The Ivalon® Eye Drain 80cc, Eye Drain 400cc, Surgical Spears, and Eye Wick 8-inch are used to absorb fluids from the operative field in ophthalmic surgery.
Device Story
Ivalon PVA surgical devices consist of sterile, single-use, dried, and compressed formalized polyvinyl alcohol (PVA) sponges. These devices are used by surgeons during ophthalmic procedures to absorb fluids from the operative field. The sponges are manufactured by M-PACT Corporation and are provided in various configurations, including 80cc and 400cc drains, 8-inch wicks, and surgical spears. The formalized PVA material is a condensation product of polyvinyl alcohol and formaldehyde, processed to ensure the removal of residual chemicals. The devices are packaged individually as sterile, compressed sponges.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Formalized polyvinyl alcohol (PVA) sponge material (condensation product of polyvinyl alcohol and formaldehyde). Sterile, single-use, dried, and compressed form factor. Available as 80cc/400cc drains, 8-inch wicks, and surgical spears.
Indications for Use
Indicated for use in ophthalmic surgery to absorb fluids from the operative field.
Regulatory Classification
Identification
An ophthalmic sponge is a device that is an absorbant sponge, pad, or spear made of folded gauze, cotton, cellulose, or other material intended to absorb fluids from the operative field in ophthalmic surgery.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Merocel Eye Drain 80cc
Merocel Eye Drain 400cc
Merocel Eye Wick 8-inch
Merocel Surgical Spears
Submission Summary (Full Text)
{0}------------------------------------------------
K97/832
## SUMMARY STATEMENT
Barry Hale M-PACT Corporation 1040 OCL Parkway Eudora, KS 66025 913-542-2135 Contact Barry Hale May 14, 1997
Name of Product: Sponge, Opthalmic
M-PACT Corporation, registration number 1928508, wishes to file an intent to market
And Surgination Design 90cc, Eve Drain 400cc, Eve Wick 8-inch, and Surgico M-PACT Corporation, registration number 1928-06, Wiskes With & Margical
sterile, single-use Eye Drain 80cc, Eye Drain 400cc, Eye Wick Birghcal
Sterile, single-use for a forma sterle, single-use Eye Drain 80cc, Eye Drain 400cc, Eye Wick Sheet, Will market the Spears muractured from formalized PVA sponge material. The classification name for the devices
products under the trade name of TVALON®. The classification name for the de products under ale aic aic air and CFR 886.4790).
The Eye Drain 80cc, Bye Drain 400cc, Bye Wick 8-inch, and Surgical Spears are The Eye Drain 80cc, Eye Drain 400cc, Eye Wilks Punch, and compressed sponge. They are used to
packaged individually as a sterile, dried and compressed sponge. They are used t packaged individually as a sterne, drive the start of a progety.
M-PACT Eye Drain 80cc, Eye Drain 400cc, Eye Wick 8-inch, and Surgical Spears are M-PACT Eye Drain 80cc, Eye Drain 400cc, Lye W.R. 9 min and of polyviny!
made from formalized PVA sponge, which is a condensation product of polyviny! made from formalized TVA sponge, which is a condensands promote on the foam
alcohol and formaldehyde. Once the chemical reaction has taken place and the foam
in the formalia alcohol and formaldehyde. Once the cheined clean of any residual chemicals.
M-PACT Corporation is the manufacturer of the formalized PVA sponge. M-PACT has
s of the bearly BAX Compresson (Departs, Colorado) to complete the packaging M-PACT Corporation is the manufacturer on the formation and the packaging
contracted with IPAX Corporation (Denver, Colorado) to complete the packaging contracted with IPAA Corporation (Denver) 300 and 100 sterilization services.
Although this is a new product for M-PACT, it is a product which is substantially Although this is a new product for M-PAC. It is a production in Surgical Spears.
equivalent to Eye Drain 800cc, Eye Drain 400cc, Eye Wick Sunch, were on the market p equivalent to Eye Drain 80cc, bye United Commerce and which were on the market prior that are currently marketed in interstate commerce and when of 1976 (May 28, 1976).
to the date of the enactment of the Medical Device Amendrom of 1976 (May 28, 1976).
Commen to the data of the enactment of the Medical Device Alichesters or 11, 2017
The M-PACT Eye Drain 80cc, Eye Drain 400cc, Eye Wick 8-Inch, and Surgical Spears The M-PACT Eye Drain 80cc, Eye Drain 400cc, Eye Pristiended use as the Merocel I Bye
are identical in size, shape, material composition and intended use as the Merocel Bye are identical in size, shape, material composition and Surgical Spears currently marketed
Drain 80cc, Eye Drain 400cc, Eye Wick 8-inch, and Surgical Spears currently, Florida Drain 80cc, Eye Drain 400cc, Lye Wick o-nitity (Komed, Jicks opville, Florida). Since the
by Merocel Corporation, Mystic, Connecticut (Komed, Jacksonville, Florida). Since by Merocel Corporation, Mystic, Comecucar (Nick 8-inch, and Surgical Spears are
M-PACT Eye Drain 80cc, Eye Drain 400cc, Eye Wick 8-inch, and Surgical Spears are M-PACT Bye Drain 80cc, Lye Drain 400cc, Eye WiEk of Internal I
identical to product currently marketed by Merocel, the only changes will be in regard to labeling and packaging.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo is circular and contains the department's name around the perimeter. In the center is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUL 3 | 1997
Mr. Barry Hale M-PACT Corp. 1040 OCL Parkway Fudora, KS 66025
Re: K971832
K971832
Trade Name: Ivalon PVA Surgical Spear; Ivalon Eye Drain, 80cc; Ivalon Eye Drain,
Trade Name: Ivalon PVA Surgical Spear; Ival 400cc; Ivalon Eye Wick
Regulatory Class: II Product Code: 86 HOZ Dated: May 14, 1997 Received: May 19, 1997
Dear Mr. Hale:
We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Section 510(ts) nothleakter tially equivalent (for the indications for above and we have determined the devices marketed in interstate commerce prior to May 28, 1976, use stated in the enclosure) to device Amendments, or to devices that have been
the enactment date of the Medical Device Amendarel Food. Drug, and Cosme the enactment date of the Medical Device rine Federal Food, Drug, and Cosmetic Act
reclassified in accordance with the provisions of the Federal controls provisions of reclassified in accordance with the provisions or the general controls provisions of the (Act). You may, therefore, market the device, successes and groundents for annual registration,
Act. The general controls provisions of the Act include requirements mishrandi Act. The general controls provisions of the Actual requence requence in the randing and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
n and the many and If your device is classified (see above) into exist additional controls. Existing major (Premarket Approval), it may be subject to sam ations of Federal Regulations, Title 21,
regulations affecting your device can be found in the Code of Federaliance with the regulations aftecting your device can be round as an assumes compliance with the Parts 800 to 895. A substantally couragents, as set forch in the Quality System
current Good Manufacturing Practice requirements, as set forch in the Right current Good Manufacturing Practice (General regulation (21 CFR Part 820) and than, through
Regulation (QS) for Medical Devices: (GDA) will verify such icegulation (QS) inspections, the Food Indian may result in regulation may result in regilatory ac periodic (QS) inspections, the Food and Drag Pressult in regulation may result in regulatory action. In
assumptions. Failure to comply with the GMP regulation may result an assumptions. Fallure to comply with the GMT regorifican whice in the Eederal addition, FDA may publish lurinel amour. Remarket notification submission does not affect
Register. Please note: this response to your premarket nonification submicres under Register. Please note: This response to your press. It through 542 of the Act for devices under any obligation you might have under sections or other Federal laws or regulations.
the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
## Page 2 - Mr. Barry Hale
This letter will allow you to begin marketing your device as described in your 510(k)
Dremarket notification . The EDA findi premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your slevice to proceed to the marker. " --
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Raeyl lorenthal
A. Ralph Rosenthal, M.D Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
INDICATIONS FOR USE:
INDICATIONS FOX USE:
The Ivalon® Eye Drain 400cc, Surgical Spears, and Eye Wick 8-inch are =================================================================================== The Ivalon® Eye Drain 80cc, Eye Drain 400cc, Surgical Opens, 2007, 2007, 1
used to absorb fluids from the operative field in ophthalmic surgery.
(Please do not write below this line - use another page if needed.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div align="center">my-chen Shir of</div> <div align="center">(Division Sign-Off)</div> <div>Division of Ophthalmic Devices</div> |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K971832 |
| Prescription Use | <div>✓</div> | OR | Over-The-Counter Use |
|------------------|--------------|----|----------------------|
|------------------|--------------|----|----------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.