635 nm laser diode provides the visible aiming beam in NGL.
## 5.3. Indications for Use
The NGL is indicated for use in photocoagulation of both anterior and posterior segments of the eye including:
- Retinal Photocoagulation, panretinal photocoagulation and intravitreal . endophotocoagulation of vascular and structural abnormalities of the retina and choroid including:
- Proliferative and nonproliferative retinopathy (including diabetic); ।
- Choroidal neovascularization secondary to age-related macular degeneration; -
- Retinal tears and detachments; -
- Macular Edema; -
- Retinopathy of prematurity; -
- Choroidal neovascularization; -
- Leaking microaneurysms. -
- Iridotomy/Iridectomy for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG), Acute Angle Closure Glaucoma (AACG), and Refractory Glaucoma.
- Trabeculoplasty for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG) and Refractory Glaucoma.
- And other laser treatments including: .
- Internal sclerostomy; -
- Lattice degeneration; -
- Central and Branch Retinal Vein Occlusion; -
- -Suturelysis;
- Vascular and pigmented skin lesions. -
#### 5.4. Brief Summary of Nonclinical Tests and Results
The device will comply with applicable sections of 21 CFR 1040 and the following standards:
| Standard # | Title |
|------------------------------|-------------------------------------------------------------------------------------|
| 11135 AAMI/ISO | Medical Devices – Validation and Routine Control of Ethylene<br>Oxide Sterilization |
| 14971 ISO | Medical Devices: Application of Risk Management to Medical<br>Devices |
| 10993-1 AAMI /<br>ANSI / ISO | Biological evaluation of medical devices -- Part 1: Evaluation and<br>testing |
| 10993-5 AAMI / | Biological evaluation of medical devices -- Part 5: Tests for In Vitro |
{2}------------------------------------------------
| Standard # | Title |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI / ISO | cytotoxicity |
| 10993-7 AAMI /<br>ANSI / ISO | Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide<br>Sterilization Residuals |
| 10993-10 AAMI /<br>ANSI / ISO | Biological evaluation of medical devices -- Part 10: Tests for<br>irritation and sensitization |
| 10993-11 AAMI /<br>ANSI / ISO | Biological Evaluation of Medical Devices - Part 11: Tests for<br>systemic toxicity |
| EN 60601-1: 1990 | Medical Electrical Equipment, Part 1 - General Requirements for<br>Safety. (Including A1:1992, A2:1995 and A13:1995) |
| EN 60601-1-2: 2001 | Medical electrical equipment Part 1: General requirements for safety<br>2. Collateral Standard: Electromagnetic compatibility-Requirements<br>and test. |
| EN 60601-2-22: 1996 | Medical electrical equipment Part 2: Particular requirements for the<br>safety of diagnostic and therapeutic laser equipment |
| EN 60601-1-4:1996 | Medical Electrical Equipment, Part 1: General Requirements for<br>Safety. 4. Collateral standard: Programmable electrical medical<br>systems. (Including A1: 1999) |
| UL 60601-1: 2003 | Medical Electrical Equipment, Part 1 - General Requirements for<br>Safety |
| IEC 60825-1:1993 | Radiation safety of laser products, equipment classification,<br>requirements and user's guide. (Including A1:1997 and A2:2001) |
Technological characteristics affecting clinical performance are similar to that of predicate devices previously listed. The Next Generation Laser system will be developed and manufactured in compliance with FDA and ISO quality system requirements. Testing will demonstrate that the functional requirements have been met and that the system specifications have been met prior to commercial product release.
The NGL console is provided non-sterile, and is not intended to be sterilized.
The Laser Probes are provided sterile. They are intended for single use.
Sterilization method for the Laser Probes will be a validated EtO sterilization cycle.
The Laser Probes will utilize the same packaging concept as used previously for similar products of adhesive backed Tyvek lidding sealed to HIPS (high impact polystyrene) trays. These packs can withstand validated sterilization processes of EtO sterilization to achieve a 10 6 SAL (sterility assurance level).
## 5.5. Trademark Reference
Oculight is a registered trademark of Iris Medical Laserex is a registered trademark of Ellex Medical Pty. Ltd.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alcon Research, Ltd. % Terry J. Dagnon Director, Regulatory Affairs 6201 South Freeway Fort Worth, Texas 76134
NOV - 5 2007
Re: K062624
Trade/Device Name: Next Generation Laser Regulation Number: 21 CFR 886.4390 Regulation Name: Opthalmic laser Regulatory Class: II Product Code: HQF Dated: August 15, 2007 Received: August 16, 2007
Dear Terry Dagnon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Terry J. Dagnon
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely ve
follon Dr
Mark N. Melkerson
Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4. INDICATIONS AND USE STATEMENT
Page 1 of 1
510(k) Number (if known): K062624
Device Name: Next Generation Laser
Indications for Use:
The NGL is indicated for use in photocoagulation of both anterior and posterior segments of the eye including:
- Retinal Photocoagulation, panretinal photocoagulation and intravitreal . endophotocoagulation of vascular and structural abnormalities of the retina and choroid including:
- Proliferative and nonproliferative retinopathy (including diabetic); ।
- I romeral re and many to age-related macular to age-related macular degeneration; ー
- Retinal tears and detachments; -
- Macular Edema; -
- Retinopathy of prematurity; -
- Choroidal neovascularization; -
- Leaking microaneurysms.
- Iridotomy/Iridectomy for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG), Acute Angle Closure Glaucoma (AACG), and Refractory Glaucoma.
- Trabeculoplasty for treatment of Chronic/Primary Open Angle Glaucoma . (COAG,POAG) and Refractory Glaucoma.
- And other laser treatments including: .
- Internal sclerostomy; -
- Lattice degeneration; -
- Central and Branch Retinal Vein Occlusion; -
- Suturelysis; -
- Vascular and pigmented skin lesions. -
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | <b>X</b> | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|-------------------------------------------------|----------|--------|------------------------------------------------|
|-------------------------------------------------|----------|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign Off)
Division of General, Restorative,
Concurrence of CDRH, Office of Device Evaluation (ODE)
and Neurological Devices
| 510(k) Number | 11062624 |
|---------------|----------|
|---------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.