K172994 · Moria SA · HNO · Oct 26, 2017 · Ophthalmic
Device Facts
Record ID
K172994
Device Name
EPI K Console
Applicant
Moria SA
Product Code
HNO · Ophthalmic
Decision Date
Oct 26, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.4370
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Epi-K is intended for use in the separation of the epithelium from the cornea in preparation for subsequent surgical procedures on denuded cornea.
Device Story
Evolution 3E Console is a power unit for keratome handpieces used in corneal surgery. It provides vacuum and power (electric motor or turbine) to motorized handpieces to separate the epithelium from the cornea. Operated by clinicians in a surgical setting, the console features a front panel for monitoring vacuum levels, pressure status, and operating modes, and a control knob for gas pressure. It connects to handpieces, footswitches, and gas/power supplies. The device now operates via mains power with a battery back-up, replacing the previous battery-only design. Output is mechanical action of the keratome; the clinician uses this to perform corneal denudation. The device facilitates surgical preparation, benefiting patients by enabling precise epithelial removal for subsequent procedures.
Clinical Evidence
Bench testing only. Verification and validation activities included testing of back-up battery performance, transformer power supply, vacuum/turbine performance, connection performance, and software verification. The device demonstrated conformity to IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
AC-powered console with battery back-up; provides vacuum and power (electric motor or turbine) to keratome handpieces. Features include vacuum/pressure displays, mode selection, and speed switches. Complies with IEC 60601-1:2012 and IEC 60601-1-2:2014. Software revisions implemented for console control.
Indications for Use
Indicated for patients requiring corneal epithelial separation in preparation for subsequent surgical procedures on denuded cornea.
Regulatory Classification
Identification
A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 26, 2017
MORIA S.A. c/o Cherita James Regulatory Consultant M Squared Associates, Inc 901 King Street. Suite 200 Alexandria, VA 22314
Re: K172994
Trade/Device Name: EPI K Console Regulation Number: 21 CFR 886.4370 Regulation Name: Keratome Regulatory Class: Class I Product Code: HNO Dated: September 26, 2017 Received: September 27, 2017
Dear Cherita James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events)
(21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K172994
Device Name Epi-K
Indications for Use (Describe)
The Epi-K is intended for use in the separation of the cornea in preparation for subsequent surgical procedures on denuded cornea.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the Moria Evolution 3E Console Special 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
MORIA S.A. Applicant: 15 Rue Georges Besse 92160 Antony France Establishment Registration Number: 9615659
Contact: Cherita James M Squared Associates, Inc. 901 King Street, Suite 200 Alexandria, Virginia 22314 Ph: 703-562-9800 xet 257 Fax: 702-562-9797 CJames@MSquaredAssociates.com
Date of Submission: October 25, 2017
Proprietary Name: Moria Evolution 3E Console
Common Name: Keratome power unit
Regulatory Class: I
Regulation: 886.4370 keratome, AC powered
Product Codes: HNO
Predicate Device: K043183 Epi-K
Device Description : The Evolution 3E Console includes pumps for producing vacuum for use with a keratome in the separation of the epithelium from the cornea in preparation for subsequent surgical procedures on denuded cornea.
The Evolution 3E Console power unit has been designed to operate the keratome by means of electric motor or turbine.
{4}------------------------------------------------
The front panel includes several displays and features:
Various indicators-Vacuum level, low vacuum, pump indicator, battery level, . mains power, manual mode
- Various displays- Operating mode, pressure status .
- Connection for the motorized handpieces- Epi-K connector, One Use-Plus . connector, M2 connector
- . Turbine and aspiration tubing connectors
- . Low vacuum, mode selection, speed switches
- The back panel includes connectors for footswitches and battery charger
- Control knob to adjust gas pressure to the turbine .
- . Connectors for Nitrogen/Air supply, Main Power Supply and Footpedal
- . Mains power switch, Residual Gas Drain Switch
- . Volume adjustment for Epi-K Footpedal Steps
- Device required labeling .
All accessory components in contact with the patient are unchanged in regard to materials and manufacturing/processing methods from the predicate device. The changes in design and function between the Evolution 3E Console power unit and predicate device have not altered sterilization, biocompatibility, or expiration date.
Indication for Use: The Epi-K is intended for use in the separation of the epithelium from the cornea in preparation for subsequent surgical procedures on denuded cornea.
of Technological Characteristics: The Evolution 3E Summary -Console modifications include a change from a battery powered device to a Mains powered device with a battery back-up, as well as minor software revisions.
Summary of Nonclinical Testing: The verification and validation activities, as identified by the risk analysis including verification of back-up battery performance, transformer power supply, vacuum/turbine and connection performance, verification of performance with accessory components of the with keratome systems, and software verification and validation were conducted to ensure that the modified device is as safe and effective as
{5}------------------------------------------------
### K172994
the predicate device, have been completed and demonstrate that the predetermined acceptance criteria have been met. The Evolution 3E Console demonstrates conformity to the following standards:
- . IEC 60601-1:2012 Medical electrical equipment - part 1: general requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
### Substantial Equivalence Discussion:
## Similarities and Differences
There is no change to the intended use or indications for use as a result of modifications to the console. The console power unit functions and operating principles are the same as the predicate device. All compatible components are the same as the predicate device. The Evo3E Console will produce the same clinical outcomes as the predicate device.
The Evolution 3E differs from the Evo2 console cleared in K043183 in the power source coming from either a mains power supply or back-up battery system when compared to the Evo2 being battery operated only. The EVO2 battery was charged via an external battery charger connected to a wall electrical supply while not in use, and disconnected during surgery. The Evo3E console operates while connected to the wall supply, and the battery back-up is available in the event of power loss at the wall supply.
Electrical safety, EMC testing, and applicable collateral standards confirm that the Evo 3E performs as safely for it intended use in the intended environment as the predicate device. The Declarations of Conformity with Design Controls provided confirm that applicable testing was conducted and support the equivalent operation of the Evo 3E Console when compared to the predicate. Subsequent software updates have also been implemented to the Evo 3E Console, and have been verified to perform as intended.
Conclusion: The Evolution 3E Console does not alter the fundamental scientific technology of the Epi-K device because it does not change the operating principle. The modifications to the console do not impact the device for its intended for use in the
{6}------------------------------------------------
### K172994
separation of the epithelium from the cornea in preparation for subsequent surgical procedures on denuded cornea.
The modified Evolution 3E Console of the Epi-K microkeratome described in this submission is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.