← Product Code [LPN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN) · K031753

# KARATS MULTIPURPOSE SOLUTION (K031753)

_Ciba Vision Corporation · LPN · Oct 28, 2004 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K031753

## Device Facts

- **Applicant:** Ciba Vision Corporation
- **Product Code:** [LPN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN.md)
- **Decision Date:** Oct 28, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5928
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

AQuify Multipurpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting and storing soft (hydrophilic)lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner. AQuify Multipurpose Solution is also indicated for daily protein removal when used with the no rub directions for use.

## Device Story

AQuify Multipurpose Solution is a sterile aqueous chemical solution for soft contact lens care. It contains dexpanthenol, polyhexanide (0.0001%), pluronic F127, edetate disodium dehydrate, sorbitol, tromethamine, and sodium phosphate. Used by patients at home for cleaning, rinsing, disinfecting, and storing lenses. The solution functions by chemical disinfection and protein removal; it also increases lens water-holding capabilities, improving comfort and reducing dryness. Effectiveness is verified via microbiological efficacy testing and clinical trials comparing patient-reported comfort and lens awareness against predicate solutions. The solution is intended for use with soft lenses on a conventional replacement schedule.

## Clinical Evidence

Three-month prospective, randomized, single-masked clinical trial (n=not specified) comparing AQuify to ReNu Multiplus. Primary endpoints: safety, efficacy, and patient preference. Results: AQuify showed statistically significant superiority in overall comfort, dryness, and lens awareness scores at two weeks, one month, and three months. Preference for AQuify reached a 4:1 ratio by the final visit. Dark-field image analysis confirmed clinically acceptable lens cleanliness for both groups. Microbiology and toxicology data previously reviewed under K021635 confirm compliance with FDA and USP standards.

## Technological Characteristics

Sterile aqueous solution. Ingredients: dexpanthenol, polyhexanide 0.0001%, pluronic F127, edetate disodium dehydrate, sorbitol, tromethamine, sodium phosphate, purified water. Chemical disinfection principle (non-heat).

## Regulatory Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

## Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

## Predicate Devices

- KARATS Multipurpose Solution
- ReNu Multipurpose Solution
- Complete Comfort Plus Multipurpose Solution
- Opti-Free Express Multipurpose Solution

## Reference Devices

- ReNu Multiplus Multipurpose Solution

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 8 2004

Image /page/0/Picture/2 description: The image shows the logo for CIBA Vision, a Novartis company. The logo is in black and white and features the word "CIBA" in large, bold letters above the word "Vision". The "i" in "Vision" is replaced with an eye symbol. Below the word "Vision" is the text "A Novartis Company" in a smaller font.

## 510(k) SUMMARY

In response to the requirements addressed by the Safe Medical Devices Act (SMDA) of 1990, a summary follows with the safety and effectiveness information upon which the substantial equivalence determination is based.

## 510(k) SUMMARY for AQuify Multipurpose Solution

- 1. Submitter Information CIBA Vision Corporation 11460 Johns Creek Parkway Duluth, Georgia 30097 Contact Person: Steven Dowdley Telephone No. 678-415-3897
- 2. Proprietary Name: AQuify Multipurpose Solution Classification Name: Soft (hydrophilic) Contact Lens Solution

#### ന് Predicate Device(s)

KARATS Multipurpose Solution (for lenses replaced in less than 30 days) ReNu Multipurpose Solution Complete Comfort Plus Multipurpose Solution Opti-Free Express Multipurpose Solution

#### 4. Description of the Device

AQuify Multi-Purpose Solution is a sterile aqueous solution containing dexpanthenol, preserved with polyhexanide 0.0001%. pluronic F127, edetate disodium dehydrate, sorbitol, tromethamine, sodium phosphate dihydrogren, and purified water.

#### 5. Indications for Use

AQuify Multipurpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting and storing soft (hydrophilic)lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner. AQuify Multipurpose Solution is also indicated for daily protein removal when used with the no rub directions for use.

#### 6. Description of Safety and Substantial Equivalence

# Cleaning Studies

Previous studies have demonstrated the cleaning capacity of AQuify Multipurpose Solution with soft contact lenses. The study previously submitted and reviewed under K021635 demonstrated that AQuify MPS is substantially equivalent in terms of daily protein removal.

# Dehydration Effect

Using a dynamic vapor sorption system, lenses treated with AQuify Multipurpose Solution showed an effect on the steady state water content at a relative humidity of 50%. AQuify 

{1}------------------------------------------------

MPS was statistically better than control lens multipurpose solution tested at increasing the steady-state water content within a lens group. All groups of lenses indicate that AQuify had a higher water content after 60 minutes at a relative humidity of 50%. This demonstrates that AQuify increased the water holding capabilities of the hydrogel lenses tested.

### Microbiology

A series studies were also completed to demonstrate the microbiological efficacy of AQuify Multipurpose Solution. These studies were submitted and reviewed under K021635. These studies demonstrate that AQuify Multipurpose Solution meets the stand-alone criteria of the disinfection efficacy test of the FDA May 1, 1997 Guidance for Industry; Premarket Notification 510(k) Guidance Document for Contact Lens Care Products. The product also meets the USP Modified criteria for Preservative Effective Testing and USP Sterility Test requirements.

### Toxicology

A series toxicology studies were completed, submitted and reviewed under K021635.

### Clinical Evaluation

The trial was a three-month prospective, randomized, single masked trial consisting of baseline, two-week, one-month and final three-month visits. Subjects compared the Aquify Multipurpose Solution and ReNu Multipurpose Solution on a contra-lateral basis. Study participants were all randomized for which eye used the test and which eye used the control product.

The primary objective of this clinical trial was to evaluate safety, efficacy, and preference between AQUIFY Multipurpose Solution and ReNu Multipurpose Solution. Both solutions used a pre-rinse and a minimum 4 hour to overnight soak.

### Symptoms

There were no significant differences in the AQUIFY Multipurpose Solution and control solution with regards to symptoms.

### Efficacv at Two Weeks

Statistically significant differences in favor of AQuify MPS were found for overall comfort and particularly overall dryness and overall lens awareness 1-10 scores and for lens awareness preference at the two-week visit. There were also strong trends that were almost statistically significant for overall comfort and lens awareness in favor of AQuify MPS.

### Efficacy at One Month

Statistically significant differences in favor of AQuify MPS were found for overall comfort and dryness 1-10 scores and for overall comfort, dryness, and lens awareness preferences.

#### Efficacy at 3 Months

Statistically significant differences in favor of AQuify MPS were found for insertion and overall comfort 1-10 scores and for insertion comfort, overall comfort, dryness, and lens awareness preferences

### Efficacy Conclusions

Overall comfort 1-10 scores and preferences were consistently statistically significantly better with AQuify MPS over ReNu Multiplus throughout the trial. AQuify MPS was also statistically significantly preferred for less dryness throughout the trial having higher mean 1-10 scores for all subjective categories for all visits over ReNu Multiplus. AQuify

{2}------------------------------------------------

MPS was increasingly preferred as the trial progressed to almost a 4:1 ratio (42% versus 11%) by the final visit with over 50% of the subjects expressing a preference.

### Dark Field Image Analysis

With reqards to dark field image analysis, there were no statistically significant differences in the investigator appraisal of lens deposits, front surface wetting and objective dark-field image analysis of lenses worn for 90 days. Both the test and control product provided clinically acceptable lens cleanliness when used according to their directions for use as measured by dark filed image analysis of the returned lenses.

#### 7. Conclusion Drawn from Data Supporting Equivalence Determination:

We conclude that AQUIFY Multipurpose Solution is substantially equivalent to ReNu Multiplus Multipurpose Solution when used to clean, rinse, chemical (not heat) disinfect and store soft (hydrophilic) contact lenses on a conventional replacement schedule of three months or longer

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 8 2004

CIBA Vision Corporation c/o Mr. Steven Dowdley 11460 Johns Creek Parkway Duluth, Georgia 30097

Re: K031753

Trade/Device Name: AQuify Multipurpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Solution Regulatory Class: Class II Product Code: LPN Dated: September 7, 2004 Received: September 8, 2004

Dear Mr. Dowdley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2- Mr. Steven Dowdley

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A kalgi korenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

510(k) Number: K031753

Device Name: AQuify Multi-Purpose Solution

# Indications for Use:

AQuify Multipurpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting and storing soft (hydrophilic)lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner. AQuity Multipurpose Solution is also indicated for daily protein removal when used with the no rub directions for use.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use: or over-the-counter: 内 Avision Sign-Off Division of Ophthalmi Nose and Throat De 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K031753](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K031753)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
