Liquid Enzyme ID 90133 Liquid Enzyme ID 90133 is indicated for use with daily wear and extended wear soft (hydrophilic) contact lens to simultaneously enzymatically clean them while they are being disinfected (soaked) in, OPTI-FREE® Rinsing, Disinfecting and Storage Solution, or OPTI-FREE® EXPRESS® Multi-Purpose Solution. Use as recommended by your eye care practitioner. OPTI-FREE® EXPRESS® Multi-Purpose Solution OPTI-FREE® EXPRESS® Multi-Purpose Solution is for use in chemical (not heat) disinfection and for cleaning, rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses. OPTI-FREE® EXPRESS® Multi-Purpose Solution can also be used as a diluent for Liquid Enzyme ID 90133. OPTI-FREE® Rinsing, Disinfecting and Storage Solution OPTI-FREE® Rinsing, Disinfecting and Storage Solution is for use in chemical (not heat) disinfection and for rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses. OPTI-FREE® Rinsing, Disinfecting and Storage Solution can also be used as a diluent for Liquid Enzyme ID 90133.
Device Story
Liquid Enzyme ID 90133 is a sterile, trypsin-based enzymatic cleaner; formulated for dilution in OPTI-FREE or OPTI-FREE EXPRESS Multi-Purpose Solution. Device used by contact lens wearers at home; one drop of enzyme added to 5 mL of disinfecting solution. Simultaneous enzymatic cleaning and chemical disinfection occurs during lens soaking. Benefits include removal of protein deposits from soft (hydrophilic) contact lenses; improved lens hygiene; maintenance of lens clarity. Healthcare providers recommend use as part of lens care regimen.
Clinical Evidence
Clinical study evaluated safety and efficacy of Liquid Enzyme ID 90133/OPTI-FREE EXPRESS regimen. Study included 100 patients (200 eyes) over 3 months. Results demonstrated regimen is safe and effective for daily simultaneous enzymatic cleaning and disinfection of all soft (hydrophilic) contact lenses.
Technological Characteristics
Sterile, trypsin-based enzymatic cleaner; buffered with sodium borate. Liquid form factor. Designed for dilution in multi-purpose contact lens solutions. No electronic components or software.
Indications for Use
Indicated for users of daily and extended wear soft (hydrophilic) contact lenses requiring enzymatic cleaning during chemical disinfection. Compatible with clear and tinted lenses.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Alcon OPTI-FREE® SUPRACLENS® Daily Protein Remover (PMA 820001/S18)
OPTI-FREE® EXPRESS® (PMA 830034/S27)
OPTI-FREE® Rinsing, Disinfection and Storage Solution (PMA 830034/S03)
Submission Summary (Full Text)
{1}------------------------------------------------
.1111 2 4 1998
April 30, 1998
Image /page/0/Picture/2 description: The image shows the number 71561 and the word "Alcon" in a bold, sans-serif font. Below "Alcon" is the word "LABORATORIES" in a smaller font. The text appears to be part of a logo or heading, possibly for a document or publication.
510(K) SUMMARY
Submitted by:
ALCON LABORATORIES, INC 6201 SOUTH FREEWAY FORT WORTH, TEXAS 76134-2099 (817) 293-0450
Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76134-2099 (817) 551-4702 (Phone) (817) 551-4630 (Fax)
## Device Name:
Contact Lens Cleaning Solution Common Name:
Proprietary Name: Liquid Enzyme ID 90133
## Indications for Use:
## Liquid Enzyme ID 90133
Liquid Enzyme ID 90133 is indicated for use with daily wear and extended wear soft (hydrophilic) contact lens to simultaneously enzymatically clean them while they are being disinfected (soaked) in, OPTI-FREE® Rinsing, Disinfecting and Storage Solution, or OPTI-FREE® EXPRESS® Multi-Purpose Solution. Use as recommended by your eye care practitioner.
# OPTI-FREE® EXPRESS® Multi-Purpose Solution
OPTI-FREE® EXPRESS® Multi-Purpose Solution is for use in chemical (not heat) disinfection and for cleaning, rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses.
OPTI-FREE® EXPRESS® Multi-Purpose Solution can also be used as a diluent for Liquid Enzyme ID 90133.
## OPTI-FREE® Rinsing, Disinfecting and Storage Solution
OPTI-FREE® Rinsing, Disinfecting and Storage Solution is for use in chemical (not heat) disinfection and for rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses.
{2}------------------------------------------------
OPTI-FREE® Rinsing, Disinfecting and Storage Solution can also be used as a diluent for Liquid Enzyme ID 90133.
## Description:
Liquid Enzyme ID 90133 is a sterile, trypsin-based enzymatic cleaner buffered with sodium The liquid enzymatic cleaner is formulated for one drop to be diluted in 5 mL of borate. OPTI-FREE® or OPTI-FREE® EXPRESS® Multi-Purpose Solution. Also, this submission provides for labeling changes for OPTI-FREE® EXPRESS® Multi-Purpose Solution and OPTI-FREE® Rinsing, Disinfecting and Storage Solution to be used as diluents for Liquid Enzyme ID 90133.
## Substantial Equivalence:
This product is substantially equivalent, in terms of its actions and indications for use, to Alcon OPTI-FREE® SUPRACLENS® Daily Protein Remover (PMA 820001/S18), OPTI-FREE® EXPRESS® (PMA 830034/S27) and OPTI-FREE® Rinsing, Disinfection and Storage Solution. (PMA 830034/S03). Liquid Enzyme ID 90133 meets the guidelines set forth in FDA's May 1, 1997 Guidance for Industry; Premarket Notification (510(k) Guidance Document for Contact Lens Care Products.
### Safety and Effectiveness:
### A. Non-Clinical Data
### Microbiological Studies
The combination Liquid Enzyme ID 90133 in OPTI-FREE® Rinsing, Disinfecting and Storage Solution was evaluated for disinfection efficacy using the FDA guidelines for contact lens solutions. The results demonstrate that the antimicrobial activity of OPTI-FREE® Rinsing, Disinfecting and Storage Solution is not reduced by the addition of Liquid Enzyme ID 90133 against bacteria or fungi.
### Preclinical
Comprehensive preclinical toxicology tests have been conducted to evaluate the acute toxicity, ocular toxicity and irritation potential of Liquid Enzyme ID 90133 in conjunction with OPTI-FREE® Rinsing, Disinfecting and Storage Solution. Based on the result of these studies, Liquid Enzyme ID 90133 is safe for its intended use with OPTI-FREE® Rinsing, Disinfecting and Storage Solution in the simultaneous cleaning and disinfection of soft (hydrophilic) contact lenses (Groups I-IV) and should not present an ocular hazard to the consumer under the recommended lens treatment regimen or under conditions of accidental or intentional misuse.
{3}------------------------------------------------
## Compatibility/Cleaning Efficacy
Studies were conducted to determine product compatibility with soft contact lenses and The studies demonstrated the its ability to clean laboratory deposited lenses. compatibility and cleaning efficacy of the Liquid Enzyme ID 90133/OPTI-FREE® EXPRESS® Multi-Purpose Solution regimen.
#### B. Clinical
A study was conducted to clinically evaluate the safety and efficacy of Liquid Enzyme ID 90133/OPTI-FREE® EXPRESS® Multi-Purpose Solution regimen for cleaning, rinsing, and disinfection of all soft (hydrophilic) contact lenses (100 patients/200 eyes -3 months). This clinical study demonstrated the Liquid Enzyme ID 90133/OPTI-FREE® EXPRESS® Multi-Purpose Solution regimen is safe and effective for the daily simultaneous enzymatic cleaning and disinfection of soft (hydrophilic) contact lenses.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 4 1998
Mr. Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76134-2099
Re: K981561
Trade Name: Liquid Enzyme ID 90133, OPTI-FREE ® EXPRESS ® Multi-Purpose Solution, and OPTI-FREE ® Rinsing, Disinfecting and Storage Solution Regulatory Class: II Product Code: 86 LPN Dated: April 30, 1998 Received: May 1, 1998
Dear Mr. Larsen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 - Mr. Ralph H. Larsen
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Roerl Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K981561 510(k) Number (if known): __
Device Name: Liquid Enzyme ID 90133
Indications for Use:
Liquid Enzyme ID 90133
Liquid Enzyme ID 90133 is indicated for use with daily wear and extended wear soft (hydrophilic) contact lens to simultaneously enzymatically clean them while they are being disinfected (soaked) in. OPTI-FREE® Rinsing, Disinfecting and Storage Solution, or OPTI-FREE® EXPRESS® Multi-Purpose Solution. Use as recommended by your eye care practitioner.
# OPTI-FREE® EXPRESS® Multi-Purpose Solution
OPTI-FREE® EXPRESS® Multi-Purpose Solution is for use in chemical (not heat) disinfection and for cleaning, rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses.
OPTI-FREE® EXPRESS® Multi-Purpose Solution can also be used as a diluent for Liquid Enzyme ID 90133.
## OPTI-FREE® Rinsing, Disinfecting and Storage Solution
OPTI-FREE® Rinsing, Disinfecting and Storage Solution is for use in chemical (not heat) disinfection and for rinsing, disinfecting and storing of clear and tinted, daily and extended wear soft (hydrophilic) contact lenses.
OPTI-FREE® Rinsing, Disinfecting and Storage Solution can also be used as a diluent for Liquid Enzyme ID 90133.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel W. C. Brown, Ph.D.
(Division Sign-Off) Division of Ophthalmic Devices K981561 510(k) Number
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use X
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.