K051186 · Trillium Medical, Inc. · HQC · Aug 18, 2005 · Ophthalmic
Device Facts
Record ID
K051186
Device Name
MICROPULSE PHACOEMULSIFICATION SYSTEM
Applicant
Trillium Medical, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Aug 18, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Phacoemulsification/extraction with ultrasonic power of cataractous lens from the eye.
Device Story
MicroPulse Phacoemulsification Console is an ophthalmic surgical system used for cataract removal. It functions by emulsifying cataractous lenses using ultrasonic power and removing them via irrigation/aspiration. The system includes a console and foot switch; it utilizes third-party accessories for the actual phaco handpiece, diathermy, and vitrectomy functions. Operated by surgeons in clinical settings, the device provides modes for diathermy, U/S phaco, irrigation/aspiration, and vitrectomy. The console features an LCD display and membrane switches for parameter control. Unlike some predicate devices, it lacks remote control, voice annunciation, and remote computer interfaces to reduce costs for international markets. The system uses a peristaltic pump for fluidics. The surgeon controls the procedure via the foot switch, which manages irrigation and driver activation. The device is designed to provide surgeons with manual control over irrigation and driver functions, even in the event of microprocessor failure, enhancing safety during the surgical procedure.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, design comparison, and adherence to electrical safety standards (EN60601-1, EN60601-2, UL2601-1, CAN/CSA C22.2 No. 601.1).
Technological Characteristics
Console and foot switch system; peristaltic pump (0-500 mmHg vacuum, 0-50 cc/min aspiration); LCD display with membrane switches; external fluid path; fluid vent. Electrical safety per EN60601-1, EN60601-2, UL2601-1. Embedded microprocessor for front-panel UI, switch matrix, and memory presets. Hardware-based foot switch control for driver activation. No patient-contacting materials in console/foot switch; accessories are third-party.
Indications for Use
Indicated for phacoemulsification and extraction of cataractous lenses from the eye using ultrasonic power.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
M. Imonti and Associates Vitrectomy Handpiece (K950799)
Circuit Tree Medical Reusable Tubing pak (K962430)
Submission Summary (Full Text)
{0}------------------------------------------------
8051186
AUG 1 8 2005
Trillium Medical, Inc. PO Box 2601 Poulsbo WA 98370
510(k) Summary of Safety and Effectiveness Trillium Medical, Inc. PO Box 2601 19880 Caldart Avenue NE Poulsbo, WA 98370 360 286 1341 trillium_medical@ix.netcom.com
Michael Curtis President
Contact:
510(k): K051186
| Common Name: | MicroPulse Phacoemulsification Console |
|--------------|--------------------------------------------|
| Device Name: | Unit, Phacofragmentation (21 CFR 886.4670) |
Phacoemulsification/extraction with ultrasonic power of Indications for Use: cataractous lens from the eye.
The Wave Digital Phaco System (K981989) Predicate Devices: Circuit Tree Phacoemulsifier (K954242)
Description of Device: The MicroPulse Phacoemulsification Console and foot switch, with third party accessories, will emulsify a cataractous lens and remove it from the eye, using the same technology that the predicate devices employ by using ultrasonic power. The MicroPulse Phacoemulsification Console modes of operation are similar to the predicate devices such as: Diathermy (Coagulation), U/S Phaco, Irrigation/Aspiration (I/A) and Vitrectomy. The component parts of the MicroPulse Phacoemulsification console include the console and foot switch. The U/S handpiece and tubing set are specified third party accessories. The Diathermy, Irrigation/ Aspiration, and Vitrectomy devices are universal type handpieces that are also considered additional third party accessories to the
Page 1 of 2
\$p_2\$
{1}------------------------------------------------
MicroPulse Phacoemulsification console. Trillium Medical will manufacture the console and foot switch only. Trilliums Distributers will add the third party accessories (see Exhibit B) to the console and foot switch to assemble a complete Phacoemulsification system.
Comparison: Sec Attachment Exhibit "A"
Conclusion: The MicroPulse Phacoemulsification Consolc and foot switch, when third party accessories are included(see Exhibit B), is comparable to the features of the predicated devices including technogical specifications, performance and intended use therefore we conclude that the subject device is substantially equivalent to the predicate devices listed above.
{2}------------------------------------------------
#### Exhibit A
## Comparison Chart
| Device Characteristics | Subject Device | Predicate Device | Predicate Device |
|---------------------------------|------------------------------------------------------------|---------------------------------------------|-----------------------------------------|
| | K051186<br>MicroPulse Phacoemul-<br>sification System | K981989<br>The Wave Digital Phaco<br>System | K954242<br>Circuit Tree Phacoemulsifier |
| Console | | | |
| Display | LCD, Membrane<br>switches | LCD, Membrane<br>switches | LCD, Membrane<br>switches |
| Pump | Peristaltic, no pulsation | Peristaltic, no pulsation | Peristaltic, no pulsation |
| Pump Vacuum Range | 0 to 500 mmHg | 0 to 500 mmHg | 0 to 500 mmHg |
| Aspiration Rate | 0 to 50 cc/min | 0 to 50 cc/min | 0 to 50 cc/min |
| Fluidics | External Fluid Path | External Fluid Path | External Fluid Path |
| System Tubing | Reusable Tubing set | Disposable/Reusable Tubing<br>cartridge | Reusable Tubing set |
| Vent | Fluid Vent | Fluid Vent | Fluid Vent |
| Modes | Irrigation,<br>Diathermy, U/S Phaco,<br>I/A,<br>Vitrectomy | Same | Same |
| Programmable User<br>Parameters | Yes | Yes | Yes |
#### Foot Switch
| Operational control | Freeholder and and and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>3 mode position, activates | -------------<br>l same | same |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------|
| | Reflux | | |
- Subject device has the same intended use as the predicate Intended Use: devices.
- All three devices were designed by the same engineers. The Design: MicroPulse and The WAVE are almost identical in design. The differences have to do mostly because of their respective target markets. The MicroPulse is basically The WAVE with reduced features/Modes, to lower its costs for the low cost
{3}------------------------------------------------
international market.
Hardwarc/Software: All threc devices share a great deal of the same hardware. The MicroPulse and The WAVE share almost identical hardware. The Circuit Tree device is hardware only, no software. The MicroPulse and The WAVE both have different embeded microprocessors, thus entirely different software.
## Material specifications:
The Console and footswitch do not have any direct or indirect (e.g. fluids) patient contacting materials. Any such patient contact, direct or indirect, would be only through the third party accessories listed in 'Exhibit B'.
Mechanical and Electrical specifications:
The MicroPulse and The WAVE share the same mechanical and electrical specifications. See the MicroPulse operators manual.
## Differences:
| Features: | The Wave has a remote control. The MicroPulse and Circuit Tree<br>devices do not. |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Wave has a remote computer interface. The MicroPulse and<br>Circuit Tree devices do not. |
| | LCD Display. The Wave is 4 lines X 40 characters. The<br>MicroPulse is 4 lines X 20 characters. The Circuit Tree has LED<br>displays. |
| | The WAVE has voice annunciation upon keypress. The<br>MicroPulse and Circuit Tree devices do not. |
| Electronics: | The circuitry of the MicroPulse and The WAVE are almost<br>identical, with the following exceptions; They both use different<br>embeded microprocessors, because the one used on The Wave is no |
Page 2 of 3
{4}------------------------------------------------
longer sold. The MicroPulse does not have voice playback clectronics for cost saving reasons. The remote computer interface (serial) is too expensive for the low cost market therefore not included. In the MicroPulse the foot switch position hardware was added to replace the function that was performed in software on The WAVE. This change allows the surgeon using the MicroPulse console to have more control, such as keeping Irrigation control and the ability to continue to control the drivers if the microprocessor fails. The design in The WAVE only allowed the surgeon to shut down the drivers by taking his foot off of the foot switch if the microprocessor failed.
The major differences in software design are as follows; In the Software: MicroPulse, the microprocessor is only used for front panel user interface. This involves reading the switch matrix, displaying data on the LCD, sctting the modes, setting the driver levels, and memorizing memory presets.
> The Sofware design of The WAVE does not only the above front panel user interface, but, also interfaces between the foot switch and the drivers (this is done real time in hardware on both Micropulse and Circuit Tree). Software routines in The WAVE are included for the remote computer interface (scrial) and voice playback, but, not on the MicroPulse or Circuit Tree. The Circuit Tree does not have a microprocessor, therefore, no software.
The circuits affecting electrical safety, Input power, power Electrical Safety: supplies, driver outputs and patient/doctor connections are identical in design with The WAVE. Electrical safety tests required by safety agency are identical and all devices are fully tested to the same requirements. Both the MicroPulse and The WAVE were designed to meet the following standards; EN60601-1, EN60601-2,UL2601-1, and CAN/CSAC22.2 No. 601.1.
Page 3 of 3
{5}------------------------------------------------
#### Exhibit B
# RECOMMENDED REUSABLE ACCESSORIES
The following accessories have been tested to be compatible with the MicroPulse console and foot switch.
### Phaco Handpiece and tips
Handpiece Part Number: PH-10140, Microsurgical Technology (K953959) Frequency (kHz) 37.5 to 40.5 Bandwidth (Hz) 250 to 400 Capacitance (pF) 1400 Stroke at full power (u) 69 to 96 Connector - 9 pin Fisher 104 scries sealed to facilitate autoclaving.
30 degree tip Part Number: PT-11030, Microsurgical Technology (K943672) 45 degree tip Part Number: PT-11045, Microsurgical Technology (K943672)
Diathermy Handpiece and Cable
Bi-Polar Pencil Part Number: 14-7000, Kirwan Surgical Products (K962678) Bi-Polar Cord Part Number: 10-5000, Kirwan Surgical Products (K913514) The Bipolar Diathermy Generator in the MicroPulse console is designed to use all of the currently available handpiece and cordsets designed for the phaco procedure.
Irrigation/Aspiration Handpiece and tips
Handpiece Part Number: IH-90200, Microsurgical Technology Connectors - Irrigation, Female Luer. Aspiraton, Male Luer.
Straight tip Part Number: IH-90301, Microsurgical Technology Curved tip Part Number: IH-90302, Microsurgical Technology Aspiration Port size - 0.3 mm.
Page 1 of 2
{6}------------------------------------------------
# Vitrectomy Handpicce and Infusion sleeve_
Handpicce Part Number: 12-101, M. Imonti and Associates (K950799) Connectors - Pressure line, Male Luer with integral lock ring. Aspiration line, Male Luer.
Cutting speed - 50 to 600 cuts/min.
Pressure drive – 20 psi. Must operate at 10 psi with a residual pressure of 5 psi at 600 cuts/min.
Infusion Sleeve Part Number: 12-150-100, M. Imonti and Associates
Tubing set
Reusable pak Part Number: 013-0000-007, Circuit Tree Medical (K962430)
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 8 2005
Trillium Medical, Inc. c/o Mr. Michael Curtis President PO Box 2601 19880 Caldart Avenue NE Poulsbo, WA 98370
Re: K051186
Trade/Device Name: MicroPulse Phacoemulsification Console Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation Unit Regulatory Class: Class II Product Code: HQC Dated: July 26, 2005 Received: August 2, 2005
Dear Mr. Curtis:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use surve in the energene in the enactment date of the Medical Device Amendments, or to conimered processions) in accordance with the provisions of the Federal Food, Drug, de vices that have boon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, ateres, many of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I ederal barated uns requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
# Page 2- Mr. Michael Curtis, President
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
| 510(k) Number: | K051186 |
|----------------------|----------------------------------------------------------------------------------------|
| Device Name: | MicroPulse Phacoemulsification Console |
| Indications for Use: | Phacoemulsification/extraction with ultrasonic power of cataractous lens from the eye. |
Prescription Use __ كل_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The Counter Use _ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ML B hicholas
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises KOSII 86
510(k) Number _
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.