K012132 · Cantor & Nissel , Ltd. · NIC · May 9, 2002 · Ophthalmic
Device Facts
Record ID
K012132
Device Name
CHROMAGEN V3.0 READING AID SOFT CONTACT LENS
Applicant
Cantor & Nissel , Ltd.
Product Code
NIC · Ophthalmic
Decision Date
May 9, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ChromaGen v3.0 Reading Aid Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes. Theses lenses may also be prescribed as a colored filter to aid individuals who experience reading discomfort not related to binocular vision problems or uncorrected refractive error. The lenses are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.
Device Story
ChromaGen v3.0 Reading Aid Soft Contact Lenses are lathe-cut, hydrophilic soft contact lenses featuring precision-tinted pupil areas. Lenses are tinted using FDA-listed reactive color additives (e.g., reactive black 5, blue 21, red 11) bonded to the lens matrix. Intended for daily wear, lenses are prescribed by eyecare professionals to correct refractive errors and serve as colored filters to alleviate reading discomfort. Patients select optimal color filters from a diagnostic set based on subjective improvement in reading ease. Lenses are disinfected via hydrogen peroxide systems. The tinting process does not alter the lens's optical or physical performance characteristics. By providing a specific color filter, the device aims to improve subjective reading comfort for patients without underlying optometric causes for their reading difficulties.
Clinical Evidence
Prospective clinical study of 47 subjects (ages 9-40) with reading difficulties. Subjects compared ChromaGen lenses, placebo (clear) lenses, and no lenses. Primary endpoint: subjective 'ease of reading' grading. Results: 72.5% of subjects preferred ChromaGen lenses over placebo. Median 'ease of reading' score was 7.0 for ChromaGen vs 4.5 for placebo. Wilcoxon analysis showed ChromaGen lenses were rated significantly higher (p<0.001).
Technological Characteristics
Lathe-cut hydrophilic soft contact lens (poly hydroxyethyl methacrylate). Tinted with FDA-listed reactive color additives (e.g., reactive black 5, blue 21, red 11, yellow 15). Disinfection via hydrogen peroxide system. Frequent replacement program. No alteration to optical/performance characteristics of the base lens.
Indications for Use
Indicated for daily wear for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons with non-diseased eyes; also indicated as a colored filter for individuals experiencing reading discomfort not related to binocular vision problems or uncorrected refractive error.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K960291 — TINTED OCU-FLEX 53 (OCCUFILCON B) SOFT HYDROPHILIC CONTACT LENS · Ocu-Ease Optical Products, Inc. · Mar 25, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 0 9 2002
# 510 (k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K012132 |
|--------------------------------|---------|
|--------------------------------|---------|
Applicant information:
| Date Prepared: | April 12, 2001 |
|-----------------|--------------------------------------------------------|
| Name: | Cantor & Nissel Limited |
| Address | Market Place<br>Brackley Northants<br>England NN13 7DP |
| Contact Person: | Mr. David Cantor<br>Managing Director |
| USA Consultant: | Martin Dalsing, |
| | Medvice Consulting, Inc. |
| | Consultant and US Agent for Cantor & Nissel, Ltd. |
| | 623 Glacier Drive |
| | Grand Junction, CO 81503 |
| | (970) 243-5490 |
| | Fax #: (970) 243-5501 E-mail: mdalsing@FDApproval.com |
Device Information:
| Device Classification: | Class II |
|------------------------|----------|
| Classification Number: | 886,5925 |
Trade Name:
Reason for 510(k) submission:
Classification Name:
ChromaGen v3.0 Reading Aid Soft Contact Lens
Expanded Indication to K990757, K994320
Lens, Soft Contact, Daily Wear
{1}------------------------------------------------
### Substantially Equivalent Devices:
The ChromaGen v3.0 Reading Aid, Soft Contact Lenses are substantially equivalent to the following legally marketed devices:
- 1. Cantor & Nissel "CANTOR & NISSEL 5X Tinted", tinted contact lens Re: K990757
- Cantor & Nissel "ChromaGen v2.0 Color Discrimination Enhancement, Soft Contact Lenses" 2. tinted contact lens Re: K994320
### Device Descriptive Characteristics:
The ChromaGen v3.0 Reading Aid, Soft Contact Lenses are a range of soft lenses with precision tinted I ho on onlying hue and saturation which, when used in combination, have been shown to be of use for patients experiencing visual discomfort when reading. The ChromaGen v3.0 Reading Aid, Soft Contact Lenses are tinted with FDA "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the encoufacturing process, the lens containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed reactive color additive on that portion of the anterior (front) surface of the lens that corresponds to the iris.
The ChromaGen color additive effect is formed by reacting one or more of the reactive color additives listed in this paragraph with (poly hydroxyethyl methacrylate). The reactive color additives that may be used in and paragraph with (pation are: reactive black 5, reactive blue 21, reactive blue 19, reactive blue 4, reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15, reactive yellow 86, or reactive orange 78. The color additives used are not removed by lens handling or approved cleaning/disinfecting procedures. The ChromaGen v3.0 Reading Aid, Soft Contact Lenses tinting process does not alter the optical and/or performance characteristics of the finished tinted soft contact lens.
### INDICATIONS FOR USE:
The ChromaGen v3.0 Reading Aid Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes.
Theses lenses may also be prescribed as a colored filter to aid individuals who experience reading discomfort not related to binocular vision problems or uncorrected refractive error. The lenses are disiofected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.
{2}------------------------------------------------
### Preclinical Studies
The biocompatibility testing, compatibility testing, physical parameters of the lens can be referenced in K990757 & K994320.
#### Clinical Testing
The ChromaGen Reading Aid, Soft Contact Lens are lenses that have the pupil area tinted with one of a variety of colors. A clinical study was carried out to evaluate these colored lenses as an aid for patients who ewxperience visual discomfort while reading. Fifty-three subjects who had reading difficulties entered into the study. They ranged in age from 9 to 40. (Patients were selected from those responding to press coverage.) Subjects were excluded if their own eyecare professional indicated that they had any underlying optometric cause for reading difficulties. Forty-seven subjects completed the study.
Subjects were tested to determine which color filters from the ChromaGen diagnostic set gave the least distortion or made the print "easiest to read" (subjectiviely). The study used placebo lenses, which were clear (light handling tint) contact lenses that the patients were told had been specially treated to make the color invisible to the naked eye. Patients were given various short paragraphs to read aloud. These consisted of simple words placed in random order (not sentences). Each patient read some of these paragraphs: (1) without lenses (2) with ChromaGen lenses, and (3) with placebo contact lenses. The order of lens use was randomized. Subjects were asked to grade the subjective improvement in "ease of reading" and were asked which type of leas they preferred. Of the 40 subjects who espressed an opinion, 29 (72.5 per cent) preferred ChromaGen lenses to the placebo lenses. Six subjects chose not to give a response and 1 rated the two lenses equal. The median "ease of reading" grading (scale from -10 to +10) was 7.0 for the ChromaGen lenses and was 4.5 for the placebo lenses. Analysis of these results (Wilcoxon) shows that ChromaGen lenses were rated very significantly more highly than the placebo lenses (p<0.001)."
{3}------------------------------------------------
## Table of Substantial Equivalence
| | Characteristic | ChromaGen<br>Reading Aid lenses<br>(subject device) | Cantor & Nissel<br>ChromaGen Color<br>Discrimination lenses<br>(K994320)<br>(predicate device) | Cantor & Nissel<br>Tinted lenses<br>(K990757 )<br>(predicate device) |
|-----|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | INDICATION<br>for USE | The ChromaGen V3.0 Reading Aid<br>Soft Contact Lenses are indicated for<br>daily wear for the correction of<br>refractive ametropia (myopia,<br>hyperopia and astigmatism) in aphakic<br>and not-aphakic persons with non-<br>diseased eyes. Theses lenses may also<br>be prescribed as a colored filter to aid<br>individuals who experience reading<br>discomfort not related to binocular<br>vision problems or uncorrected<br>refractive error. The lenses are<br>disinfected using a hydrogen peroxide<br>lens care system only and are available<br>in a frequent replacement program. | The ChromaGen V2.0 Color<br>Discrimination Enhancement Soft<br>Contact Lenses are indicated for daily<br>wear to enhance color discrimination in<br>patients with protan or deutan (red-<br>green) color vision deficiencies. The<br>lens may also be prescribed for the<br>correction of refractive ametropia<br>(myopia, hyperopia and astigmatism)<br>in aphakic and not aphakic persons<br>with non-diseased eyes. The lenses are<br>disinfected using a hydrogen peroxide<br>lens care system only and are available<br>in a frequent replacement program. | The Cantor & Nissel 5X Tinted Soft<br>Contact Lenses are indicated for daily<br>wear to enhance or alter the apparent<br>eye color, including ocular masking,<br>either in sighted or non-sighted eyes.<br>The lens may also be prescribed for the<br>correction of refractive ametropia<br>(myopia, hyperopia and astigmatism)<br>in aphakic and not aphakic persons.<br>The lenses are disinfected using a<br>hydrogen peroxide lens care system<br>only and are available in a frequent<br>replacement program. |
| 2.) | Efficacy | Established, Clinical<br>study | Established, Clinical<br>study | Established Pre-clinical<br>studies |
| 3.) | Device | Colored Filter Contact<br>Lens (lathe-cut) | Colored Filter Contact<br>Lens (lathe-cut) | Colored Filter Contact<br>Lens (lathe-cut) |
| 4.) | Device Material | Soft Contact Lens<br>(hydrophilic) | Soft Contact Lens<br>(hydrophilic) | Soft Contact Lens<br>(hydrophilic) |
| 5.) | FDA Listed<br>Color Additives | The reactive colored additives consist<br>of reactive black 5, reactive blue 4,<br>reactive blue 19, reactive 21, reactive<br>blue 163, reactive yellow 15, reactive<br>yellow 86, reactive orange 78, reactive<br>red 11 and reactive red 180. | The reactive colored additives consist<br>of reactive black 5, reactive blue 4,<br>reactive blue 19, reactive 21, reactive<br>blue 163, reactive yellow 15, reactive<br>yellow 86, reactive orange 78, reactive<br>red 11 and reactive red 180. | The reactive colored additives consist<br>of reactive black 5, reactive blue 4,<br>reactive blue 19, reactive 21, reactive<br>blue 163, reactive yellow 15, reactive<br>yellow 86, reactive orange 78, reactive<br>red 11 and reactive red 180. |
| 6.) | Color Additive<br>Characteristics | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are not<br>altered by the lens coloring process. | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are not<br>altered by the lens coloring process. | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are not<br>altered by the lens coloring process. |
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### MAY 0 9 2002
Cantor & Nissel, Ltd. c/o Martin Dalsing Medvice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503
#### Re: K012132/S002
Trade/Device Name: ChromaGen.js , Reading Aid Soft Contact Lens Regulation Number: 886.5925 Regulation Name: Soft (hydrophilic) contact Lens Regulatory Class: Class II Product Code: NIC; LPL Dated: April 9, 2002 Received: April 16, 2002
Dear Mr. Dalsing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
K012i32
### INDICATIONS FOR USE STATEMENT
ChromaGen v3.0 Reading Aid, Soft Contact Lens Device Name:
### INDICATIONS FOR USE:
The ChromaGen v3.0 Reading Aid Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes.
Theses lenses may also be prescribed as a colored filter to aid individuals who experience reading discomfort not related to binocular vision problems or uncorrected refractive error. The lenses are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.
(PLEASE DO OT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Khur Wakunta
Division Sian-Off) Division of Ophthalmic Ear. Nose and Throat Devises
510(k) Number K012132
Prescription Use (Per 21 CFR 801.109) or
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.