Bausch + Lomb Boston Original Cleaner, Bausch + Lomb Boston Original Conditioning Solutions

K173089 · Bausch & Lomb, Incorporated · MRC · Nov 27, 2017 · Ophthalmic

Device Facts

Record IDK173089
Device NameBausch + Lomb Boston Original Cleaner, Bausch + Lomb Boston Original Conditioning Solutions
ApplicantBausch & Lomb, Incorporated
Product CodeMRC · Ophthalmic
Decision DateNov 27, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5918
Device ClassClass 2

Indications for Use

Boston Original Cleaner is indicated for use to clean silicone acrylate and fluoro silicone acrylate gas permeable contact lenses after each removal and before conditioning (wetting, soaking, disinfecting). Boston Original Conditioning Solutions is indicated for wetting, disinfecting and soaking (after cleaning and rinsing) of fluoro silicone acrylate and silicone acrylate gas permeable contact lenses.

Device Story

Device consists of two solutions: Boston Original Cleaner (surfactant solution with alkyl ether sulfate, silica gel, titanium dioxide) and Boston Original Conditioning Solutions (aqueous buffered solution with cellulosic viscosifier, cationic cellulose derivative polymer, polyvinyl alcohol, chlorhexidine gluconate, EDTA). Used by contact lens wearers at home. Cleaner removes debris from lens surface; conditioning solution provides wetting, disinfection, and soaking. Modified regimen replaces tap water rinse with Bausch + Lomb Sensitive Eyes Plus Saline Solution. Benefits include effective lens maintenance and disinfection, reducing risk of ocular complications. Healthcare providers recommend use to ensure lens hygiene and patient comfort.

Clinical Evidence

Bench testing only. Microbiology studies confirmed antimicrobial preservative efficacy per ISO 14730 and regimen performance per ISO 14729:2001/Amd. 1:2010. Lens compatibility verified per ISO 11981. Regimen testing (10-second rub, 5-second rinse) demonstrated effectiveness of the modified saline rinse procedure. No clinical studies were conducted; safety and efficacy are supported by 20 years of commercial history.

Technological Characteristics

Sterile, aqueous-based chemical solutions. Cleaner: surfactant-based (alkyl ether sulfate, silica gel, titanium dioxide). Conditioning solution: buffered, containing cellulosic viscosifier, cationic cellulose derivative polymer, polyvinyl alcohol, chlorhexidine gluconate (0.006%), and EDTA (0.05%).

Indications for Use

Indicated for users of silicone acrylate and fluoro silicone acrylate gas permeable contact lenses requiring cleaning, wetting, disinfecting, and soaking.

Regulatory Classification

Identification

A rigid gas permeable contact lens care product is a device intended for use in the cleaning, conditioning, rinsing, lubricating/rewetting, or storing of a rigid gas permeable contact lens. This includes all solutions and tablets used together with rigid gas permeable contact lenses.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 27, 2017 Bausch & Lomb Incorporated Melissa Thomas Senior Manager Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609 Re: K173089 Trade/Device Name: Bausch + Lomb Boston Original Cleaner, Bausch + Lomb Boston Original Conditioning Solutions Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable Contact Lens Care Products Regulatory Class: Class II Product Code: MRC Dated: October 1, 2017 Received: October 5, 2017 Dear Melissa Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Denise L. Hampton -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173089 Device Name Bausch + Lomb Boston Original Cleaner Bausch + Lomb Boston Original Conditioning Solutions #### Indications for Use (Describe) Boston Original Cleaner is indicated for use to clean silicone acrylate and fluoro silicone acrylate gas permeable contact lenses after each removal and before conditioning (wetting, soaking, disinfecting). Boston Original Conditioning Solutions is indicated for wetting, disinfecting and soaking (after cleaning and rinsing) of fluoro silicone acrylate and silicone acrylate gas permeable contact lenses. | Type of Use (Select one or both, as applicable) | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------| | <table border="0"><tr><td><span></span> Registration Use (Part 21 CFR 801 Subpart D)</td><td><span></span> Over-The-Counter Use (21 CFR 801 Subpart G)</td></tr></table> | <span></span> Registration Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart G) | | <span></span> Registration Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart G) | | _ | Prescription Use (Part 21 CFR 801 Subpart D) |X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY ## Bausch + Lomb Boston Original Cleaner & Boston Original Conditioning Solutions #### 1. Submitter Information | Primary | Alternate | |-----------------------------------------------------------------|---------------------------------------------------------------| | Melissa Thomas | Glenn Davies, O.D. | | Senior Manager, Regulatory Affairs<br>1400 North Goodman Street | Sr. Director Regulatory Affairs,<br>1400 North Goodman Street | | Rochester, NY 14609 | Rochester, NY 14609 | | Tel. (585) 338-6045 | Tel. (585) 338-8215 | | Fax (585) 338-0702 | Email: Glenn.Davies@bausch.com | | Email: Melissa.Thomas@bausch.com | | Summary Prepared: September 27, 2017 ### 2. Device Name | Trade Name: | Bausch + Lomb Boston Original Cleaner<br>Bausch + Lomb Original Conditioning Solutions | |------------------------|----------------------------------------------------------------------------------------| | Classification: | Accessories, Contact Lens Care Products | | Device classification: | Class II | | Regulation Number: | 886.5918 Rigid Gas Permeable Contact Lens Care<br>Products | | Product Code: | MRC | #### 3. Predicate Device Bausch + Lomb Boston Original Cleaner (P820069) Bausch + Lomb Boston Original Conditioning Solutions (K980133) Bausch + Lomb Sensitive Eyes Plus Saline Solution (K170483) #### 4. Description of the Device Boston Original Cleaner is a sterile, concentrated, homogeneous surfactant solution containing alkyl ether sulfate and silica gel as cleaning agents, titanium dioxide and fragrance. Boston Original Conditioning Solutions is a sterile, aqueous buffered, solution containing a cellulosic viscosifier, cationic cellulose derivative polymer and {4}------------------------------------------------ polyvinyl alcohol as wetting and cushioning agents; preserved with chlorhexidine gluconate (0.006%) and edetate disodium (EDTA) (0.05%). #### 5. Intended Use Boston Original Cleaner is indicated for use to clean silicone acrylate and fluoro silicone acrylate gas permeable contact lenses after each removal and before conditioning (wetting, soaking, disinfecting). Boston Original Conditioning Solutions is indicated for wetting, disinfecting and soaking (after cleaning and rinsing) of fluoro silicone acrylate and silicone acrylate gas permeable contact lenses. #### 6. Description of Safety and Substantial Equivalence A series of preclinical testing was performed to demonstrate the safety and effectiveness of the modified regimen which replaces the water rinse with Bausch + Lomb Sensitive Eyes Plus Saline Solution as described in Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997. A brief summary of the test results is provided below: #### Microbiology Studies were previously performed to establish discard dating for both Boston Original Cleaner and Boston Original Conditioning Solutions. The test articles were evaluated initially and again following simulated use for the proposed 90 day discard dates. The results of these evaluations demonstrate that both solutions meet the requirements of ISO 14730, Annex C, Ophthalmic optics - Contact lens care products - Antimicrobial preservative efficacy testing and guidance on determining discard date. Additionally a modified regimen procedure allowing for a rinse with Sensitive Eyes Plus Saline Solution in place of tap water was conducted. The results demonstrate that modified regimen with Sensitive Eyes Plus Saline Solution meets the FDA performance criteria for regimen evaluation as described in the May 1, 1997 Guidance for Industry, Pre-market Notification (510(k)) Guidance Document for Contact Lens Care Products when used in a 10 second rub, 5 second rinse (per lens side) regimen. The products used in this regimen also meet the performance criteria established in ISO Standard 14729: 2001/Amd. 1:2010: (E) for regimen testing. Regimen testing was previously reviewed as part of K170483. #### Biocompatibility Biocompatibility tests were unnecessary for this application. Previous data submitted is still applicable, please reference K980133 for Boston Original Conditioning Solutions, and P820069 for Boston Original Cleaner. #### Lens Compatibility The results of lens compatibility studies per ISO 11981 demonstrate that replacing water with Bausch + Lomb Sensitive Eyes Plus Saline Solution as part of the regimen is compatible with gas permeable contact lenses. {5}------------------------------------------------ #### Clinical Data Clinical studies involving the modified regimen for Boston Original Cleaner and Conditioning Solutions were unnecessary for this application. This Cleaner -Conditioning System has been commercially available in the market for almost 20 years with well demonstrated safety and efficacy. # 7. Substantial Equivalence The cumulative results of laboratory testing sponsored by Bausch + Lomb demonstrate that the safety, effectiveness and performance of the modified regimen for Boston Original Cleaner and Conditioning Solutions to include Bausch + Lomb Sensitive Eyes Plus Saline Solution when used with gas permeable contact lenses are substantially equivalent to the current regimen with a water rinse.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...