CHROMAGEN V2.0 HAPLOSCOPIC SYSTEM & COLOR VISION ENHANCEMENT SOFT CONTACT LENSES
K994320 · Cantor & Silver , Ltd. · NCZ · Oct 20, 2000 · Ophthalmic
Device Facts
Record ID
K994320
Device Name
CHROMAGEN V2.0 HAPLOSCOPIC SYSTEM & COLOR VISION ENHANCEMENT SOFT CONTACT LENSES
Applicant
Cantor & Silver , Ltd.
Product Code
NCZ · Ophthalmic
Decision Date
Oct 20, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The ChromaGen v2.0 Haploscopic System & Color Discrimination Enhancement Soft Contact Lenses are indicated for daily wear to enhance color discrimination in patients with protan or deutan (red-green) color vision deficiencies. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons with non-diseased eyes. The lenses are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.
Device Story
ChromaGen v2.0 consists of hydrophilic soft contact lenses with precision-tinted pupils; tints utilize FDA-listed reactive color additives (e.g., reactive blue 4, red 11, yellow 15) affixed to the anterior surface corresponding to the iris. Prescribed by eyecare professionals for daily wear; intended to improve color discrimination in patients with red-green color vision deficiencies; also corrects refractive ametropia. Lenses are lathe-cut; require hydrogen peroxide-based disinfection. Device does not cure colorblindness; may cause alterations in spatial perception, reduced low-contrast acuity, or night-vision distortions. Clinical benefit derived from selective light filtration via tinted pupil to aid color differentiation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on descriptive characteristics, material safety, and manufacturing process comparisons to legally marketed predicate devices.
Technological Characteristics
Hydrophilic soft contact lens; lathe-cut manufacturing; tinted pupil using FDA-listed reactive color additives (e.g., reactive black 5, blue 4/19/21/163, red 11/180, yellow 15/86, orange 78). Daily wear; frequent replacement program; hydrogen peroxide disinfection only. No electronic components or software.
Indications for Use
Indicated for daily wear in patients with protan or deutan (red-green) color vision deficiencies to enhance color discrimination. Also indicated for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons with non-diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
X-Chrom RED Tinted PMMA Contact Lens
CANTOR & SILVER 5X Tinted
Related Devices
K012132 — CHROMAGEN V3.0 READING AID SOFT CONTACT LENS · Cantor & Nissel , Ltd. · May 9, 2002
K022373 — CHROMAGEN V3.0 READING AID & COLOR DISCRIMINATION ENHANCEMENT SPECTACLE LENS · Cantor & Nissel , Ltd. · Oct 7, 2002
K984219 — COLORMAX COLOR VISION ENHANCEMENT LENSES · Color Vision Technologies, Inc. · Nov 18, 1999
K991995 — SOFTCHROME TINTS, TRANSPARENT TINTED SOFT CONTACT LENS · Softchrome, Inc. · Aug 27, 1999
{0}------------------------------------------------
## OCT 2 0 2000
## 510 (k): SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### K994320 The assigned 510(k) number is:
### Applicant information:
| Date Prepared: | October 1, 2000 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name: | Cantor & Silver Limited |
| Address | Market Place<br>Brackley Northants<br>England NN13 7DP |
| Contact Person: | Mr. David Cantor<br>Managing Director/President |
| Phone Number:<br>Fax: | 011 44 1280 702002<br>011 44 1280 703003 |
| USA Consultant: | Martin Dalsing,<br>Med-Vice Consulting, Inc.<br>Consultant for Cantor & Silver, Inc.<br>623 Glacier Drive<br>Grand Junction, CO 81503<br>(970) 243-5490<br>Fax #: (970) 243-5501 E-mail: mdalsing@gj.net |
### Device Infor
| Device Classification: | Class II |
|------------------------|----------------------------------------------------------------------------------------------|
| Classification Number: | LPL, NCZ |
| Trade Name: | ChromaGen v2.0 Haploscopic System &<br>Color Discrimination Enhancement Soft Contact<br>Lens |
| Classification Name: | Lens, Soft Contact, Daily Wear |
{1}------------------------------------------------
### SUMMARY BASIS for Substantial Equivalence ~
### Substantially Equivalent Devices:
The ChromaGen v2.0 Color Discrimination Enhancement Soft Contact Lenses are substantially equivalent to the "X-Chrom" RED Tinted PMMA Contact Lens" and the Cantor & Silver "CANTOR & SILVER 5X Tinted", the predicate devices.
### NOTE: The "X-Chrom" RED Tinted lens is classified as a pre-amendment device.
### Device Descriptive Characteristics:
The ChromaGen v2.0 Color Discrimination Enhancement filters are designed specifically to improve discrimination between colors that are normally confused by people with protan or deutan (red-green) color vision deficiencies
The ChromaGen v20 Color Discrimination Enhancement Soft Contact Lenses are a range of soft contact lenses with precision tinted pupils of varying hue and saturation which, when used monocular and binocularly, have been shown to be of use in patients with defective color vision. Use of this product may enhance the discrimination of certain colors and may reduce the discrimination of certain colors. This device in not a cure for color vision deficiencies or colorblindness.
The ChromaGen v2.0 Color Discrimination Enhancement Soft Contact Lenses are tinted with FDA "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lens containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed reactive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The ChromaGen v2.0 Color Discrimination Enhancement Soft Contact Lenses are tinted to the eyecare professional instructions.
The ChromaGen color additive effect is formed by reacting one or more of the reactive color additives listed in this paragraph with (poly hydroxyethy) methacrylate). The reactive color additives that may be used either alone or in combination are: reactive black 5, reactive blue 21, reactive blue 19, reactive blue 4, reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15, reactive yellow 86, or reactive orange 78. The color additives used are not removed by lens handling or approved cleaning/disinfecting procedures. The ChromaGen v2.0 Haploscopic System & Color Discrimination Enhancement Soft Contact Lenses tinting process does not alter the optical and/or performance characteristics of the finished tinted soft contact lens.
{2}------------------------------------------------
# PRECAUTION FOR CHROMAGEN LENSES
### Special precaution for ChromaGen practitioners regarding the possible alterations in spatial perception.
Patients using the darker shades of tint in their ChromaGen lenses may experience some or all of the following:
- reduced low contrast acuity, -
- reduced illumination at night, -
- distortions in distance perception of moving objects or while driving, »
- distortions of apparent velocity. -
Thus, wearing these darker lenses may make driving difficult, especially at night, or under foggy, misty, or other adverse conditions.
### INDICATIONS FOR USE STATEMENT
The ChromaGen v2.0 Haploscopic System & Color Discrimination Enhancement Soft Contact Lenses are indicated for daily wear to enhance color discrimination in patients with protan or deutan (red-green) color vision deficiencies. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons with non-diseased eyes. The lenses are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.
{3}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE TABLE
The following table summarizes Cantor & Silver Limited claim of substantial equivalency in terms of safety and efficacy to the predicate devices previously mentioned.
| | Characteristic | ChromaGen V2.0 | X-CHROM LENS<br>(pre-amendment device) | Cantor & Silver<br>5X Prosthetic |
|-----|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | INDICATION<br>for USE | The ChromaGen v2.0<br>Haploscopic System & Color<br>Discrimination Enhancement<br>Soft Contact Lenses are<br>indicated for daily wear to<br>enhance color discrimination in<br>patients with protan or deutan<br>(red-green) color vision<br>deficiencies. The lens may also<br>be prescribed for the correction<br>of refractive ametropia (myopia,<br>hyperopia and astigmatism) in<br>aphakic and not aphakic persons<br>with non-diseased eyes. The<br>lenses are disinfected using a<br>hydrogen peroxide lens care<br>system only and are available in<br>a frequent replacement program. | Correction of color-blindness particularly<br>red-green color blindness.<br>NOTE: Not regulated (pre-amendment<br>device) | The Cantor & Silver 5X<br>Prosthetic Tinted Soft Contact<br>Lenses are indicated for daily<br>wear to enhance or alter the<br>apparent eye color, including<br>ocular masking, either in sighted<br>or non-sighted eyes that require a<br>prosthetic contact lens for<br>cosmetic management of<br>conditions such as corneal, iris or<br>lens abnormalities. The lens<br>may also be prescribed for the<br>correction of refractive ametropia<br>(myopia, hyperopia and<br>astigmatism) in aphakic and not<br>aphakic persons or for occlusive<br>therapy for conditions such as<br>diplopia, amblyopia or extreme<br>photophobia. The lenses are<br>disinfected using a hydrogen<br>peroxide lens care system only<br>and are available in a frequent<br>replacement program. |
| 2.) | Production<br>Method | Lathe-Cut | Lathe-cut | Lathe-cut |
| 3.) | Contact Lens<br>Material | hydrophilic | PMMA | hydrophilic |
| 3.) | FDA Listed<br>Color Additives | The reactive colored additives<br>consist of reactive black 5,<br>reactive blue 4, reactive blue 19,<br>reactive 21, reactive blue 163,<br>reactive yellow 15, reactive<br>yellow 86, reactive orange 78,<br>reactive red 11 and reactive red<br>180. | Not Available | The reactive colored additives<br>consist of reactive black 5,<br>reactive blue 4, reactive blue 19,<br>reactive 21, reactive blue 163,<br>reactive yellow 15, reactive<br>yellow 86, reactive orange 78,<br>reactive red 11 and reactive red<br>180. |
| 4.) | Color Additive<br>Characteristics | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are<br>not altered by the lens coloring<br>process. | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are not<br>altered by the lens coloring process. | The color additives used are not<br>removed by lens handling and<br>approved cleaning/disinfecting<br>procedures. The optical and<br>performance characteristics are<br>not altered by the lens coloring<br>process. |
| 5.) | Use and<br>restrictions | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably<br>required to accomplish the<br>intended prosthetic effect. | Not Available | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably<br>required to accomplish the<br>intended prosthetic effect. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 2 0 2000
Mr. Martin Dalsing Consultant for Cantor & Silver Limited c/o Medvice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503
Re: K994320 Trade Name: The ChromaGeny20 Haploscopic System & Color Discrimination Enhancement Soft Lenses Regulatory Class: II Product Code: 86 NCZ Dated: August 21, 2000 Received: August 24, 2000
Dear Mr. Dalsing:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
### Page 2 – Mr. Martin Dalsing
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noctication " THE I Dr I milling sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-6413. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Device Name:
ChromaGen v2.0 Haploscopic System & Color Discrimination Enhancement Soft Contact Lens
### INDICATIONS FOR USE:
The ChromaGen v2.0 Haploscopic System & Color Discrimination Enhancement Soft Contact Lenses are indicated for daily wear to enhance color discrimination in patients with protan or deutan (red-The lens may also be prescribed for the correction of refractive green) color vision deficiencies. green) color vision deliciences. - The tenism) in aphakic and not aphakic persons with non-diseased antenopia (nyopia, ifyoropa and mings.hudrogen peroxide lens care system only and are available in a frequent replacement program.
(PLEASE DO OT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Heve Hilmanto
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number. K994370
Prescription Use
(Per 21 CFR 801.109)
or
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.