← Product Code [LPN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN) · K030278

# SAUFLON MULTIPURPOSE SOLUTION - NO RUB (K030278)

_Sauflon Pharmaceuticals, Ltd. · LPN · Jul 30, 2003 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K030278

## Device Facts

- **Applicant:** Sauflon Pharmaceuticals, Ltd.
- **Product Code:** [LPN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN.md)
- **Decision Date:** Jul 30, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5928
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Sauflon Multipurpose Solution is indicated for use in the daily cleaning, rinsing, chemical (not heat) disinfection, removal of proteins and storage and soft (hydrophilic) contact lenses, as recommended by the eye care practitioner

## Device Story

Sauflon Multipurpose Solution is a sterile, isotonic chemical solution for soft (hydrophilic) contact lens care. It contains poloxamer, sodium phosphate buffer, sodium chloride, and disodium edetate, preserved with polyhexanide 0.0001%. It is free of chlorohexidine and thimerosal. The solution is used by patients at home for daily lens maintenance, including cleaning, rinsing, disinfection, and storage. It functions as a chemical disinfectant (non-heat). The device is used to maintain lens hygiene and comfort, potentially reducing the risk of ocular complications associated with contaminated lenses.

## Clinical Evidence

Clinical trial of 50 subjects wearing group II and IV soft contact lenses over 3 months, compared to 21 control subjects using the predicate device. The study evaluated safety and acceptability, confirming substantial equivalence to the predicate.

## Technological Characteristics

Sterile, isotonic solution. Ingredients: poloxamer, sodium phosphate buffer, sodium chloride, disodium edetate, polyhexanide 0.0001% preservative. No chlorohexidine or thimerosal. Container components meet USP <661> requirements for ophthalmic preparations. Chemical disinfection principle.

## Regulatory Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

## Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

## Predicate Devices

- Complete Brand Multi-Purpose Solution - No Rub ([K974485](/device/K974485.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SAUFLON MULTIPURPOSE SOLUTION 510(k)

7.0

# 510(k) SUMMARY

# SAUFLON MULTIPURPOSE SOLUTION

| 1.               | Submitted by:             | Sauflon Pharmaceuticals Ltd  49-53 York Street  Twickenham  TW1 3LP  UK                                                                                                                                                                         |
|------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                  | Official correspondent:   | Ligia Delacruz, PhD  Regulatory Affairs Manager                                                                                                                                                                                                 |
| 2.               | Device name               |                                                                                                                                                                                                                                                 |
|                  | Common Name:              | Multipurpose Solution                                                                                                                                                                                                                           |
|                  | Proprietary Name:         | Sauflon Multipurpose Solution                                                                                                                                                                                                                   |
| 3.               | Classification:           | Class II  (Performance Standards)  21 CFR 886.5928  Soft (hydrophilic) contact lens solution                                                                                                                                                    |
| 4.               | Substantial  Equivalence: | The product is substantially equivalent to the currently  marketed Complete Brand Multi-Purpose Solution - No  Rub                                                                                                                              |
|                  | 5. Device Description:    | Sauflon Multipurpose Solution is a sterile, isotonic  solution that contains poloxamer, sodium phosphate  buffer, sodium chloride, and disodium edetate;  preserved with polyhexanide 0.0001%. Contains no  chlorohexidine or thimerosal.       |
|                  |                           | Sauflon Multipurpose Solution remains unchanged  from the previously approved product in K974485,  except for the revised directions for use.                                                                                                   |
| 6. Intended use: |                           | The Sauflon Multipurpose Solution is indicated for use  in the daily cleaning, rinsing, chemical (not heat)  disinfection, removal of proteins and storage and soft  (hydrophilic) contact lenses, as recommended by the  eye care practitioner |

{1}------------------------------------------------

#### 7. Pre-clinical Testing

### Solution compatibility

Details of the compatibility study on the Sauflon Multipurpose Solution are contained in K974485

#### Toxicology

The toxicological testing of Sauflon Multipurpose Solution is contained in K974485

### SAUFLON MULTIPURPOSE SOLUTION CONTAINER

The Sauflon Multipurpose solution container components meet the requirements of the USP<661> for containers and closures for ophthalmic preparations, and it has been confirmed by the appropriate tests (i.e the cytotoxicity, occular irritancy, and systemic toxicity tests).

### Microbiology

#### Sterility

The Sauflon Multipurpose Solution meets the requirements of sterility testing as per K974485.

#### Preservative efficacy

The Sauflon Multipurpose Solution meets the requirements of the preservative efficacy test with rechallenge at 14 days as per K974485.

### Disinfection Efficacy

The Sauflon Multipurpose Solution meets the requirements of both the stand-alone with organic load disinfection test and the regimen test.

#### Clinical Studies 8.

A clinical trial of 3 months usage of the Sauflon Multipurpose Solution by 50 subjects, wearing soft (hydrophilic) contact lenses of either group II and IV, compared to 21 control subjects using to the AMO Complete Brand Multi-Purpose Solution - No Rub, showed the safety, acceptability and substantial equivalence of the Sauflon Multipurpose Solution to the predicate device for its intended use.

#### 9. Conclusions

The Sauflon Multipurpose Solution for use in a no rub regimen for lenses replaced in 30 days or less is substantially equivalent to the AMO Complete Brand Multi-Purpose Solution - No Rub.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with wavy lines below them.

Public Health Service

JUL 3 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Sauflon Pharmaceuticals Ltd. c/o Dr. Ligia Delacruz Regulatory Affairs Manager 49-53 York St. Twickenham. Middlesex TW1 3LP United Kingdom

Re: K030278

Trade/Device Name: Sauflon Multipurpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: June 11, 2003 Received: June 13, 2003

Dear Dr. Ligia Delacruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Dr. Ligia Delacruz

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Levy Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

### 510(k) Number (if known):

Device Name:

Indications For Use:

Sauflon Multipurpose Solution

The Sauflon Multipurpose Solution is indicated for use in the daily cleaning, rinsing, chemical (not heat) disinfection, removal of proteins and storage of soft (hydrophilic) contact lenses, as recommended by the eye care practitioner

hipe Smith

(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K030278

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of the CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The Counter

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K030278](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPN/K030278)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
