FLUOREX ULTRA (FLUSILFOCON E) RIGID GAS PERMEABLE CONTACT LENS
K991781 · G.T. Laboratories, Inc. · HQD · Jun 29, 1999 · Ophthalmic
Device Facts
Record ID
K991781
Device Name
FLUOREX ULTRA (FLUSILFOCON E) RIGID GAS PERMEABLE CONTACT LENS
Applicant
G.T. Laboratories, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Jun 29, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
The Fluorex® Ultra (flusilfocon E) Spherical Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic and which may exhibit astigmatism of 1.50 diopters or less that does not interfere with visual acuity. The Fluorex® Ultra (flusilfocon E) Translating Bifocal and Trifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased presbyopic eyes that are myopic or hyperopic and which may exhibit astigmatism of 3.50 diopters or less that does not interfere with visual acuity.
Device Story
Fluorex® Ultra (flusilfocon E) is a rigid gas permeable contact lens; material is a fluorine-containing polyacrylate-silicone copolymer. Lens acts as a refractive medium when placed on the cornea to focus light rays onto the retina. Intended for daily wear to correct visual acuity in myopic, hyperopic, and presbyopic patients. Device is prescribed by eye care professionals. Benefits include vision correction for specified refractive errors.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including gas permeability, Shore D hardness, wetting angle, water absorption, refractive index, specific gravity, and light transmission. Biocompatibility supported by non-cytotoxic results in agar overlay testing.
Technological Characteristics
Material: fluorine-containing polyacrylate-silicone copolymer (flusilfocon E). Standards: ANZI/ASTM D542-20 (refractive index), ANZI/ASTM 280.6.4.5 (light transmission), ANZI Z80.6; 5.6.1 (water absorption), ANZI/ASTM (Shore D hardness). Gas permeability (Dk): 60.0. Refractive index: 1.458. Specific gravity: 1.103. Wetting angle: 13.9°. Form factor: rigid gas permeable contact lens.
Indications for Use
Indicated for daily wear in non-aphakic, non-diseased eyes. Spherical lenses for myopia, hyperopia, and astigmatism ≤1.50D. Bifocal/trifocal lenses for presbyopia with myopia or hyperopia and astigmatism ≤3.50D.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Fluorex® 700 (flusilfocon A) Rigid Gas Permeable Contact Lenses (P880001)
Related Devices
K022128 — BOSTON EQUALENS II (OPRIFOCON A) RIGID GAS PERMEABLE CONTACT LENS · Bausch & Lomb, Incorporated · Aug 22, 2002
K071043 — BOSTON XO (HEXAFOCON A), BOSTON EO (ENFLUFOCON B) AND BOSTON ES (ENFLUFOCON A) RIGID GAS PERMEABLE CONTACT LENSES · Bausch & Lomb, Inc. · Aug 2, 2007
K991780 — FLUOREX 700 -UV (FLUSILFOCON A), FLUOREX 500-UV (FLUSILFOCON B) AND FLUOREX 300-UV (FLUSILFOCON C) RIGID GAS PERMEABLE C · G.T. Laboratories, Inc. · Jun 29, 1999
K221768 — Oxfore®100 (hexafocon A) Rigid Gas Permeable Contact Lenses · Paragon Vision Sciences, Inc. · May 5, 2023
K201194 — Acuity 200 (fluoroxyfocon A) Rigid Gas Permeable Contact Lens · Acuity Polymers, Inc. · Sep 8, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
# 510 (K) SUMMARY FOR FREEDOM OF INFORMATION
JUN 2 9 1999 Fluorex® Ultra (flusilfocon E) Material for Rigid Permeable Contact Lenses
- G. T. Laboratories, Inc 1. Submitted by: 2007 Johns Drive Glenview, IL 60025-1616
Contact Person:
- John M. Szabocsik, Ph.D. Szabocsik and Associates 203 N. Wabash, Ste 1200 Chicago, IL 60601 TEL: 312-553-0828
2. Date Prepared
- Contact Lens 3. Common/Usual
- Fluorex® Ultra (flusilfocon E) 4. Proprietary Name Rigid Gas Permeable Contact Lenses
May 17, 1999
- The lens material (flusilfocon E), 5. Device Description is a fluoro-silicate-acrylic-The Fluorex® Ultra copolymer. (flusilfocon E) Rigid Gas Permeable Contact Lens when placed on a cornea acts as a refracting medium to focus light rays onto the retina.
- The spherical lens is intended to 6. Intended Use be used in the correction of visual acuity in eyes that are myopic or Bifocal and trifocal hyperopic. designs are available for the correction of visual acuity in eyes that are also presbyopic.
- The sponsor considers these lenses 7. Equivalence: to be substantially equivalent to Fluorex® 700 (flusilfocon A) Rigid Gas Permeable Contact Lenses, approved under PMA #P880001. The attached table summarizes the comparison of the new material to the predicate device.
{1}------------------------------------------------
Fluorex® Ultra, 510(k) FOI, Pg 2
# Fluorex® Ultra compared to Fluorex® 700
ﻟﻢ ﺗﻢ ﻳﺘﻢ ﺗﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
| Material Type | Fluorex® 700 | Fluorex® Ultra |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Material Description | Fluoro-silicate-acrylic-copolymers. Tinted lenses contain one or more of the following dyes: D&C Green dye #6, D&C Yellow dye #10. D&C Violet #2, D&C Red dye #17, CI Solvent Yellow 18. | |
| Gas permeability (D k) (polarographic method) | 70.0 | 60.0 |
| Hardness (Shore D hardness ANZI/ASTM) | 85.5 | 86.0 |
| Wetting Angle (Wilhelmy plate method) | 15.3° | 13.9° |
| Water absorption (ANZI Z80.6; 5.6.1) | <1.0% | <1.0% |
| Refractive Index (ANZI/ASTM D542-20) | 1.457 | 1.458 |
| Specific Gravity | 1.097 | 1.103 |
| Light Transmission, (ANZI/ASTM 280.6.4.5) | 88.1% (clear)<br>75.0% (gray)<br>74.9% (blue)<br>72.6% (green) | >70% (aqua) |
| Actions | When placed on the human cornea, the hydrated lens acts as a corrective refracting medium to focus light rays on the retina. | |
| Chord Diameter* | 7.00 to 10.50 mm | |
| Center thickness | 0.10 to 1.00 mm | |
| Base Curve | 6.00 to 9.00 mm | |
| Powers | -20.00 to +12.00 D | |
*Parameters for spherical lenses
{2}------------------------------------------------
## Fluorex® Ultra, 510(k) FOI, Pg 3
### CHEMISTRY
The Fluorex® Ultra (flusilfocon E) material is a rigid gas permeable material for making contact lenses, consisting of a fluorine-containing polyacrylate-silicone blend, identical to the already approved Fluorex® 700 (flusilfocon A) material in component monomers, with differences only in the relative proportion of each component, which results in the difference in gas permeability (Dx). No monomers were found to be extracted from the material.
#### TOXICOLOGY
The Fluorex® Ultra (flusilfocon E) material was found to be noncytotoxic in an agar overlay cytotoxicity test.
#### MICROBIOLOGY
No testing was required, based on the similarity to the predicate device
#### CLINICAL
No testing was required, based on the similarity to the predicate device
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 9 1999
G.T. Laboratories, Inc. C/O John M. Szabocsik, Ph.D. 203 North Wabash Avenue Suite 1200 Chicago, IL 60601
Re: K991781 Trade Name: Fluorex® Ultra (flusilfocon E) Rigid Gas Permeable Contact Lens Regulatory Class: II Product Code: 86 HQD Dated: May 17, 1999 Received: May 25, 1999
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{4}------------------------------------------------
Page 2 - John M. Szabocsik, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Roerl. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) NUMBER (IF KNOWN)
Fluorex® Ultra (flusilfocon E) Rigid Gas DEVICE NAME Permeable Contact Lens
INDICATIONS FOR USE
The Fluorex® Ultra (flusilfocon E) Spherical Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic and which may exhibit astigmatism of 1.50 diopters or less that does not interfere with visual acuity.
The Fluorex® Ultra (flusilfocon E) Translating Bifocal and Trifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased presbyopic eyes that are myopic or hyperopic and which may exhibit astigmatism of 3.50 diopters or less that does not interfere with visual acuity.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use (Per 21 CFR 801.109)
Over-The-Counter-Use OR (Optional Format 1-2-96)
J. W. C. Brown, Ph.D.
(Division Sign-Off) Division of Ophthalmic Devices K991781 510(k) Number.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.