← Product Code [ORZ](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/ORZ) · K172645

# iLux Instrument, iLux Disposable (K172645)

_Tear Film Innovations, Inc. · ORZ · Dec 26, 2017 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/ORZ/K172645

## Device Facts

- **Applicant:** Tear Film Innovations, Inc.
- **Product Code:** [ORZ](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/ORZ.md)
- **Decision Date:** Dec 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5200
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The iLux System is indicated for the application of localized heat and pressure therapy in adult patients with chronic diseases of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye.

## Device Story

iLux System is a handheld device for eye care professionals (ECP) to treat meibomian gland dysfunction. It consists of a handheld instrument and a single-use sterile disposable component. The ECP positions the disposable behind the eyelid and uses a built-in magnifier to view the eyelid margin. The instrument uses lime-green and infrared LEDs to heat eyelid tissue to 40–42 °C, melting meibum. The ECP then applies manual compression to express the meibum. The instrument displays real-time inner eyelid temperature and compression force, allowing the ECP to titrate therapy. The device is powered by a rechargeable lithium-ion battery. By unclogging glands, the device aims to alleviate symptoms of evaporative dry eye.

## Clinical Evidence

A randomized, 1:1 controlled clinical study of 142 subjects (284 eyes) compared iLux to LipiFlow. Primary endpoints (MGS, TBUT) and secondary endpoint (OSDI) showed significant improvement from baseline in both groups with no significant difference between groups at weeks 2 and 4. Safety endpoints included pain, discomfort, ocular surface staining, IOP, and BSCVA. Four device-related AEs occurred in the iLux arm, all self-limited and transient. Results support non-inferiority.

## Technological Characteristics

Handheld instrument with single-use sterile disposable. Heating via lime-green and infrared LEDs. Powered by rechargeable lithium-ion battery. Disposable contact material: silicone. Sterilization: Ethylene Oxide (disposable). Software Level of Concern: Moderate. Connectivity: Standalone. Safety: Complies with IEC 60601-1 and 60601-2.

## Regulatory Identification

An eyelid thermal pulsation system is an electrically-powered device intended for use in the application of localized heat and pressure therapy to the eyelids. The device is used in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye. The system consists of a component that is inserted around the eyelids and a component to control the application of heat and pressure to the eyelids.

## Special Controls

In addition to the general controls of the Act, the LipiFlow® Thermal Pulsation System is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing should validate electromagnetic compatibility (EMC) and safety of exposure to non-ionizing radiation;
(2) Design, description, and performance data should validate safeguards related to the temperature and pressure aspects of the device, including during fault conditions;
(3) Performance data should demonstrate the sterility of patient-contacting components and the shelf-life of these components;
(4) The device should be demonstrated to be biocompatible; and
(5) Performance data should demonstrate that any technological changes do not adversely effect safety and effectiveness.

## Predicate Devices

- LipiFlow Thermal Pulsation System ([K133127](/device/K133127.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 26, 2016

Tear Film Innovations, Inc. John Slate Vice President, R&D 12625 High Bluff Drive, Suite 107 San Diego, CA 92130

Re: K172645

Trade/Device Name: iLux System Regulation Number: 21 CFR 886.5200 Regulation Name: Eyelid Thermal Pulsation System Regulatory Class: Class II Product Code: ORZ Dated: November 28, 2017 Received: November 29, 2017

Dear John Slate:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Srinivas Nandkumar -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172645

Device Name iLux System

Indications for Use (Describe)

The iLux System is indicated for the application of localized heat and pressure therapy in adult patients with chronic diseases of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Tearfilm Innovations. The logo features a stylized eye on the left, with a blue and green gradient. To the right of the eye is the word "tearfilm" in blue and green, with "tear" in blue and "film" in green. Below "tearfilm" is the word "innovations" in blue.

## 510(K) SUMMARY - K172645

#### Submitter Information

| Company Name:                 | Tear Film Innovations, Inc.                                                         |
|-------------------------------|-------------------------------------------------------------------------------------|
| Company Address:              | 12625 High Bluff Drive, Suite 107<br>San Diego, CA 92130                            |
| Company Phone:                | (844) 458-9776                                                                      |
| Company Facsimile:            | N/A                                                                                 |
| Contact Person:               | John Slate PhD<br>Vice President of Research and Development<br>jslate@tearfilm.com |
| Date:                         | December 12, 2017                                                                   |
| Device Identification         |                                                                                     |
| Device Trade Name:            | iLux® System                                                                        |
| Common Name:                  | N/A                                                                                 |
| Classification Name(s):       | Eyelid Thermal Pulsation System                                                     |
| Classification Regulation(s): | 21 CFR 886.5200                                                                     |
| Device Class:                 | Class II                                                                            |
| Product Code(s):              | ORZ                                                                                 |
| Advisory Panel:               | Ophthalmic                                                                          |

#### Identification of Predicate Devices

The iLux System is substantially equivalent to the following device, which was cleared for commercial distribution in the United States:

- The LipiFlow Thermal Pulsation System manufactured by TearScience, Inc. and cleared for commercial distribution under 510(k): K133127.
#### Device Description

The iLux System is a medical device intended for use by Licensed Eye Care Professionals (ECP) to apply localized heat and pressure therapy to adult patients' lower and/or upper eyelids. The il.ux System consists of a hand-held Instrument coupled to a single-use sterile Disposable component that is positioned behind the eyelid. The iLux System allows the ECP to view the eyelid margin through a magnifier, then warm the eyelid tissue to a target range of 40 to 42 °C to melt the meibum blocking the gland orifices and then apply compression to the eyelid to express

{4}------------------------------------------------

the melted meibum through the orifices. The inner eyelid temperature and the amount of force applied to the eyelid are both displayed on the Instrument allowing the clinician to titrate the ideal amount of heating and compression to optimize unclogging of the blocked glands.

The Disposable is a sterile, single-use component that is attached to the iLux Instrument. It includes all parts intended to contact the patient's eyelid.

The iLux Instrument is a handheld device that allows the ECP to view, heat, and compress the portion of the eyelid that is in contact with the Disposable. The source of heating is lime-green and infrared optical radiation produced by LEDs in the Instrument. The ilux Instrument is powered by a rechargeable lithium ion battery. The software Level of Concern is Moderate.

Ancillary accessories include the charging stand with AC power supply.

#### Indications for Use

The iLux System is indicated for the application of localized heat and pressure therapy in adult patients with chronic diseases of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye.

| Comparison Feature                                                                                      | The iLux® System<br>Tear Film Innovations Inc. | LipiFlow® Thermal Pulsation System<br>TearScience, Inc. |
|---------------------------------------------------------------------------------------------------------|------------------------------------------------|---------------------------------------------------------|
| Single Patient Use?                                                                                     | Yes                                            | Yes                                                     |
| Provided Sterile?                                                                                       | Yes                                            | Yes                                                     |
| Method of sterilization, SAL<br>(Pertinent to Disposable)                                               | Ethylene Oxide, SAL 10-6                       | Gamma, SAL 10-6                                         |
| Biocompatibility Testing (ISO<br>10993-1), of Patient Contact<br>Materials<br>(Pertinent to Disposable) | Yes, Silicone Cover                            | Yes, Silicone and UV Cured Adhesive                     |
| Packaging<br>(Pertinent to Disposable)                                                                  | Sealed Pouch                                   | Sealed Tray                                             |
| Controller Component<br>Characteristics                                                                 | Instrument, Model 2020                         | Console, Model LFTP-1000                                |
| Operating Orientation                                                                                   | Handheld                                       | Desktop                                                 |
| Temperature accuracy                                                                                    | +/- 1.0°                                       | +/- 0.5°                                                |
| Target temperature range                                                                                | 40 to 42 °C                                    | 40 to 43 °C                                             |
| Maximum Inner Eyelid Surface<br>temperature limit                                                       | 44 °C                                          | 44 °C                                                   |
| Maximum Outer Eyelid<br>Surface temperature limit                                                       | 45 °C                                          | No outer eyelid sensor or<br>temperature limit          |
| Heat Source                                                                                             | LEDs (lime-green and IR<br>wavelengths)        | Resistive (plastic) electric heater                     |
| Rate of heating<br>(time to target temperature)                                                         | 15 to 50 seconds                               | 10 to 60 seconds                                        |
| Glands can be viewed during<br>procedure                                                                | Yes, through magnifier                         | Not visible                                             |

#### Comparison of Technological Characteristics with Predicate Device

Tear Film Innovation, Inc. Traditional 510(k) Premarket Notification iLux System

{5}------------------------------------------------

| Comparison Feature                      | The iLux® System<br>Tear Film Innovations Inc.                                                             | LipiFlow® Thermal Pulsation System<br>TearScience, Inc.                               |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Pressure Control                        | Manual: Eye Care Professional<br>determines (based on patient<br>feedback and direct viewing of<br>glands) | Automatic: Pre-programmed with some<br>adjustment allowed by Eye Care<br>Professional |
| Pressure type                           | Compression, repeated as<br>necessary                                                                      | Massage                                                                               |
| Treatment of upper and lower<br>evelids | Sequential                                                                                                 | Concurrent                                                                            |
| Treatment Time                          | Typically, 8-12 minutes                                                                                    | 12 minutes                                                                            |
| Software                                | Yes                                                                                                        | Yes                                                                                   |
| Power Source                            | Batteries, DC Power                                                                                        | AC Power                                                                              |
| Safety per IEC 60601-1 & -2             | Meets Requirements                                                                                         | Meets Requirements                                                                    |

### Summary of Testing Performed

A program of design verification and validation testing was performed that includes the following:

- . Biocompatibility
- . Sterility and EO Residual
- Packaging Integrity (i.e., Sterile Barrier) ●
- Transportation
- Optical Radiation Safety ●
- . Electromagnetic Compatibility and Electrical Safety
- Stabilitv/Shelf-Life .
- Performance/Functionality/Safety ●
- Software
- Clinical performance ●

Results of the evaluations demonstrate that the iLux System met the safety and performance requirements as it relates to its indication for use.

#### Clinical Study Summary

A clinical study was conducted to verify that the technological differences between the iLux System and the LipiFlow System (i.e., predicate) do not adversely affect safety and effectiveness as it relates to indications for use.

142 subjects (284 eyes) participated in the study, comprised of 101 women and 41 men, ages 19 to 86 years (mean = 54.9 ± 15.3 years). The subjects were randomized for treatment, in a 1:1 ratio into the treatment groups (i.e., iLux, LipiFlow).

The iLux arm of the study met the criteria for non-inferiority relative to the LipiFlow arm for the coprimary effectiveness endpoints - Meibomian Gland Score (MGS) and Tear Break-Up Time (TBUT) and the secondary effectiveness endpoint - Ocular Surface Disease Index (OSDI).

- . MGS improved significantly from baseline in both treatment groups at both week 2 and week 4. MGS improvements did not differ significantly between the two treatment groups at either follow-up visit.

{6}------------------------------------------------

- . TBUT improved significantly from baseline in both treatment groups at both week 2 and week 4. TBUT improvements did not differ significantly between the two treatment groups at either follow-up visit.
- OSDI improved significantly from baseline in both treatment groups at both week 2 and . week 4. OSDI improvements did not differ significantly between the two treatment groups.

There was a total of four device/procedure-related adverse events (AEs) of any type. All AEs were observed in the iLux arm and consisted of: burning sensation without skin findings (n=2), petechial hemorrhage in lower lids (n=1), and transient decrease in BSCVA with findings of superficial punctate keratitis (n=1). All were self-limited, transient, and resolved without sequelae.

The secondary safety endpoints were: 1) discomfort and pain, and 2) changes from baseline for ocular surface staining, Intraocular pressure (IOP), and BSCVA (Best Spectacle Corrected Visual Acuity).

- . Pain scores were significantly lower from the baseline for both treatment groups, immediately post-treatment, and at day 1, week 2 and week 4. Immediately posttreatment, the mean pain score was significantly lower for the LipiFlow arm than for the iLux. Pain scores did not differ significantly between the two treatment groups at day 1, week 2, or week 4.
- . Discomfort scores were significantly lower from the baseline for both treatment groups, immediately post-treatment, and at day 1, week 2 and week 4. Immediately posttreatment, the mean discomfort score was significantly lower for the LipiFlow arm than for the iLux. Improvements in discomfort scores did not differ significantly between the two treatment groups at day 1 or week 2. At week 4, however, the discomfort score was significantly lower for the iLux group.
- Ocular surface staining was significantly increased immediately following treatment in both . treatment groups. At day 1, week 2, and week 4, the ocular surface staining significantly reduced relative to baseline in both treatment groups however did not differ significantly between the two treatment groups.
- . No clinically significant changes in IOP were observed immediately post-treatment or throughout follow-up in either treatment group.
- BSCVA was significantly reduced relative to baseline immediately post-treatment in both . treatment groups but not to a clinically meaningful level. At week 2 and week 4, BSCVA was significantly improved relative to baseline in both treatment groups, but the change was not different between treatment groups.

### Conclusions Drawn from Nonclinical & Clinical Evaluation

The results of the evaluation demonstrate that the iLux System is substantially equivalent to the LipiFlow Thermal System as it pertains to the indications for use and device performance.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/ORZ/K172645](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/ORZ/K172645)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
