TOTAL30™ (lehfilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity. TOTAL30™ Asphere (lehfilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity. TOTAL30™ for Astigmatism (lehfilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to 6.00 diopters (D) of astigmatism. TOTAL30™ Multifocal (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity. TOTAL30™ Multifocal Toric (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and byperopia) in phakic persons with non-diseased eves who may require a reading addition of +3.00 diopters (D) or less and who may have up to 6.00 diopters (D) of astigmatism. The lenses are to be prescribed for daily wear, with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, or disposal, as recommended by the eve care professional. Lenses should be discarded with a new pair each month, or more often, if recommended by the eye care professional.
Device Story
TOTAL30 soft contact lenses are silicone hydrogel lenses (lehfilcon A) designed for daily wear with monthly replacement. Lenses feature surface treatment to achieve high surface water content (≥90%) and incorporate benzotriazole monomers for UVA/UVB and high-energy visible light protection. Lenses are cast-molded and supplied sterile in phosphate-buffered saline. Prescribed by eye care professionals for refractive error correction; patients perform daily cleaning and disinfection. Output is optical correction of vision. Benefits include improved comfort and oxygen permeability compared to traditional hydrogels. Clinical performance is comparable to existing silicone hydrogel lenses.
Clinical Evidence
Prospective, randomized, double-masked, parallel-group study (n=118; 78 test, 40 control). Primary endpoint: Snellen distance visual acuity. Results showed similar clinical performance in vision, comfort, fit, handling, and ocular health between lehfilcon A and comfilcon A control lenses.
Technological Characteristics
Material: Lehfilcon A (Group 5B silicone hydrogel); 55% water content; ≥90% surface water content. Oxygen permeability: 123 x 10^-11 (cm2/sec)(ml O2/ml.mmHg). Refractive index: 1.40. UV/HEV blocking monomers included. Manufacturing: Cast molded. Sterilization: Steam autoclave. Packaging: Blister packs with PBS.
Indications for Use
Indicated for phakic/aphakic persons with non-diseased eyes for correction of myopia, hyperopia, presbyopia, and astigmatism (up to 6.00D).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K182782 — DAILIES TOTAL1, DAILIES TOTAL1 Asphere, DAILIES TOTAL1 for ASTIGMATISM, DAILIES TOTAL1 Multifocal, DAILIES TOTAL1 Multifocal Toric · Alcon Laboratories, Inc. · Jan 18, 2019
K212806 — Alcon (serafilcon A) soft contact lenses for daily wear · Alcon Laboratories, Inc. · Dec 28, 2021
K160212 — ACUVUE (senofilcon C) Soft Contact Lens · Johnson & Johnson Vision Care, Inc. · Mar 28, 2016
K244009 — Pure Plus UV Aspheric (Otufilcon A) Silicone Soft (hydrophilic) Contact Lens for Daily Wear (Daily Wear); Pure Plus UV Aspheric (Otufilcon A) 1-Day Silicone Soft (hydrophilic) Contact Lens (Daily Disposable) · Yung Sheng Optical Co., Ltd. · Sep 25, 2025
K130342 — SAUFLON CLARITI (SOMOFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER · Sauflon Pharmaceuticals, Ltd. · Aug 5, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2021
Alcon Laboratories Inc. Alicia Plesnarski, RAC Director, Global Regulatory Affairs 6201 South Freeway Fort Worth, TX 76131-2099
Re: K210436 Trade/Device Name: TOTAL30TM brand soft contact lenses: TOTAL30TM TOTAL30TM Asphere, TOTAL30TM for Astigmatism, TOTAL30TM Multifocal. TOTAL30™ Multifocal Toric Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 11, 2021 Received: February 12, 2021
Dear Alicia Plesnarski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# J Angelo Green -S
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K210436
### Device Name
TOTAL30™ soft contact lenses: TOTAL30, TOTAL30 Asphere, TOTAL30 for Astigmatism, TOTAL30 Multifocal, TOTAL30 Multifocal Toric
### Indications for Use (Describe)
TOTAL30™ (lehfilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30™ Asphere (lehfilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
TOTAL30™ for Astigmatism (lehfilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to 6.00 diopters (D) of astigmatism.
TOTAL30™ Multifocal (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30™ Multifocal Toric (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and byperopia) in phakic persons with non-diseased eves who may require a reading addition of +3.00 diopters (D) or less and who may have up to 6.00 diopters (D) of astigmatism.
The lenses are to be prescribed for daily wear, with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, or disposal, as recommended by the eve care professional. Lenses should be discarded with a new pair each month, or more often, if recommended by the eye care professional.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Alcon" in a bold, sans-serif font. The word is a dark blue color. The letters are evenly spaced and the word is centered in the image. The background is white.
### 510(k) Summary
This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.
#### I. Submitter of the 510(k)
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
|-------------------------|----------------------------------------------------------------------------------|
| Primary Contact Person: | Alicia M. Plesnarski, RAC |
| Phone: | 678 415-3924 |
| Fax: | N/A |
| Email: | alicia.plesnarski@alcon.com |
| Back up Contact Person: | Sherri Lakota |
| Phone: | 817 615-5472 |
| Fax: | 817 302-4574 |
| Email: | sherri.lakota@alcon.com |
| Date Prepared: | April 12, 2021 |
#### II. Devices Subject to this 510(k)
| Trade name: | TOTAL30TM brand soft contact lenses<br>TOTAL30TM,<br>TOTAL30TM Asphere,<br>TOTAL30TM for Astigmatism,<br>TOTAL30TM Multifocal,<br>TOTAL30TM Multifocal Toric |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name: | (lehfilcon A) contact lens |
| Classification name: | Soft (hydrophilic) Contact Lens |
| Device classification: | 21 CFR 886.5925 (b) (1) |
| Product code: | LPL; MVN |
{4}------------------------------------------------
### III. Predicate Device
The predicate device is the CooperVision® Biofinity® (comfilcon A) soft contact lens. Comfilcon A represents a Group 5C silicone hydrogel contact lens material ('enhanced oxygen permeable materials') according to ISO 18369-1:2017, Ophthalmic optics - Contact lenses - Part 1: Vocabulary, classification system and recommendations for labelling specifications, as follows:
| Group Suffix | Hydrogel Material | Description |
|--------------|--------------------|------------------------------------------------------------------------------------------------------------------|
| 5C | Low water subgroup | A subgroup of Group 5 which contains less than<br>50 % water and no ionic monomer or oligomer<br>at pH 6 to pH 8 |
The predicate device has US FDA Premarket Notification 510(k) clearance for daily wear (K052560, December 6, 2005).
### IV. Device Description
The subject device, TOTAL30™ (lehfilcon A) soft contact lenses, are made from a surface treated, silicone containing hydrogel lens material that is approximately 55% water and 45% lehfilcon A. The color additive Reactive Blue 247 is added to the lens material to create a light blue-green edgeto-edge color to make it easier to see when handling. The lenses contain two benzotriazole monomers to block UVA and UVB radiation and, additionally, reduce transmittance in the range from 380 nm to 450 nm.
Lehfilcon A represents a Group 5B silicone hydrogel contact lens material ("enhanced oxygen permeable materials') according to ISO 18369-1:2017, Ophthalmic optics - Contact lenses - Part 1: Vocabulary, classification system and recommendations for labelling specifications, as follows:
| Group Suffix | Hydrogel Material | Description |
|--------------|---------------------|------------------------------------------------------------------------------------------------------------|
| 5B | High water subgroup | A subgroup of Group 5 which contains more than 50 % water and no ionic monomer or oligomer at pH 6 to pH 8 |
{5}------------------------------------------------
Lehfilcon A lens designs include spherical, asphere, toric, multifocal toric contact lenses in the following parameter ranges:
- . Diameter Range: 13.0 to 15.0 mm
- Base Curve Range: 8.0 to 9.2 mm .
- . Power Range: -20.00 D to +20.00 D
- Center Thickness: varies with design and power .
(Example: 0.08 mm for -3.00 D spherical)
- Cylinder Power (Toric) -0.25 D to -10.00 D .
- ADD Power (Multifocal) LO, MED, HI .
Lenses have the following properties:
- . 1.40 (hydrated) Refractive index:
- . Water content: 55% by weight in normal saline
- . Surface water content: ≥ 90%
- Oxygen permeability: 123 x 10 -11 .
[(cm2/sec)(ml O2 /ml.mmHg)] measured at 35 °C (normalized Dk-Polarographic method)
- Light transmittance: . >85%
- . UV Transmittance: The transmittance characteristics are less than 1% in the UVB range of 280 nm to 315 nm and less than 10% in the UVA range of 315 to 380 nm for the entire power range.
Lehfilcon A contact lenses are supplied sterile in sealed blister packs containing isotonic phosphate buffered saline solution (PBS) with polymeric wetting agents. The compatibility and package integrity of the blister pack packaging system have been demonstrated and successfully used for other Alcon marketed contact lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility).
{6}------------------------------------------------
#### V. Indications for Use
TOTAL30 (lehfilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30 Asphere (lehfilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30 for Astigmatism (lehfilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes with up to 6.00 diopters (D) of astigmatism.
TOTAL30 Multifocal (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
TOTAL30 Multifocal Toric (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to 6.00 diopters (D) of astigmatism.
The lenses are to be prescribed for daily wear with removal for cleaning and disinfection (chemical not heat) prior to reinsertion, or disposal, as recommended by the eve care professional. Lenses should be discarded and replaced with a new pair each month, or more often, if recommended by the eye care professional.
{7}------------------------------------------------
# VI. Comparison to Technological Characteristics with the Predicate Device
Table 1 provides a side-by-side comparison of the device as compared to the predicate device in terms of intended use and technological information.
| | Predicate Device<br>(CooperVision) | Subject Device<br>(Alcon) |
|--------------------------------------------------|------------------------------------------------------------|-------------------------------------------|
| Trade Name<br>(brand) | Biofinity | TOTAL30 |
| Submission<br>Number | 510(k) K052560 | 510(k) K210436 |
| Device<br>Classification Name | Daily Wear Soft Contact Lens<br>21 CFR 886.5925 (b)(1) | Same |
| Intended Use | Vision correction | Same |
| Wearing Modality | Daily wear | Same |
| Replacement<br>Schedule | Monthly | Same |
| Material<br>Classification<br>(ISO 18369-1:2017) | Group 5C, low water silicone<br>hydrogel | Group 5B, high water silicone<br>hydrogel |
| Lens Material | Comfilcon A | Lehfilcon A |
| Surface Treatment | No | Yes |
| Visibility Tint | Light blue | Light blue-green |
| UV/UV vis<br>absorbing<br>monomers | No | Yes* |
| Manufacturing<br>Method | Cast molded | Cast molded<br>(DSM FLEX) |
| Lens Designs | Spherical, asphere, toric,<br>multifocal, multifocal toric | Same |
| Parameter Ranges: | | |
Table 1. Substantial Equivalence Comparison
{8}------------------------------------------------
| Table 1. Substantial Equivalence Comparison | |
|---------------------------------------------|--|
|---------------------------------------------|--|
| | Predicate Device<br>(CooperVision) | Subject Device<br>(Alcon) |
|------------------------------------|---------------------------------------------|---------------------------------------------------|
| Power: | +20.00 to -20.00 D | Same |
| Base Curve: | 8.0 to 9.5 mm | 8.0 to 9.2 mm |
| Diameter: | 13.0 to 15.0 mm | Same |
| Cylinder Power<br>(toric): | -0.25 to -10.00 D | Same |
| Add Power<br>(multifocal): | +0.50 to +3.00D | LO, MED, HI |
| Water Content | 48% | 55% |
| Surface Water<br>Content | -- | ≥ 90%** |
| Refractive Index | 1.40 | 1.40 |
| Oxygen<br>Permeability | 128***<br>(coulometric method) | 123***<br>(polarographic method) |
| Sterilization | Steam sterilization,<br>validated autoclave | Same |
| Packaging | Cartons containing<br>sealed blister packs | Same |
| Package Storage<br>Saline Solution | Isotonic saline | Phosphate buffered saline (PBS)<br>with additives |
*UV absorbers are commonly found in numerous US legally commercialized contact lenses (examples K151918; K100349; K131378; K182782)
** High surface water content property similar to other US legally commercialized contact lenses with a surface treatment (example K113168)
***units 10 -11 (cm2/sec)(ml O2 /ml x mm Hg)
Lehfilcon A soft contact lenses are equivalent to the predicate lens and similar to other daily wear soft contact lenses in terms of technological characteristics and intended use.
Any differences which may exist between the subject device (lehfilcon A soft contact lenses) and the predicate device or other silicone hydrogel soft contact lenses for daily wear do not adversely affect the safety and effectiveness of the subject device.
{9}------------------------------------------------
### VII. Performance Data
A series of nonclinical tests and a clinical study were performed to demonstrate the substantial equivalence of lehfilcon A contact lenses to the predicate device. All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses and ISO standards, as applicable. In addition, nonclinical biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58).
## Biocompatibility Testing
As listed below, a series of in vitro and in vivo biocompatibility evaluations, including cytotoxicity, ocular irritation, sensitization, systemic toxicity and genotoxicity testing, confirms that lehfilcon A contact lenses are non-toxic and biocompatible.
- . Cytotoxicity Studies (lens, lens extracts, and package saline)
- Ocular Irritation / toxicity (Ocular) in Rabbits .
- o Lens wear study in rabbits per ISO 9394 (lens)
- Primary ocular irritation studies per ISO 19993-10 (lens extracts and package o saline)
- Sensitization Studies (lens extracts and package saline) .
- Acute Systemic Toxicity Study in Mice (lens extracts) .
- . Genotoxicity Studies (lens extracts and package saline)
The primary packaging materials used for lehfilcon A lenses are the same as those used for other Alcon contact lenses with FDA 510(k) clearance (K113168, K180669); therefore, existing biocompatibility data for the packaging materials applies.
Biocompatibility testing was conducted in accordance with the US Good Laboratory Practice (GLP) for Nonclinical Laboratory Studies regulation (21 CFR Part 58) and relevant FDA guidance and ISO 10993 series of biocompatibility standards.
{10}------------------------------------------------
# Physical-Chemical Testing
The following nonclinical bench testing, conducted using GxP conditions, and, where applicable, ISO 18369-2, -3, -4 standards, established the physico-chemical properties of lehfilcon A contact lenses:
- · Refractive Index
- · Oxygen Permeability
- · Ion Permeability
- · Mechanical Properties
- · Wetting Contact Angle
- Transmittance Properties .
- Percent Water Content .
- Residuals and Extractables .
- Lens Parameters .
- Package Saline Properties (pH and osmolality) .
# Solution Compatibility Testing
Lehfilcon A contact lenses are for daily wear use with daily removal for cleaning and disinfection and one-month replacement. Therefore, the compatibility of lehfilcon A contact lenses with commonly available cleaning and disinfection solutions (peroxide and MPDS), and preserved or unpreserved commercial contact lens saline and rewetting drops was confirmed following the methodology described in ISO 11981:2017, Ophthalmic optics - Contact lenses and contact lens care products – Determination of physical compatibility of contact lens care products with contact lenses.
Lehfilcon A contact lenses were also analyzed for uptake and release of preservatives found in various lens care products. Testing was conducted according to ISO 11986:2017, Ophthalmic optics - Contact lenses and contact lens care products - Determination of preservative uptake and release. Uptake and release profiles of lehfilcon A contact lenses for Polyguad (PQ), Aldox, Polyhexamethylene biguanide (PHMB) and alexidine were comparable to the control lenses tested.
# Sterilization and Stability Testing
Lehfilcon A contact lenses in buffered saline solution are provided sterile in sealed blister packs. Results of an ongoing stability study demonstrate that package lenses remain sterile and stable for the labeled expiration date.
The results of all non-clinical testing demonstrate:
{11}------------------------------------------------
- The lens material, lens extracts and package saline of the device are non-toxic, non-irritating ● and non-sensitizing.
- Lens physical and material properties of the device are consistent with industry-marketed . lenses and equivalent to the predicate lens.
- Like the predicate device is compatible with commonly available contact lens . cleaning and disinfecting solutions, saline solutions and rewetting drops.
- Successful stability testing supports the labeled expiration date for the device. .
### Clinical Performance Testing
A three-month clinical study, conducted according to the May 1994 FDA 510(k) contact lens guidance document for a new contact lens material and ISO 11980:2012, Ophthalmic optics -Contact lenses and contact lens care products - Guidance for clinical investigations, assessed the safety and performance of lehfilcon A soft contact lenses for daily wear as compared to a predicate control lens (2:1 ratio test to control). Product safety was assessed based on adverse events, device deficiencies and biomicroscopy findings.
Eight (8) study sites in the US enrolled a total of 119 subjects of whom 118 were randomized and exposed to study lenses in this prospective, randomized, stratified (by corneal curvature radius), controlled, double-masked, parallel group study. Of the 118 subjects (236 eyes) enrolled into the study, 78 subjects were dispensed Phoenix (lehfilcon A) test lenses and 40 subjects were dispensed Biofinity control lenses (2:1 ratio). Through the course of the study, 3 subjects discontinued early (2 subjects from the control group withdrew and 1 subject, from the test group, moved due to COVID-19). Lenses were worn in a daily wear modality and were replaced monthly. The primary effectiveness endpoint was Snellen VA at distance. The demographic characteristics for all enrolled subjects were similar between the lehfilcon A and comfilcon A groups. Overall, the maiority of subjects were female (74.6%), predominantly White, and most were not of Hispanic or Latino ethnicity. The overall mean age of the subjects was 33.9 years.
Additional endpoints include manifest refraction, keratometry, lens fit, surface characteristics and subjective ratings of vision, comfort and handling.
{12}------------------------------------------------
The study results showed similar clinical performance between the test lehfilcon A contact lenses and control comfilcon A contact lenses in the clinically relevant areas of vision, comfort, fit, handling and health when worn on a daily wear basis.
The clinical study demonstrated the substantial equivalence of the subject device with the predicate, control lens.
### Risk and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of silicone hydrogel contact lenses on a daily wear basis. The benefits to the patient are the same as those for other silicone hydrogel contact lenses.
### VIII. Conclusions
Lehfilcon A soft contact lenses are substantially equivalent to the predicate lens and similar to other daily wear soft contact lenses in terms of technological characteristics and intended use. Nonclinical and clinical data demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Any differences which may exist between the subject device (lehfilcon A soft contact lenses) and the predicate device or other daily wear silicone hydrogel soft contact lenses do not adversely affect the safety and effectiveness of the subject device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.