SUNSOFT (METHAFILCON A) SOFT (HYDROPHILIC) DAILY WEAR CONTACT LENS
Device Facts
| Record ID | K964696 |
|---|---|
| Device Name | SUNSOFT (METHAFILCON A) SOFT (HYDROPHILIC) DAILY WEAR CONTACT LENS |
| Applicant | Sunsoft Corp. |
| Product Code | LPL · Ophthalmic |
| Decision Date | Feb 20, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.5925 |
| Device Class | Class 2 |
Indications for Use
The Sunsoft (methafilcon A) Soft (Hydrophilic) Contact Lens is a random copolymer of 2-hydroxyethylmethacrylate and methacrylic acid. When hydrated, the lens consists of 45% methafilcon A and 55% water by weight when immersed in normal saline. It is indicated for the correction of visual acuity.
Device Story
Sunsoft (methafilcon A) soft hydrophilic contact lens; random copolymer of 2-hydroxyethylmethacrylate and methacrylic acid; 45% polymer/55% water content. Used for visual acuity correction. Manufactured via static molding and lathing process. Evaluated for biocompatibility (cytotoxicity, systemic toxicity, ocular irritation) and physical/optical properties (light transmittance, refractive index, oxygen permeability, tensile strength). Process validation confirmed consistency across toric, asphere, and sphere designs with Cpk values ≥ 1.2. Clinician-prescribed device.
Clinical Evidence
Bench testing only. Biocompatibility studies (cytotoxicity, acute systemic toxicity, ocular irritation) showed no adverse effects. Physical/optical characterization (transmittance, refractive index, oxygen permeability, tensile strength) and process validation (axis, cylinder, base curve, diameter, thickness, dioptric power) confirmed equivalence to approved manufacturing methods.
Technological Characteristics
Material: methafilcon A (random copolymer of 2-hydroxyethylmethacrylate and methacrylic acid). Hydrated state: 45% polymer, 55% water. Designs: toric, asphere, sphere. Manufacturing: static molding and lathing. Properties: light transmittance, refractive index, oxygen permeability, tensile strength.
Indications for Use
Indicated for correction of visual acuity in patients requiring contact lenses.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
- Sunsoft (methafilcon A) Soft (Hydrophilic) Contact Lens
Related Devices
- K011577 — OCU-FLEX 55 SPHERICAL, TORIC,THIN ZONE TORIC,ASPHERICAL,TORIC ASPHERICAL AND THIN ZONE TORIC ASPHERICAL (METHAFILCON A)S · Ocu-Ease Optical Products, Inc. · Jun 19, 2001
- K030167 — MIERU (ETAFILCON A) DAILY WEAR CONTACT LENS CLEAR AND VISIBILITY TINT, WITH UV BLOCKER · Clearlab Pte , Ltd. · Apr 2, 2003
- K030748 — AQUASOFT (OCUFILCON D) DAILY WEAR CONTACT LENS CLEAR AND VISIBILITY TINT, WITH UV BLOCKER · Clearlab Pte , Ltd. · May 14, 2003
- K180004 — Menicon ASRB (asmofilcon A) Silicone Hydrogel Soft Contact Lens · Menicon Co, Ltd. · Feb 22, 2018
- K962185 — HYDRASOFT (EAST) AND HYDRASOFT TORIC(EAST) (METHAFILCON B) DAILY WEAR SOFT HYDROPHILIC CONTACT LENS · CooperVision, Inc. · Dec 17, 1996