← Product Code [MML](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MML) · K990224

# LABTICIAN LID LOAD EXTERNAL EYELID SIZING SET (K990224)

_Labtician Ophthalmics, Inc. · MML · Apr 12, 1999 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MML/K990224

## Device Facts

- **Applicant:** Labtician Ophthalmics, Inc.
- **Product Code:** [MML](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MML.md)
- **Decision Date:** Apr 12, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5700
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

Labtician Lid Load External Eyelid Sizing Weights are used by the physician to determine the correct gold eyelid weight to be implanted in the treatment of lagophthalmos. Labtician Lid Load External Eyelid Sizing Weights are used by the physician who affixes a succession of external eyelid weights to the outer surface of the eyelid using until the proper eyelid closure is attained. Because the levator muscle seems to strengthen after the weight is added, the optimal weight for implantation is ususally that which holds lid about 1.0 mm lower than the normal lid as the patient looks straight ahead. The corresponding implantable gold eyelid weight is then prescribed for implantation. Sizing weights are not designed to be implanted, and are provided nonsterile and are re-usable following cleaning with soap and water and/or alcohol and rinsing with tap water.

## Device Story

Device consists of set of 12 tantalum strips (0.6g to 2.8g) used to determine optimal gold eyelid implant weight for lagophthalmos treatment. Physician manually affixes weights to patient's upper eyelid using adhesive tape; assesses closure until optimal weight identified (typically holding lid 1.0 mm lower than normal). Device is non-sterile, reusable, and cleaned with soap/water or alcohol. Benefits patient by ensuring correct sizing for subsequent surgical implantation of gold weight, facilitating gravity-assisted eyelid closure.

## Clinical Evidence

No clinical data provided; substantial equivalence based on identical design, materials, and intended use to the predicate device.

## Technological Characteristics

Material: 99.5% pure tantalum. Form factor: Rectangular strips with 12.7 mm spherical radius of curvature. Dimensions: 12 sizes (0.6g to 2.8g in 0.2g increments). Non-sterile. Reusable.

## Regulatory Identification

An eyelid weight is a prescription device made of gold, tantalum, platinum, iridium, or surgical grade stainless steel that is rectangular in shape and contoured to the shape of the eye. The device is intended for the gravity assisted treatment of lagophthalmos (incomplete eyelid closure).(1) The external eyelid weight is adhered to the outer skin of the upper eyelid. (2) The implantable eyelid weight is implanted into the upper eyelid.

## Special Controls

*Classification.* (1) Class II (special controls) for the external eyelid weight. The external eyelid weight is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9. The special controls for the external eyelid weight are:(i) Testing demonstrating the biocompatibility of the device; and
(ii) Labeling must include the following information:
(A) Specific instructions regarding the proper placement, sizing, and removal of the device; and
(B) A warning stating that the patient should be instructed to remove the device prior to entering a magnetic resonance environment.
(2) Class II (special controls) for the implantable eyelid weight. The special controls for the implantable eyelid weight are:
(i) Testing demonstrating the biocompatibility of the device;
(ii) Testing demonstrating the sterility and shelf life of the device;
(iii) Nonclinical testing evaluating the compatibility of the device in a magnetic resonance environment.
(iv) Patient labeling to convey information regarding the safety and compatibility of the device in a magnetic resonance environment, the conditions under which a patient with the device can be safely scanned, and a mechanism for a healthcare provider to obtain detailed information about magnetic resonance safety and compatibility if needed.

## Predicate Devices

- EyeClose External Eyelid Weights ([K940974](/device/K940974.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4/9/99

### 510(k) Summary of Safety and Effectiveness 8. In accordance with CFR 807.92 (April 26, 1992), the following information is submitted:

- Labtician Ophthalmics. Inc. 1. Name: 2140 Winston Park Dr. Unit #6 Address: Oakville, ON L6H 5V5 Canada
Telephone: 905 829-0055 905 829-0056 Fax

R. David Sculati Contact:

Date of Summary Preparation: January 22, 1999

- Labtician Lid Load™ External Eyelid 2. Name of Device: Sizina Weights External Eyelid Sizing Weights Common Name: None known Classification Name:
- Predicate Device: EyeClose External Eyelid Weights, K940974. 3.
- Device Description: 4.

Labtician Lid Load External Eyelid Sizing Weights are spherically radiused strips of pure tantalum (99.5%), constructed in twelve sizes ranging from 0.6 grams to 2.8 grams in 0.2 gram increments.

A Labtician Lid Load External Eyelid Sizing Weight is attached to the outer skin of the upper eyelid with a double coated adhesive tape strip or other suitable adhesive.

The physical specifications and materials used in the construction of Labtician Iid Load External Eyelid Sizing Weights are identical to the EyeClose External Eyelid Weights, with the exception that Labtician Lid Load External Eyelid Sizing Weights are unpainted.

- ട. Intended Use:
Sizing weights are used to determine the correct gold eyelid weight to be implanted in the treatment of lagophthalmos. The sizing weights are used by the physician who affixes a succession of external eyelid weights to the outer surface of the eyelid until the proper evelid closure is attained. The corresponding implantable gold eyelid weight is then prescribed for implantation.

Today, the most commonly used and successful surgical treatment of lagophthalmos is implantation of the gold evelid load weight. This device is used to allow the upper lid to close by force of gravity when the levator palpebrae muscle is relaxed.

{1}------------------------------------------------

- 6. Technological Characteristics of the Device: Labtician Lid Load External Eyelid Sizing Weights are produced to the same specifications as the EyeClose External Weights with the exception that the Labtician Lid Load External Sizing Weights are unpainted.
Labtician Lid Load External Eyelid Sizing Weights are constructed of pure tantalum (99.5%). The weights are designed in a rectangular shape with a spherical radius of curvature of 12.7 mm which conforms to the shape of the eye. All edges are smoothly rounded.

# SUBSTANTIAL EQUIVALENCE COMPARISON

| Labtician Lid Load | Med Dev           |
|--------------------|-------------------|
| Extenal Eyelid     | EyeClose External |
| Sizing Weight      | Eyelid Weight     |

| Indications for Use     | Same*   | Same** |
|-------------------------|---------|--------|
| Target Population       | Same    | Same   |
| Design                  | Same    | Same   |
| Materials               | Same*** | Same** |
| Performance             | Same    | Same   |
| Sterility (non sterile) | Same    | Same   |
| Biocompatibility        | Same    | Same   |
| Mechanical Safety       | Same    | Same   |
| Anatomical Site         | Same    | Same   |
| Human Factors           | Same    | Same   |

★ affixed externally-used for determining the correct size gold weight to be implanted in treatment of lagophthalmos.

** affixed externally-used for treatment of lagophthalmos

***pure tantalum-unpainted ****pure tantalum-painted on outer surface

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The caduceus is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters. The text is arranged to follow the curve of the circle, with the department name at the bottom and "USA" at the top.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 2 1999

R. David Sculati Vice President & General Manager Labtician Ophthalmics, Inc. 2140 Winston Park Drive, Unit 6 Oakville, Ontario, Canada, L6H 5V5

Re: K990224 Trade Name: Labtician Lid Load™ External Eyelid Sizing Weight Set Regulatory Class: Unclassified Product Code: 86 MML Dated: January 22, 1999 Received: January 25, 1999

Dear Mr. Sculati:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{3}------------------------------------------------

Page 2 - Mr. David Sculati

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

ARoepl Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| Page                      | of                                         |
|---------------------------|--------------------------------------------|
| 510(k) Number (if known): | K990224                                    |
| Device Name:              | Labtician Lid Load External Sizing Weights |
| Indications For Use:      |                                            |

## Statement of Indications for Use

Labtician Lid Load External Eyelid Sizing Weights are used by the physician to determine the correct gold eyelid weight to be implanted in the treatment of lagophthalmos.

Labtician Lid Load External Eyelid Sizing Weights are used by the physician who affixes a succession of external eyelid weights to the outer surface of the eyelid using until the proper eyelid closure is attained. Because the levator muscle seems to strengthen after the weight is added, the optimal weight for implantation is ususally that which holds lid about 1.0 mm lower than the normal lid as the patient looks straight ahead.

The corresponding implantable gold eyelid weight is then prescribed for implantation.

Sizing weights are not designed to be implanted, and are provided nonsterile and are re-usable following cleaning with soap and water and/or alcohol and rinsing with tap water.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Youzé

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use_

(Optional Format 1-2-96)

Adiley Boudreaux for DEC

(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K990224

R

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MML/K990224](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MML/K990224)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
