Vitreq disposable laser probes, light fibers and Chandelier
Applicant
VitreQ BV
Product Code
HQB · Ophthalmic
Decision Date
May 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VitreQ Disposable Laser Probes, 90° Directional Laser Probes and (90°) Directional Illuminated Laser Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments at operating wavelengths of 500nm to 900nm. The (90°) Directional Illuminated Laser Probes, Light Fibers and Chandelier are for illumination during ophthalmic surgery and should only be used with the light wavelength range of 425nm to 700nm.
Device Story
Ophthalmic instruments for vitreoretinal surgery; includes disposable laser probes, directional laser probes, illuminated laser probes, light fibers, and chandeliers. Input: laser energy from external generator (500-900nm) or light from external source (425-700nm). Operation: fiber optic transmission via total internal reflection. Directional probes feature bendable metal tips (up to 90°) for peripheral access, reducing need for eye indentation and protecting the lens. Output: directed laser energy for photocoagulation or illumination of surgical field. Used in OR by surgeons. Benefits: enables precise photocoagulation and visualization, potentially shortening surgical intervention time. Single-use design.
Clinical Evidence
Bench testing only. Performance testing included verification of laser transmission, power efficiency, and output tolerances. Biocompatibility testing conducted per ISO 10993-1. Light hazard protection validated per ANSI Z80.36-2016. Human factors testing performed.
Indicated for use in vitreoretinal surgery for endo-ocular laser photocoagulation (500-900nm) and surgical site illumination (425-700nm).
Regulatory Classification
Identification
An ophthalmic photocoagulator is an AC-powered device intended to use the energy from an extended noncoherent light source to occlude blood vessels of the retina, choroid, or iris.
Peregrine 23 ga and 25 ga Adjustable Chandelier Illuminator (K151604)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Vitreq B.V. % Genci Omari, Official Correspondent Beaver-Visitec International, Inc. 500 Totten Pond Rd - 10 CityPoint Waltham, Massachusetts 02451
Re: K182646
Trade/Device Name: Vitreq disposable laser probes, light fibers and Chandelier Regulation Number: 21 CFR 886.4690 Regulation Name: Ophthalmic Photocoagulator Regulatory Class: Class II Product Code: HQB, HQF, MPA Dated: March 18, 2019 Received: March 22, 2019
Dear Denise Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182646
#### Device Name
VitreQ Disposable Laser Probes, 90° Directional Laser Probes, 90° Directional Illuminated Laser Probes, Light Fibers and Chandelier.
Indications for Use (Describe)
VitreQ Disposable Laser Probes, 90° Directional Laser Probes and 90° Directional Illuminated Laser Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments at operating wavelengths of 500nm to 900nm.
The 90° Directional Illuminated Laser Probes, Light Fibers and Chandelier are for illumination during ophthalmic surgery and should only be used with the light wavelength range of 425nm to 700mm.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <table style="border:none;"><tr><td><div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td><td><div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary VitreQ Disposable Laser Probes, Light Fibers and Chandelier
## Submitter of 510(k):
| Company name: | VitreQ B.V. |
|----------------------|---------------------------------------------------------|
| Registration number: | 3012037425 |
| Address | Seggelant-Noord 2, 3237 MG Vierpolders, The Netherlands |
| Phone: | +31 181 745090 |
| Fax: | +31 181 478583 |
| Correspondent: | JanKees Wouts |
| Email: | RA@vitreq.com |
## Device Name:
| Device Trade Name: | VitreQ Disposable Laser Probes, 90° Directional Laser Probes and<br>(90°) Directional Illuminated Laser Probes<br>VitreQ Disposable Light Fibers, Standard Light Fibers, Wide View<br>Light Fibers, 45 Degree Shielded Light Fibers and Chandelier. |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 886.4390 |
| Regulatory Name: | Ophthalmic Photocoagulator |
| Regulatory Class: | II |
| Product Code: | HOF, HQB, MPA |
## Predicate Devices
Our device is substantially equivalent to the predicate device cited in the table below:
| 510(k) | Applicant | Device Name | Product Code | Claim |
|---------|-------------------------|-------------------------------------------------------------|---------------|------------------------------|
| K121187 | Katalyst Surgical, LLC | Katalyst Laser Probes & Illuminated Laser Probes | HQF, HQB, MPA | Predicate Device (primary) |
| K151604 | Peregrine Surgical Ltd. | Peregrine 23 ga and 25 ga Adjustable Chandelier Illuminator | MPA | Predicate Device (secondary) |
## Description
The VitreQ Disposable Laser Probe is an ophthalmic instrument, which is used in posterior segment eye surgery and is appropriate for photocoagulation. The device can be connected to an ophthalmic laser generator, which is not a part of this application.
The VitreQ Disposable Laser Probe is introduced into the posterior segment of the eye through an incision or an ophthalmic entry port, after the vitreous is removed during vitreoretinal surgery.
The Disposable Laser Probe guides the laser energy to the intended surgical site to provide photocoagulation treatment. The laser enerqy intensity or power output is not controlled, altered or significantly reduced by the disposable laser probe. The laser spot size on the target tissue can be varied by altering the distance between the tissue and the probe tip.
To reach the target tissue in the periphery section of the 90° Directional Laser Probe tip can be directed due to bending the metal tip. The angle of the tip in the most directed position is about 90°. An angled laser probe reduces the need of a strong indentation of the eye and is intended to prevent accidently touching the crystal lens.
{4}------------------------------------------------
The VitreQ Disposable Laser Probes, 90° Directional Laser Probes and (90°) Directional Illuminated Laser Probes are constructed with an optical laser connector, a glass fiber optic covered by a protective sheath and one handle for surgeon manipulation, metal tubing is extending from the handle which penetrates the surgical site. The device is intended for singleuse only.
The VitreQ Disposable Laser Probes can only be used with a medical laser at an operating wavelength range of 500nm to 900nm.
For the VitreQ Disposable Laser Probes, (90°) Directional Illuminated Laser Probe the functionality is combined with illumination. To achieve the illumination of the surgical site, the probe is constructed with an additional plastic fiber, which has a connector to attach the fiber to an ophthalmic light source, which is not a part of this application.
The VitreQ Disposable Light Fibers are constructed with a handle, the plastic illumination fiber, and the connector to attach the fiber to an ophthalmic light source. To match the connection to the ophthalmic light source, there are several reusable light source adaptors available.
The VitreQ (90°) Directional Illuminated Laser Probes, Light Fibers and Chandelier can only be used with the light wavelength range of 425nm to 700nm.
## Indications for Use:
VitreQ Disposable Laser Probes, 90° Directional Laser Probes and (90°) Directional Illuminated Laser Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments at operating wavelengths of 500nm to 900nm. The (90°) Directional Illuminated Laser Probes, Light Fibers and Chandelier are for illumination during ophthalmic surgery and should only be used with the light wavelength range of 425nm to 700nm.
## Summary of technological characteristics
#### Comparison of identifications for Use
The indications for use of the VitreQ Disposable Laser Probes, Light Fibers and Chandelier are identical to the indications for use of the predicate devices.
The subject device and primary predicate device are both intended for use in vitreoretinal surgery to perform endo-ocular laser photocoaqulation treatments at operating wavelengths of 500nm to 900nm.
Although the secondary predicate device is only used to support similar technology. Both, the subject device and secondary predicate device, are for illumination during ophthalmic surgery. The subject device is more specification of the wavelengths.
#### Comparison of device characteristics
The table below shows the substantial equivalence with regards to the characteristics of the subject device and predicate device.
| Characteristics | Subject Device | | | Predicate Device (primary) | Predicate Device (secondary) | SE Discussion |
|---------------------------------------|-------------------------------------|-----------------------------|-----------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| | Directional Laser Probes | Illuminated Laser Probes | Light Fibers & Chandelier | | | |
| Laser Probe | Nitinol | | Not applicable | Nitinol | Not applicable | Not applicable |
| Laser Optical Fiber | Glass | | Not applicable | Glass | Not applicable | Not applicable |
| Characteristics | Subject Device | | | Predicate Device (primary) | Predicate Device (secondary) | SE Discussion |
| | Directional Laser Probes | Illuminated Laser Probes | Light Fibers & Chandelier | K121187 Laser Probes & Illuminated Laser Probes | K151604 23 ga and 25 ga Adjustable Chandelier Illuminator | |
| Illumination Fiber | Not applicable | | PMMA Fluorinated Polymer Cladding | PMMA Fluorinated Polymer Cladding | PMMA Fluorinated Polymer Cladding | Not applicable |
| Handpiece | ABS | | PMMA Fluorinated Polymer Cladding | 3D-printed Nylon | Acetal | Substantial Equivalent. Vitreq handpiece is made of ABS, it concerns a part which is not in contact with the patient |
| Laser Jacket | Thermoplastic rubber | | Not applicable | Elastomer | Not applicable | Substantial Equivalent. Vitreo jacket is made from thermoplastic rubber, it concerns a part which is not in contact with the patient |
| Light jacket | Not applicable | | LDPE/EVA/ PVC | Unknown | PTFE | Substantial Equivalent. Vitreo jacket is made from EVA/PVC/LDPE, it concerns a part which is not in contact with the patient |
| Needle | Not applicable | | Stainless Steel / MP35N | Not applicable | Stainless Steel | Substantial Equivalent. Vitreq needle is made from Stainless Steel/ MP35N, it is biocompatible. See Tab 15. |
| Laser connector | Nickel plated brass Stainless Steel | | Not applicable | Nickel plated brass Stainless Steel | Not applicable | Not applicable |
| Light connector | Not applicable | Stainless Steel | | Unknown | Acetal with Stainless Steel | Substantial Equivalent. VitreQ connector is from stainless steel only, it concerns a part which is not in contact with the patient |
| Laser Type | Solid state diode | | Not applicable | Solid state diode | Not applicable | Not applicable |
| Laser Wavelength | 500nm to 900nm | | Not applicable | 500nm to 900nm | Not applicable | Not applicable |
| Max Threshold of Laser Fiber | 3 Watt | | Not applicable | 3 Watt | Not applicable | Not applicable |
| Transmission of Treatment Laser | 99.8% | | Not applicable | 99.6% | Not applicable | Substantial Equivalent. The difference between predicate and subject device are negligible. See Tab 18. |
| Laser Power Efficiency | 0.998 | | Not applicable | 0.996 | Not applicable | Substantial Equivalent. The difference between predicate and subject device are negligible. |
| Laser Output Power | $500 \pm 15%$ mW | | Not applicable | $500 \pm 20%$ mW | Not applicable | Substantial Equivalent. The Vitreq device has mor strict tolerances than the predicate device. |
| Laser Size | NA value between 0.1 and 0.2 | | Not applicable | NA value between 0.1 and 0.2 | Not applicable | Not applicable. Bench testing is performed see Tab 18 |
| Light Type | Not applicable | LED, Xenon or Mercury | | Xenon or Mercury | Xenon or Mercury | Substantial Equivalent. Vitreq also tested the LED, |
| Characteristics | Subject Device | | | Predicate<br>Device<br>(primary) | Predicate<br>Device<br>(secondary) | SE Discussion |
| | Directional<br>Laser Probes | Illuminated<br>Laser Probes | Light Fibers &<br>Chandelier | K121187<br>Laser Probes &<br>Illuminated Laser<br>Probes | K151604<br>23 ga and<br>25 ga<br>Adjustable<br>Chandelier<br>Illuminator | |
| | | | | | | art technology. The values<br>Xenon and LED are<br>substantial equivalent.<br>Bench testing is performed<br>see Tab 18. |
| Wavelength | Not<br>applicable | 425nm to 700nm | | 425nm to 700nm | 425nm to 700nm | Not applicable |
| Transmission of<br>Illumination Light | Not<br>applicable | 12.5 % | 39% | Unknown | 39% | Identical, transmission of<br>illuminated laser probes is<br>less than light fibers/<br>chandeliers |
{5}------------------------------------------------
## K182646
{6}------------------------------------------------
## Substantial Equivalence Discussion
The VitreQ Disposable Laser Probes, 90° Directional Laser Probes and (90°) Directional Illuminated Laser Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments at operating wavelengths of 500nm to 900nm. The (90°) Directional Illuminated Laser Probes, Light Fibers and Chandelier are for illumination during ophthalmic surgery and should only be used with the light wavelength range of 425nm to 700nm.
The VitreQ Disposable Laser Probes, Light Fibers and Chandelier make use of the same principle of operation, which is transmission of radiation energy through a fiber based on the principle of total reflection within a multimode fiber (consisting of a core and cladding).
The differences in technological characteristics are accompanied by information that demonstrates that the device is as safe and effective as the predicate device.
#### Design Differences:
The VitreQ Directional Laser probes have a different tip design same surgical result will be achieved.
The VitreQ light fibers have a different tip design, which decrease the hazard of phototoxicity, which is confirmed by test results according to standard ANSI Z80.36.
There are differences in available product sizes between the subject and predicate device, this is not related to safety and efficacy.
The subject and predicate device have a different handle design. Both solutions are considered equivalent to initiate movement of the tip. There are differences in shape and dimensions, which are considered related to the look and feel of the instrument.
The purpose of the jacket of the laser fiber and the optical fiber is identical. There are small differences in dimensions and material.
The lengths of the subject and predicate device are different, but the length of the device is not related to safety and effectiveness.
Reorientation of the subject and predicate device is based on a different principle. Providing a means to direct the device to the surgical site does not influence the safety and effectiveness but can improve performance and allow bimanual surgery to shorten the surgical intervention.
{7}------------------------------------------------
## Differences in materials
The VitreQ handle, jackets (light and light connector have differences with regards to material, but these items are not a patient contact material.
### Differences in performance (related to laser only)
The laser power output value of the VitreQ device and the predicate device are identical. The VitreQ device has more strict tolerances than the predicate device.
### Differences in performance (related to illumination only)
VitreQ has tested Light Hazard Protection according to the FDA recognized standard ANSI Z80.36-2016. The predicate device delivers information in the Instructions for Use according the standard ISO 15004-2, which is no longer FDA recognized. Information to the user is provided according to standard ANSI Z80.36-2016.
### Substantial Equivalence Conclusion
There are differences with regards to design, handle construction, materials, packaging, shelf life and labeling. These differences do not affect the similarity in principal technology. function and operational characteristics of the devices. As a result, it is determined that the Vitreq Disposable Laser Probes, Light Fibers and Chandelier is substantially equivalent to the predicate device.
## Sterilization and shelf life
The Vitrea Disposable Laser Probes, Light Fibers and Chandelier have been successfully adopted to the validated Ethylene Oxide sterilization method. Sterilant residuals and bacterial endotoxins are kept below the allowable limits. A Shelf life study is performed to establish an expiration date of 5 years after sterilization.
## Biocompatibilitv
The Vitrea Disposable Laser Probes. Light Fibers and Chandelier have been evaluated and tested for biological safety according to the standard ISO 10993-1 and the related FDA quidance for industry.
## Performance Testing Summary
The VitreQ Disposable Laser Probes, Light Fibers and Chandelier have been tested to meet the product requirements and requirements from (safety) standards. Verification covers performance testing and human factors testing.
## Conclusion
The VitreQ Disposable Laser Probes, Light Fibers and Chandelier were shown to be substantially equivalent to the predicate devices in intended use and fundamental technology and technological characteristics to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.