OPTI-FREE EXPRESS MULTI-ACTION CLEANING AND REWETTING DROP
Applicant
Alcon Laboratories
Product Code
LPN · Ophthalmic
Decision Date
Sep 16, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Intended Use
OPTI-FREE® EXPRESS® LENS DROPS may be used to lubricate and rewet daily, extended wear and disposable soft (hydrophilic) contact lenses as follows: - Moistening lenses as needed during the day to reduce discomfort. . - Moistening extended wear lenses prior to retiring at night and upon . awakening.
Device Story
OPTI-FREE EXPRESS Lens Drops is a sterile, buffered, isotonic, aqueous solution for soft (hydrophilic) contact lens care. The solution contains a citrate buffer, sodium chloride, edetate disodium (0.05%), Polyquad (polyquaternium-1, 0.001%), and Clens. The product is applied topically by the patient to the contact lens surface to provide lubrication and moisture. It is intended for over-the-counter use to alleviate lens-related discomfort and maintain lens hydration during daily wear or extended wear cycles. The solution also functions to reduce lens lysozyme levels.
Clinical Evidence
Clinical studies evaluated safety and efficacy. Results demonstrated the solution is safe and effective for its intended use and significantly reduces lens lysozyme levels on soft contact lenses.
Indicated for lubrication and rewetting of daily, extended wear, and disposable soft (hydrophilic) contact lenses to reduce discomfort during the day or prior to retiring/upon awakening.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
K983150 — COMPLETE BRAND LUBRICATING AND REWETTING DROPS · Allergan, Inc. · Nov 25, 1998
K980147 — YOUR CHOICE MULTI-PURPOSE CONTACT LENS CLEANING SOLUTION · Automatic Liquid Packaging, Inc. · Apr 13, 1998
K032030 — BLINK CL LUBRICANT EYE DROPS · Advanced Medical Optics, Inc. · Sep 25, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
# SEP 1 6 1999
# 510(K) SUMMARY
Submitted by:
Ralph H. Larsen Manager, Requlatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76134-2099 (817) 551-4702 (Phone) (817) 551-4630 (Fax)
### Device Name:
| Common Name:<br>and the first of the state of the program and the first of the programment of the first of the first of the first of the first of the first of the first of the first of the f | Contact Lens Care Rewetting Solution | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--|
#### OPTI-FREE® EXPRESS® Lens Drops Proprietary Name:
### Indications for Use:
OPTI-FREE® EXPRESS® LENS DROPS may be used to lubricate and rewet daily, extended wear and disposable soft (hydrophilic) contact lenses as follows:
- Moistening lenses as needed during the day to reduce discomfort. .
- Moistening extended wear lenses prior to retiring at night and upon . awakening.
#### Description:
OPTI-FREE EXPRESS Lens Drops is a sterile, buffered, isotonic, aqueous solution that contains a citrate buffer and sodium chloride with edetate disodium 0.05% and Polyguad® (polyguaternium-1) 0.001% as preservatives and Clens® ~ 100.
# Substantial Equivalence:
OPTI-FREE EXPRESS Lens Drops is substantially equivalent, in terms of its actions and indications for use, to OPTI-FREE Rewetting Drops approved under PMA 830034/S16. OPTI-FREE EXPRESS Lens Drops meets the guidelines set forth in FDA's May 1, 1997 Guidance for Industry; Premarket Notification 510(k) Guidance Document for Contact Lens Care Products.
{1}------------------------------------------------
# Safety and Effectiveness:
# A. Non-Clinical Data
OPTI-FREE EXPRESS Lens Drops was evaluated for microbiological safety and effectiveness using the FDA guidelines for contact lens solutions. Results of all tests were in compliance with the FDA testing guidelines
OPTI-FREE EXPRESS Lens Drops has been shown to be safe in comprehensive toxicology studies. OPTI-FREE EXPRESS Lens Drops should not present a hazard to the consumer when used under the recommended treatment regimen for soft (hydrophilic) contact lenses, or under conditions of accidental or intentional misuse.
## B. Clinical
Studies were conducted to clinically evaluate the safety and efficacy of OPTI-FREE EXPRESS Lens Drops These clinical studies demonstrate that OPTI-FREE EXPRESS Lens Drops is safe and effective and also significantly reduce lens lysozyme levels on soft contact lenses when used as indicated.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of a stylized eagle with three lines representing its wings and body. The eagle is facing to the right.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 6 1999
Mr. Ralph H. Larsen, RAC Manager, Regulatory Affairs Alcon Laboratories 6201 South Freeway Fort Worth, Texas 76134-2099
Re: K984573 Trade Name: OPTI-FREE® EXPRESS® Lens Drops Regulatory Class: II Product Code: 86 LPN Dated: July 14, 1999 Received: July 15, 1999
Dear Mr. Larsen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{3}------------------------------------------------
Page 2 - Mr. Ralph H. Larsen, RAC
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Alver L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K984573
Device Name: OPTI-FREE® EXPRESS® Lens Drops
Indications for Use:
OPTI-FREE® EXPRESS® LENS DROPS may be used to lubricate and rewet daily, extended wear and disposable soft (hydrophilic) contact lenses as follows:
- Moistening lenses as needed during the day to reduce discomfort. .
- Moistening extended wear lenses prior to retiring at night and upon awakening.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
Division Sign-Off) Division of Ophthalmic Devices 510(k) Number_K 984573
Prescription Use (Per 21 CFR 801.109) OR
Over-the-Counter Use X
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.