K111446 · Abbott Medical Optics, Inc. · HQC · Dec 21, 2011 · Ophthalmic
Device Facts
Record ID
K111446
Device Name
SOVEREIGN COMPACT PHACOEMULSIFICATION SYSTEM
Applicant
Abbott Medical Optics, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Dec 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sovereign Compact Phacoemulsification System is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
Device Story
Modular ophthalmic microsurgical system for anterior segment cataract surgery; facilitates lens disruption and extraction. System console powers and controls ultrasonic phacoemulsification, diathermy, irrigation/aspiration, and vitrectomy. Inputs include user settings via key panel, remote control, or footpedal; ultrasonic energy delivered via compatible handpiece (e.g., 690880). System transforms electrical energy into ultrasonic vibration at the handpiece tip to emulsify lens nuclei. Output includes vacuum/pressure (mmHg, kPa) and fluidic control. Used in surgical settings by ophthalmologists. Enhancements include optional feature-activated handpiece compatibility, Surgical Media Center (SMC) support, and multilingual support. Benefits include improved surgical control and efficiency during cataract removal.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed that functional requirements and specifications were met, demonstrating performance equivalent to predicate devices in phacoemulsification, irrigation, diathermy, and vitrectomy modes.
Technological Characteristics
AC-powered modular console with active color graphic user interface. Features peristaltic aspiration pump, ultrasonic phacoemulsification, diathermy, irrigation/aspiration, and vitrectomy. Connectivity includes receptacles for footpedal, remote control, and automated IV pole. Software supports Surgical Media Center (SMC) and multilingual interface. Handpiece 690880 utilizes ultrasonic vibration for lens disruption.
Indications for Use
Indicated for use in cataract surgery to disrupt a cataractous lens with ultrasound and extract the cataract in patients requiring such surgical intervention.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
DEC 2 1 2011
### SOVEREIGN COMPACT PHACOEMULSIFICATION SYSTEM K111446
# 4. 510(K) SUMMARY
This 510(k) summary is being submitted in accordance with the Medical Device Amendments of 1976, the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92.
Applicant:
Abbott Medical Optics Inc. 1700 E. St. Andrew Place P.O. Box 25162 Santa Ana, CA 92799-5162, USA
Contact Person:
Rosanne M. Yetemian, PhD, MSRS Regulatory Affairs Specialist 1700 E. St. Andrew Place Santa Ana, CA 92705 Tel: (714) 247-8282 Fax: (714) 247-8487 Email: rosanne.yetemian@amo.abbott.com
R.H
Date of 510(k) Summary Preparation:
November 28, 2011
Device that is the subject of this notification:
| Trade/Proprietary Name: | The Sovereign Compact Phacoemulsification System |
|-------------------------|--------------------------------------------------|
| Classification Name: | Phacofragmentation System |
| Regulation Number: | 21 CFR 886.4670 |
| Regulation Name: | Phacofragmentation system |
| Regulatory Class: | Class II |
| Product Code: | HQC |
The devices to which substantial equivalence is claimed:
| Table 4-1: Predicate Devices to Which Substantial Equivalence is Claimed | | | | |
|--------------------------------------------------------------------------|----------------------|---------------|---------------|----------------|
| Predicate Device Name | Predicate Trade Name | 510(k) Holder | 510(k) Number | Clearance Date |
| Sovereign Compact Phacoemulsification System Predicate Devices | | | | |
| Mojave Cataract<br>Extraction System | Sovereign Compact | AMO | K003638 | 02/13/2001 |
| AMO Ophthalmic<br>Surgical System | WhiteStar Signature | AMO | K060366 | 04/07/2006 |
| AMO Ultrasonic Handpiece 690880 Predicate Device | | | | |
| Sovereign Phaco<br>Handpiece | Sovereign Handpiece | AMO | K981116 | 05/19/1998 |
ABBOTT MEDICAL OPTICS INC.
510(k) SUMMARY
{1}------------------------------------------------
#### SOVEREIGN COMPACT PHACOEMULSIFICATION SYSTEM K111446
#### 4.1 DEVICE DESCRIPTION SUMMARY
The Sovereign Compact Phacoemulsification System) is a modular ophthalmic microsurgical system that facilitates anterior segment (cataract) surgery for the disruption and extraction of a cataractous lens. As with other phacofragmentation systems, The System is identified in 21 CFR 886.4670 as a "phacofragmentation system that is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract."
The main performance functions of The System include ultrasonic phacoemulsification, diathermy, irrigation and aspiration, and vitrectomy. Ultrasonic phacoemulsification provides vibration at an ultrasonic frequency when used in conjunction with a compatible AMO ultrasonic handpiece. The System is controlled and powered by The System console, which includes an active color graphic user interface and receptacles for an automated IV pole, footpedal, remote control, and drainage packs.
The materials, basic scientific concepts, physical properties and intended use of The System are identical to those of The Mojave Cataract Extraction System (branded as Sovereign Compact) (K003638) predicate device. The AMO Ophthalmic Surgical System was also listed as a predicate because it shares the same available software modes as The System. Both predicate devices are manufactured by Abbott Medical Optics Inc. (AMO).
The System will be marketed as either a complete unit or an optional software upgrade kit for current Sovereign Compact Systems in the field.
In addition, the enhancements made to The System include an optional feature-activated capability for the use of the AMO Ultrasonic Handpiece 690880 (Handpiece 690880), which is substantially equivalent to the Sovereign Phaco Handpiece cleared under the Sovereign Cataract Extraction System 510(k) (K981116). Handpiece 690880 is intended for use during the phacoemulsification procedure to break up (emulsify) the nucleus of the cataractous lens and remove the remaining nuclear fragments.
#### 4.2 INDICATIONS FOR USE
The Sovereign Compact Phacoemulsification System is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
#### 4.3 TECHNOLOGICAL CHARACTERISTICS
For anterior segment (cataract) surgical procedures, the system provides the following main diathermy, phacoemulsification, irrigation and aspiration, and vitrectorny. The modes: technological characteristics of the device are substantially equivalent to those cleared under the 510(k)s of the Mojave Cataract Extraction System (K003638) and the AMO Ophthalmic Surgical System (K060366). The technological characteristics of the AMO Ultrasonic Handpiece are substantially equivalent to those of the Sovereign Phaco Handpiece cleared under 510(k) of the Sovereign Cataract Extraction System (K981116).
The System, which is the subject device, is an upgrade to the existing predicate device, the Mojave Cataract Extraction System (branded as Sovereign Compact and cleared on February 13, 2001 under K003638). Since the clearance of this device, The System has been upgraded
{2}------------------------------------------------
to include software modes that are also available with the second predicate, the AMO Ophthalmic Surgical System (K060366).
The differences between The System and its two predicate devices include:
- AMO Ultrasonic Handpiece 690880 compatibility 0
- Support for the latest version of the Surgical Media Center (SMC) o
- The ability to view vacuum units in kilopascals (kPa) o
- Support for an IV pole extension o
A comparison summary of The System and the two predicate devices is provided in Table 4-2. All unique features available with The System with respect to both predicates are underlined for clarity.
| Features and<br>Characteristics | Proposed Device<br>The Sovereign Compact<br>Phacoemulsification<br>System | Predicate Device<br>Mojave Cataract<br>Extraction System<br>(K003638) | Predicate Device<br>AMO Ophthalmic-<br>Surgical System<br>(K060366) |
|---------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------|
| Intended Use | Anterior Segment<br>Ophthalmic Surgery | Anterior Segment<br>Ophthalmic Surgery | Anterior and Posterior<br>(optional) Segment<br>Ophthalmic Surgery |
| Technological<br>Characteristics | | | |
| User Interface | Key panel, remote<br>control, footpedal | Key panel, remote<br>control, footpedal | Touch screen, remote<br>control, footpedal |
| Hardware:<br>-Console | Compact system console<br>with display screen and<br>key entry | Compact system console<br>with display screen and<br>key entry | Upper-Tier system<br>console with display<br>screen and key entry |
| -Footpedal | Wired<br>(closed or open-toe) | Wired<br>(closed toe) | Wired or wireless<br>(open-toe) |
| -Remote control unit | Wired | Wired | Wireless |
| -Console stand | Cart with mayo tray and<br>articulating arm | Cart with mayo tray and<br>articulating arm | Cart with mayo tray and<br>articulating arm |
| -IV Pole | Yes<br>(with optional extension) | Yes | Yes |
| Software:<br>-OS for Safety Critical<br>Functions | Same | Same | Same |
| Available Modes: | Same | No | Same |
| Specifications:<br>-Aspiration Pump Type | Peristaltic | Peristaltic | Peristaltic and Venturi |
| - Handpiece 690880<br>compatibility | Yes | No | No |
| -SMC Support | Yes | No | No |
| -Multilingual support | Yes | No | Yes |
| -Vacuum/pressure | mmHg, kPa | mmHg | mmHg |
#### Table 4-2 Comparison Overview of The System to the Predicate Devices
{3}------------------------------------------------
#### SOVEREIGN COMPACT PHACOEMULSIFICATION SYSTEM K111446
| Features and<br>Characteristics<br>Accessories | Proposed Device<br>The Sovereign Compact<br>Phacoemulsification<br>System | Predicate Device<br>Mojave Cataract<br>Extraction System<br>(K003638) | Predicate Device<br>AMO Ophthalmic<br>Surgical System<br>(K060366) |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Disposable Accessories | • Tips and Sleeves<br>• Fluidics Packs<br>• Administration Set<br>• Sterilization Tray<br>• Phaco Handpieces | • Tips and Sleeves<br>• Fluidics Packs<br>• Administration Set<br>• Sterilization Tray<br>• Phaco Handpieces | • Tips and Sleeves<br>• Fluidics Packs<br>• Administration Set<br>• Sterilization Tray<br>• Phaco Handpieces |
| Reusable Accessories | • Fluidics Pack<br>• Irrigation/Aspiration<br>Handpiece<br>• Diathermy Cord with<br>Bipolar Forceps<br>• Vitrectomy Handpiece | • Fluidics Pack<br>• Irrigation/Aspiration<br>Handpiece<br>• Diathermy Cord with<br>Bipolar Forceps<br>• Vitrectomy Handpiece | • N/A (Disposable Only)<br>• Irrigation/Aspiration<br>Handpiece<br>• Diathermy Cord with<br>Bipolar Forceps<br>• Vitrectomy Handpiece |
#### 4.4 SUMMARY OF NON-CLINICAL. TESTS
The System and Handpiece 690880 have undergone testing and are in compliance with applicable safety standards. The subject device was found to perform equivalently to the predicate device during the following modes of anterior segement ophthalmic surgery: phacoemulsification, irrigation, diathermy and vitrectomy. Therefore, the subject device and the predicate devices have similar safety, effectiveness, and performance profiles.
#### 4.5 SUMMARY OF CLINICAL TESTS
No clinical studies were deemed necessary to determine the safety and effectiveness or substantial equivalence of The System and Handpiece 690880 to their predicate devices.
#### 4.5 CONCLUSIONS
The technological characteristics that determine the functionality and performance of The Sovereign Compact Phacoemulsification System and the AMO Ultrasonic Handpiece 690880 are substantially equivalent to those of the predicate devices listed. The Sovereign Compact Phacoemulsification System and the AMO Ultrasonic Handpiece will be manufactured in . compliance with FDA and ISO quality systems requirements. The data presented from the nonclinical tests demonstrate that the device is safe and effective, and performs as safely and effectively as the legally marketed predicate devices. Validation and verification demonstrates that the functional requirements and specifications have been met prior to commercial release.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Abbott Medical Optics, Inc. c/o Rosanne Yetemian, PhD, MSRS Regulatory Affairs Specialist 1700 E. St. Andrew Place Santa Ana, CA 92705
DEC 2 1 2011
Re: K111446
Trade/Device Name: Sovereign Compact Phacoemulsification System Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation system Regulatory Class: Class II Product Code: HQC Dated: December 19, 2011 Received: December 20, 2011
Dear Dr. Yetemian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. Rosanne Yetemian
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K111446
Device Name: The Sovereign Compact Phacoemulsification System
Indications For Use:
The Sovereign Compact Phacoemulsification System is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
CLC
(Division Sign-Off) (Division Sign-Off)
Division of Ophthalmic, Neurological as
The Cources of Devices Ear. Division of OF
Nose and Throat Devices
K118146
510(k) Number
Sovereign Compact Phacoemulsification System Abbott Medical Optics Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.