← Product Code [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL) · K972303

# OSCULAR SCIENCES/ AMERICAN HYDRON, INC  DAILY WEAR SOFT CONTACT LENS (K972303)

_Ocular Sciences Puerto Rico, Inc. · LPL · Aug 20, 1997 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K972303

## Device Facts

- **Applicant:** Ocular Sciences Puerto Rico, Inc.
- **Product Code:** [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL.md)
- **Decision Date:** Aug 20, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Hydron Biomedics 55 (ocufilcon D) visibility tint soft (hydrophilic) contact lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit astigmatism of 2.00 D or less that does not interfere with visual acuity; it is available in powers from -10 to + 10 diopters. Eyecare practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement (see WEARING SCHEDULE). When prescribed for frequent/ planned replacement wear, the lens may be disinfected using a chemical disinfection system.

## Device Story

Hydron Biomedics 55 is a soft (hydrophilic) contact lens designed for daily wear. It functions as a corrective refractive device, fitting over the corneal surface to address myopia, hyperopia, and minor astigmatism. The lens is a hemispherical shell manufactured from polymerized HEMA and other monomers, crosslinked with EGDMA. It features an optical zone, base curves, and beveled edges to aid centration and comfort. The device is prescribed by eyecare practitioners for either single-use disposable or frequent/planned replacement wear. When used for planned replacement, the lens requires cleaning and chemical disinfection. By providing corrective refractive power, the lens improves visual acuity for the patient.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on identical material, manufacturing process, and physical/chemical properties to previously cleared/approved predicates.

## Technological Characteristics

Soft (hydrophilic) contact lens; material: polymerized HEMA and other monomers crosslinked with EGDMA; Lens Group IV (high water ionic); base curves 6.50-10.8 mm; diameters 12.5-18.0 mm; powers -10.00 to +10.00 D; center thickness 0.025-0.27 mm; in-monomer visibility tint.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Hydron Biomedics 55 ([K942214](/device/K942214.md))
- Hydron Biomedics 55 (PMA [P890023](/device/P890023.md)/S4)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K972303

## 510(k) Summary

AUS 20 88

## SUBMITTER:

## Submitted on behalf of:

| Company Name: | Ocular Sciences/American Hydron Inc                |
|---------------|----------------------------------------------------|
| Address:      | 475 Eccles Avenue<br>South San Francisco, CA 95014 |
| Phone:        | (415) 583-1400                                     |
| Fax:          | (415) 583-1108                                     |

CONTACT PERSON:

David Marcus Ph.D.

DATE SUMMARY PREPARED: August 7, 1977 Hydron Biomedics 55 (ocufilcon D) visibility TRADE NAME: tint molded daily wear soft contact lens

contact lens COMMON NAME:

## SUBSTANTIALLY EQUIVALENT TO:

The Hydron Biomedics 55 (ocufilcon D) visibility tint Ocular Sciences/ American Hydron contact lens with in-monomer tint is equivalent to the tinted daily wear lens of the same material cleared in K942214 for daily wear and is identical to the in-monomer tinted ocufilcon D contact lens approved in PMA P890023/S4 for extended wear currently marketed by Ocular Sciences/ American Hydron, Inc. in the U.S.

The Hydron Biomedics 55 (ocufilcon D) visibility tinted Ocular Sciences/ American Hydron contact lens for daily wear is identical to the Hydron Biomedics 55 in-monomer tinted contact lens marketed for use in the U.S. by Ocular Sciences/ American Hydron approved in PMA P890023/S4 on August 28, 1996; it is manufactured by the same manufacturing process. Additionally, the subject contact lens is substantially equivalent to the tinted Hydron Biomedics 55 contact lens cleared under K942214. This lens is in the Lens Group IV high water ionic group as established by the FDA and located in the Guidance Document for Daily Wear Contact Lenses, Revised Edition May 1994. The physical, optical, and chemical properties of the Hydron Biomedics 55 (ocufilcon D) visibility tinted cast molded contact lens are equivalent to the clear and identical to the tinted versions of the Hydron Biomedics 55 (ocufilcon D cast molded contact lens. 08/07/97

Page 000020

{1}------------------------------------------------

# K972303

### DESCRIPTION OF THE DEVICE:

Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA and other monomeric ingredients crosslinked with EGDMA and other components which yield the appearance of lenses which are designed to fit over the corneal surface of the eye. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness) and hyperopia (farsightedness). Each lens provides corrective power which corresponds to the refractive power of the eye to which it is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6 mm. Secondary and tertiary curves as well as beveled edge configurations are built into the lens for the purpose of aiding in lens centration and comfort.

## INDICATIONS FOR USE:

The Hydron Biomedics 55 (ocufilcon D) visibility tint soft (hydrophilic) contact lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit astigmatism of 2.00 D or less that does not interfere with visual acuity; it is available in powers from -10 to + 10 diopters.

Eyecare practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement (see WEARING SCHEDULE). When prescribed for frequent/ planned replacement wear, the lens may be disinfected using a chemical disinfection system.

## PARAMETERS AVAILABLE:

Base Curves: 6.50 mm to 10.8 mm Diameters: 12.5 mm to 18.0 mm Powers: -10.00 to + 10.00 Diopters sphere Center Thickness: 0.025 mm to 0.27 mm depending on power

08/07/97

Page 000021

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & H. SAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administ 9200 Corporate Boulevard Rockville MD 20850

AUG 20 1997

Ocular Sciences/ American Hydron, Inc. c/o David Marcus, Ph.D. 22210 Quinterno Court Cupertino, CA 95014

Re: K972303

> Trade Name: HYDRON BIOMEDICS 55 (Ocufilcon D) Soft (Hydrophil Wear Contact Lens (In Monomer Visibility Tint, Cast Mole

Regulatory Class: II Product Code: 86 LPL Dated: June 17, 1997 Received: June 19, 1997

#### Dear Dr. Marcus:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2 - David Marcus, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permiss your device to proceed to the market. " " " i

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications Statement

AUG 20 1997

#### 510(k) Number (if known): K972303

## Device Name: Hydron Biomedics 55 (ocufilcon D) Visibility Tint Daily Wear Contact Lens

Indications For Use:

The Hydron Biomedics 55 (ocufilcon D) visibility tint soft (hydrophilic) contact lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit astigmatism of 2.00 D or less that does not interfere with visual acuity; it is available in powers from -10 to + 10 diopters.

Eyecare practitioners may prescribe the lens for either single-use disposable wear or frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement (see WEARING SCHEDULE). When prescribed for frequent/ planned replacement wear, the lens may be disinfected using a chemical disinfection system.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use______________________________________________________________________________________________________________________________________________________________ . OR

Over-The Counter Use_

(Optional Format 1-2-96)

08/07/97

Myra Smith 8/11/97

(Division Sign-Off) Division of Ophthalmic De 510(k) Number_K9 22 303 Page 000013

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K972303](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K972303)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
