K080803 · American Optisurgical, Inc. · HQC · Sep 22, 2008 · Ophthalmic
Device Facts
Record ID
K080803
Device Name
VISUAL PHACOEMULSIFICATION SYSTEM, MODEL 551
Applicant
American Optisurgical, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Sep 22, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vizual™ Phacoemulsification System is intended for use in the disruption and extraction of cataractous lens material from the eye with ultrasonic energy, a process known as phacoemulsification.
Device Story
Vizual™ Phacoemulsification System is an anterior segment surgical system for cataract removal. It provides irrigation, ultrasonic (U/S) phacoemulsification, irrigation/aspiration (I/A), vitrectomy, and coagulation. System includes phaco unit, U/S handpiece, tubing cartridge, and foot pedal; compatible with standard phaco tips, sleeves, vit cutters, and cautery tools. Features non-invasive vacuum sensing with external tubing cartridge that vents to bottle to minimize contamination. Operated via LCD touch panel; supports up to 24 user profiles. Optional cart with electric I/V pole allows software-controlled bottle height. Used by surgeons in clinical settings to perform phacoemulsification, facilitating cataract extraction and potentially improving patient visual outcomes.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to the predicate device.
Technological Characteristics
System includes LCD touch screen, peristaltic pump (low pulsation), and external fluid path. Vacuum range: 0-500 mmHg; Aspiration rate: 0-50 cc/min. Modes: Irrigation, Phaco, I/A, Vitrectomy, Coagulation. Features non-invasive vacuum sensing and external tubing cartridge. Software-controlled I/V pole height. Standalone or cart-mounted. Sterilization/biocompatibility assumed via compliance with standard surgical device requirements.
Indications for Use
Indicated for the disruption and extraction of cataractous lens material from the eye via phacoemulsification.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
SEP 2 2 2008
. . . . .
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
:
.
K080803
:
## 510(k) Summary
| 510(k) Submitter/<br>Owner | Herbert Cameron III<br>President<br>American Optisurgical, Inc.<br>Email: cameron@optisurgical.com |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | David Salzberg<br>Director of Regulatory Affairs<br>American Optisurgical, Inc.<br>Email: regulatory@optisurgical.com |
| Date prepared | 03/20/2008 |
| Trade name | Vizual™ Phacoemulsification System, Model 551 |
| Common name | Phacoemulsification System |
| Classification | Phacofragmentation system (21 CFR 886.4670, Product<br>Code HQC) |
| Predicate device | K020527, Horizon Phacoemulsification System<br>(manufactured by American Optisurgical, Inc.) |
| Device description | The Vizual™ Phacoemulsification system is a complete<br>anterior segment surgical system, offering Irrigation, Phaco<br>or Ultrasonic (U/S), Irrigation/Aspiration (I/A),<br>Vitrectomy, and Coagulation modes of operation. The<br>device can be operated as a standalone unit, or connected to<br>an optional cart with electric I/V pole where bottle height<br>can be controlled by the software.<br>The Vizual™ Phacoemulsification System includes the<br>Phaco unit, U/S handpiece, tubing cartridge, and foot pedal.<br>The device has been designed to work with commercially<br>available (and 510k cleared) components and accessories<br>including Phaco tips and sleeves, vit cutters, cautery<br>pencils, and cautery forceps.<br>The Vizual™ Phacoemulsification System utilizes a non-<br>invasive vacuum sensing system including an external<br>tubing cartridge that vents to the bottle versus air, thus |
| | minimizing the possibility of introducing contamination<br>into the fluid path. The external tubing cartridge design is<br>advantageous as the fluid path is visible to the user, and can<br>be easily installed or removed in a single step. |
| | The Vizual™ Phacoemulsification System uses a LCD<br>touch panel interface to allow access to all modes of<br>operation. The system software allows for storing up to<br>twenty-four different user profiles. |
| Intended use | The Vizual™ Phacoemulsification System is intended for<br>use in the disruption and extraction of cataractous lens<br>material from the eye with ultrasonic energy, a process<br>known as phacoemulsification. |
| Technological<br>characteristics | The Vizual™ Phacoemulsification System has the same<br>technological characteristics as the predicate device cleared<br>under 510(k) K020527. A summary is as follows: |
{1}------------------------------------------------
| Device<br>Characteristics | Subject Device | Predicate Device |
|---------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Display | LCD Touch Screen | LCD Touch Screen |
| Pump | Peristaltic, low pulsation | Peristaltic, low pulsation |
| Pump Vacuum<br>Range | 0 to 500 mmHg | 0 to 500 mmHg |
| Aspiration Rate | 0 to 50 cc/min | 0 to 50 cc/min |
| Fluidics | External Fluid Path | External Fluid Path |
| System Tubing | Tubing/Cartridge unit (reusable<br>and disposable models) | Tubing/Cartridge unit (reusable<br>and disposable models) |
| Vent | Fluid Vent | Fluid Vent |
| Modes | Irrigation, Phaco,<br>Irrigation/Aspiration (I/A),<br>Vitrectomy, and Coagulation | Irrigation, Phaco,<br>Irrigation/Aspiration (I/A),<br>Vitrectomy, and Coagulation |
| Programmable User<br>Parameters | Yes | Yes |
Table II – Summary of Device Characteristics
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 2 2008
American Optisurgical, Inc. c/o David Salzberg 25501 Arctic Ocean Lake Forest, CA 92630
Re: K080803
Trade/Device Name: Vizual™ Phacoemulsification System, Model 551 Regulation Number: 21 CFR 886.4670 Regulation Name: Phacoemulsification System Regulatory Class: II Product Code: HQC Dated: August 20, 2008 Received: August 20, 2008
Dear Mr. Salzberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10gally marketed predicate device results in a classification for your device and thus, perror to your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Presmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems on 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Maling B. Egleston, und
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K080803
Device Name: Vizual™ Phacoemulsification System
Indications for Use: The Vizual™ Phacoemulsification System is intended for use in the disruption and extraction of cataractous lens material from the eye with ultrasonic energy; a process known as phacoemulsification.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K080803
Page 4-1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.