MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty. MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity. MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic persons with non-diseased eyes who may have - 10.00 diopters of astigmatism or less.
Device Story
MyDay (stenfilcon A) soft contact lenses; daily disposable; refractive medium placed on cornea to focus light on retina; corrects myopia, hyperopia, astigmatism, and presbyopia. Available in Asphere, Toric, Multifocal, and Multifocal Toric designs. Prescribed by eye care professionals for patient use. Lens material is a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide; includes UV blocker (benzotriazolyl methacrylate) and blue visibility tint. Provides optical correction for ametropic patients; improves visual acuity.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing of physical, mechanical, and optical properties per ANSI/ISO standards (e.g., ANSI Z80.20-2016, ISO 18369). Biocompatibility confirmed via extractable residual analysis per EN ISO 10993-18.
Technological Characteristics
Silicone hydrogel (stenfilcon A); random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. Water content 54%; oxygen permeability 80x10-11. UV blocker: benzotriazolyl methacrylate. Sterilization: moist heat. Manufacturing: molded. Dimensions: 13.0-15.5mm diameter, 8.4-8.7mm base curve. Standards: ANSI Z80.20-2016, ISO 18369, ASTM D1708-18.
Indications for Use
Indicated for correction of ametropia (myopia/hyperopia), astigmatism, and presbyopia in aphakic and non-aphakic persons with non-diseased eyes. Power range -20.00D to +20.00D; astigmatic correction up to -10.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
CooperVision SUS (stenfilcon A) Soft (hydrophilic) Contact Lenses for Single Use Daily Wear (K131378)
Related Devices
K191763 — MyDay (Stenfilcon A) ASPHERE, MyDay (Stenfilcon A) TORIC, MyDay (Stenfilcon A) MULTIFOCAL, MyDay (Stenfilcon A) MULTIFOCAL TORIC · CooperVision, Inc. · Mar 23, 2020
K210930 — ACUVUE® OASYS MAX 1-DAY Contact Lenses (senofilcon A) · Johnson & Johnson Vision Care, Inc. · Nov 29, 2021
K172692 — Tangible Hydrogel with Tangible Polymer (etafilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens · Tangible Science, LLC · Dec 13, 2017
K240477 — Saview SH 38 UV (Senofilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens; Saview SH 38 UV toric (Senofilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens; Saview SH 38 UV multifocal (Senofilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens · St. shine Optical Co., Ltd. · Jul 1, 2024
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 29, 2019
CooperVision, Inc. Marie E. Dutton Senior Regulatory Affairs Specialist 5870 Stoneridge Drive. Suite 1 Pleasanton, CA 94588
Re: K190965
Trade/Device Name: MyDay (stenfilcon A) Soft (hydrophilic) Daily Disposable Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: April 11, 2019 Received: April 12, 2019
Dear Ms. Dutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens
### Indications for Use (Describe)
MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.
MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic persons with non-diseased eyes who may have - 10.00 diopters of astigmatism or less.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### I. SUBMITTER:
CooperVision, Inc. 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588
## Contact Person:
Marie Dutton Senior Regulatory Affairs Specialist CooperVision, Inc. 5870 Stoneridge Drive. Suite 1 Pleasanton, CA 94588 Phone: (925) 251-6645 (925) 251-6643 Fax: E-mail: MDutton@coopervision.com
# Date Prepared:
April 11, 2019
#### II. DEVICE:
| Trade Name: | MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable<br>Contact Lens |
|-----------------------|--------------------------------------------------------------------------|
| Common Name: | Soft (hydrophilic) Contact Lens |
| Classification Name: | Lens, Contact, (Disposable) [21 CFR 886.5925 (b) (1)] |
| Regulatory Class: | II |
| Product Code: | LPL, MVN |
| Classification Panel: | Ophthalmic |
#### III. PREDICATE DEVICE:
CooperVision's SUS (stenfilcon A) Soft (hydrophilic) Contact Lenses for Single Use Daily Wear, K131378
#### IV. DEVICE DESCRIPTION:
MyDay Contact Lenses are available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.
The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.
When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.
{4}------------------------------------------------
MyDay (stenfilcon A) contact lens parameters are:
| • Chord Diameter: | 13.0 mm to 15.5mm |
|---------------------|----------------------------------------|
| • Base Curve: | 8.4 ± 0.5 mm and 8.7 ± 0.5 mm |
| • Center Thickness: | 0.08 mm to 0.218 mm (varies with power |
| • Powers: | -20.00D to +20.00D |
| • Cylinder Powers: | -0.25D to -10.00D |
| • Axis: | 0° to 180° in 10° increments |
| • Add Power Range: | +0.50 to +4.00 |
The physical/optical properties of the lens are:
- Specific Gravity: 1.033 Refractive Index: 1.401 ● 96%
- Light Transmittance:
Surface Character:
- Hydrophilic
- Water Content: 54%
- Oxygen Permeability: 80x10-11 [(cm2/sec)x(ml O2)/(ml x mm Hg)]
#### V. INDICATIONS FOR USE:
●
●
MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.
{5}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The technological characteristics of the subject device and the predicate device are compared in the table below.
| Technology/Material Comparison | | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| | Predicate Device | Subject Device |
| Product Name | CooperVision<br>SUS (stenfilcon A) | CooperVision<br>MyDay (stenfilcon A) |
| Material USAN Name | stenfilcon A | Same |
| 510(k) Number | K131378 | TBD – Current Submission |
| FDA Category (Group) | Silicone Hydrogel | Same |
| Manufacturing Method | Molded | Same |
| Sterilization | Moist Heat | Same |
| Packaging Materials | Injection molded<br>polypropylene blisters<br>covered by aluminum foil<br>laminate; blister strips are<br>packed into printed cartons | Same |
| Packaging Solution | Phosphate Buffered Saline<br>Solution with Tween | Same |
| Visibility Tint | Reactive Blue #246 (RB246) | Same |
| UV Blocker | Norbloc | Same |
#### VII. PERFORMANCE DATA
Results from non-clinical studies were provided in support of the substantial equivalence determination.
## Performance testing:
In accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, May 12, 1994 amended June 28, 1994, the following battery of testing was performed. All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols. Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated:
- Contact angle per ANSI Z80.20-2016 ●
- Light transmittance per ISO 18369-2:2017 and ISO 18369-3:2017 ●
- Water content per ISO18369-2:2017 and ISO 18369-4:2017
- Mechanical properties per ANSI Z80.20-2016 and ASTM D1708-18 ●
- Refractive index per ISO 18369-2:2017 and ISO 16369-4:2017
- Total extractables per ANSI Z80.20-2016 and ISO 18369-4:2017 ●
- Non-polymeric residuals in lens and packaging solution, solvents chosen o per EN ISO 18369-4:2017
{6}------------------------------------------------
### Biocompatibility testing:
An extractable residual analysis, in accordance with EN ISO 10993-18:2009, Annex C. was performed on both the current stenfilcon A and the modified stenfilcon A formulations. Based on the comparison of the extractable residual profiles, the residual concentration is maintained. Therefore, the similarity in the extractable residual levels is established, and additional biocompatibility testing was not required to complete the substantial equivalent determination.
## Clinical testing:
Test results provided in this 510(k) are sufficient to adequately characterize the subject device in terms of its physical/mechanical/optical characteristics when compared to the predicate. The results are equivalent, determining that additional clinical performance data was not required to complete the substantial equivalent determination. Additionally, the technical characteristics and manufacturing and sterilization processes of the subject lens are equivalent to stenfilcon A contact lens currently marketed by Cooper Vision; therefore, it was confirmed that no clinical data is required.
#### VIII. CONCLUSIONS:
Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physical/mechanical/optical properties of the lens. Based on the performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). The performance testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.