iTEAR Neurostimulator

DEN190026 · Olympic Ophthalmics, Inc. · QKV · May 1, 2020 · Ophthalmic

Device Facts

Record IDDEN190026
Device NameiTEAR Neurostimulator
ApplicantOlympic Ophthalmics, Inc.
Product CodeQKV · Ophthalmic
Decision DateMay 1, 2020
DecisionDENG
Submission TypeDirect
Regulation21 CFR 886.5305
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The iTEAR100 Neurostimulator is an electromechanical nerve stimulator device, indicated for temporary use (up to 30 days) to increase acute tear production during vibratory stimulation of the external nasal nerve in adults, under prescription of an eye care provider.

Device Story

iTEAR100 Neurostimulator is an electromechanical nerve stimulator; provides vibratory stimulation to external nasal nerve to increase acute tear production. Device is non-implantable; used by adult patients under prescription of eye care provider. Operation involves mechanical stimulation; device parameters include vibration amplitude, frequency, and pressure. Output is increased tear production. Clinical benefit is temporary relief of symptoms related to tear production. Safety and effectiveness supported by clinical performance testing, non-clinical performance testing (mechanical output, safety, use life), biocompatibility evaluation, and software verification/validation.

Clinical Evidence

Two clinical studies (CLP-002, CLP-007). CLP-007 (n=60, randomized, sham-controlled) showed statistically significant increase in Schirmer's test scores (mean change 21.7mm active vs -0.5mm sham, p<0.05). CLP-002 (n=108, prospective, open-label) demonstrated sustained acute tear production increase over 30 days. Safety profile included mild, transient adverse events (headache, sneezing, nose soreness); one serious unanticipated event (nausea/dizziness) reported.

Technological Characteristics

Hand-held electromechanical actuator; ABS housing/cantilever; lithium-ion battery. Operates at 250-270 Hz, 0.5-0.6 mm amplitude. Biocompatible per ISO 10993-5/10. Electrical/thermal safety per IEC 60601-1; EMC per IEC 60601-1-2. Firmware-controlled; non-sterile.

Indications for Use

Indicated for adults requiring increased acute tear production via vibratory stimulation of the external nasal nerve; temporary use up to 30 days; prescription use only.

Regulatory Classification

Identification

An electromechanical tear stimulator is a non-implantable device intended to increase tear production via mechanical stimulation.

Special Controls

In combination with the general controls of the FD&C Act, the electromechanical tear stimulator is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under anticipated conditions of use must evaluate tear production and all adverse events, including tissue damage, pain, headache, and discomfort. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted: (i) An assessment of mechanical output specifications, including vibration amplitude and frequency, pressure and force, and acoustic (noise level) properties; (ii) Mechanical safety testing to validate safeguards related to the pressure aspects of the device; and (iii) Use life testing. (3) Performance data must demonstrate the electrical safety, thermal safety, and electromagnetic compatibility (EMC) of all electrical components of the device. (4) All patient-contacting components of the device must be demonstrated to be biocompatible. (5) Software verification, validation, and hazard analysis must be performed. (6) Physician and patient labeling must include: (i) A detailed summary of the device's technical parameters; (ii) Instructions for use, including an explanation of all user-interface components and information regarding proper device placement; (iii) Information related to electromagnetic compatibility classification; (iv) Instructions on how to clean and maintain the device; (v) A summary of the clinical performance testing conducted with the device; (vi) Language to direct end users to contact the device manufacturer and MedWatch if they experience any adverse events with this device; and (vii) Information on how the device operates and the typical sensations experienced during treatment.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### DE NOVO CLASSIFICATION REQUEST FOR ITEAR100 NEUROSTIMULATOR #### REGULATORY INFORMATION FDA identifies this generic type of device as: Electromechanical tear stimulator. An electromechanical tear stimulator is a nonimplantable device intended to increase tear production via mechanical stimulation. NEW REGULATION NUMBER: 21 CFR 886.5305 CLASSIFICATION: Class II PRODUCT CODE: QKV #### BACKGROUND DEVICE NAME: iTEAR100 Neurostimulator SUBMISSION NUMBER: DEN190026 DATE DE NOVO RECEIVED: May 15, 2019 Olympic Ophthalmics, Inc. CONTACT: 400 NW Gilman Blvd #1370 Issaquah, WA 98027 #### INDICATIONS FOR USE The iTEAR100 Neurostimulator is an electromechanical nerve stimulator device, indicated for temporary use (up to 30 days) to increase acute tear production during vibratory stimulation of the external nasal nerve in adults, under prescription of an eye care provider. #### LIMITATIONS The sale, distribution, and use of the iTEAR100 Neurostimulator is restricted to prescription use in accordance with 21 CFR 801.109. Patient training is required on the proper use of the device before home use. The safety and effectiveness of the iTEAR100 Neurostimulator for the treatment of dry eye disease or for the improvement in dry eye symptoms have not been established. {1}------------------------------------------------ The device increases tear production during neurostimulation, i.e., tearing was assessed only during stimulation. The clinical study was not designed to evaluate any changes in nerve sensitivity. Clinical study results demonstrate a trend of decreased effectiveness (tear production) over time. The mechanism for this decrease has not been identified and was not analyzed as part of the study. # PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS. ## DEVICE DESCRIPTION iTEAR100 Neurostimulator is a hand-held, portable electromechanical actuator intended to acutely increase tear production through vibratory stimulation of the external nasal nerve in patients. The device is battery-operated with a single vibratory tip. As the effective tip is pushed against the tissue, the beam deflects inward until it is deflected fully into the device. The device is comprised of a direct current vibration motor, vibrating beam(cantilever), a control circuit with micro SD card, tactile switch. The motor is attached to a cantilever beam made from acrylonitrile butadiene styrene (ABS), a control circuit to charge the battery and deliver power to the motor, and a housing also made from ABS. The motor and cantilever beam move together to produce a linear vibration at the tip of the cantilever (oscillating tip). A lithium-ion rechargeable battery is supplied inside the device, and is not removable or replaceable. A micro USB port on the device allows for charging. The device is locked by firmware from use after a predetermined number of stimulation days has been triggered. The frequency of operation without any force applied is approximately 270 Hz. The amplitude of movement is approximately 0.5 mm without force applied. In the optimal force range, the cantilever tip is depressed approximately 1 mm and the frequency is approximately 250 Hz. Below is a list of stimulation parameters for the device: | Parameter | Control - In air | With Force | |---------------------------------------|------------------|------------| | Frequency | 270 Hz | 250 Hz | | Amplitude | 0.6 mm | Not tested | | Acceleration (g) | 50 | 5 | | Tip Retraction (maximum force) (100%) | <6N | N/A | | Acoustic | <35 db | <35 db | {2}------------------------------------------------ Image /page/2/Picture/0 description: A hand is holding a white, triangular-shaped device with a blue oscillating tip. The device has several black dots on its surface. The text "Oscillating Tip" is written next to the device, with a blue arrow pointing to the tip. The device appears to be in a white box. # Figure 1. iTEAR100 Neurostimulator The device activates tear production through stimulation of the nasolacrimal reflex. Stimulation mechanically activates external nasal nerve and initiates the nasolacrimal reflex, resulting in tear secretion. To produce the intended effect, the vibratory tip of the device should be applied to the lateral aspect of the nose for several seconds. Image /page/2/Picture/3 description: The image shows two diagrams of a human face. The diagram on the left shows a side view of a person's face with a blue circle around the nose. The diagram on the right shows a frontal view of a person's face with the nerves labeled. The nerves labeled are the infratrochlear nerve, supratrochlear nerve, external branch of anterior ethmoidal nerve, and infraorbital nerve. Figure 2 is a schematic showing correct use of the device ## SUMMARY OF NONCLINICAL/BENCH STUDIES ## BIOCOMPATIBILITY/MATERIALS All patient contacting portions (all outer surfaces) of the device were assessed for cytotoxicity, skin irritation, and skin sensitization potential. Evaluation of these biocompatibility endpoints for devices that contact intact skin is consistent with recommendations in the June 2016 FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process." These biocompatibility tests were performed per ISO 10993-5:2009 (MEM elution cytotoxicity) and ISO 10993-10:2010 (intracutaneous skin irritation, guinea pig maximization skin sensitization). For intracutaneous skin irritation and guinea pig maximization testing, both polar and non-polar {3}------------------------------------------------ test article extracts were evaluated. For the intracutaneous skin irritation testing, observations were recorded to 72 hours after injection. For the guinea pig maximization testing, observations were made out to 72 hours after the challenge phase. Results demonstrated acceptable performance. # SHELF LIFE/STERILITY The iTEAR100 Neurostimulator is provided non-sterile. Cleaning and maintenance instructions of the stimulator components of the device are included in the labeling. # ELECTROMAGNETIC CAPABILITY (EMC) AND ELECTRICAL SAFETY Testing for the iTear100 device conformed to the following electromagnetic compatibility, electrical, mechanical, and thermal safety standards: - ES 60601-1:2005/(R)2012. A1:2012. C1:2009/(R)2012 and A2:2010/(R)2012 o (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance (IEC 60601-1:2005, MOD) - IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment Part 1-2: General o Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests 5-89 - . 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability - . IEC 62133:2012 (Second Edition) Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications ## PERFORMANCE TESTING - BENCH The sponsor has provided the test methods and results on multiple samples of their device for the following verification tests. This testing is intended to address the risks of physical injury to the local muscles, nerves and skin due to repeated stimulation. Excessive acceleration, force, amplitude and frequency could all cause injury and or abrasions. | Test | Purpose | Method | Acceptance Criteria | Results | |-----------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------|---------| | Acoustic Levels | To assess the acoustic output | Sound level meter in acoustic isolation chamber | (b)(4) | passed | | Safeguards related to pressure/force against tissue | To assess force output performance when in operation | Load cell was lowered onto vibrating tip of device until tip depressed in housing | | passed | {4}------------------------------------------------ | Vibrational<br>amplitude/ frequency | To assess the motion of<br>the applicator tip of<br>device | Use strobe to determine<br>frequency and direct<br>measurement of amplitude | (b)(4) | passed | |-------------------------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------|--------| | Thermal<br>performance | To determine operating<br>temperature of device | Thermocouples were<br>applied to device and<br>temperature measured with<br>device in operation | | passed | | Use Life * | To assess fatigue of the<br>cantilever | Cantilever deflection<br>loading measured before<br>and after life cycle test | | passed | * The sponsor has performed fatigue testing on their device for " " hours which simulates (b)(4) bilateral treatments. This resulted in no change in mechanical parameters or physical characteristics of the device. #### SOFTWARE All components of the device are controlled/monitored by software, which is responsible for the functionality, user interface, safety checks and performance accuracy. The software is considered to be a moderate level of concern (LOC) because inadvertent software errors could result in injury to the patient. The following device functions of the iTEAR100 Neurostimulator are controlled by the software: - 1. Software detects the state of the switch and operate the vibration motor. - 2. When in use the software modulates the motor on/off signal creating a stutter in the motor motion. This stuttering serves as a treatment duration indicator for the user. - 3. Logs use data to memory SD Card. - 4. Detect battery level. - 5. Provide USB detect for charging. This disables the unit from being used when connected to USB. - 6. LED control. LED is flashed signaling battery state. - 7. Disable device after 30 Day of use The submission contained all the elements of software documentation corresponding to the "Moderate" level of concern, as outlined in the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Adequate documentation describing the software/firmware, software specifications, architecture design, software development environment, traceability, revision level history, unresolved anomalies and cybersecurity provide the foundation that the software will operate in a manner as described in the specifications. Hazard analysis was performed to characterize software risks including device malfunction, measurement related errors and cybersecurity. The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory results. ## SUMMARY OF CLINICAL INFORMATION {5}------------------------------------------------ Two clinical studies. CLP-OO2 and CLP-OO7. were completed with the iTEAR100 Neurostimulator™ and are discussed below. # CLP-002 Study CLP-OO2 was a prospective, open-label, single-arm, multi-center (8 US sites), 2-stage study. The first stage of this pivotal trial included 30 days of follow-up while using the device at home and included all enrolled subjects. After 30 days, subjects were given the option to participate in an additional 150 days of follow-up while using the device (180 days of follow-up in total). The subject was provided a take-home device to perform the study treatment per the instructions for use. It was recommended that subjects use the study device on each side of the nose for 30 seconds on each side twice per day for the 30-day follow-up period. In the extended follow up, subjects were asked to use the device at least once every other day for the 150 days in the extended follow-up period. Below are the inclusion and exclusion criteria: Inclusion Criteria - 1. Twenty-one (21) years of age or older - 2. Schirmer test with anesthetic of (b)(4) in at least one eve - 3. Ability to produce tears upon training with (b)(4) change in Schirmer score compared to baseline in at least one eye (the technique for this assessment is provide below) - 4. In the opinion of the investigator, subject is in good general health and free of any condition that could impair study participation or ocular evaluation - 5. Subject is willing and able to give written informed consent and commits to comply with study requirements Exclusion Criteria - 1. Sjogren's syndrome or other rheumatologic condition - 2. Intraocular surgery within 6 months of Visit 1 - 3. Intraocular or periocular injection within 6 months of Visit 1 - 4. Used intranasal neurostimulation within two months of Visit 1 or plans to use it during the study - 5. Lid function abnormalities - 6. Any acute infectious or non-infectious ocular condition of the anterior or posterior segments in either eye within 30 days of Visit 1 - 7. Diseases/conditions of ocular surface associated with clinically significant scarring/destruction of conjunctiva/cornea - 8. Subject has any of the following conditions: - · History of facial nerve palsy - · History of neuromuscular disorder - · Uncontrolled ocular or systemic disease - · Other clinically significant local skin condition (e.g., skin infection) at target treatment site - 9. Participation in any clinical trial with a new active substance or a new device within 30 days of Visit 1 (with the exception of the devices to be used in the study described herein). {6}------------------------------------------------ Study Visits were performed at Baseline (Day 0). Follow-up Call (day 3). Day 30. Dav 90, Day 180. Study Visits were performed at Baseline (Day 0), Follow-up Call (day 3), Day 14, Day 30, Day 90, Day 180. Clinical assessments included: - o Corrected distance visual acuity (Baseline, Day 30, Day 180) - Corneal and conjunctival staining (Baseline, Day 14, Day 30, Day 90, Day 180) ● - o Schirmer test (Baseline, Day 14, Day 30, Day 90, Day 180) - o Neurological exam (Baseline, Day 30, Day 180) - o Adverse event recording (all visits) The primary endpoint was statistically significant improvement from baseline to Day 30 on the Schirmer Index (SI). Briefly, anesthesia was applied. followed by placement of the Schirmer strip in the subject's eye and measurement of the distance a subject's tears traveled along it. After 5 minutes, iTEAR100 stimulation was applied for 30 seconds on each side (with the strip in place), and the was re-measured 5 minutes afterward. The SI is the difference between the 2 measurements (pre- and post- stimulation). Subjects were allowed to maintain medication regimens if the regimens were considered to be stable and unlikely to change over the course of 180 days. Subjects were chosed on their ability at day 0 to respond to neurostimulation and the company recommends a trial in the office prior to writing a prescription. The study enrolled 108 subjects, of which 101 subjects completed the 30 days of follow-up. Eighty (80) subjects agreed to continue in the extended follow-up phase of the study, of which 58 reached the final 180-day endpoint. Demographic data and subject accountability data for the study are provided below: Table 1- Demographic Characteristics | | | n | |-----------|------------------------|-----| | Sex | F | 82 | | | M | 26 | | Age | <30 | 6 | | | 30-50 | 18 | | | 50-70 | 60 | | | >70 | 24 | | Ethnicity | Hispanic/Latino | 5 | | | Not<br>Hispanic/Latino | 101 | | | NA | 2 | | Race | White | 87 | | | Black | 2 | | | Other | 3 | | | NA | 16 | | | All | 108 | {7}------------------------------------------------ # Table 2 - Subject Accountability | | Main Study | | | | Optional Extended<br>Follow-up:<br>At Day 30, 80<br>subjects chose to<br>enroll in the<br>extended follow-up | | |--------------------------------------|------------|-----------------|--------|--------|--------------------------------------------------------------------------------------------------------------|---------| | | Baseline | Day 3<br>(call) | Day 14 | Day 30 | Day 90 | Day 180 | | Theoretical | 108 | 108 | 108 | 108 | 80 | 80 | | Death | 0 | 0 | 0 | 0 | 0 | 0 | | Withdrawal<br>(cumulative) | 0 | 0 | 4 | 6 | 6 | 9 | | Lost to<br>Follow Up<br>(cumulative) | 0 | 0 | 1 | 1 | 7 | 13 | | Missing | 0 | 1 | 2 | 0 | 0 | 0 | | Expected | 108 | 108 | 104 | 102 | 74 | 71 | | ActualA | 108 | 103 | 97 | 88 | 51 | 40 | | ActualB | 108 | 107 | 101 | 101 | 67 | 58 | | % Follow Up<br>ActualA | 100% | 95.4% | 89.8% | 81.5% | 63.8% | 50.0% | | % Follow Up<br>ActualB | 100% | 99.1% | 93.5% | 93.5% | 83.8% | 72.5% | Expected is equal to Theoretical minus Deaths minus Withdrawals Actual is all subjects with data available for primary endpoint and within window is all subjects with data available for primary endpoint Follow-up Actual is equal to ActualA divided by Theoref Follow-up Actual® is equal to Actual® divided by Theoretical The stimulated Schirmer's test score was greater than the unstimulated test score at each visit during the 30-day follow-up period (Table 3). There was also an increase in the pre-stimulation Schirmer score during the course of the study. {8}------------------------------------------------ Image /page/8/Figure/0 description: The image is a bar graph that shows pre-stimulation and post-stimulation values at baseline, day 14, and day 30. The graph also shows the difference between pre- and post-stimulation. At baseline, the pre-stimulation value is around 6, and the post-stimulation value is around 28. At day 30, the pre-stimulation value is around 9, and the post-stimulation value is around 19. Table 3 - Change in tear production (Schirmer score) Safety information for the entire 180-day follow-up period is provided below. Table 4 - Ophthalmic Adverse Events | Event<br>Description | Day<br>to<br>AE | Duration<br>(days) | Severity | Impact<br>of<br>Device | Action Taken | |-----------------------------------------------|-----------------|--------------------|----------|------------------------|--------------------------------| | Inflammation due to Schirmer strip | 13 | 14 | Mild | No change | Drug Rx | | Blurred vision | 1 | 232* | Mild | No Change | New Contact Lenses | | Eye Floater | 92 | 1 | Mild | No Change | None | | Corneal Medicamentosa | 91 | 60 | Mild | No change | Decrease artificial tear usage | | Corneal Neovascularization | 180 | 21 | Moderate | No change | None | | Dry Eye Pain | 2 | 19 | Mild | No change | Artificial Tears | | Eyelid irritation due to Schirmer strip | 0 | 1 | Mild | No change | None | | Stye on lower lid OD | 1 | | Mild | No change | Drug Rx | | Vitreous Hemorrhage OS | 64 | 16 | Mild | Discontinued | None | | Severe itching and sensitive eyes papillae OU | 5 | 8 | Moderate | Discontinued | Drug Rx | | Eye infection OU | 55 | 7 | Mild | No change | None | | Eye infection OU | 74 | 8 | Mild | No change | None | | Eye infection OU | 95 | 0 | Mild | No change | None | | Corneal Abrasion due to Schirmer strip | | 23 | Moderate | No change | Drug Rx | | Sjogren's syndrome | 30 | Ongoing | Mild | No change | Referred to rheumatologist | *The cause of the blurry vision was poor contact lens fit and the days to resolution represents the time of the phone call inquiry that was made outside the study follow-up period, since the subject could not recall the exact date of the new lens fitting. {9}------------------------------------------------ There were 2 events that were definitely related to the device: one event described as headache, sneezing, and ticking sensation (onset at 1 day from study enrollment); the other was described as intermittent nose soreness (onset at 85 days from study enrollment). There were 7 events that were possibly related to the device (Table 5). | Event<br>Description | Day<br>to<br>AE | Duration<br>(days) | Severity | Impact<br>of<br>Device | Action Taken | |--------------------------------------------------------|-----------------|--------------------|----------|------------------------|-----------------------------------------| | Inflammation due<br>to Schirmer strip | 13 | 14 | Mild | No change | Drug Rx | | Blurred vision | 1 | 232* | Mild | No Change | New<br>Contact<br>Lenses | | Eye Floater | 92 | 1 | Mild | No Change | None | | Corneal<br>Medicamentosa | 91 | 60 | Mild | No change | Decrease<br>artificial<br>tear<br>usage | | Corneal<br>Neovascularization | 180 | 21 | Moderate | No change | None | | Dry Eye Pain | 2 | 19 | Mild | No change | Artificial Tears | | Eyelid irritation<br>due to Schirmer<br>strip | 0 | 1 | Mild | No change | None | | Stye on lower lid<br>OD | 1 | | Mild | No change | Drug Rx | | Vitreous<br>Hemorrhage OS | 64 | 16 | Mild | Discontinued | None | | Severe itching and<br>sensitive<br>eyes<br>papillae OU | 5 | 8 | Moderate | Discontinued | Drug Rx | | Eye infection OU | 55 | 7 | Mild | No change | None | | Eye infection OU | 74 | 8 | Mild | No change | None | | Eye infection OU | 95 | 0 | Mild | No change | None | | Corneal Abrasion<br>due to Schirmer<br>strip | | 23 | Moderate | No change | Drug Rx | | Sjogren's<br>syndrome | 30 | Ongoing | Mild | No change | Referred<br>to<br>rheumatologist | Table 5 - Adverse Events Possibly Related to Device *The cause of the blurry vision was poor contact lens fit and the days to resolution represents the time of the phone call inquiry that was made outside the study follow-up period, since the subject could not recall the exact date of the new lens fitting. Headache and dizziness were the most common related events reported in the study. Seven (7) of the 9 related adverse events were considered mild, 1 was considered moderate, and 1 was considered severe. Most related events were transient and resolved within approximately 1 week. After a single device treatment, one subject experienced continuous nausea, dizziness, {10}------------------------------------------------ lightheaded and headache for 30-day duration. The subject was instructed to seek neurological evaluation, stop device use, and discontinue from the study. This adverse event was considered to be a serious unanticipated adverse event, possibly related to the device. Four subjects experienced a loss of 2 lines or more in visual acuity during treatment. For 3 of the 4 subjects, loss of visual acuity was attributed to fluctuations in the disease, and/or the fact that subjects were asked not to use artificial tears or other treatments when follow-up visits were scheduled. Pinhole correction and manifest refraction was not attempted or used for any of these subjects. The subjects' vision returned to baseline or 20/20 as fast as 2 weeks for 1 subject and within 180 days for the other 2 subjects. The last subject experienced a loss of visual acuity when the 30 day follow-up visit was performed without correction, but demonstrated no change in visual acuity when the exam was performed with correction, at 180 days. Therefore, this subject's change in visual acuity was attributed to a lack of correction at the 30 day visit. All 4 subjects' vision returned to baseline during the follow-up period. The investigators believed that the alterations in vision were due to temporary worsening of dry eye disease, however, causality could not be definitively determined due to limitations in the study design. # CLP-007 Study CLP-007 was a multi-center, nonsignificant-risk, prospective, double-masked, randomized, sham controlled, single visit clinical trial that enrolled 60 subjects. The sham group received an iTEAR100 Neurostimulator device which looked identical to and made noise similar to a fully functional device but had a tip that did not vibrate. All subjects received cross-over treatment with the active device at the conclusion of the study procedures. Regardless of whether they were using the sham or active device, subjects were instructed to apply the device tip against the skin of the nose at the junction of the nasal cartilage and the nasal bone. The application duration was 30 seconds or less for each side of the nose. The inclusion and exclusion criteria are provided below: Inclusion Criteria - 1. 18 years of age or older; - 2. Willing and able to provide an English language written informed consent; - 3. Able to safely use the study device and be free of any condition that, in the opinion of the investigator, could impair study participation. # Exclusion Criteria - 1. Presence of clinically significant nasal/facial skin conditions (e.g., infection, ulceration, wound): - 2. Under arrest or was otherwise in custody; - 3. Presence of any other condition, which in the judgment of the principal investigator would prevent a potential subject from safely completing the study or tolerating device use, such as mental illness, dementia, severe agitation, etc.). No study-mandated tests were required for enrollment. The schedule of assessments and procedures for the study is shown in the table below. {11}------------------------------------------------ Table 6 - Clinical Assessments | | Schedule | |---------------------------------------------------------------------------------------------|---------------------| | Assessment | Screening/Day<br>0a | | Eligibility screening & IC | X | | OSDI questionnaire | X | | Demographic information | X | | Focused DES history | X | | Baseline nasal skin exam & facial neurological exam | X | | Baseline visual acuity (OS) | X | | Baseline slit lamp examination (SLE) with emphasis on comeal edema (OS) | X | | Baseline blood pressure (BP), heart rate (HR), pulse oximetry | X | | Baseline Schirmer's test (OD) | X | | Baseline meibomian gland expression score (OS) | X | | Randomization & experimental device training/treatmentb | X | | Repeat Schirmer's test (OD) with new Schirmer strip during device treatmentb,c | X | | Repeat BP, HR, pulse oximetry during Schirmer's test (at 2.5 & 5 min)b | X | | Repeat nasal skin exam & facial neurological exam | X | | Repeat visual acuity (OS) | X | | Repeat SLE with emphasis on corneal edema (OS) | X | | Repeat meibomian gland expression score (OS) | X | | Cross-over treatment with fully active iTEAR device & qualitative assessment of<br>response | X | | PI usability survey | X | | Subject usability survey | X | | Record adverse events | X | a Screening and Day 0 were the same b Second investigator to perform; initial investigator was out of the room. · Device treatment begins immediately after placement of Schirmer's test strip The primary endpoint of this study was the mean within-subject change in the Schirmer's test. score post vs. pre neurostimulation in the active treatment vs. sham groups in the primary analysis population (i.e., subjects with baseline Schirmer's test score ≤10mm). The replacement Schirmer technique was used for this study, under which a Schirmer strip is applied to the subject's eye to assess pre-stimulation tear production, and then a second strip is inserted to assess post-stimulation tear production. {12}------------------------------------------------ The safety of the device was evaluated by the incidence of device-related adverse events and serious adverse events, as well as changes in the following measures: - Best corrected Snellen distance visual acuity (VA); . - . Ophthalmic exam with emphasis to determine the presence of corneal edema; - . Nasal skin examination and facial neurologic examination; - . Hemodynamic parameters (blood pressure, heart rate, pulse oximetry). Below are the study demographics: Table 8 - Baseline Demographics | Summaryª | Randomized Treatment | | | |----------------------------------------|----------------------|--------------|---------------------| | | Active (N=28) | Sham (N=32) | All Subjects (N=60) | | Sex, n (%) | 16 (57.1%) | 24 (75.0%) | 40 (66.7%) | | Female | 16 (57.1%) | 24 (75.0%) | 40 (66.7%) | | Male | 12 (42.9%) | 8 (25.0%) | 20 (33.3%) | | Age (Years) | | | | | Mean (SD) | 49.7 (15.99) | 50.0 (16.25) | 49.9 (16.00) | | Median | 51 | 50 | 51 | | Min, Max | 24, 79 | 21, 77 | 21, 79 | | Ethnicity, n (%) | | | | | Hispanic or Latino | 3 (10.7%) | 7 (21.9%) | 10 (16.7%) | | Not Hispanic or Latino | 25 (89.3%) | 25 (78.1%) | 50 (83.3%) | | Race, n (%) | | | | | Asian | 5 (17.9%) | 7 (21.9%) | 12 (20.0%) | | Black | 2 (7.1%) | 1 (3.1%) | 3 (5.0%) | | Caucasian | 18 (64.3%) | 21 (65.6%) | 39 (65.0%) | | Other | 3 (10.7%) | 3 (9.4%) | 6 (10.0%) | | Subjects using Eye<br>Treatment, n (%) | 21 (75.0%) | 25 (78.1%) | 46 (76.7%) | Each randomized subject underwent all study assessments, thus subject accountability was 100%. In the primary analysis population, there was a statistically significant increase in Schirmer's test scores in the active group (mean within-subject change pre- to post-stimulation = 21.7mm) compared to sham (mean within-subject change pre- to post-stimulation = - 0.5mm) with a between group difference of 22.2mm (SD 2.61) favoring the active device, as shown in in the figure below. The bars included in the figure show the standard deviation for the mean change reported, which indicates the degree of variability for the with-in subject change scores. {13}------------------------------------------------ Image /page/13/Figure/0 description: The image is a bar graph titled "Mean Change in Schirmer From Baseline". The y-axis is labeled "Schirmer Change Post-stimulation (mm)" and ranges from -10 to 35. There are two bars, one labeled "Active, 21.7" which has a value of approximately 21.7, and the other labeled "Sham, -0.5" which has a value of approximately -0.5. Figure 9 - Schirmer's Test Score Change in Active Compared to Sham Groups There were no device-related adverse events or serious adverse events reported in this study. One unrelated adverse event involved a near-syncopal episode that occurred following the administration of topical proparacaine anesthetic to the subject's right eye prior to placement of the baseline Schirmer's test strip. The subject was randomized but had not received treatment with either a sham or an active study device. She was terminated from any further study participation and was monitored by the investigator at the site until her symptoms had resolved. Additionally, there were no observed significant changes in post vs. pre-test visual acuity, corneal edema, facial neurologic examinations, or hemodynamic parameters. # Pediatric Extrapolation In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population. # LABELING The professional and patient labeling are adequate and meet the requirements of 21 CFR 801.109. The labels summarize the clinical trial results that characterized the probable benefit and the identified risks of the device. including tissue damage, pain. headache, and discomfort. Both guides contain requirements for use by prescription only and proper patient training, Indications for Use, contraindications, device description, technical parameters, warnings, precautions, potential complications, instructions for use (including an explanation of all userinterface components and information regarding proper device placement), recommended stimulation schedule, instructions for device maintenance/cleaning, summary of clinical trials, information related to electromagnetic compatibility, language to direct end users to contact the device manufacturer and MedWatch if they experience any adverse events with this device, expected service life, disposal & replacement, environmental operating conditions, electrical specifications, and symbols & markings. # RISKS TO HEALTH The table below identifies the risks to health that may be associated with use of an electromechanical tear stimulator and the measures necessary to mitigate these risks. {14}------------------------------------------------ | Identified Risks | Mitigation Measures | |------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Tissue damage due to<br>overstimulation/understimulation<br>or mechanical injury, device<br>breakage | Clinical performance testing<br>Non-clinical performance testing<br>Software verification, validation, and hazard analysis<br>Labeling | | Adverse tissue reaction | Biocompatibility evaluation<br>Labeling | | Electrical shock or burn | Electrical, thermal, and mechanical safety testing<br>Software verification, validation, and hazard analysis<br>Labeling | | Interference with other devices | Electromagnetic compatibility (EMC) testing<br>Software verification, validation, and hazard analysis<br>Labeling | | Pain, headache, or discomfort | Clinical performance testing<br>Non-clinical performance testing | | Insufficient tear production | Clinical performance testing | Table 9 - Identified Risks to Health and Mitigation Measures ## SPECIAL CONTROLS In combination with the general controls of the FD&C Act, the electromechanical tear stimulator is subject to the following special controls: - 1. Clinical performance testing under anticipated conditions of use must evaluate tear production and all adverse events, including tissue damage, pain, headache, and discomfort. - 2. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted: - a. An assessment of mechanical output specifications including vibration amplitude and frequency, pressure and force, and acoustic (noise level) properties; - b. Mechanical safety testing to validate safeguards related to the pressure aspects of the device; and - c. Use life testing. - 3. Performance data must demonstrate the electrical safety, thermal safety, and electromagnetic compatibility (EMC) of all electrical components of the device. - 4. All patient-contacting components of the device must be demonstrated to be biocompatible. - 5. Software verification, validation and hazard analysis must be performed. - 6. Physician and patient labeling must include: - a. A detailed summary of the device's technical parameters; {15}------------------------------------------------ - Instructions for use, including an explanation of all user-interface components and b. information regarding proper device placement; - Information related to electromagnetic compatibility classification: C. - d. Instructions on how to clean and maintain the device; - A summary of the clinical performance testing conducted with the device; e. - Language to direct end users to contact the device manufacturer and MedWatch if f. they experience any adverse events with this device; and - g. Information on how the device operates and the typical sensations experienced during treatment. ## BENEFIT-RISK DETERMINATION The risks of the device are based on non-clinical laboratory data, as well as data collected in clinical trials described above. The device exhibited an acceptable safety profile for up to 30 day use in the clinical studies which were conducted. One device-related serious adverse event was observed; this event resolved without treatment. Non-serious device-related adverse events were few in number and all were self-limited. The probable benefits of the device are also based on nonclinical laboratory data, as well as data collected in clinical trials. Compared to a sham control, the device was shown to provide a statistically significant increase in acute tear production during a single stimulation. Results from an additional clinical study suggest that a temporary increase in acute tear production may be seen in response to repeated device stimulation up to 30 days. ## Patient Perspectives This submission did not include specific information on patient perspectives for this device. #### Benefit/Risk Conclusion In conclusion, given the available information above, the probable benefits outweigh the probable risks for the iTEAR100 Neurostimulator. The device provides benefits, and the risks can be mitigated by use of general controls and the identified special controls. #### CONCLUSION The De Novo request for the iTEAR100 Neurostimulator is granted and the device is classified as follows: Product Code: OKV Device Type: Electromechanical tear stimulator Class: II Regulation Number: 21 CFR 886.5305
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...