FREQUENCY 55 (METHAFILCON A) HYDROPHILIC CONTACT LENS FOR DAILY WEAR
K973063 · Aspect Vision Care, Ltd. · LPL · Aug 26, 1997 · Ophthalmic
Device Facts
Record ID
K973063
Device Name
FREQUENCY 55 (METHAFILCON A) HYDROPHILIC CONTACT LENS FOR DAILY WEAR
Applicant
Aspect Vision Care, Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Aug 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens (clear and tinted) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity. Eyecare practitioners may prescribe the lens for daily wear in a Frequent Replacement Program. The lenses may be disinfected using chemical or hydrogen peroxide disinfection systems.
Device Story
Hemispherical soft contact lens; manufactured from polymerized methafilcon A (HEMA crosslinked with EGDMA); designed to fit over corneal surface. Provides corrective refraction for myopia, hyperopia, and astigmatism. Features base curves, optical zone, and beveled edges for centration and comfort. Prescribed by eyecare practitioners for daily wear, including frequent replacement programs. Patient disinfects lenses using chemical or hydrogen peroxide systems. Device serves as refractive correction; improves visual acuity for patients with ametropia.
Clinical Evidence
No clinical data provided; substantial equivalence based on material, design, and manufacturing process similarity to the predicate device.
Technological Characteristics
Material: methafilcon A (HEMA crosslinked with EGDMA); FDA Group 4 (ionic, high water content). Form factor: hemispherical soft contact lens, 14.0 mm diameter, 8.6 mm base curve, 0.07 mm center thickness. Power range: +20.00 to -20.00D. Tinting: Reactive Blue No. 4 in-monomer process.
Indications for Use
Indicated for daily wear correction of myopia and hyperopia in aphakic and non-aphakic persons with non-diseased eyes, including astigmatism up to 2.00 Diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens for Daily Wear (K971164)
Reference Devices
LifeStyle FREQUENCY Progressive (methafilcon A) Soft Contact Lens for Daily Wear
Related Devices
K971164 — FREQUENCY 55 (METHAFILCON A) HYDROPHILIC CONTACT LENS · Aspect Vision Care, Ltd. · May 21, 1997
K971049 — FREQUENCY 38 (POLYMACON) HYDROPHILIC CONTACT LENS FOR DAILY WEAR (CLEAR AND TINTED) · Aspect Vision Care, Ltd. · May 21, 1997
K973192 — FLEXLENS, FLEXLENS TORIC · Specialty Ultravision, Inc. · Nov 20, 1997
K982947 — BIOMEDICS 55 (OCUFILCON D) UV BLOCKING SOFT (HYDROPHILIC) CONTACT LENS · Ocular Sciences Puerto Rico, Inc. · Jan 29, 1999
{0}------------------------------------------------
# AUG 26 ES K973063
## 510 (k) Summary
#### SUBMITTER:
#### Submitted on behalf of:
| Company Name:<br>Address: | Aspect Vision Care, Ltd.<br>South Point, Hamble<br>Southampton S031 4RF<br>United Kingdom |
|---------------------------|-------------------------------------------------------------------------------------------|
| Phone:<br>Fax: | 011 44 1703 455 567<br>011 44 1703 455 523 |
| CONTACT PERSON: | Martin S. Knopf |
| DATE SUMMARY PREPARED: | August 14, 1997 |
| TRADE NAME: | FREQUENCY 55 TM (methafilcon A) H |
thafilcon A) Hydrophilic Contact TRADE NAME: Lens for Daily Wear
contact lens COMMON NAME:
SUBSTANTIALLY EQUIVALENT TO:
FREQUENCY 55™ (methafilcon A) Hydrophilic Contact Lens for Daily Wear are equivalent to the FREQUENCY 55™ (methafilcon A) Hydrophilic Contact Lens for Daily Wear cleared pursuant to K971164.
FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens for Daily Wear are substantially equivalent to the indications for use, method of manufacture and packaging to the company's predicate device, FREQUENCY 55™ (methafilcon A) Hydrophilic Contact Lens for Daily Wear. This predicate device received marketing clearance pursuant to K971164. The purpose of this premarket notification is to request clearance to tint FREQUENCY 55™ (methafilcon A) Hydrophilic Contact Lens for Daily Wear using Reactive Blue No. 4 in an alternative, inmonomer tinting process. The lens molding manufacturing process will otherwise remain the same as that currently cleared pursuant to K971164. In addition, Aspect Vision Care, Ltd. will continue to manufacture these devices at the same manufacturing location as the predicate devices.
This lens is in Group 4, lonic, high water content polymers as established by the FDA and located in the Guidance Document for Daily Wear Contact Lenses, Revised Edition May 1994. The physical, optical, and chemical properties of the FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens for Daily Wear are equivalent to those of the LifeStyle FREQUENCY Progressive (methafilcon A) Soft Contact Lens for Daily Wear.
{1}------------------------------------------------
#### DESCRIPTION of the DEVICE:
Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA crosslinked with EGDMA. which vield the appearance of lenses which are designed to fit over the comeal surface of the eye. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness), hyperopia (farsightedness) and astigmatism (multiple foci). Each lens provides corrective power. which is to correspond to the refractive power of the eye to which it is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6mm. Secondary and tertiary curves as well as beveled edge configurations are built into the lens for aiding in lens centration and comfort.
#### INDICATIONS FOR USE:
### Device Name: FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens for Daily Wear
The FREQUENCY 55™ (methafilcon A ) Hydrophilic Contact Lens (clear and tinted ) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
Evecare practitioners may prescribe the lens for daily wear in a Frequent Replacement Program. The lenses may be disinfected using chemical or hydrogen peroxide disinfection svstems.
#### PARAMETERS AVAILABLE:
The FREQUENCY 55™ (methafilcon A ) Hydrophilic Contact Lens
| Powers: | +20.00 to -20.00D |
|-------------------|-------------------|
| Center Thickness: | 0.07 mm |
| Diameter: | 14.0 mm |
| Base Curve: | 8.6 mm |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Aspect Vision Care, Ltd. c/o Mr. Martin S. Knopf President and CEO of Knopf Associates, Inc. 84 West Main Street Freehold, NJ 07728
AUG 26 1997
Re: K973063
> Trade Name: Frequency™55 (Methafilcon A) Soft (Hydrophilic) Daily Wear Contact Lens (Clear and Visibility tint, Spherical and Cost-molded) Regulatory Class: II Product Code: 86 LPL Dated: August 14, 1997 Received: August 18, 1997
Dear Mr. Knopf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (HDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Martin S. Knopf
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS STATEMENT
## Device Name: FREQUENCY™ 55 (methafilcon A) (Hydrophilic Contact Lens for Daily Wear (clear and tinted)
The FREQUENCY 55 (methafilcon A) Hydrophilic Contact Lens (clear and tinted) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
Eyecare practitioners may prescribe the lens for daily wear in a Frequent Replacement Program. The lenses may be disinfected using chemical or hydrogen peroxide disinfection systems.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR Overthe-Counter Use Prescription Use (Optional Format 1-2-96)
0-6 - 7 -
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.