HYDRASOFT (EAST) AND HYDRASOFT TORIC(EAST) (METHAFILCON B) DAILY WEAR SOFT HYDROPHILIC CONTACT LENS
K962185 · CooperVision, Inc. · LPL · Dec 17, 1996 · Ophthalmic
Device Facts
Record ID
K962185
Device Name
HYDRASOFT (EAST) AND HYDRASOFT TORIC(EAST) (METHAFILCON B) DAILY WEAR SOFT HYDROPHILIC CONTACT LENS
Applicant
CooperVision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Dec 17, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The HYDRASOFT (East) lens is intended for use as a daily wear lens for the correction of refractive ametropia (myopic, hyperopic and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes.
Device Story
Hydrasoft (East) soft contact lens; spherical or toric design; composed of methafilcon B (hydroxyethylmethacrylate and methacrylic acid copolymer). Device functions as refractive medium on cornea to focus light on retina. Used by patients for vision correction. Manufactured via alternate process; equivalent to predicate Hydrasoft lenses.
Clinical Evidence
Bench testing only. Evaluated water content, light transmittance, refractive index, and mechanical properties per 1994 FDA guidance. Biocompatibility assessed via systemic injection and cytotoxicity testing.
Technological Characteristics
Material: methafilcon B (hydroxyethylmethacrylate and methacrylic acid copolymer). Form: spherical or toric soft contact lens. Function: refractive medium. Manufacturing: alternate polymerization process.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic patients with non-diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K980418 — BENZ METHAFILCON A OPTICAL BLANK, BENZ METHAFILCON A SPHERICAL AND TORIC SOFT (HYDROPHILIC) CONTACT LENSES FOR DAILY WER · Benz Research and Development Corp. · Feb 27, 1998
K011577 — OCU-FLEX 55 SPHERICAL, TORIC,THIN ZONE TORIC,ASPHERICAL,TORIC ASPHERICAL AND THIN ZONE TORIC ASPHERICAL (METHAFILCON A)S · Ocu-Ease Optical Products, Inc. · Jun 19, 2001
K973063 — FREQUENCY 55 (METHAFILCON A) HYDROPHILIC CONTACT LENS FOR DAILY WEAR · Aspect Vision Care, Ltd. · Aug 26, 1997
{0}
K962185
CoperVision
DEC 17 1996
East Coast Office
200 WillowBrook Office Park
Fairport, New York 14450
(716) 385-6810
(800) 538-7824
Fax (716) 264-3279
West Coast Office
10 Faraday
Irvine, California 92718
(714) 597-8130
(800) 341-2030
Fax (800) 209-0154
J. 510(k) SUMMARY
Page 1 of 2
1. Submitter's Name:
CooperVision, Inc.
711 North Road
Scottsville, NY 14546
Phone: (716) 385-6810
FAX: (716) 889-5688
2. Contact Person:
Bernard Hallatt
Phone: (716) 264-3222
FAX: (716) 889-5688
3. Date Summary Prepared: May 1, 1996
4. Name of Device:
- Trade Name: Hydrasoft (East)
- Common Name: Soft Contact Lens
- Classification Name: Soft (hydrophilic) Contact Lens (Per 21 CFR § 886.5925)
5. Legally Marketed Device: Hydrasoft® and Hydrasoft® Toric (methafilcon B)
6. Description of Device:
The HYDRASOFT (East) (methafilcon B) Soft (hydrophilic) Contact lens is available as a spherical or astigmatic (toric) lens. The lens material, methafilcon B, is a random copolymer of hydroxyethylmethacrylate and methacrylic acid.
When placed on the cornea in its hydrated state the HYDRASOFT (East) Contact Lens act as a refracting medium to focus light rays on the retina.
77
{1}
510(k) SUMMARY (Cont'd)
Page 2 of 2
7. Intended Use:
The HYDRASOFT (East) lens is intended for use as a daily wear lens for the correction of refractive ametropia (myopic, hyperopic and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes.
8. Technological Characteristics:
The technological characteristics of the HYDRASOFT (East) spherical lens and HYDRASOFT TORIC (East) Lens are the same as the predicate device, Hydrasoft® and Hydrasoft® Toric.
9. Summary of Non-Clinical Tests:
Physical and chemical properties testing was performed on lenses made using the alternate manufacturing method and polymerization process as called or in the May 1994 Premarket Notification (510(k)) Guidance Document For Daily Wear Contact Lenses. Water content, light transmittance, refractive index and mechanical properties were tested. Toxicity was determined by the Systemic Injection Test and Cytotoxicity Test along with company data incorporated by reference to an identical manufacturing process for tetrafilcon A lenses.
10. Conclusion:
The determination of substantial equivalence is based on the results of non-clinical testing. Review of the all test data demonstrates methafilcon B lenses manufactured with the alternate manufacturing method and polymerization process are equivalent to the currently marketed Hydrasoft® and Hydrasoft® Toric contact lenses. The Hydrasoft (East) and Hydrasoft Toric (East) contact lenses have the same design, indication and directions for use as the Hydrasoft® and Hydrasoft® Toric contact lenses. CooperVision, Inc. concludes, therefore, that the Hydrasoft (East) and Hydrasoft Toric (East) contact lenses are substantially equivalent to the Hydrasoft® and Hydrasoft® Toric contact lenses.
78
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.