SEE3 (LOTRAFILCON A) SOFT CONTACT LENS

K970746 · Ciba Vision Corporation · LPL · May 9, 1997 · Ophthalmic

Device Facts

Record IDK970746
Device NameSEE3 (LOTRAFILCON A) SOFT CONTACT LENS
ApplicantCiba Vision Corporation
Product CodeLPL · Ophthalmic
Decision DateMay 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

SEE3™ (lotrafilcon A) soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia, and astigmatism) in aphakic and/or non aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear with removal for cleaning and disinfection and monthly replacement. SEE3™ (lotrafilcon A) lenses may be cleaned using a chemical, not heat, disinfection system.

Device Story

SEE3™ (lotrafilcon A) is a soft contact lens for daily wear; indicated for refractive error correction. Lens material consists of 24% water and 76% lotrafilcon A (silicone-containing hydrogel) with plasma coating. Device is spherical, 14.0 mm diameter, 8.8 mm base curve, power range -20.00D to +20.00D. Supplied sterile in sealed glass vials; steam sterilized. Patient removes lens daily for chemical cleaning and disinfection; monthly replacement schedule. Healthcare provider prescribes lens based on clinical fit and vision assessment. Benefits include optical correction of ametropia in non-diseased eyes.

Clinical Evidence

Three-month prospective clinical study conducted under Good Clinical Practices. Evaluated safety, effectiveness, and clinical performance (vision, health, comfort, fit) compared to concurrent control (predicate) lenses under daily wear conditions. Results demonstrated similar overall performance to predicate lenses.

Technological Characteristics

Material: 24% water, 76% lotrafilcon A (silicone-containing hydrogel) with plasma coating. Oxygen permeability: 140 x 10⁻¹¹ [(cm²/sec)(ml O₂/ml·mmHg)]. Refractive index: 1.43. Specific gravity: 1.08. Light transmittance: ≥ 99%. Sterilization: Steam autoclave in sealed glass vials. Form factor: Spherical, 14.0 mm diameter.

Indications for Use

Indicated for daily wear for optical correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons with non-diseased eyes.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Submission Summary (Full Text)

{0} CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, GA USA 30155 SEE3™ (lotrafilcon A) Soft Contact Lens 510(k) Summary of Safety and Substantial Equivalence 510(k) Summary K970746 1. Submitter Information: MAY - 9 1997 Company: CIBA Vision Corporation 11460 Johns Creek Parkway Duluth, Georgia 30155 -1518 Contact Person: Alicia M. Plesnarski, RAC Senior Associate, Regulatory Affairs Telephone: 770-418-3924 FAX: 770-418-3033 Date Prepared: February 28, 1997 2. Device Name: - Common Name: Soft Contact Lens - Trade/Proprietary Name: SEE3™ (lotrafilcon A) - Classification Name: Daily Wear Soft (hydrophilic) Contact Lens - Device Classification: Class II {21 CFR 886.5925 (b) (1)} 3. Predicate Device: The predicate lens was selected to address both intended use (daily wear) and material type (FDA Group I: low water, nonionic polymer): - Cibasoft® (tefilcon), FDA Group I, low water, nonionic soft contact lenses for daily wear; CIBAVision® Corporation - PMAs P810005, P820086 Premarket Notification 510(k) : clear (lotraficon A) Soft Contact Lens Part IX. Summary of Safety & Substantial Equivalence, Rev: 00, 510SUM_0.DOC, Page 1 of 4 00653 {1} 10/08/1994 20:45 PAGE 03 | CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, GA USA 30155 | | --- | | SEE3™ (lotrafilcon A) Soft Contact Lens | | 510(k) Summary of Safety and Substantial Equivalence | ## 4. Description of Device: The SEE3™ (lotrafilcon A) soft contact lens is a new soft contact lens suitable for daily wear. The lens material is 24% water and 76% lotrafilcon A, a silicone containing hydrogel treated with a plasma coating. The SEE3™ (lotrafilcon A) contact lens is available in a spherical lens design of the following dimensions: - Chord Diameter: 14.0 mm - Center Thickness: 0.05 to 0.35 mm (varies with power) - Base Curve: 8.8 mm - Power Range: -20.00D to +20.00D - Powers Available: -5.00D to +5.00D (0.25D steps), -6.00, -5.50, +5.50, +6.00D A clear lens has the following properties: - Specific gravity: 1.08 - Refractive index: 1.43 (hydrated) - Light transmittance: ≥ 99 % - Water content: 24 % by weight in normal saline - Oxygen permeability: 140 x 10⁻¹¹ [(cm²/sec)(ml O₂/ml·mmHg)] measured at 35°C, Coulometric method. Lenses are supplied sterile in sealed glass vials containing isotonic phosphate buffered saline solution. The compatibility and package integrity of the glass vial packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Vial containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Stability study data currently supports a twelve (12) month shelf-life for the SEE3™ (lotrafilcon A) soft contact lens in sealed glass vials. Shelf-life studies are ongoing to determine extension of expiration dating. ## 5. Indications for Use: SEE3™ (lotrafilcon A) soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia, and astigmatism) in aphakic and/or non aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear with removal for cleaning and disinfection and monthly replacement. SEE3™ (lotrafilcon A) lenses may be cleaned using a chemical, not heat, disinfection system. Premarket Notification 510(k) : clear (lotrafilcon A) Soft Contact Lens Part IX. Summary of Safety & Substantial Equivalence, Rev. 01, 510SUM_2.DOC, Page 2 of 4 {2} CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, GA USA 30155 SEE3™ (lotrafilcon A) Soft Contact Lens 510(k) Summary of Safety and Substantial Equivalence ## 6. Description of Safety and Substantial Equivalence: A series of nonclinical tests and clinical studies were performed to demonstrate the safety and effectiveness of the SEE3™ (lotrafilcon A) contact lens, and establish substantial equivalence to a currently marketed, predicate (control) lens. All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses and in conformance to applicable device regulations. Results demonstrate the lens is non-toxic and biocompatible, and has material characteristics comparable to or better than other currently marketed soft contact lenses. Clinically, the lens has performed satisfactorily in a daily wear investigation. Results from all tests demonstrate the substantial equivalence to previously FDA approved, and currently marketed predicate (control) lenses. ## Nonclinical Testing: A series of in vitro and in vivo preclinical toxicology and biocompatibility testing was performed to assess the safety and effectiveness of the contact lens. All nonclinical toxicology tests were conducted in accordance with the GLP regulation (21 CFR Part 58). The results of the nonclinical testing on the SEE3™ (lotrafilcon A) contact lens demonstrate that: - The lens material and extracts are not toxic and non irritating. - Lens physical and material properties are consistent with industry marketed lenses. - The lens material remains unaffected, with respect to lens properties, by exposure to chemical cleaning and disinfection systems. The amount of preservatives in the tested care regimens show similar uptake and release profiles to the predicate device. ## Clinical Testing: The SEE3™ (lotrafilcon A) contact lens was investigated in daily wear clinical study. The three month clinical evaluation was conducted in accordance with current Good Clinical Practices and published regulations (21 CFR Parts 50, 56, 312, 812). The study assessed the safety and effectiveness, and clinical performance as compared to a predicate control lens. Clinical evaluation of the SEE3™ (lotrafilcon A) lens demonstrated similar overall performance in the clinically relevant areas of vision, health, comfort and fit to compared to concurrent controls when used under daily wear conditions. Premarket Notification 510(k) : clear (lotrafilcon A) Soft Contact Lens Part IX. Summary of Safety & Substantial Equivalence, Rev: 00, 510SUM_0.DOC, Page 3 of 4 00655 {3} | CIBA Vision | CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, GA USA 30155 | | --- | --- | | SEE3™ (lotrafilcon A) Soft Contact Lens | | | 510(k) Summary of Safety and Substantial Equivalence | | ## Substantial Equivalence: The SEE3™ (lotrafilcon A) contact lens is similar to other daily wear soft contact lenses in terms of water content (24% water) and ionic characteristics (FDA Group I: low water, nonionic), clinical performance, and indications for use. In addition, the lenses may be disinfected using a chemical, not heat, disinfection regimen. Any differences which may exist between the (lotrafilcon A) contact lens and other Group I soft hydrophilic plastic contact lenses does not adversely effect the safety and effectiveness of the device. Premarket Notification 510(k) : clear (lotraficon A) Soft Contact Lens Part IX. Summary of Safety & Substantial Equivalence, Rev: 00, 510SUM_0.DOC, Page 4 of 4 00656
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...